The information in this document is subject to change without notice. Chroma ATE INC. makes no warranty of any kind with regard to this manual, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
All Chroma instruments are warranted against defects in material and workmanship for a period of one year after date of shipment. Chroma agrees to repair or replace any assembly or component found to be defective, under normal use during this period. Chroma’s obligation under this warranty is limited solely to repairing any such instrument, which in Chroma’s sole opinion proves to be defective within the scope of the warranty when returned...
Material Contents Declaration The recycling label shown on the product indicates the Hazardous Substances contained in the product as the table listed below. : See <Table 1>. : See <Table 2>. <Table 1> Hazardous Substances Lead Mercury Cadmium Hexavalent Polybrominated Polybromodiphenyl Part Name Chromium...
Page 7
“ ” indicates that the level of the specified chemical substance exceeds the threshold level specified in the standards of SJ/T-11363-2006 and EU 2005/618/EC. Chroma is not fully transitioned to lead-free solder assembly at this moment; however, most of the components used are RoHS compliant.
Failure to comply with these precautions or specific WARNINGS given elsewhere in this manual will violate safety standards of design, manufacture, and intended use of the instrument. Chroma assumes no liability for the customer’s failure to comply with these requirements.
Safety Symbols DANGER – High voltage. Explanation: To avoid injury, death of personnel, or damage to the instrument, the operator must refer to an explanation in the instruction manual. High temperature: This symbol indicates the temperature is now higher than the acceptable range of human. Do not touch it to avoid any personal injury.
Revision History The following lists the additions, deletions and modifications in this manual at each revision. Date Version Revised Sections Oct. 2008 Complete this manual. Mar. 2009 Add the contents about “RS-232 Communication Interface”. Jul. 2013 Add “Material Contents Declaration”, “CE Declaration” and “Table 1-1 Standard Accessory”.
Introduction Introduction An Overview of Product The 11022 LCR Meter is an automatic instrument used for testing and analyzing components. The unit was designed to solve the problems of low labor efficiency and low product quality that have occurred since the electronics sector began to flourish. The testing functions included in this unit are: L, C, R, and Z, which supply the perfect functions on the production line and in quality control.
LCR Meter 11022 User’s Manual • Interfaces : GPIB, Handler and RS-232 interface Incoming Inspection Upon receipt of this instrument, please check the items as following: (1) Any damages or scratches on the surface of the product. (2) Listed in Table 1-1 and Table 1-2 are accessories for this instrument.
Remove all connected wires and cables on the instrument before cleaning. Use a brush to clean the dust on it. For internal cleaning, use a low-pressure air gun to vacuum the dust inside or send it back to the distributors or agents of Chroma for cleaning.
LCR Meter 11022 User’s Manual Measurement Range Parameter Range 0.001uH ~ 99.999kH 0.001pF ~ 1.9999F 0.01mΩ ~ 99.99MΩ | Z | 0.01mΩ ~ 99.99MΩ 0.0001 ~ 9999 0.0001 ~ 9999 θ -180.00° ~ +180.00° Accuracy Within 1 year of factory calibration.
Page 21
Specifications 100mΩ | Z | accuracy (Ze [%]) = A + B × | Z | | Z | 180° θ accuracy (θe) = × Ze π Frequency 50Hz 60Hz 100Hz 120Hz 1kHz 10kHz 20kHz 40kHz 50kHz 100kHz Parameter 0.15 0.14 0.12 0.12...
Page 22
LCR Meter 11022 User’s Manual 3. D, Q Accuracy For all D value tanθe × (1 + D 2 ) Accuracy of D = ± 1 – D × tanθe For Q ≥ 10 tanθe × (1 + Q 2 ) Accuracy of Q = ±...
LCR Meter 11022 User’s Manual Zero Zero open: Eliminate measurement errors due to parasitic stray impedances of the test fixture. Zero short: Eliminate measurement errors due to parasitic residual impedances of the test fixture. Measurement Time Start from measurement, analog sampling calculation to Binning or Compare signal output measuring time.
