RWD reserves final interpretation right of this instruction manual. RWD has the right to use the instruction as confidential information. Any individual and/or organization shall not disclose the instruction of all or part of the information by any means without RWD’s written permission. Nor shall any other person or organization be allowed to obtain all or part of the information of this instruction manual by any means. No individual and/or organization shall publish, modify, reproduce, issue, rent, adapt, and translate into other languages without RWD’s written permission. ...
Please read this instruction manual and all other auxiliary materials carefully before installing and using the product, which will be helpful to work with it better. RWD has been always dedicated to improving the product function and the service quality, and will reserve the rights to revise the product itself and contents described in the instruction manual at any time without notice in advance. ...
1.2Product Features Standard open circuit breathing loop design; Applicable to rats, mice, rabbits, cats, hamsters and other species; Single breathing circuit design; Anesthesia mask and induction chamber switch adopt double independent control design. Induction and anesthesia can be performed simultaneously. With the function of rapid oxygenation, to remove the residual anesthetic mixture gas in the pipe or anesthesia induction chamber at the fastest speed; Oxygen or air is alternative as the gas source; Pour Fill style anesthetic vaporizer, with capacity ≤ 120 ml, automatic flow and temperature compensation, the precision is not less than 0.5 %, the working temperature is 10 ~ 35 °C; ...
Oxygen grade Medical grade or air Isoflurane: Settable concentration 0 ~ 0.5 ~ 1.0 ~ 1.5 ~ 2.0 ~ 2.5 ~ 3.0 ~ 3.5 ~ 4.0 ~ 5.0 % (v/v) point Sevoflurane: 0 ~ 0.5 ~ 1.0 ~ 2.0 ~ 3.0 ~ 4.0 ~ 5.0 ~ 6.0 ~ 7.0 ~ 8.0 % (v/v) Perfusion volume of The volume between max and min visible liquid level is about 120mL. anesthetic agent About 3 × oxygen flow rate (L/min) × set concentration value(%(v/v)) Consumption of e.g. When the Isoflurane concentration is set at 2% and the oxygen anesthetic agent flow rate is set at 600 mL/min, a bottle of Isoflurane of 100 mL could be consumed for about 28 hours. Loss of anesthetic 22 °C, dial at “0” , < 0.5 mL/24 h. agent Maximum pressure 50 kPa (Anesthetic vaporizer) load Max inclined angle 30° 1.5Product Configuration Table 13 Configuration Name Quantity Usage Description TAIJI Series Compact It is designed to support respiratory Standard Small Animal 1 pcs anesthesia for rodents, rabbits, cats, and Anesthesia Machine other small animals up to 7 kg. It is designed to be two meters and connect ...
Page 10
More attention should be paid to its 2 Anesthetic vaporizer maintenance and operation. We suggest that connect with RWD regularly to calibrate. The internal design of the evaporator varies with different anesthetics. Please use the correct anesthetic agent following the tag ...
Page 11
No. Item Description Induction chamber Icon “ ”, switch to “ON” allows anesthetic gas to flow to 6 outlet switch induction chamber. outlet of Anesthesia 7 Used for connecting Anesthesia mask. mask outlet of Anesthesia 8 Used for connecting Anesthesia induction chamber. induction chamber 9 Air connection Used for connecting oxygen or air pumps. 7 ...
To ensure proper function of your system, perform a pre-use checkout procedure in accordance with Section 4- Operation; The system must always be attended by a veterinarian or trained anesthesia technician; After guarantee period, annual service inspections to ensure proper operation is recommended. Only RWD-certified technicians should be permitted to service the system or replace internal parts; Keep open flames and combustibles (e.g., ether and acetone) away from the anesthesia system; Do not place any heavy things on the anesthesia system; ...
Remove the system from service if any indications of improper function exist, and contact RWD; Never pour any other fluids into the anesthetic vaporizer except the anesthetic agent for which it is designed, note the identification outside the vaporizer before use; Make sure the pressure of the oxygen is below 0.42 MPa. 2.2.2System Protection Immediately collect and clean up spilled anesthetic; Isoflurane is very caustic, avoid it contact with mask and other components if spillage occurs. If spillage should occur, allow it to ...
