Download Print this page

Applying Tolerances - HEIDENHAIN ND 1300 - V2.1.0 Quick Reference Manual

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Operation
Viewing measurement data
Probed data points with form errors
are viewed by selecting a feature in
the feature list and touching the VIEW
button.
1.
Select a feature
Touch the desired feature in the feature
list.
2.
Press the
VIEW button
Form errors are displayed as
lines extending from data
points to the feature. The
two greatest form errors are
shown in red.

Applying Tolerances

Tolerances are applied by selecting
a feature, touching the TOL button,
selecting a tolerance type and entering
tolerance data.
1.
Select a feature
Touch the desired feature in the feature
list.
2.
Press the
TOL button
Tolerance types are
displayed at the bottom
of the screen as tolerance
icons.
3.
Select a tolerance
Touch a tolerance icon to select the
desired tolerance type and then touch
the word TOLERANCE at the top left
corner of the screen to select a specifi c
tolerance.
4.
Enter tolerance data
Enter NOMINAL and TOLERANCE data
into data fi elds provided in the tolerance
screen.
5.
View the result
Green squares near features in the
feature list indicate passed tolerances.
Red squares and outlined characters
on the DRO screen indicate failed
tolerances.
Programming
Programs are recorded sequences
of measurement and other operator
activities stored by the ND 1300 to
be played back later when inspecting
identical parts. This guide discusses
recording, running, saving, loading and
deleting programs.
6
Note:
Programs can also be copied and edited.
Refer to the User's Guide for detailed
information.
1.
Recording a program
Touch the
PROGRAM tab.
Touch the round red
RECORD icon.
Enter a program name and press the
FINISH key to begin recording.
Perform part alignment, measurement
and other steps as usual. Program
recording is indicated by a red
program tab.
To end recording, press the
PROGRAM tab and then press the
square black STOP icon. The new
program will be stored.
Press the
FINISH key to end the
programming session and return to
the DRO.
2.
Running a program
Touch the
PROGRAM tab.
Touch a program name.
Press the black triangular
RUN icon.
The feature type and points probed
will be displayed as points are probed.
After establishing a datum, press the
VIEW button to see point targeting
while points are probed.
The program will stop automatically
when all program steps have been
played. A message box will be
displayed.
Touch the message box to end the
programming session and return to
the DRO.
Saving programs
Programs can be saved to a USB drive.
Plug an empty USB drive into the USB
port on the side of the ND 1300.
Touch the
PROGRAM tab and touch
the program name.
Touch the
COPY PROGRAM
icon.
Press the
FINISH key to return to the
DRO.
Loading programs
Programs can be loaded from a USB
drive.
Plug the USB drive into the USB port
on the side of the ND 1300.
Touch the C:
DRIVE icon
to change drives. The A:
(USB) DRIVE icon and list of
programs stored on the USB
drive will be shown,
Touch the desired program
name in the list and then
touch the LOAD PROGRAM
icon. The highlighted program
will be loaded to the local (C:)
drive.
Touch the
A: DRIVE icon. The
C: DRIVE will be shown with
the loaded program in the C:
DRIVE program list.
The loaded program can now be
selected and run.
3.
Deleting a program
Touch the
PROGRAM tab.
Touch a program name.
Press the
CANCEL key. The program
will be deleted.
Note:
Use caution when deleting programs,
and store a backup of the program fi rst.
Deleted programs cannot be restored.
Press the
FINISH key to end the
programming session and return to
the DRO.
Reporting Results
Reports of results can be sent to a USB
printer, USB fl ash drive or a PC. The
report type and destination are specifi ed
in the PRINT setup screen.
Note:
Refer to the ND 1300 user guide on our
web site at: www.heidenhain.de for
details.
Press the
SEND key to report
results.

Advertisement

loading

This manual is also suitable for:

Nd 1300