Chapter 4 Remote Interface Reference
Temperature Measurement Commands
[SENSe:]TEMPerature:TRANsducer:FRTD:TYPE {85 | 91}
This command sets the four-wire RTD type. The alpha is entered as
either 85 (for
= .000385) or 91 (for
= .000391).
[SENSe:]TEMPerature:TRANsducer:FRTD:TYPE?
This command queries for the type of RTD being used.
Returns either +91 or +85.
[SENSe:]TEMPerature:TRANsducer:FRTD:RESistance[:REFerence] < value >
This command sets the four-wire RTD R
value. The value is in the range
0
of 4.9
to 2.1 k .
[SENSe:]TEMPerature:TRANsducer:FRTD:RESistance[:REFerence]?
This command queries for the RTD R
value being used.
0
UNIT:TEMPerature {C | CEL | F | FAR | K}
This command sets the measurement and reporting units for
temperature measurements.
UNIT:TEMPerature?
This command queries the temperature measurement reporting units.