RWD TAIJI Series User Manual

Compact small animal anesthesia machine

Advertisement

Quick Links

 
TAIJI Series Compact  
Small Animal Anesthesia Machine 
User Manual           A 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TAIJI Series and is the answer not in the manual?

Questions and answers

Summary of Contents for RWD TAIJI Series

  • Page 1   TAIJI Series Compact   Small Animal Anesthesia Machine  User Manual           A ...
  • Page 2: Intellectual Property Right

    RWD reserves final interpretation right of this instruction manual.  RWD  has  the  right to use the instruction as confidential information. Any individual  and/or  organization  shall  not disclose the instruction of all or part of the information  by  any  means  without  RWD’s  written  permission.  Nor  shall  any  other  person  or  organization  be  allowed  to  obtain  all  or  part  of  the  information  of  this  instruction  manual by any means.  No individual and/or organization shall publish, modify, reproduce, issue, rent, adapt,  and translate into other languages without RWD’s written permission. ...
  • Page 3: Table Of Contents

    Non RWD authorized personnel repairs or alterations to the instruments;  Product may not be in accordance with the manual.  Content  1- Introduction 1  1.1 Overview 1  1.2 Product Features 2  1.3 Equipment Environment Requirements 3  1.4 Product Parameter 3  1.5 Product Configuration 4  1.6 Product Structure 5  2- Safety 8  2.1 Usage Restriction 8  2.2 Safety Instruction 8  2.3 System Malfunction 9  2.4 Emergency Shutdown 9  3- Unpacking and Assembly 10 ...
  • Page 4 8.1 Pressure Equivalents 25  8.2 Pressure Unit Conversions 25  8.3 Reference for Maintaining Anaesthetic Concentration 25     ...
  • Page 5: 1- Introduction

    Please  read  this  instruction  manual  and  all  other  auxiliary  materials  carefully  before installing and using the product, which will be helpful to work with it better.  RWD  has  been  always  dedicated  to  improving  the  product  function  and  the  service  quality,  and  will  reserve  the  rights  to  revise  the  product  itself  and  contents  described in the instruction manual at any time without notice in advance. ...
  • Page 6: Product Features

    1.2Product Features  Standard open circuit breathing loop design;  Applicable to rats, mice, rabbits, cats, hamsters and other species;  Single breathing circuit design;  Anesthesia  mask  and  induction chamber switch adopt double independent  control design. Induction and anesthesia can be performed simultaneously.  With  the  function  of  rapid  oxygenation,  to remove the residual anesthetic  mixture  gas  in  the  pipe  or  anesthesia  induction  chamber  at  the  fastest  speed;  Oxygen or air is alternative as the gas source;  Pour Fill style anesthetic vaporizer, with capacity ≤ 120 ml, automatic flow  and  temperature  compensation,  the  precision  is  not  less  than  0.5  %,  the  working temperature is 10 ~ 35 °C; ...
  • Page 7: Equipment Environment Requirements

    1.3Equipment Environment Requirements  Please  prepare  equipment  operation  status  as  Table  1-1  to  ensure  reliable  operation and safety.  Table 11  Environment  Detail Description  Pressure: ≤ 0.42 MPa  Gas tightness  Leakage of machine to the atmosphere: less than 25 ml/min  Operating temperature: 10°C ~ 35°C  Operating  Operating humidity: 5% ~ 90% (Non-condensing)  environment  Working pressure: 57 kPa ~ 106 kPa  Storage and transportation temperature: -40°C ~ 55°C  Storage and  Storage and transportation humidity: 5% ~ 95% (Non-condensing)  transportation  environment  Storage and transportation pressure: 50 kPa ~ 106 kPa  1.4Product Parameter  Table 12  Parameter  Specifications  Material  Mainly Metal  Anesthesia machine  (without outer  Length ≤ 280 mm, width ≤ 230 mm, height ≤ 320 mm  package)  Weight of anesthesia ...
  • Page 8: Product Configuration