Installation Installation Ambient Environment (1) Do not use the meter in a dusty or vibrating Keep from following cases: location. Do not expose it to sunlight or Direct sunlight corrosive gas. Be sure that the ambient temperature is 10 ~ 40°C and that the Dust relative humidity is below 90%.
LCR Meter 11022 User’s Manual Fuse There is one fuse installed in the rear panel. When replacing the fuse, please turn off the power and pull the plug from the power supply. Fuse spec. AC 100V ~ 120V → T1.0A 250V AC 220V ~ 240V →...
Page 28
LCR Meter 11022 User’s Manual read. For example, test voltage, parameter, speed, series or parallel…etc. (6) Main index key Pressing this key allows entry to the main index screen. In this screen you may select what you want to test, for example, the analysis of unknown test result, open test, short test, compare function…etc.
Panel Description Rear Panel (1) AC Line Socket This socket is an approved (by the International Electrotechnial Commission) three-line socket 320. Please use the correct wire, such as Belden SPH-386 or similar (W12 010130). (2) Fuse 1.0A or 0.5A slow flow fuse to prevent the instrument being affected by excess current in 90 ~ 125V or 190 ~ 250V.
Page 30
LCR Meter 11022 User’s Manual panel. • Notice the marked polarity on the panel. (7) Guard Terminal This terminal connects the cover of the instrument directly, then connect to the ground to avoid exterior obstruction that makes instrument connection incomplete and can compromise safety.
Page 32
LCR Meter 11022 User’s Manual < SYSTEM SETUP > CALIBRATION To start the system calibration MEM MANAGE To start the memory management To start the system parameter setting SYSTEM CONFIG screen 4. Press [F3] (System Configuration) key to enter system parameter setting screen as follows: <SYSTEM CONFIG>...
Page 33
Panel Description Note C.C: Constant Current AVERAGE TIME: Average time selection range is 1 ∼ 256. The factory default value is set as “1.” CABLE LENGTH: There are 4 kinds of cable length available 0M, 1M, 2M, and 4M. Default is 0M. After changing the cable length please run the open/short calibration CAUTION to get a more accurate results.
LCR Meter 11022 User’s Manual Select GPIB interface position. The factory default value is set as “17”, with a range of 00 ∼ 30. TRIGGER DELAY: When using with auto machinery, external trigger EXT timing will need to be adjusted because measuring speed and auto machinery ON speed must match.
Panel Description Operation Instruction 4.4.1 Open Correction 1. After turning the instrument on and ensuring every event is OK, press [Main Index] to enter main index as following. < MAIN INDEX > Binning setting BINNING Compare setting COMPARE Zeroing correction. CORRECTION To the second page NEXT PAGE 1/2...
LCR Meter 11022 User’s Manual [PASS]. Press any key to escape to another index as following. < MAIN INDEX-CORRECT OPEN > SINGLE Single frequency FREQ: 100KHz RANGE: 4 CALIB: 2 Multi frequency MULTI 0.000pF MEASURE PASS… 100% PRESS ANY KEY TO ESCAPE...
Panel Description The test cable is connected as the above figure shows, connect H to L Press Trig key to process short test. The display shows FAIL as short test error, it indicates short test action with problem. Check if the test cable is broken or the clip with bad contact, re-execute short test after inspecting.
Page 38
LCR Meter 11022 User’s Manual < MEAS. DISPLAY > Control bias source ON/OFF BIAS I : OFF Cp : 0.02 pF Control bias source value setting BIAS I 000.000A D : 0.0001 Enter to the first page NEXT PAGE 4/4 2.
Page 39
Please see 4.3.1 System Setup. BIAS I function can be used only when make a connection of 50 Pin Handler interface and 1320 Bias Current Source 0-10A or 1320 Bias Current Source 0-20A by using Chroma 1320 control line in Table 1-2 Optional Accessory. The connection is shown as below. To avoid control signal interferes each other, please be careful that 24 Pin Handler connector (M) and 9 Pin D-SUB connector (F) of Chroma 1320 control line can’t be connected...
