Siemens Healthineers Uroskop Omnia Max Operator's Manual

Table of Contents

Advertisement

Quick Links

Uroskop Omnia Max
Operator Manual – VF11 and higher
siemens-healthineers.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Uroskop Omnia Max and is the answer not in the manual?

Questions and answers

Summary of Contents for Siemens Healthineers Uroskop Omnia Max

  • Page 1 Uroskop Omnia Max   Operator Manual – VF11 and higher siemens-healthineers.com...
  • Page 2 WARNING consists of the following elements: • Information about the nature of a hazardous situation • Consequences of not avoiding a hazardous situation • Methods of avoiding a hazardous situation VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 3: Table Of Contents

    3.1.5 Standard working position (recumbent patient) 3.1.6 Patient registration 3.1.7 Selecting the exam set/organ program 3.1.8 Inserting/retracting the grid 3.1.9 Setting image reversal 3.1.10 Checking/setting the patient position VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 4 4.3.1 Postprocessing Documentation 4.4.1 Documentation Ending the examination 4.5.1 Closing the study Technical description Technical Specification 5.1.1 Uroskop Omnia Max system 5.1.2 Digital fluoro-radiography system 5.1.3 Network information 5.1.4 Electromagnetic compatibility (EMC) Labels 5.2.1 Overview 5.2.2 Uroskop Omnia Max system 5.2.3...
  • Page 5 Radiation protection, removable (option) Room accessories 7.3.1 Monitor desk for the control room (option) 7.3.2 Accessories cart (option) 7.3.3 "MoodLight" illuminated panel (option) List of accessories Glossary Index VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 6: Safety

    The electronic Operator Manuals for the system are available on the internet: https://doclib.healthcare.siemens.com. Validity This Operator Manual is valid for Uroskop Omnia Max. Target group This Operator Manual is intended for qualified personnel with the necessary knowledge and expertise to operate X-ray equipment in accordance with the relevant country- specific regulations.
  • Page 7: Storage Location

    Supplements and addenda to the Operator Manual Please read the supplements and addenda to the Operator Manual necessitated by technical developments. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 8: Laws And Regulations

    Regulations in the EU Due to the Medical Device Directive 93/42/EEC and EU regulation 2017/745 (MDR), Uroskop Omnia Max complies with the standards of the EN 60 601-1 series. Incident reporting According to EU regulation 2017/745 (MDR), any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the EU Member State in which the user and/or patient is established.
  • Page 9: Explosion Protection

    These products are marked by the adjacent symbol (on a green triangle). Products with anesthesia equipment test may be repaired only by the manufacturer or by a Siemens Service department authorized by the manufacturer. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 10: Fire Safety

    Please talk to your Siemens Representative about these measures early enough! VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 11: Detector Maintenance

    If you should require additional information, please contact the qualified people specified by the manufacturer. 1.1.8 Malfunctions System malfunction In the case of a system malfunction, call Siemens Service. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 12: Personal Safety Measures

    EMERGENCY SHUTDOWN button off again and start up the system; refer to ( Page 58 Switching the system on/off). Re-initialize the application program. In all other cases, e.g. in the case of a malfunction, please contact Siemens Service. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 13 EMERGENCY SHUTDOWN buttons to shut down the entire system. In this case, the system must not be operated again. Please notify Siemens Service immediately. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 14: Impact And Crushing Hazard

    In particular, ensure that the safety accessories, such as the arm shield, are always attached. (1) Arm shield VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 15 Operator's head hits the monitor support arm Risk of head injury. ◆ Be aware of the risk of injury from hitting the edges of the support arm carrying the display monitors. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 16 The places marked in the illustrations indicate hazard points where the patient or operating personnel could suffer injury by crushing or collision. The hazard points are marked as shown below: Crush hazard points Edges UKGOGPU Uroskop Omnia Max Basic unit VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 17: Collision Hazard

    Risk of collision, damage and injury (for example bruises, contusions, cuts). ◆ Check that no part of the patient's body is in the hazard area of the patient table when an acoustic signal sounds. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 18 In the following situations, when tilting the unit by -90° (Trendelenburg), collisions can occur between the tube support arm and the TFT support arm. The collision hazards are marked as shown below. Possible collision points VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 19: Equipment Damage

    ◆ Do not allow the patient to sit at the head end of the table. ◆ Call Siemens Service and have the tabletop checked immediately if it is possible that it might have been damaged. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 20: Safety During Patient Examination

    ◆ Maintenance work should be performed by trained technical personnel only. ◆ Purchasing a maintenance contract will keep the system in optimum condition. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 21 – ensure that the footboard or the micturition seat is securely locked together with the tabletop on both sides prior to tilting up the unit; – check that the footboard/micturition seat is attached firmly prior to tilting up the unit. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 22 ◆ Make sure there is no uninsulated contact between the patient and the table tops or between the patient, the user and the table tops when using this equipment. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 23 During collimation, the used measuring field is automatically changed to the center measuring field. Outlines of the active measuring fields are visible in the monitor image. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 24 ◆ If the Video Manager fails, make sure that the relevant system (endoscopy, ultrasound) provides another means of display which functions properly. Behavior in the event of an Observe error messages error VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 25: Radiation Protection