3- Unpacking and Assembly 3.1Unpack TAIJI Compact Small Animal Anesthesia Machine of RWD has been carefully packed in our factory for safe delivery to you. When you have received your product, please do the as follow: Check the shipping documents to ensure that all boxes have been received; Examine the shipping boxes for damage. Immediately make a damage claim to the carrier if there is serious damage and contact RWD. Taking photos is suggested for keeping the evidence; Carefully open each box and remove each individual component. Save all boxes and packing materials for future shipments; Check the packing list or invoice to ensure all components ordered are included. In case there is any doubt or need any help, contact RWD or local dealer immediately. 3.2System Assembly 3.2.1 Equipment Items Table 31 Machine S/N Product name Specification Configuration Small Animal Anesthesia TAIJI-IE/ 1 Standard Machine TAIJI-SE Anesthesia machine 2 Silicone tube / Standard 3 Intubation Connector / Standard 4 ...
Page 15
Corrugated tube 14 Corrugated tube - ID 22.0 mm R-CT-18 Optional 3.2.2Adjustment of System 1) Take out and place a machine on a stable workbench; Please handle some parts of the machine carefully due to heavy weight! Figure 31 Connect the gas outlet of air pump to the gas inlet of anesthesia machine ① through the screw connection and ensure that the gas source is off as shown in Figure 3-2; ② Tighten by rotation ④ Figure 32 11 ...
Page 16
Connect the Air connection to the gas inlet of anesthesia machine and ② ensure that the gas source is off as shown in Figure 3-3; Figure 33 3) Ensure that the knob on vaporizer is in “0” position. Figure 34 12 ...
Page 17
3.2.3System Connection Connect the anesthesia induction chamber (optional) and the anesthesia mask (optional) to the corresponding ends of anesthesia machine, as shown in Figure 3-5 and Figure 3-6. ⑩ ① ⑪ ⑭ ② Insert to secure ⑩ Tighten by clamping ③ ⑦ ⑧ Figure 35 Connection for passive recovery ⑫ Connection A ⑩ ① B ⑭ Insert to secure ② ③ Insert to the rear card slot ⑬ ⑥ ⑨ ...
3.3Anesthetic Fill The user’s choice of vaporizer differs in the way of filling anesthetics. TAIJI small animal anesthesia machines are equipped with vaporizers subject to Easy-Fill. Make sure to pour the correct anesthetic agent into the filler port. Filling with the incorrect anesthetic agent could cause serious injury or death to the animal! If anesthetic agent spills on system surfaces, allow it to evaporate. Rubbing spilled anesthetic agent with a cloth may damage the system finish. ...
Figure 39 Figure 310 4- Operation 4.1Pre-use Inspection RWD recommends the following inspections prior to application of TAIJI to ensure their stable use: Compact Small Animal Anesthesia Machine Ensure that Anesthetic vaporizer have been added to vaporizer; Ensure that the knob on vaporizer is in “0” position; Ensure that the oxygen source (or air pump) is tightly connected to anesthesia machine; Ensure that the pressure of gas source does not exceed 0.42 MPa and that the use time covers the entire experiment; ...
Page 20
3) Place the animal in the induction chamber and turn the outlet switch of induction chamber “ ” to the right until “ON”, as shown in the figure below. In this case, the anesthesia gas will flow to the anesthesia induction chamber; Figure 42 Press the lock button and rotate the vaporizer dial to adjust anesthetic concentration to an appropriate value, or rotate the dial directly to adjust the concentration back; ...
6) Take out the animal from the anesthesia induction chamber and place its head/nose in the anesthesia mask (detailed operation subject to the type of anesthesia mask used), and then carry out subsequent experimental operations upon the confirmation of proper anesthesia of animal. Note: RWD recommends pressing the O flush button to flush out the high concentration of anesthetic gas in the induction chamber before taking out the animal from the anesthesia induction chamber, so as to reduce the amount of anesthetic gas volatilized when the induction chamber is opened. ...