LCR Meter 11022 User’s Manual 11022 LCR METER BIAS I 000.000A: It is the function for setting the value of 1320 Bias Current Source 0-10A or 1320 Bias Current Source 0-20A. Press [Δ], [∇], [<] and [>] to adjust the current. The factory default value is 000.000A. Please notice that the maximum current value only can be set to 10A as connecting a set of 1320 Bias Current Source 0-10A.
Page 41
Panel Description < MAIN INDEX-BINNING > BIN SET Bin setting. PARAMETER : Cp – D STORE Store value setting NOMINAL : 00000.0000pF AUX_HI : +99999.9999- RECALL Recall value setting AUX_LO : +00000.0000- BIN COUNT Bin test count Press [F1] then enter Bin Set screen as follows: <...
Page 42
LCR Meter 11022 User’s Manual BIN COUNT BIN COUNT SPEED : M Measurement speed TRIG. : INT Trigger mode Range Auto/Hold setting RANGE: AUTO 1736 1736 RESET Clear count value TOTAL: Press [MEAS DISPLAY] key to enter main measurement screen. Then press [F4] to enter the third page for BINNING function enabling, the screen will show the current binning result as below.
Panel Description 4.4.5 Description of Compare Setting Operation After turning on main unit power display and ensuring all is normal, press [Main Index] key to enter main function list as the following: < MAIN INDEX > Binning test setting. BINNING Compare test setting.
LCR Meter 11022 User’s Manual < MEAS. DISPLAY – ΔABS> BIAS V: OFF External DCV setting Cp : 0.0001nF BINNING: OFF Binning setting : 0.0001 Compare setting COMPARE: ON-Δ : HI : HI NEXT PAGE 3/4 To the 4 page Press [F3], meanwhile compare setting function will show difference value percentage of primary parameter and COMPARE nominal value.
Page 45
To avoid control signal interferes each other, please be careful that 24 Pin Handler connector (M) and 9 Pin D-SUB connector (F) of Chroma 1320 control line can’t be connected simultaneously. Meanwhile, 24 Pin Handler connector (M) should be connected and power off the two units and then perform connection.
Page 46
LCR Meter 11022 User’s Manual <MAIN INDEX-BIAS COMP. TEST > SPEED: M Measurement speed setting LsO : 0.6450mH TRIG.: EXT Trigger mode setting 87.6 FAIL LsB : 0.6040mH R1:A 10Ω The primary measurement range setting 86.5 R2: A 10Ω The secondary measurement range setting...
Page 47
The connection is shown as below. To avoid control signal interferes each other, please be careful that 24 Pin Handler connector (M) and 9 Pin D-SUB connector (F) of Chroma 1320 control line can’t be connected simultaneously. Meanwhile, 9 Pin D-SUB connector (F) should be connected and power off the two units and then perform connection.
Page 48
LCR Meter 11022 User’s Manual <MAIN INDEX-BIAS COMP. SET > DIGIT UP The digit cursor on for adjusting upward Ls O_NOMINAL: 600.000uH DIGIT DOWN The digit cursor on for adjusting downward Hi: 630.000uH Lo: 570.000uH QO_Lo_LIMIT: 070.000 DIGIT Move the cursor to right for one digit Ls B_Hi : 550.000uH...
Page 49
Panel Description <SYSTEM CONFIG> CONTINUE Test continuously BIAS C. MODE : MASTER TEST FAIL STOP STOP Fail and stop test BAUDRATE : NONE <MAIN INDEX-BIAS COMP. TEST > SPEED: M Measurement speed setting LsO : 0.6450mH TRIG.: MAN Trigger mode setting 87.6 LsB : R1: A...
5. The 24 Pin Handler connector (M) is able to input control signal or output judgment signal as using MASTER MODE with Chroma 1320 control line for performing BIAS COMPARE measurement. It is for the use of automatic equipment, the pins are shown below.