    Oxygen grade  Medical grade or air  Isoflurane:  Settable concentration  0 ~ 0.5 ~ 1.0 ~ 1.5 ~ 2.0 ~ 2.5 ~ 3.0 ~ 3.5 ~ 4.0 ~ 5.0 % (v/v)  point  Sevoflurane:  0 ~ 0.5 ~ 1.0 ~ 2.0 ~ 3.0 ~ 4.0 ~ 5.0 ~ 6.0 ~ 7.0 ~ 8.0 % (v/v)  Perfusion volume of  The volume between max and min visible liquid level is about 120mL.  anesthetic agent  About 3 × oxygen flow rate (L/min) × set concentration value(%(v/v))  Consumption of  e.g. When the Isoflurane concentration is set at 2% and the oxygen  anesthetic agent  flow rate is set at 600 mL/min, a bottle of Isoflurane of 100 mL could  be consumed for about 28 hours.  Loss of anesthetic  22 °C, dial at “0” , < 0.5 mL/24 h.  agent  Maximum pressure  50 kPa (Anesthetic vaporizer)  load  Max inclined angle  30°  1.5Product Configuration  Table 13  Configuration  Name  Quantity  Usage Description  TAIJI Series Compact  It is designed to support respiratory  Standard  Small Animal  1 pcs  anesthesia for rodents, rabbits, cats, and  Anesthesia Machine  other small animals up to 7 kg.  It  is  designed  to  be  two  meters  and  connect ...
  • Page 9: Product Structure

    1.6Product Structure  Front view  1  4  2  5  6  3    Figure 14  Rear view  5 ...
  • Page 10 More  attention  should  be  paid  to  its 2  Anesthetic vaporizer  maintenance  and  operation.  We  suggest  that  connect  with  RWD regularly to calibrate.    The  internal  design  of  the  evaporator  varies  with  different anesthetics.  Please  use  the  correct  anesthetic  agent  following  the tag ...
  • Page 11 No.  Item  Description  Induction chamber  Icon  “ ”,  switch  to  “ON”  allows  anesthetic  gas  to  flow  to 6  outlet switch  induction chamber.  outlet of Anesthesia  7  Used for connecting Anesthesia mask.  mask  outlet of Anesthesia  8  Used for connecting Anesthesia induction chamber.  induction chamber  9  Air connection  Used for connecting oxygen or air pumps.    7 ...
  • Page 12: 2- Safety

    To  ensure  proper function of your system, perform a pre-use checkout  procedure in accordance with Section 4- Operation;  The  system  must  always  be  attended  by  a  veterinarian  or  trained  anesthesia technician;  After  guarantee  period,  annual  service  inspections  to  ensure  proper  operation  is  recommended. Only RWD-certified technicians should be  permitted to service the system or replace internal parts;  Keep  open  flames  and  combustibles  (e.g.,  ether  and  acetone)  away  from the anesthesia system;  Do not place any heavy things on the anesthesia system; ...
  • Page 13: System Malfunction

    Remove  the  system  from  service  if  any  indications  of  improper  function exist, and contact RWD;  Never  pour  any  other  fluids  into  the  anesthetic  vaporizer  except  the  anesthetic agent for which it is designed, note the identification outside  the vaporizer before use;  Make sure the pressure of the oxygen is below 0.42 MPa.  2.2.2System Protection  Immediately collect and clean up spilled anesthetic;  Isoflurane  is  very  caustic,  avoid  it  contact  with  mask  and  other  components  if  spillage  occurs.  If  spillage  should  occur,  allow  it  to ...
  • Page 14: 3- Unpacking And Assembly

    3- Unpacking and Assembly  3.1Unpack  TAIJI Compact Small Animal Anesthesia Machine of  RWD has been carefully  packed in our factory for safe delivery to you. When you have received your product,  please do the as follow:  Check the shipping documents to ensure that all boxes have been received;  Examine the shipping boxes for damage. Immediately make a damage claim to  the carrier if there is serious damage and contact RWD. Taking photos is  suggested for keeping the evidence;  Carefully open each box and remove each individual component. Save all  boxes and packing materials for future shipments;  Check the packing list or invoice to ensure all components ordered are  included. In case there is any doubt or need any help, contact RWD or local  dealer immediately.  3.2System Assembly  3.2.1 Equipment Items  Table 31  Machine  S/N  Product name  Specification  Configuration  Small Animal Anesthesia  TAIJI-IE/  1  Standard  Machine  TAIJI-SE  Anesthesia  machine  2  Silicone tube  /  Standard  3  Intubation Connector  /  Standard  4 ...
  • Page 15 Corrugated tube  14  Corrugated tube - ID 22.0 mm  R-CT-18  Optional  3.2.2Adjustment of System  1) Take out and place a machine on a stable workbench;  Please handle some parts of the machine carefully due to  heavy weight!      Figure 31  Connect the gas outlet of air pump to the gas inlet of anesthesia machine  ①  through the screw connection and ensure that the gas source is off as shown  in Figure 3-2;  ②  Tighten by rotation  ④                  Figure 32  11 ...
  • Page 16 Connect the Air connection to the gas inlet of anesthesia machine and  ②  ensure that the gas source is off as shown in Figure 3-3;                          Figure 33  3) Ensure that the knob on vaporizer is in “0” position.    Figure 34  12 ...
  • Page 17 3.2.3System Connection  Connect the anesthesia induction chamber (optional) and the anesthesia mask (optional) to  the corresponding ends of anesthesia machine, as shown in Figure 3-5 and Figure 3-6.      ⑩  ①  ⑪  ⑭  ②  Insert to secure  ⑩  Tighten by clamping  ③  ⑦     ⑧      Figure 35 Connection for passive recovery      ⑫      Connection A  ⑩  ①  B  ⑭  Insert to secure  ②  ③  Insert to the rear card slot  ⑬  ⑥                  ⑨   ...
  • Page 18: Anesthetic Fill

