Test Failures (With Cable Id Unit) - Fluke 620 User Manual

Cablemeter
Table of Contents

Advertisement

620
Users Manual
Failure
Miswire
Open
Pair Length
Split Pair
Fault
*Flashes the number of the wires incorrectly connected.
**These non-specific faults are as follows:
A cable with a resistive path between the wires. Displays measured resistance.
Shield connected with UTP cable selected.
UTP cable connected with STP or COAX selected.
14
Table 7. Test Failures (with Cable ID Unit)
Display
Displays the incorrect wiring of the end connectors.*
Displays the broken wire(s) and the distance to the break.
Indicates that the length of the pairs within a cable are
abnormally different. Use LENGTH to check pair lengths.
A portion of the cable assembly has split pairs or a poor quality
cable such as telephone wire was used instead of LAN type
cabling which may cause excessive crosstalk.
Detects anomalies that do not fit a specific category or failure.**