Panel Description Enable LOAD function < MAIN INDEX-CORRECT LOAD> LOAD: OFF Disable LOAD function FREQUENCY: 1.0KHz PARAMETER: Rs-Xs MEAS. LOAD Measure parameter actual value REF. A VAL: +00000.0000-Ω Ω MEAS A VAL: REF. B VAL: +00000.0000-Ω Ω MEAS B VAL: Set the frequency and parameter requires along with parameter reference value desire to set then turn LOAD function on.
Page 52
LCR Meter 11022 User’s Manual BIN SET Bin setting < MAIN INDEX-BINNING > PARAMETER: Cp – D STORE Store setting value NOMINAL : 00000.0000pF RECALL Recall setting value AUX_HI : +99999.9999- AUX_LO : +00000.0000 Press [F1] then enter Bin Setting screen is as below: The digit which cursor on for adjusting upward.
Panel Description 4.4.9 Dual Frequency Function (Option) 1. Dual frequency function is mainly operating DUT capacitor parameter of multi-bandwidth. For example, Multi-layer Ceramic Capacitor (MLCC) applies to integrated circuit and filtering circuit its operation bandwidth is extra widely. The capacitance value should be measured in lower operation frequency, whereas ESR value should be noticed in higher operation frequency.
Page 54
LCR Meter 11022 User’s Manual <SYSTEM CONFIG> 25 OHM 25 OHM mode VM/IM DISPLAY : OFF CONST RSOURCE : 100/25 OHM AVERAGE TIME : 001 CABLE LENGTH : 0M BEEPER : LOW KEY LOCK : OFF Next. P 2/2 To the 1 page The calculation of trigger delay time is decided by DUT capacity.
Page 55
Panel Description < MAIN INDEX-DUAL FREQ. > SETTING Dual frequency function setting STORE Store setting 1ST PARA: Cp-D FREQ: 1.0KHz NOMINAL: +00000.0000-F RECALL Recall setting 2ND PARA : D FREQ: 1.0KHz TEST Dual frequency function test Press [F1] to go to modification setting screen is shown as below. <MAIN INDEX-DUAL FREQ.
Description of GPIB/RS-232 Interface Description of GPIB/RS-232 Interface Overview Using the IEEE-488.2 interface, the users can operate the 11022 by remote control, transfer data and etc. The RS-232 interface is for option. This chapter is mainly for providing the description of the 11022 interface Bus (GPIB) or RS-232 command for the users’ writing program to control the 11022 in order to do test data process.
Description of GPIB/RS-232 Interface 5.2.6 Response of Interface Messages • The meter is capable of responding to the following messages. Interface Message Response ‧ Only addressed devices that receive this command are set to local mode. GTL (Go To Local) ‧...
Description of GPIB/RS-232 Interface Command Description 5.4.1 Command Configuration Command is with tree configuration level. All of 11022 related commands can see totally from the following figure. Root ABORt CALCulate FORMat LIMit BEEPer CONDition [STATe] LOWer [DATA] STATe UPPer [DATA] MATH EXPRession CATalog NAME...
Description of GPIB/RS-232 Interface :CALCulate:MATH:EXPRession:NAME Then can get NAME command correctly. Moreover, if send (Example: NAME and CATalog) two commands simultaneously. Be able to use the following method make the instruction more easily and clearly. :CALCulate : MATH: EXPRession:NAME;CATalog Its’ meaning is the same as the following two instructions. Isn’t it more briefly? :CALCulate:MATH:EXPRession:NAME :CALCulate:MATH:EXPRession:CATalog The commands need using colon ( : ) to separate between level and level.
LCR Meter 11022 User’s Manual (3) Before the first parameter in every command need a blank character. To end character To end character has three kinds: [CARRIAGE RETURN] (0Dh), [NEW LINE] (0Ah) and [CARRIAGE RETURN] (0Dh) + [NEW LINE] (0Ah) 5.4.4 Common Command...
Description of GPIB/RS-232 Interface Main Board Test Key Control (15) *STB It queries the contents of the Status Bit Register. 5.4.5 Command Description ABORt Command (1) Command: ABORt Parameter: None Return: None Function: It terminates processing trigger system immediately, that is say to reset trigger system.