    3.3Anesthetic Fill  The  user’s  choice  of  vaporizer  differs  in  the  way  of  filling  anesthetics.  TAIJI  small animal anesthesia machines are equipped with vaporizers subject to Easy-Fill.  Make  sure  to  pour  the  correct  anesthetic  agent  into  the filler port. Filling with the incorrect anesthetic agent could cause serious injury or death to the animal!    If  anesthetic  agent  spills  on  system  surfaces,  allow  it  to evaporate.  Rubbing  spilled  anesthetic  agent  with  a  cloth may damage the system finish. ...
  • Page 19: 4- Operation

    Figure 39  Figure 310  4- Operation  4.1Pre-use Inspection  RWD  recommends  the  following  inspections  prior  to  application  of  TAIJI   to ensure their stable use:  Compact Small Animal Anesthesia Machine Ensure that Anesthetic vaporizer have been added to vaporizer;  Ensure that the knob on vaporizer is in “0” position;  Ensure  that  the  oxygen  source  (or  air  pump)  is  tightly  connected  to  anesthesia machine;  Ensure  that  the  pressure  of  gas  source does not exceed 0.42 MPa and that  the use time covers the entire experiment; ...
  • Page 20 3) Place  the  animal  in  the  induction  chamber  and  turn  the  outlet  switch  of  induction  chamber  “ ”  to  the  right  until “ON”, as shown in the figure  below. In this case, the anesthesia gas will flow to the anesthesia induction  chamber;       Figure 42  Press  the  lock  button  and  rotate  the  vaporizer  dial  to  adjust  anesthetic  concentration  to  an  appropriate  value,  or  rotate  the  dial  directly  to adjust  the concentration back; ...
  • Page 21: Maintenance After The Operation

    6) Take  out  the  animal  from  the  anesthesia  induction  chamber  and  place its  head/nose in the anesthesia mask (detailed operation subject to the type of  anesthesia  mask  used),  and  then  carry  out  subsequent  experimental  operations upon the confirmation of proper anesthesia of animal.    Note:  RWD  recommends  pressing  the  O   flush  button  to  flush  out  the  high  concentration of anesthetic gas in the induction chamber before taking out the animal  from  the  anesthesia  induction chamber, so as to reduce the amount of anesthetic gas  volatilized when the induction chamber is opened. ...
  • Page 22: 5- Trouble Shooting

    5- Trouble Shooting  5.1Safety  Certain hazardous materials may be used during the troubleshooting. Wastes and  contaminants should be disposed in accordance with local laws, regulations, and rules,  along with proper and basic personal protection (for example: gloves, masks, goggles,  etc.).  5.2Machine Status  Unless  otherwise  specified,  the  anesthesia  system  may  be  connected  to  gas  supplies  during  the  performance  of  the  troubleshooting  procedures  described  in this  manual. Make sure the gas supplies and vaporizer are turned off before the procedure.  5.3Record Keeping  A  record  of  problems  and  their  resolution  should  be  kept.  Such records should  include the date, the nature of the problem encountered, and the actions that resolved ...
  • Page 23 Fill the reservoir with appropriate  anesthetic agent.  b. Anesthetic agent   Do not add anesthetic in use, please  reservoir is empty  No or low  place the vaporizer for 15 minutes  1  anesthetic vapor  before use after adding the medicine.  output  Turn the dial on the vaporizer and adjust  c. Vaporizer is turned off  it to the desired volume-percent of  anesthetic.  d. Leak around the  Make sure the sealing cap on the  vaporizer filler port  vaporizer is fully closed.  e. Vaporizer malfunction -  Contact RWD or local dealer for  internal fault   servicing.  Fill vaporizer with the appropriate  anesthetic agent.  a. Vaporizer is empty   Do not add anesthetic in use.  b. Anesthetic concentration  Adjust the dial to increase the volume- Animal sleep  is set too low  percent of Anesthetic vaporizer.  2  level seems too  Check breathing system components for  light  c. Leak in animal breathing  leaks, tears, holes, etc. Ensure the mask  circuit  tube makes a good seal with the animal. ...
  • Page 24 2. Verify that inlet port is not obstructed  by animal hair, etc.  Valves need cleaning or replacement,  O2 flush button  6  The internal valve is faulty  please contact Raywald for after service  is stuck  support.  Float in oxygen  7  Dirty flow tube  Flow tube needs cleaning.  flowmeter sticks  Oxygen flow  Flow valve needs cleaning or  control valve  Dirty or damaged needle  8  replacement. Contact RWD or local  knob is hard to  valve  dealer for servicing.  turn  Oxygen flow  Contact RWD or local dealer for  9  Internal fault  failure  servicing.  a. Loose oxygen hose  Tighten the connection with a crescent  Audible leakage  connection  wrench.  10  around oxygen  hose connector  b. Tube does not fit. ...
  • Page 25: 6- Maintenance

