Instruction Manual
8.
SPECIFICATIONS / FEATURES
Conductivity
Ranges
(Resolution)
TDS
Ranges
(Resolution)
Salinity %
Resolution
Salinity ppt
Resolution
Accuracy
Temperature
Resolution
Accuracy
Cell Constant
Temperature
Compensation
Temperature
Coefficient
Normalization
Temperature
TDS factor
Calibration
points
Auto- or
Manual-ranging
Auto standard
recognition
0 to 20.00 (0.01) µS/cm
20.0 to 200.0 (0.1) µS/cm
200 to 2000 (1) µS/cm
2.01 to 20.00 (0.01) mS/cm
20.1 to 200.0 (0.1) mS/cm
0 to 10.00 (0.01) ppm
10.0 to 100.0 (0.1) ppm
100 to 1000 (1) ppm
1.01 to 20.00 (0.01) ppt
20.1 up to 200.0* (0.1) ppt
*depending on TDS factor used
0.10 to 5.00 %
0.01%
1.0 to 50.0 ppt
0.1 ppt
±1% full scale
-10.0 to 110.0 °C
0.1 °C
± 0.5 °C
Selectable
Automatic / Manual
(from 0 to 50 °C)
0.0 to 3.0% per °C
25.0 , 20.0 °C (selectable)
0.4 to 1.0
Maximum 1 per range
Selectable (except Salt 6+)
Selectable (except Salt 6+)
COND 6+ TDS 6+
0.1, 1, 10
5
-25-
COND 6+, TDS 6+, SALT 6+
SALT 6+
0.1, 1, 10
0.1, 1, 10
5
1