    Siemens Healthineers undertakes no liability for diagnoses, which are performed on monitors of other vendors. 1.2.5 Radiation protection The X-ray equipment Uroskop Omnia Max with radiation protection in accordance with IEC 60601-1-3: 2008. Important information Automatic format collimation and/or automatic dose rate control contribute considerably to reducing the radiation dose to the patient and operating personnel.
  • Page 26 • Keep the source-to-skin distance as large as possible since there is the possibility of causing a deterministic effect. Remove all radiopaque parts from the fluoroscopic or radiographic field, if possible. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 27 ◆ Make sure there is no danger of collision with the foot switch when tilting or lowering the unit. Main operating areas for left-hand/right-hand version Location and extent of the Type of examination: ureterorenoscopy relevant operating areas A and B VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 28 1 Safety Right-hand version VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 29 1 Safety Left-hand version Main operating area, dimensions in cm Location and extent of the Type of examination: percutaneous nephrolitholapaxy (PNL) relevant operating areas B, C, and D VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 30 1 Safety Right-hand version VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 31 1 Safety Left-hand version VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 32 – Phantom from left edge: ≈ 17.5 cm centered in beam – X-Ray tube position max. left Height above the Measurement A Measurement B Measurement C Measurement D floor [cm] [µGy/h] [µGy/h] [µGy/h] [µGy/h] VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 33 • Position of PMMA Phantom (25x25x15 cm): – Table Height: ≈ 72 cm; – Phantom from left edge: ≈ 17.5 cm centered in beam – X-Ray tube position max. left VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 34 3467 3676 1634 1248 2891 3084 1509 1192 2478 2668 1374 1109 2081 2030 1249 1024 1692 1649 (1) Height above floor [cm] (2) Air kerma rate [mGy/h] VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 35: Equipment Safety Measures

    Objects on top of the generator cabinet block ventilation Generator becomes inoperable because it is too warm. ◆ Do not store objects on top of the generator cabinet. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 36: Power Connection

    Risk of malfunction of nearby equipment. ◆ Reorient or relocate the equipment or system or shield its location. ◆ Test the operation of the equipment in its new location. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 37: Maximum Patient Weight

    • Move especially – Monitor support systems – OR lights – Gurneys – Beds – Instrument tables out of the tilting range of the table. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 38: Heating Up Of The X-Ray Tube Assembly

    To prevent the X-ray tube from being touched unintentionally, the label "Caution, hot surface" is affixed to both sides of the OPTITOP X-ray tube, see ( Page 119 X-ray tube assembly). VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 39: Software

    Use of unapproved or manipulated software or hardware components Risk of malfunctions posing a hazard to patients and equipment ◆ Only use software or hardware components approved by Siemens Healthineers. ◆ Perform repairs only with our expressed written approval. Software installed inside the Virtual Workstation is exempt from this rule.
  • Page 40: Network Security

    • Devices shall be placed in network zones which fulfill and are appropriate for their information security requirements; • The system must be used by persons with necessary specialist knowledge after training supplied by Siemens Healthineers Representative; • The hospital is responsible for authorized access to the system; VF11 and higher | Operator Manual Uroskop Omnia Max Print No.
  • Page 41: Original Accessories

    • Siemens Healthineers will provide timely patches or mitigations for serious software security vulnerabilities; • Siemens Healthineers releases a security patch for the system regularly (90 days) to repair general security vulnerabilities. 1.3.8...
  • Page 42 ◆ Please notify Siemens Service. The settings of the TFT display have to be adapted to the supplied video output signal of the endoscopy system or the endoscopy system may have to be adjusted. List of disposables Disposable: Drain bag VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 43: Installation And Repair

    • Batteries and packaging material must be disposed of in an environmentally safe manner according to national regulations. • For further information on the disposal of this system, refer to the technical documentation. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 44: System Overview/System Operation

    Uroskop Omnia Max is not for mammography examinations. Patient group Uroskop Omnia Max is not approved for use with patients with a weight (incl. accessories) > 272 kg. Contraindications There are no known contraindications for use of the system.
  • Page 45: Side Effects

    Please note that risks inherently connected to the use of this device (e.g. radiation or moving parts) can not be avoided due to the nature of this device. Physical functionality Universal diagnostic X-ray system Uroskop Omnia Max with overtable X-ray tube assembly and free patient access in urology. Intended user profile...
  • Page 46: System Overview

    IT network characteristics Uroskop Omnia Max may only be run in an environment approved or authorized by the manufacturer. The system may not be operated in public networks without firewalls or isolation.
  • Page 47 X-ray/endoscopic/ultrasound images on the reference image monitor. The spring-loaded articulated arm can be used to position the monitors as required during treatment without restricting access to the patient. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 48 The Video Manager interface allows the connection of several external video sources at the same time, for example: • Endoscopy • Ultrasound • PACS workstation (display of multimodality images) VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 49: Overview Of Control Units