5- Trouble Shooting 5.1Safety Certain hazardous materials may be used during the troubleshooting. Wastes and contaminants should be disposed in accordance with local laws, regulations, and rules, along with proper and basic personal protection (for example: gloves, masks, goggles, etc.). 5.2Machine Status Unless otherwise specified, the anesthesia system may be connected to gas supplies during the performance of the troubleshooting procedures described in this manual. Make sure the gas supplies and vaporizer are turned off before the procedure. 5.3Record Keeping A record of problems and their resolution should be kept. Such records should include the date, the nature of the problem encountered, and the actions that resolved ...
Page 23
Fill the reservoir with appropriate anesthetic agent. b. Anesthetic agent Do not add anesthetic in use, please reservoir is empty No or low place the vaporizer for 15 minutes 1 anesthetic vapor before use after adding the medicine. output Turn the dial on the vaporizer and adjust c. Vaporizer is turned off it to the desired volume-percent of anesthetic. d. Leak around the Make sure the sealing cap on the vaporizer filler port vaporizer is fully closed. e. Vaporizer malfunction - Contact RWD or local dealer for internal fault servicing. Fill vaporizer with the appropriate anesthetic agent. a. Vaporizer is empty Do not add anesthetic in use. b. Anesthetic concentration Adjust the dial to increase the volume- Animal sleep is set too low percent of Anesthetic vaporizer. 2 level seems too Check breathing system components for light c. Leak in animal breathing leaks, tears, holes, etc. Ensure the mask circuit tube makes a good seal with the animal. ...
Page 24
2. Verify that inlet port is not obstructed by animal hair, etc. Valves need cleaning or replacement, O2 flush button 6 The internal valve is faulty please contact Raywald for after service is stuck support. Float in oxygen 7 Dirty flow tube Flow tube needs cleaning. flowmeter sticks Oxygen flow Flow valve needs cleaning or control valve Dirty or damaged needle 8 replacement. Contact RWD or local knob is hard to valve dealer for servicing. turn Oxygen flow Contact RWD or local dealer for 9 Internal fault failure servicing. a. Loose oxygen hose Tighten the connection with a crescent Audible leakage connection wrench. 10 around oxygen hose connector b. Tube does not fit. ...
A record of system service and maintenance should be kept. Such records should include service and maintenance dates, part numbers of any replaced parts, dates when consumables are replenished, and other pertinent data. 6.5Parts and Materials Contact RWD or your local dealer if parts or materials are needed during the maintenance. 6.6Cleaning the Equipment Maintenance of the anesthesia system involves possible contact with biological and chemical hazards. Wear gloves, mask, and goggles during all cleaning! ...
Page 26
6.6.1Required Materials and Supplies Germicidal cleaner(Cleaning solution with sterilizing function, such as 3% hydrogen peroxide or 75% alcohol); Water and clean cloth; Personal Protective Equipment (gloves, masks, goggles, etc). 6.6.2Weekly Maintenance RWD recommends that the anesthesia machine be fully maintained once a week. Follow these steps: Wipe all surfaces of the system with a cloth dampened with germicidal cleaning solution; Remove hair, dust, and debris from all vaporizer surfaces; especially around the volume-percent dial and the anesthetic fill port; Check whether the air filter tank is overweight. If it is overweight, ...
Page 27
3) Attach a drain tube to the drain port of anesthetic vaporizer. Place the other end of the tube in a receptacle to catch the drained anesthetic agent; Figure 62 4) Loosen the sealing cap, then loosen the bottom screw with a screwdriver to make the anesthetic agent flows naturally; Figure 63 5) When the anesthetic agent stops draining from the vaporizer, tighten the bottom screw and sealing cap; ...
The warranty of this equipment starts from the date of leaving the factory. During the warranty period, the equipment cannot be used normally due to problems such as materials and process defects. RWD undertakes after-sales service such as equipment maintenance and parts replacement. Any damage caused by improper use or over-range use is not covered by the ...