Braun ExactFit 1E Instruction Manual
Braun ExactFit 1E Instruction Manual

Braun ExactFit 1E Instruction Manual

Upper arm blood pressure monitor
Hide thumbs Also See for ExactFit 1E:

Advertisement

Quick Links

1E
ExactFit
BUA4000US

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ExactFit 1E and is the answer not in the manual?

Questions and answers

Summary of Contents for Braun ExactFit 1E

  • Page 1 ExactFit ™ BUA4000US...
  • Page 2 Braun ExactFit ™ BUA4000 Upper arm blood pressure monitor...
  • Page 3: Product Description

    Product description 1 LCD display 2 Power (start/stop) button 3 Hose port 4 Battery compartment 5 Connector 6 Arm cuff 7 Air hose...
  • Page 4: What You Should Know About Blood Pressure

    Intended use of Braun ExactFit™ 1E This device is intended for home use, to intermittently measure systolic and diastolic blood pressure and pulse rate in adults. This device has not been tested for and hence is not intended for neonatal and infant subjects, pregnant, including pre-eclamptic users.
  • Page 5: Inserting Batteries

    The use of this blood pressure monitor is not intended as a substitute for a consultation with your doctor. Avoid compression or restriction of the connection tubing. Continuous cuff pressure due to the connection tubing kinking can affect blood flow and cause possible injury to the user.
  • Page 6: Taking A Measurement

    Artery mark Main arteries Fig. 4 Fig. 5 Fig. 6 4. Pull the cuff so that the top and bottom edges are tightened around your arm. Make sure not to wrap the cuff very tightly. Allow space for one finger to fit between the cuff and the arm.
  • Page 7: Battery Charge Indicator

    Battery charge indicator Batteries discharged—replacements required When the batteries are fully discharged, the empty battery symbol is displayed. You cannot take any further measurement and must replace the batteries. Refer to the section called “Inserting batteries”. Storage and cleaning DO NOT put the item directly under the sunlight, in high temperature or •...
  • Page 8 Error Description Potential cause and remedy «Err254» Device internal error Remove and then reinsert the batteries. If the problem persists, contact customer support. «Err255» Calibration error Remove and then reinsert the batteries. If the problem persists, contact customer support. Blood pressure is a Reminder: always use the same arm fluctuating value.
  • Page 9: Specifications

    Specifications Model: BUA4000US Operation principle: Oscillometric method Display: Liquid crystal display Range of measurement: Rated Cuff pressure : 0 mmHg~295mmHg Measurement pressure : SYS: 60-240mmHg; DIA:40-190mmHg Pulse 40~199 beats/minute Laboratory accuracy: pressure: +/-3 mmHg (cuff pressure) Pulse rate : +/- 5% max Battery life: 200 measurements Auto power off:...
  • Page 10: Warranty

    The warranty DOES NOT cover defects or damage resulting from abuse or failure to follow the userinstructions. The warranty becomes void if the device is opened, tampered with, or used with non-Braun branded parts or accessories, or if repairs are undertaken by unauthorized persons.
  • Page 11 Contact Us Questions or Comments? Call us toll-free at: 1-800-327-7226 Contact us at: www.BraunHealthcare.com/contact-us Or visit our website at: www.BraunHealthcare.com Please be sure to specify the model number. Certain trademarks used under license from The Procter & Gamble Company or its affiliates. ExactFit™...
  • Page 12 PN A009103R1 022924...
  • Page 13 LEGEND INFO DO NOT PRINT THIS PAGE Braun BUA4000US Artwork Part #: A009103R1 Dieline Part #: NA Subject: Owners Manual Region: US Flat Size: MM: W180 x H160 Folded Size: IMM: W90 x H160 Scale: 1/1 Material: Cover: 80# matte text...