    Any such cable can be rolled up at the rear of the unit when it is not used. 2.1.3 Overview of control units Control console The control console is located on the monitor desk in the control room. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 50 Radiation release - fluoroscopy Hand-held control and advanced multifunctional foot switch Hand-held control The hand-held control attaches to the patient table or the X-ray column with a magnetic holder. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 51 With the Advanced multifunctional foot switch, you control the table movements, foot switch radiation release, display switchover of the reference monitor, and image storage as well as longitudinal movements of the X-ray system. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 52 Save video frame (endoscopy or ultrasound) During fluoroscopy: • Save fluoroscopy image • Save Snapshot image (if licensed) After fluoroscopy: • Save fluoroscopy image (LIH) (press briefly) • Save fluoroscopy series (press longer) VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 53 Select a different organ program Reset fluoroscopy alarm/Acknowledge error messages with OK button/Override motion blocking Vertical diaphragm opens Vertical diaphragm closes Horizontal diaphragm opens Horizontal diaphragm closes Switch on light localizer VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 54 Move grid into park/working position (LED is off when grid is in park position) Raise table Lower table Tilt table -90° (configurable by Siemens Service) Tilt table +90° (configurable by Siemens Service) VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 55 Memory key M1 for recalling stored unit position Memory key M2 for recalling stored unit position Memory key M3 for recalling stored unit position Foot switch Foot switch for releasing radiation for fluoroscopy and acquisition. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 56 30 - 250 s (configurable) Patient handle (max. 40 kg) Accessory rail (for inserting the dose measuring chamber) Tape measure for source-object measurement (10) Slide for covering the laser line light localizer VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 57: Acoustic Signals

    • Skin dose threshold alert • X-ray tube too hot • User information display • Error message requiring user confirmation The sounds can be configured by Customer Service. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 58: System Operation

    No buttons should be pressed and no entries made during the initialization phase! If the endoscopy unit is already switched on before the Uroskop Omnia Max is switched on, the reference monitor automatically switches to the endo image (--> black monitor) after the Uroskop Omnia Max startup is completed.
  • Page 59 If the power network at the site is very unstable, it may be advisable to install an uninterruptible power supply to prevent data loss. Contact Siemens Healthineers for more information. Functioning in the event of In the event of power failure, all indicators and displays are dark. Only the imaging...
  • Page 60 As default, or after 5 minutes of inactivity, the LCD displays the screen saver. The backlight LCD automatically dims after 10 minutes of inactivity. Press any button to restore the screen. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 61 ◆ Correct the alarms and restart if necessary. Starting the UPS on battery Before using this feature, the UPS must have been powered by utility power with output enabled at least once. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 62 • The audio alarm beeps every 3 seconds. The remaining battery power is low. Shut down all applications on the connected system because automatic UPS shutdown is imminent. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 63: Moving The Grid

    The grid can be moved into a park position by means of a motor drive, e.g. to work with a reduced dose when examining children. In the examination of adults, a non-inserted grid leads to impaired image quality due to an increase in scattered radiation. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 64: Limiting The Radiation Field

    Full format Zoom 1 Zoom 2 Zoom 3 42 cm x 42 cm 30 cm x 30 cm 22 cm x 22 cm 15 cm x 15 cm VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 65: Using Digital Zoom

    Full format Zoom 1 Zoom 2 Zoom 3 42 cm x 42 cm 30 cm x 30 cm 22 cm x 22 cm 15 cm x 15 cm VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 66: Moving The X-Ray System And X-Ray Table

    If the tube assembly is not in an end position when the system is switched on, the tube assembly is moved to the exposure position the first time this key is pressed. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 67 Moving the X-ray system longitudinally The X-ray system consists of: SIEMENS Uroskop Omnia Max (1) X-ray tube assembly with collimator (2) Detector VF11 and higher | Operator Manual Uroskop Omnia Max Print No.
  • Page 68: Positioning The Patient Table

    • For table tilting of > 15°, the maximum load capacity on the table is 180 kg. It is not recommended to tilt the table by more than 15° while it is holding a patient during the examination. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 69 1 Hold the hand-held control so that the patient symbol corresponds to the patient position. 2 Press the key on the hand-held control or on the Advanced multifunctional foot switch. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 70 Hand-held Function control The patient table is raised. The patient table is lowered. ◆ Press the key on the hand-held control or on the Advanced multifunctional foot switch. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 71 Moving the table to the transfer position The transfer position allows the patient to get on and off the table. This position can be configured to the customer's needs. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 72 The LEDs of all memory keys (M1, M2, M3) flash. The LEDs of the transfer position and standard working position flash. 2 Press the standard working position key. The current standard working position is stored. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 73: Using The Memory Keys

    If no position is stored under the key that is actuated, the following standard position is set: Table centered longitudinally/transversely, Tube assembly at the far foot end, Full format collimation VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 74: Video Manager (Option)

    Video Manager that are being checked are continuously displayed at the bottom of the screen. Once the configuration run is completed, the previously displayed image is shown again. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 75: Other Control Functions

    • 0.0 mm Cu • 0.1 mm Cu • 0.2 mm Cu • 0.3 mm Cu to reduce the air kerma and air kerma rate, e.g. for children. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 76 Do not touch the lamp with bare fingers. 8 Insert the heat shield all the way to the stop and tighten the screws. 9 Attach the lamp housing. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 77: Examination