Page 66
LCR Meter 11022 User’s Manual (7) Command: BINning: NO: STATe Parameter: {ON(1) | OFF(0)} Return: {1 | 0} Function: It sets if BIN NO>9 function in SYSTEM CONFIG is enabled. (8) Command: BINning: NO Parameter: {1~99} Return: {1~99} Function: It sets BIN NO when BIN NO>9 function in SYSTEM CONFIG is enabled.
Page 67
Description of GPIB/RS-232 Interface (2) Command: CALCulate2:FORMat Parameter: {IMAGinary|PHASe|D|Q|REAL|RS|XS|LP|INV} Return: {PHASe|D|Q|REAL|MLINear|LP|INV} Function: It sets or queries the current measurement secondary parameter. Description: IMAGinary Impedance imaginary PHASe Phase Dissipation factor Quality factor REAL Impedance real Equivalent series resistance Imaginary series equivalent value MLINear Absolute value of impedance Parameter...
Page 68
LCR Meter 11022 User’s Manual Description: ON(1) Turn on comparison function OFF(0) Turn off comparison function (7) Command: CALCulate{1|2}:LIMit:UPPer[:DATA] Parameter: {The high limit value | MAXimum | MINimum Return: The high limit value, the format is<NR3> Function: It sets or queries the high limit value.
Page 69
Description of GPIB/RS-232 Interface Parameter: {Primary, secondary parameter the nominal value in COMPARE (reference value) (NR3)} Return: None Function: It sets primary, secondary parameter the nominal value in COMPARE mode. If the nominal value of primary parameter set on 100, command is DATA [: DATA] REF1, 100;...
Page 70
LCR Meter 11022 User’s Manual Equivalent series capacitance Equivalent parallel inductance Equivalent series inductance Equivalent series impedance Equivalent series resistance Equivalent parallel resistance (2) Command: DUAL:FIRSt:CALCulate2:FORMat Parameter: {IMAGinary|PHASe|D|Q|REAL|RS|XS} Return: {IMAGinary|PHASe|D|Q|REAL|RS|XS} Function: It sets or queries the 1 measurement secondary parameter in dual frequency function.
Page 71
Description of GPIB/RS-232 Interface Function: The 2 measurement setting or query in dual frequency function if auto range selection mode is enabled. (7) Command: DUAL: SECOnd [:SENSe]:FIMPedance:RANGe[:UPPer] Parameter: {Measurement range value |UP|DOWN} Unit: [MOHM | kOHM | OHM | MAOHM], if the unit isn’t given then it will be defined as OHM.
Page 72
LCR Meter 11022 User’s Manual function, i.e. the high limit value of measurement secondary parameter. (14) Command: DUAL:FIRSt:BINning:LOWer:AUX Parameter: {Low limit (NR3) of the 1st measurement secondary parameter.} Return: {Low limit (NR3) of the 1st measurement secondary parameter. Function: The AUX-LO value of 1st measurement setting Binning in dual frequency function, i.e.
Page 73
Description of GPIB/RS-232 Interface NOGO (19) Command: DUAL: TRIGger:SOURce Parameter: {BUS|EXTernal|INTernal|MANual} Return: {BUS|EXTernal|INTernal|MANual} Function: It sets or queries trigger mode currently in dual frequency function. Description: BUS Bus trigger EXTernal External trigger INTernal Internal trigger MANual Manual trigger (20) Command: DUAL [:SENSe]:FIMPedance:APERture Parameter: 0.025 (Fast), 0.065 (Medium), 0.500 (Slow) Return: 0.025 (Fast), 0.065 (Medium), 0.500 (Slow) Function: It sets or queries measurement speed currently in dual frequency function.
Page 74
LCR Meter 11022 User’s Manual FORMat Command (1) Command: FORMat[:DATA] Parameter: {ASCii} Return: ASK Function: It sets or queries the data format of transmission numeral. INITiate Command Set (1) Command: INITiate[:IMMediate] Parameter: None Return: None Function: All sequences escape from idle status and enter initial status.
