Specifications
7 SPECIFICATIONS
Measurement Method
Light Source
Wavelength
Detector
Absorbance Range
Photometric Precision
Automatic Range Selection
Resolution
Accuracy
Calibration Points
Display
Sample Cells
Sample Required
Operating Temperature Range
Sample Temperature Range
Operating Humidity Range
Power Supply
Battery Life
Electromagnetic Compliance
(EMC)
Enclosure Rating
Insulation Rating
Weight
Dimensions
Photometric
Light Emitting Diode (LED)
525 nm
Silicon photodiode
0 – 2.5 Abs
+/- 0.0015 Abs
0 – 1.99 ppm (Free & Total Chlorine)
2.0 – 6.0 ppm (Free & Total Chlorine)
5.9 – 8.2 pH
5 – 90 ppm (Cyanuric Acid)
0.01 ppm (0 – 1.99 ppm; Free & Total Chlorine)
0.1 ppm (2.0 – 6.0 ppm; Free & Total Chlorine)
0.1 pH (5.9 – 8.2 pH)
1 ppm (5 – 90 ppm; Cyanuric Acid)
+/- 0.02 ppm (0 – 1.99 ppm; Free & Total Chlorine)
+/- 0.2 ppm (2.0 – 6.0 ppm; Free & Total Chlorine)
+/- 0.1 pH (5.9 – 8.2 pH)
+/- 4 ppm; Cyanuric Acid
User selectable; 1 point per colorimetric test
4-digit 14-segments customized liquid crystal
display with annunciators
Borosilicate glass with screw caps, fill line and
indexing mark
Height x diameter: 51 x 25 mm ( 2 x 1 in)
10 ml (0.33 oz)
0 – 50°C (32°F to 122°F)
0 – 50°C (32°F to 122°F)
0-90% RH non-condensing at 30°C (86°F)
4 AAA alkaline batteries
> 3000 tests
Emitted Interference - EN 61326
Immunity to Interference - EN 61326
IP67
Pollution degree 2
Instrument: 200 g (7 oz)
Instrument in carrying case: 1.25 kg (2.75 lb)
Instrument: 6.8(W) x 15.5(L) x 4.6(H) cm;
(2.7 x 6.1 x 1.8 in)
Carrying case: 16(W) x 35(L) x 12(H) cm;
(6.3 x 13.8 x 4.7 in)
38
C401/301/201/102/101
Need help?
Do you have a question about the WATERPROOF C401 COLORIMETER and is the answer not in the manual?