    X-ray beam and scattering effects). 3.1.3 Moving the X-ray tube assembly into the park position For easier patient positioning, move the X-ray tube assembly into the park position. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 78: Patient Transfer Position (Recumbent Patient)

    Selecting a patient from the HIS/RIS 1 Click the Get worklist/Update HIS/RIS button in the Patient subtask card to update the HIS/RIS worklist. The Get Worklist Dialog is displayed: VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 79 Manual patient registration 1 Click the Registration button in the Patient subtask card. 2 Enter all the relevant data in the data entry screen displayed; mandatory fields appear in bold. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 80 1 Select the relevant patient in the Examined Patients list. 2 Click the Add study button on the Patient subtask card. The Data Entry Dialog is displayed: VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 81 3 Enter the required additional data. 4 Click Examination. The data is transferred to the Examination task card. Emergency patient 1 Click the Registration button on the Patient subtask card. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 82: Selecting The Exam Set/Organ Program

    The system comes with factory-installed preferred organ programs. Pulsed fluoroscopy (fluoro) The system enables pulsed fluoroscopy with pulse rates of 3 p/s, 7.5 p/s, 10 p/s, 15 p/s, 30 p/s. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 83: Inserting/Retracting The Grid

    Image left/right mirror key. The image is mirrored vertically. The LED lights up when this function is active. 2 Press the Image top/down flip key. The image is flipped horizontally. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 84: Checking/Setting The Patient Position

    When the collimator is opened again, the originally selected measuring field is activated again and displayed. When a different exam set/organ program is selected, the collimator position is reset to the preferred value specified in the organ program. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 85: Examination/Acquisition

    (6) Prefiltration at the X-ray tube assembly (copper filter) (7) Applied dose Shown as a percentage of 2 Gy. Please note that radiation is stopped automatically after 10 minutes of continuous (uninterrupted) fluoroscopy. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 86: Performing Fluoroscopy

    Changing the frame rate The programmed frame rate for pulsed fluoroscopy can be changed by pressing the arrow button. ◆ Press the f/s button and select the required value. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 87: Acquisition

    The image series is saved in the patient study. 3.2.3 Acquisition Acquisition - DR The exposure data of kV, mA, time, and area dose product are displayed on the monitor. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 88 (7) Copper filter (8) Focus (9) Applied dose, tube load Acquisition - RAD The exposure data are displayed on the monitor and can be changed before an exposure. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 89 For a radioscopy of 0.5 s or less, X-ray is terminated within 0.5 s after releasing the foot switch pedal or hand switch button. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 90: Reference Images

    "Archived" (A) flag. A detailed description of all functions can be found in the Operator Manual of the FLUOROSPOT Compact imaging system. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 91: Ending The Examination

    3 Click the Close study button in the Patient subtask card. The study is closed, i.e. no new images can be added. If a follow-up examination of the same patient becomes necessary, the patient must be reregistered. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 92: Basic Workflow

    Patient positioning can be performed at the same time as patient registration. Prior to the examination the monitors have to be positioned for the procedure to be performed. 4.1.2 Registering a patient Patient task card VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 93: Further Preparations

    ◆ Move the grid into the park position if an organ program for pediatrics has been selected. Selecting the image orientation 1 Select vertical/horizontal image flip/mirror according to the patient positioning. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 94: Examination

    The acquired fluoroscopic/X-ray images are displayed on the left monitor. Acquired images can also be stored in the reference folder and displayed on the right monitor. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 95: Storing Fluoroscopic Images

    Video Manager. You can do this using either the hand-held control or the foot switch. The live image of the selected video source is displayed on the reference monitor. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 96: Saving Video Images (Hd Endostore)

    ◆ If you press the button for Saving video frames, the currently displayed endoscopic or ultrasound image of the current study is stored (chronologically along with any stored X-ray images). VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 97: Postprocessing

    2 Select all images that are to be sent, printed, or archived. To send the images to a workstation or to the archive (destination nodes are preconfigured), click the Send button on the Send subtask card. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 98: Ending The Examination

    3 We recommend that you activate the Close study function so that no further images can be added to the study. 4 Move the monitor support arm into the park position and move the patient table into the transfer position. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 99: Technical Description