Page 75
Description of GPIB/RS-232 Interface [:SENSe]:CORRection:DATA:SHORt:RANGe:1OHM? Parameter: None Return: SHORT correction value of RANGE 1Ω, the format is < NR3 > Function: It sets or queries SHORT correction value of RANGE 1Ω, that is R, X (7) Command: [:SENSe]:CORRection:DATA:SHORt:RANGe:100MOHM [:SENSe]:CORRection:DATA:SHORt:RANGe:100MOHM? Parameter: None Return: SHORT correction value of RANGE 100mΩ, the format is <...
Page 76
LCR Meter 11022 User’s Manual (3) Command: SOURce:BIAS:CURRent:CONDition Parameter: {ON(1)|OFF(0)} Return: {1|0} Function: It returns and sets the output condition of bias current source. Description: ON(1) Enable bias current source output. OFF(0) Disable bias current source output. (4) Command: SOURce:VOLTage[:LEVel][:IMMediate] :OFFSet:STATe...
Page 77
Description of GPIB/RS-232 Interface (3) Command: SYSTem:ERRor? Parameter: None Return: It returns error message in queue, the fixed return is 0. (4) Command: SYSTem:KLOCk Parameter: {ON(1)|OFF(0)} Return: {1|0} Function: It sets or queries whether 11022 key is locked or not. (5) Command: SYSTem:LOCal Parameter: None Return: None...
LCR Meter 11022 User’s Manual Function: It sets LINK1320 ON/OFF. TRIGger Command Set (1) Command: TRIGger:DELay Parameter: Trigger delay time Unit: [MS|S] if there is no command unit the auto define S. Return: Trigger delay time, the format is <NR3>.
Description of GPIB/RS-232 Interface (2) When users give TRIGger command, the sequential other setting commands (non FETCh ? commands) will be given. The program needs to write the function of time delay. This delay time is the measurement time of 11022, otherwise the part of sequential commands can’t normal function.
LCR Meter 11022 User’s Manual Status Reporting Structure Service Request Figure 5-1 Status Reporting Structure The 11022 LCR Meter can send an SRQ (Service Request) control signal when it requires the controller to perform a task. When the 11022 LCR Meter generates an SRQ, it also sets Bit 6 of the status Byte Register, RQS (Request Service) bit.
Description of GPIB/RS-232 Interface Status Byte Register The Status Byte Register is composed of eight bits that summarize an overlaying status data structure. The Status Byte Register can be read using either *STB? Or SPOLL, which return a decimal expression of the contents of the register (equal to the sum of the total bit weights of all the bits set to “1”).
LCR Meter 11022 User’s Manual Standard Event Status Register The Standard Event Status Register is frequently used and is one of the simple. Standard Event Status Register Read by *ESR? Standard Event Status Enable Register Summary Message Event Set by *ESE...
Description of GPIB/RS-232 Interface Bit No Bit Weight Description Power on Bit – This bit is set when the 11022 LCR Meter has been turned OFF and then ON since the last time this register was read. Always zero (0) Command Error Bit- This bit is set if the following command errors occur.
Description of Handler Interface Description of Handler Interface Binning, Compare and Bias Compare in 11022 as well as LINK 1320(1320 Bias Current Source 0-10A and 1320 Bias current Source 0-20A) are all connected to external unit by Handler interface. There are connectors 50 Pin (Standard) and 24 Pin (Optional) in Binning, Compare and Bias Compare, LINK 1320 only has 50 Pin connector.
LCR Meter 11022 User’s Manual 6.1.2 24 Pin Name Description /EXT External trigger 3.20 BIN 7 BIN 7, primary parameter test value within BIN 7 sets range 4.24 BIN 8 BIN 8, primary parameter test value within BIN 8 sets range Ground 9.13...
