FV1000/FV1200
Document ID: DID-06397
Set Up Manual
(License D)
FLUOVIEW Series
17.00

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FLUOVIEW FV1200 and is the answer not in the manual?

Questions and answers

Summarization of Contents

Introduction and Safety Precautions
Laser Safety Guidelines
Details on safe laser operation, including enclosed paths, safety glasses, controlled areas, and handling laser components.
System Assembly
IX83 Microscope Setup and Securing
Steps to secure the IX83 microscope, including placing rings, mounting L-type fittings, and handling the unit during transport.
Microscope Setup Procedures
IX81 Microscope Mounting
Step-by-step guide for mounting the IX81 microscope, including placing parts, securing the frame, and using specific screws.
BX61WI Microscope Mounting
Procedure for mounting the BX61WI microscope, including setting it on the table, installing the waterproof cover, and securing components.
Mounting the IX83 Microscope Unit
Steps for mounting the IX83 microscope unit, including attaching screws, removing dummies, and assembling the stage.
FV Unit Setup
Mounting FV10-SU with IX81 Microscope
Step-by-step instructions for mounting the FV10-SU/FV12-SU unit to the IX81 microscope, including attaching feet, inserting parts, and securing screws.
M-Ar Laser and LD559 Power Supply Setup
Procedures for setting up power supplies for M-Ar lasers and LD559, including volume adjustment, interlock plug connection, and ferrite core attachment.
Cable Connection Procedures
Connecting FV10-PC and FV12-CTL Cables
Instructions for connecting cables between FV10-PC, FV12-CTL, and microscopes, noting compatibility and connection types (COM, USB, IEEE1394).
Laser Power Check Procedure
Optical Adjustment for Laser Alignment
Procedures for aligning laser angle and position using optics, adjusting based on laser power checks and overlapping beams.
System Adjustments
Main SU Scan Field Adjustment
Adjusts scan range and centers scan images by tuning Galvano motor Gain and Offset. Covers setup, scanning, and parameter adjustments.
Scanner Delay1 Adjustment (1 kHz X-axis Scan)
Adjusts Scanner Delay1 for 1 kHz bi-directional scan to minimize center position difference in the X-axis Galvano.
ASU Scan Field Adjustment
Adjusts ASU image to match SU image by tuning Gain and Offset for the ASU Galvano motor. Requires specific laser and scan settings.
OSR Performance Check Method
OSR Performance Check: Preparation and Acquisition Settings
Details preparation steps for OSR check, including attaching spectral cubes, microscope objectives, and configuring ASW acquisition settings.
OSR Performance Check: Intensity Profile and Resolution Calculation
Calculates intensity profiles, checks bead diameter FWHM, and ensures it is within specified limits for performance evaluation.
HDD Backup and Recovery Procedures
Creating a System Image Backup
Step-by-step instructions for creating a system image backup on HP PCs, involving Control Panel, backup options, and selecting the destination drive.
HDD Recovery Procedures
Outlines procedures for HDD recovery, directing users based on whether the C-drive or D-drive has failed.
FV12-CTL-SET (Z440) Connectivity
Keyboard and Mouse Connection
Illustrates how to connect USB mouse, USB keyboard, and MOXA devices using the available ports.