    5 Technical description 5 Technical description Technical Specification 5.1.1 Uroskop Omnia Max system Basic system System Optionally available as a right-hand or left-hand version Tube assembly/flat detector Longitudinal travel range of 15 cm (with recumbent patient) Tilting range Motorized tilt of ± 90°...
  • Page 100 3/PE ~ 380/400/440/480 V (± 10 %) at 50 Hz or 60 Hz (± 1 Hz); Internal line resistance according to project planning guide Internal fuse 3 x 50 A slow-blow Ambient conditions (operation) Temperature range +15 °C to +35 °C VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 101 Max. output during exposure 126 kVA Internal line resistance To achieve full generator power, the measured internal line resistance may not exceed the following values: Max. R Line Line VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 102 1280 x 1024 Maximum brightness, typical 700 cd/m MAX dynamic integrated flat detector Detector technology Amorphous silicon flat detector with Cesium iodide scintillator High-performance fiber-optic connection to digital imaging system VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 103 Modulation depth at the Nyquist fre‐ Approx. 13% quency Weight 25 kg Anti-scatter grid Stationary, Pb 15:1, 80 lines/cm, f = 125 cm, removable Tabletop-detector distance 6.3 cm VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 104 500 ms Exposure time tolerance ± (10 % +1 ms) kV tolerance ± 8 % mAs tolerance ≤ ± (10% + 0.2 mAs) mA tolerance ≤ ± 20 % VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 105 Smaller radiation field sizes down to 3 cm x 3 cm can be set as required. Laser line light localizer Power < 1 mW, wave length 540 - 700 nm, laser product class II VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 106: Digital Fluoro-Radiography System

    Marking, annotation, image comment, R/L labeling Graphics: Quantification with angle/distance measurement Digital fluoro-radiography system FLUOROSPOT Compact (FLC) Exam Manager Selection of exams; addition, deletion, or replacement of organ programs VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 107 Measurement chamber integrated in the collimator housing to measure the dose area product and the standard patient entrance surface dose The dose area product is displayed on the live image monitor (mandatory in some countries) VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 108 DICOM format and stored in the same patient file together with all the patient images acquired; images of all modalities can be transferred to a PACS as a study; supports HD (high definition) endoscopy VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 109 2 x 1; 3 x 1; 3 x 2; 1 x 1; 1 x 2; 1 x 3; 2 x 2; 2 x 3 Up to 3 network nodes at the same time and one laser camera configurable VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 110 • The measured values then have to be converted to the patient entry reference point at a height of 30 cm above the tabletop. • Measurement at patient entrance reference point. • All measurements were performed with the patient positioning aid. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 111 5 Technical description Fluoro programs Fluoro presets are not selectable on Uroskop Omnia Max high dose low dose always preset 2 is automatically used Organ pro‐ 1_URO_Standard_DR 4_Uro_Pae‐ gram name diatry_DR Fluoro program set CP Standard CP_Paediat‐ Fluoro curve 73 kV Plateau...
  • Page 112: Network Information

    (6) Switch on Hospital Network (7) Any RIS or PACS (8) Ethernet Connection (9) Fibre optics Uroskop Omnia Max uses the Windows firewall to restrict access. The following ports need to be opened in the hospital network firewall: Application Functionality...
  • Page 113: Electromagnetic Compatibility (Emc)

    Shut down the system and turn it on again. Pay attention to error messages during restart. Declaration of immunity and The product complies with following test levels of IEC 60601-1-2:2014: emission compliance VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 114 Declaration of EMC XP product is suitable for professional healthcare facility environment according environment IEC60601-1-2 ed. 4. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 115: Labels

    30 cm (12 inches) to any part of the Uroskop Omnia Max, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result. Note for CISPR11 class A...
  • Page 116: Uroskop Omnia Max System

    (3) Endoscopy shelf (4) Tube stand (5) Hand-held control (6) Tabletop (7) Monitor support arm (8) Monitor desk 5.2.2 Uroskop Omnia Max system SIEMENS SYSTEM (1) - (14) (1) Supplementary label Uroskop Omnia Max VF11 and higher | Operator Manual Uroskop Omnia Max Print No.
  • Page 117 (2) Manufacturer's identification label (2a) MDR Combi Label (6) SGS product certification label (7) Follow the instructions in the Operator Manual SECOND SET OF LABELS TUBE HOUSING ASSEMBLY (8) Supplementary label VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 118 (9) Tube unit identification label (10) X-ray tube identification label (11) Approval label SECOND SET OF LABELS FLATFIELD DETECTOR (12) Supplementary label (13) Identification label (14) Approval label VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 119: X-Ray Tube Assembly

    10 01 Siemens Aktiengesellschaft, Wittelsbacherplatz 2 , DE-80333 Muenchen Made in Germany (1) Unit identification label (1) - (7) (2) Revision status (3) Approval label (4) Approval label VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 120: Multileaf Collimator

    (S) Serial No. 10 01 2005 Siemens AG, Wittelsbacherplatz 2, DE-80333 Muenchen Made in Germany (2) Manufacturer's identification label FILTRATION 1.00 mm Al / 70 kV (3) Filter value label VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 121: Tabletop

    Siemens Part # 83 75 545 G2107 (8) Identification label for localizer light 5.2.6 Tabletop SIEMENS Uroskop Omnia Max (1), (2) (1) Warning label indicating risk of crushing (2) Applied part type B label VF11 and higher | Operator Manual Uroskop Omnia Max Print No.
  • Page 122: Imaging System Container

    10 01 (1) - (3) Siemens Aktiengesellschaft, Wittelsbacherplatz 2 , DE-80333 Muenchen Made in Germany (1) Manufacturer's identification label (2) Revision status label IP X8 (3) Manufacturer's supplementary label VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 123: Flc Keyboard