Description of Handler Interface EOT End of Test 45-46 COM COMMON 47-48 49-50 GND Connect with cover 6.2.2 24 Pin Name Description /EXT External trigger 3.20 A LO Primary parameter test value too low 4.24 A HI Primary parameter test value too high GND Ground 9.13 A NG Primary parameter test value not in specification...
LCR Meter 11022 User’s Manual Description for Bias Compare Handler Interface Pin 6.3.1 50 Pin Name Description VEXT External DC voltage acceptable voltage range is 5V ~ 24V. 3-18 /EXT External trigger 20-21 VEXT External DC voltage acceptable voltage range is 5V ~ 24V.
Description of Handler Interface Description for Link 1320 Handler Interface Pin 6.4.1 50 Pin Name Description External DC voltage acceptable voltage range is 5V ∼ 24V. VEXT IBCLK COMMON IBDATA XB20ARDY IBSTR IBIRDY COMMON XBC1320 VINT Internal DC voltage +5V 14-19 External DC voltage acceptable voltage range is 5V ∼...
LCR Meter 11022 User’s Manual The secondary parameter will over high/low limit setting value if BIN number is judged as “0”. Therefore, the binary string data is “00000000” outputted by Handler interface. The primary parameter test value isn’t in all setting specifications if BIN result is judged as “OUT”. Thus, the binary string data is “01100100”...
Description of Handler Interface 6.6.2 24 Pin Signal Name Description /EXT External trigger The secondary parameter test value with second frequency is out 3, 20 SFQ NOGO of setting range The secondary parameter test value with second frequency is in 4, 24 SFQ GO setting range...
Page 92
LCR Meter 11022 User’s Manual /EXT Previous BIN value Bin value currently Measurement FAST MEDIUM SLOW Speed T1 ≥ 1.25μs 1.25μs 1.25μs Measurement T2 < Time T3 < 21ms 51ms 360ms Set Trigger Edge to Rising state in system parameter operation then set INTEG. CYCLE to a cycle.
Description of Handler Interface Measurement Speed FAST MEDIUM SLOW T1 ≥ 1.25μs 1.25μs 1.25μs Measurement T2 < Time T3 < 21ms 51ms 360ms 6.7.2 Setting with Trigger Delay Time (Freq. ≥ 100Hz) Set Trigger Edge to Falling state in system parameter operation then set INTEG. CYCLE to a cycle.
Page 94
LCR Meter 11022 User’s Manual /EXT Previous BIN value BIN value currently Measurement FAST MEDIUM SLOW Speed T1 ≥ 1.25μs 1.25μs 1.25μs Measurement T2 < Time T3 < (21+T ) ms (51+T ) ms (360+T ) ms DELAY DELAY DELAY T1 is /EXT external trigger signal, this external trigger time should be more than 1.25μs to...
Optional Accessory Appendix A Optional Accessory A110211 Component Test Fixture This fixture is specially designed for testing DIP passive component impedance. It adopts 4-terminal measured method which can test resistance range extensively. DIP passive component suitable maximum pitch is evaluated as 73mm and minimum pitch is 5mm.
Page 96
10mm. The rest are evaluated as more suitable dimensions are 4mm, 5mm, 6.3mm and 8mm respectively. A110236 Rack Mountain Kit This kit is specially designed for Chroma LCR 11022 fixes on large-sized system rack conveniently such as Chroma 8800 ATS system.
Page 97
Optional Accessory A110244 High Capacitance Capacitor Test Fixture This fixture is specially designed for testing high capacitance capacitor. High capacitance capacitor junction is M5 screw hole and outer diameter over 6mm. The DUT is with very low impedance when clip the junction by using test clamp isn’t easy to be fixed.
Page 99
Kuei-shan Hwaya Technology Park Taoyuan County 33383, Taiwan 33383 台灣桃園縣龜山鄉 華亞科技園區華亞一路 66 號 T +886-3-327-9999 F +886-3-327-8898 Mail: info@chromaate.com http://www.chromaate.com Copyright by CHROMA ATE INC. All Rights Reserved. All other trade names referenced are the properties of their respective companies.