Chemistry - EUTECH INSTRUMENTS WATERPROOF C401 COLORIMETER Instruction Manual

Portable colorimeter
Table of Contents

Advertisement

Measurement: Chlorine, Free & Total
C401/301/201/102/101
To re-calibrate the meter: All instruments are factory calibrated and user
calibration is not recommended. However, if a Chlorine standard can be
prepared, measured accurately and qualified by a different method
(titration, for example) or another instrument of higher precision and
accuracy, then a Chlorine standard in the 4.1-5.0 ppm Cl
range can be
2
used. Please refer to Section 3.2 Chlorine, Free & Total, for details.

2.4.6 Chemistry

Chlorine exists in water as Un-Combined (Free) Chlorine and Combined
Chlorine. The concentration of Free Chlorine can be determined directly
and the concentration of Combined Chlorine may be calculated by
subtracting the concentration of Free Chlorine from that of Total Chlorine.
In the Free Chlorine test method Un-Combined Chlorine reacts directly
with DPD (N,N-diethyl-p-phenylenediamine) in a conditioned sample to
give a pink color. In the Total Chlorine test method Combined Chlorine
reacts with potassium iodide to form tri-iodide ions. The tri-iodide ions and
the Free Chlorine present react with DPD in a conditioned sample to give
a pink color.
The depth of the pink color produced is proportional to the concentration of
Chlorine.
The amount of 525nm light absorbed by the pink color is
accurately measured by the meter and then converted into the Chlorine
concentration.