    (S) Serial No. 10 01 2005 Siemens AG, Wittelsbacherplatz 2, DE-80333 Muenchen Made in Germany (2) Manufacturer's identification label AS 00/01/02/03/04/05/06/07/08/09/10/11/12/13/14/15/16/17/18/19/20/21/22/23/24 (3) Revision status label (4) Approval label VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 124: Hand-Held Control

    (2) Revision status label 5.2.12 TFT support arm (1) Identification label (1) - (2) (2) Revision status label 5.2.13 TFT image monitors (1) Identification label (1), (2) (2) Manufacturer's identification label VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 125: Endoscopy Shelf

    On the right-hand side of the base panel SIEMENS (1P) Model No. 01234567 0123 (S) Serial No. 10 01 2005 Siemens AG, Wittelsbacherplatz 2, DE-80333 Muenchen Made in Germany (1) Manufacturer's identification label VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 126 07757607 (S) Serial No. 10 01 Siemens Aktiengesellschaft, Wittelsbacherplatz 2 , DE-80333 Muenchen Made in Germany (2) Identification label for high-voltage generator with IVK identifier (3) Approval label VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 127: Functional Checks/Cleaning

    Emergency Stop button. The movement must stop immediately, refer to ( Page 13 Red Emergency Stop buttons) and ( Page 128 Checks during the examination). VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 128: Checks During The Examination

    6 Functional checks/Cleaning SIEMENS STOP Uroskop Omnia Max STOP STOP 5 Perform a functional check of the interrupt mechanism (collision protection) by: initiating an arbitrary system movement and actuating the collision protection mechanism at the foot end of the patient table during the movement.
  • Page 129: Monthly Checks

    Risk of unnecessary radiation exposure ◆ In case of unwanted radiation, press the nearest emergency STOP button. ◆ If no cause can be found, contact Siemens Healthineers and discontinue use of the system. 6.1.3 Monthly checks Performing monthly checks 1 Perform a functional check of the automatic dose rate control.
  • Page 130: Functional Check Of The Automatic Systems

    The minimum kV value (approx. 40 kV) is displayed. 9 Switch off fluoroscopy. In Germany, the automatic dose rate control system is checked as part of the constancy test. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 131: Functional Check Of The Monitors/Tft Displays

    ◆ Open and close diaphragm. The light beam changes its height or width. When ACSS lights up at the display of the MAXtouch, the automatic collimation is active. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 132: Maintenance Plan For Checking The System

    Before starting to clean the system, move the X-ray tube into the park position. ( Page 44 System overview/System operation) Make sure that you do not accidentally operate the foot switch and thus release radiation unintentionally while cleaning the unit. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 133: Cleaning

    ◆ Use only substances for cleaning and disinfection which are compatible. ◆ Do not let cleaning liquids seep into the openings of the system (e.g. air openings, gaps between covers). ◆ Observe the cleaning and disinfection instructions. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 134: Ventilation Slots

    3 Slightly lift the tabletop on this side, then push the lever on the other side and lift the tabletop until the safety rod engages in the guide (arrow, illustration below). (1) Safety rod Ventilation slots ◆ Keep the ventilation slots of components clear. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 135: Plastic Material

    Accessory parts Please note that for some accessory parts, special instructions on cleaning are given in the corresponding chapters. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 136 Some substances contained in disinfectants are known to be hazardous to health. The concentration of such substances in the air must not exceed the statutorily defined limit. We recommend that you follow the manufacturers’ operating instructions for these products. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 137: Accessories

    ◆ Do not tilt the table beyond ± 15°. Move the table only under visual control (there is no automatic collision protection with attached tabletop extension). VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 138 An example of incorrect use with uneven weight distribution: A patient of a maximum weight sits at the end of the fully extended tabletop. To remove the extension proceed in the reverse order as for its attachment. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 139: Tabletop Extension 30 Cm (Option)

    2 Insert the shafts of the extension all the way to the end stop in the guide rails of the table and lower the tabletop extension until it is horizontal. 3 Check that the extension is attached securely by pulling on it horizontally. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 140: Coxafix Leg Support, Clip-On (Option)

    Make sure the leg supports are firmly attached (by jiggling them) before positioning the patient's legs. Use the leg supports only with a table tilt angle of up to 15°. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 141 When it locks into place, you can hear this clearly. If the leg support is attached to the accessory rail as described, it cannot fall. Check that it is positioned securely. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 142 6 Fasten the second leg support in the same way. The left and right leg supports are attached so that the leg rests are within the tabletop area. Setting the working position VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 143 2 Turn the cap screw (E) counterclockwise until the support arm can be swiveled. 3 Swivel/tilt the leg rest into the working position. 4 Turn the cap screw (E) fully clockwise. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 144 3 Press the safety button (M) to the inside with your thumb and keep it pressed. 4 Incline the leg support slightly to the inside. 5 Release the safety button. 6 Lift the leg support out of the accessory rail. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 145 3 Swivel the leg rest into the working position. 4 Turn the cap screw (E) fully clockwise. Removing the leg rest 1 Hold the leg rest firmly with one hand. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 146 2 Turn the cap screw (E) counterclockwise until the leg rest moves easily. 3 Hook your finger into the loop. 4 Pull the loop and hold it. 5 Remove the leg rest by lifting it upwards. 6 Release the loop. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 147: Coxafix Leg Support, Slide-On (Option)