    A record of system service and maintenance should be kept. Such records should  include  service  and  maintenance  dates,  part  numbers  of  any  replaced  parts,  dates  when consumables are replenished, and other pertinent data.  6.5Parts and Materials  Contact  RWD  or  your  local  dealer  if  parts  or  materials  are  needed  during  the  maintenance.  6.6Cleaning the Equipment  Maintenance of the anesthesia system involves possible contact with biological  and  chemical  hazards.  Wear  gloves,  mask,  and  goggles during all cleaning! ...
  • Page 26 6.6.1Required Materials and Supplies  Germicidal  cleaner(Cleaning  solution  with  sterilizing  function,  such as  3% hydrogen peroxide or 75% alcohol);  Water and clean cloth;  Personal Protective Equipment (gloves, masks, goggles, etc).  6.6.2Weekly Maintenance  RWD  recommends  that  the  anesthesia  machine  be  fully  maintained  once  a  week.  Follow these steps:  Wipe all surfaces of the system with a cloth dampened with germicidal  cleaning solution;  Remove  hair,  dust,  and  debris  from  all  vaporizer  surfaces;  especially  around the volume-percent dial and the anesthetic fill port;  Check  whether  the  air  filter  tank  is  overweight.  If  it  is  overweight, ...
  • Page 27 3) Attach  a  drain  tube  to the drain port of anesthetic vaporizer. Place the  other  end  of  the  tube  in  a  receptacle  to  catch  the  drained  anesthetic  agent;    Figure 62  4) Loosen  the  sealing  cap,  then  loosen  the  bottom  screw  with  a  screwdriver to make the anesthetic agent flows naturally;    Figure 63  5) When  the  anesthetic  agent  stops  draining  from  the  vaporizer,  tighten  the bottom screw and sealing cap; ...
  • Page 28: 7- Product Warranty

    The  warranty  of this equipment starts from the date of leaving the factory. During  the  warranty  period,  the  equipment  cannot  be  used  normally  due  to  problems  such as  materials  and  process  defects.  RWD  undertakes  after-sales  service  such  as  equipment  maintenance and parts replacement.  Any  damage  caused  by  improper  use  or  over-range  use  is  not  covered  by  the ...
  • Page 29: 8- Other References

    8- Other References  8.1Pressure Equivalents  1 atm = 1033 cmH O = 760 mmHg = 760 Torr = 1013 mb = 14.7 psi  1 psi = 70.3 cmH O = 51.7 mmHg =68.9mb = 6.9 kPa  1 mmHg = 1.36 cmH O = 1.33mb  1 cmH O = 0.736 mmHg = 0.981 mb  8.2Pressure Unit Conversions  Table 81   Unit  psi  kPa  millibar  mmHg  inchH O  O  psi    27.680  6.8947  68.947  70.308  51.715    0.2491  2.491  2.5400  1.8683  inchH O  3.6127x10   kPa  0.14504  4.0147   ...
  • Page 30                             RWD Life Science Co., Ltd.  Web: www.rwdstco.com  Add: 850 New Burton Road, Suite 201, Dover, DE 19904, Kent, Delaware, USA  Add:  19F,  Building  9A,  Vanke  Cloud  City  III,  Liuxin  4  Street,  Nanshan  District,  Shenzhen518000, Guangdong, P.R. China  Tel: +001-858-900-6602   +86-755-86111286  After-sales Service: +86-755-86111281  After-sales E-mail: service@rwdstco.com     ...