Chapter 4 Remote Interface Reference
Triggering Commands
TRIGger:DELay {< seconds > | MIN | MAX}
This command inserts a trigger delay between the trigger signal and each
sample that follows. If you do not specify a trigger delay, the meter
automatically selects a delay for you (see page 89). Select from 0 to 3600
seconds. MIN = 0 seconds. MAX = 3600 seconds. [volatile memory]
TRIGger:DELay? [{MIN | MAX}]
This command queries the trigger delay.
TRIGger:DELay:AUTO {OFF | ON}
This command enables or disables an automatic trigger delay. The delay
is determined by function, range, integration time, and ac filter setting.
Selecting a specific trigger delay value automatically turns off automatic
trigger delay. [volatile memory]
TRIGger:DELay:AUTO?
This command queries the automatic trigger delay setting.
Returns "0" (
) or "1" (
).
OFF
ON
4
SAMPle:COUNt {< value > | MIN | MAX}
This command sets the number of readings (samples) the meter takes per
trigger. Select from 1 to 50,000 readings per trigger.
= 1 and
= 50,000. [volatile memory]
MIN
MAX
SAMPle:COUNt? [{MIN | MAX}]
This command queries the sample count.
TRIGger:COUNt {< value > | MIN | MAX | INFinity}
This command sets the number of triggers the meter will accept before
returning to the "idle" state. Select from 1 to 50,000 triggers. The
INFinity parameter instructs the meter to continuously accept triggers
(you must send a device clear to return to the "idle" state). Trigger count
is ignored while in local operation. MIN = 1 and MAX = 50,000. [volatile
memory]
TRIGger:COUNt? [{MIN | MAX}]
This command queries the trigger count. If you specify an infinity trigger
count, the query command returns "9.90000000E+37".