    2 Insert the leg support at the end of the accessory rail and slide it into the working position. 3 Secure the leg support using the turning screw (A). VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 148: Pediatric Leg Rests (Option)

    Attachment 1 Attach the leg supports so that they are inclined towards the tabletop center (see ( Page 141 Attachment)). 2 Ensure that they are securely fastened. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 149 7 Release the loop. 8 Swivel/turn the support arm into its working position. 9 Turn the cap screw (E) fully clockwise. 10 Ensure that the support arm is securely fastened. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 150 3 Turn the cap screw (D) fully clockwise. 4 Ensure that the support arm is securely fastened. Titling/swiveling the pediatric 1 Firmly hold the support arm. leg rest VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 151 2 Turn the cap screw (E) counterclockwise until you can move the support arm. 3 Hook your finger into the loop. 4 Pull the loop and hold it. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 152: Amatech Leg Support (Option)

    The foam core pad protects against pressure points. The Amatech foot supports are designed for a maximum patient weight of 159 kg. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 153 4 Place both of the patient's legs in the stirrups simultaneously. 5 Loosen the handle on the boot locking mechanism and carefully slide the boot up or down the rod. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 154 1 Grip the handles (3) with the palm of your hands facing down and turn the handles upward so that your thumbs point to the ceiling. 2 Once the required lithotomy and/or abduction position is reached, simply let go of the handles. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 155: Arm Shield

    Table mat with head support A mattress (1) with head support (2) is supplied for comfortable positioning of the patient. The head support is simply placed on the patient table. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 156: Patient Hand Restraint (Option)

    The Velcro strap allows them to be closed quickly. The restraints can be attached to the accessory rails at any point. The Velcro strap allows them to be closed quickly. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 157 ◆ Ensure that the hand restraints are tightened prior to use. ◆ Use other positioning aids for immobilizing weak or very unstable patients. Attachment 1 Turn the hand screw (A) counterclockwise. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 158: Shoulder Support (Option)

    The shoulder supports can be attached to the accessory rails at any point. The individual shoulder supports can be adjusted longitudinally and transversely. The pads can be swiveled and height-adjusted to the object. 13 kg 13 kg VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 159 Hand screw (B): You can move the shoulder support transversely to the table. Hand screw (C): You can swivel the shoulder support and adjust it to the height of the object. 2 Tighten the hand screws again. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 160: Patient Arm Rest (Option)

    2 Tilt the arm rest to one side (see illustration) and insert it in the accessory rail. 3 Fold the arm rest downwards. 4 Turn the hand screw (A) clockwise until secure attachment is achieved. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 161 5 The arm rest can now be adjusted as follows: • vertically (3) • diagonally (4) • swivelable (2) • tiltable (1) 6 Then tighten the hand screw (B) again. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 162: Accessories For The Patient Table

    Do not let the drain pan be loaded with any weight by patients or users. Otherwise, the drain pan may be damaged. Attachment 1 Bring the table into the horizontal position. 2 Fold the elbow supports outwards. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 163 (A) and fold the rail downwards until it locks into place. 4 Mount the water supply/drain hose. Moving laterally ◆ Push the pan laterally (dark arrow). No latched position. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 164 The latching position is only at the end of the adjustment area. Detaching ◆ Lift the pin (A) (bright arrow) and detach the pan completely (dark arrow). Removal 1 Bring the table into the horizontal position. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 165: Plastic Drain Bag Assembly (Option)

    (on-site installation). Safety information Use the drain bag only once. Attachment 1 To mount the flexible drain bag assembly, first install the spring frame. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 166 3 Slide the drain bag onto the spring frame and place the table-width plastic tab towards you. 4 Push the safety bracket trhough the hollow seam of the drain bag. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 167 7 Connect the drain hose of the bag to the drain union. 8 After use, dispose of the drain bag in an environmentally compatible manner according to relevant regulations. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 168: Elbow Rests

    (patient table, lifting column) or on the variable holder (optional). (1) Magnetic holder on lifting column (2) Magnetic holder on patient table (3) Mounting for hand control unit VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 169 1 Position the mounting at an angle to the front accessory rail of the patient table. 2 Fold the mounting downwards. Removal ◆ The holder can be removed in reversed order of the attachment procedure. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 170: Micturition Seat (Option)

    ◆ When using the micturition seat, make absolutely sure that the patient arm rests are used properly. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 171 2 Move the tabletop so that it is flush with the table support frame (see unit tabletop illustration below). 3 Remove the tabletop extension. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 172 4 Switch on the light localizer to check the position. 5 Dynamic video examinations are performed moving only the X-ray system in order to position the X-ray field (resting patient)! VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 173 The arm rests (left-side/right-side version) are components of the micturition seat. They can be attached to the accessory rails at any point. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 174 Setting the working position 1 Loosen the hand screw (A) slightly. 2 Slide the arm rest into position (dark arrow). 3 Tighten the hand screw again (light arrow). VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 175 Swiveling out of the 1 Grasp the recessed grip located below the seat on the right or left side and pull the 90° position seat forward out of the stop. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 176 4 Pull the shafts of the base plate out from the guide rails of the tabletop. Note that the weight of the base plate without micturition seat is 150 N (15 kg), see ( Page 170 Micturition seat (option)). VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 177: Footboard (Option)

    1 Set the unit to a horizontal position. 2 Press the two levers on the underside of the footboard upward. Two red buttons then appear on the top of the footboard (arrows). VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 178 During examinations in which the patient stands, the two grip strips provide defined hand grip positions for the patient. Attachment ◆ Attach the grip strips (1) on both sides at hip level. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 179: Anesthesia Screen Holder (Option)

    • At a maximum load of 8.0 kg, the extension height of 1050 mm may not be exceeded. This corresponds to one half of the maximum possible extension range. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 180: Endoscopy Cable Holder (Option)

    3 The locking device for height adjustment can be unlocked by lifting the black ring (B). 7.2.9 Endoscopy cable holder (option) The endoscopy cable holders are used to ensure safe routing of the endoscopy cables from the endoscopy equipment to the patient. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 181 1 Tilt the unit footwards until you can easily get to the top side of the tube unit. 2 Remove the two screws (A). 3 Now you can fasten the two endoscopy cable holders (B) in the free holes. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 182: Endoscopy Shelf (Option)

    Used for patient compression during abdominal examinations. It is fastened with a total of four holders to the accessory rails of the patient table. Check the functioning of the compression belt at least once per year. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 183: Radiation Protection, Removable (Option)

    These pockets are riveted to the rails. The plastic pockets are easily cleaned with a mild dishwashing detergent. The lead equivalent of the double-layer lead rubber flap is 0.5 mm Pb at 80 kV. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 184 When in position in the stop, a spring lever in the left profile rail engages in a notch of the bar. 3 Check for the firm seating of the bar on the collimator. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 185: Room Accessories

    7.3.1 Monitor desk for the control room (option) The monitor desk in the control room is used for positioning up to 2 monitors, keyboard, mouse and control panel. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 186: Accessories Cart (Option)

    ◆ Please be careful when pushing cart or wheeled accessories. They are not designed to be pushed over a step, cable or other hindrance because they can tip over or suddenly stop. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 187 (2) Hand restraints (1 pair) (3) Pediatric leg rests (1 pair) (4) Table extension 95 cm (4a) and table extension 30 cm (4b) (5) Holder for hand control unit VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 188: Moodlight" Illuminated Panel (Option)

    You can use this to create a comfortable and relaxing atmosphere for the patient in the exam room. The MoodLight control panel is located on the tube stand. Operating element Function Selection of the 3 color memory locations Auto sequence VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 189 1 Press the button for cycling through the color palette to the left/right. The value last set is displayed. The color palette is cycled through in the activated direction. 2 Press one of the buttons M1, M2, M3 for longer. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 190: List Of Accessories

    List of accessories Below is a list of accessories that are approved for the Uroskop Omnia Max system. If you wish to use other accessories with the Uroskop Omnia Max system that are not contained in the list, please observe the following:...
  • Page 191 5756379 for left-hand unit version Siemens Healthcare GmbH 5756049 for right-hand unit version Siemens Healthcare GmbH Monitor desk 7454593 Siemens Healthcare GmbH Accessories cart 7453991 Siemens Healthcare GmbH VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 192 11155052 Radiation protection stand 85 cm Radiation protection, remov‐ 8375370 Siemens Healthcare GmbH able Junior PAL Stirrups F-10056 Allen Medical Systems Little PAL Stirrups F-10057 Allen Medical Systems VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 193: Glossary

    With the Urodynamics option, a DVI-I live image signal is provided for a urodynamics system. Video Manager The Video Manager interface allows the connection of several external units at the same time. VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 194: Index

    Compression belt 182 of the X-ray tube assembly 38 location 14 Conditions of use 46 High frequency surgical equipment 9 Endoscopic equipment Connection connection 42 of endoscopic equipment 42 VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 195 71 protection 35 positioning 68 Monthly checks 129 Standard working position preparing 77 MoodLight illuminated panel 188 moving to 72 raising/lowering 70 tilting 71 VF11 and higher | Operator Manual Uroskop Omnia Max Print No. XPL5-360G-MDR.620.13.01.24...
  • Page 196 Index Stop buttons 13 Uroskop Omnia Max location 14 labels 116 technical specification 99 Stray radiation in the operating area 32 User profile 45 Supplements to the Operator Manual 7 Validity Switching the system on/off 58 of the Operator Manual 6...
  • Page 197 Siemens Healthineers Headquarters Manufacturer Siemens Healthcare GmbH Siemens Healthcare GmbH Henkestr. 127 Henkestr. 127 91052 Erlangen 91052 Erlangen Germany Germany Phone: +49 9131 84-0 siemens-healthineers.com Published by Siemens Healthcare GmbH / Print No. XPL5-360G-MDR.620.13.01.24 / © Siemens Healthcare GmbH, 2022...