LCR Meter
11022
User's Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LCR Meter 11022 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Chroma LCR Meter 11022

  • Page 1 LCR Meter 11022 User’s Manual...
  • Page 3 LCR Meter 11022 User’s Manual Version 1.2 July 2013...
  • Page 4: Legal Notices

    The information in this document is subject to change without notice. Chroma ATE INC. makes no warranty of any kind with regard to this manual, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 5: Warranty

    All Chroma instruments are warranted against defects in material and workmanship for a period of one year after date of shipment. Chroma agrees to repair or replace any assembly or component found to be defective, under normal use during this period. Chroma’s obligation under this warranty is limited solely to repairing any such instrument, which in Chroma’s sole opinion proves to be defective within the scope of the warranty when returned...
  • Page 6: Material Contents Declaration

    Material Contents Declaration The recycling label shown on the product indicates the Hazardous Substances contained in the product as the table listed below. : See <Table 1>. : See <Table 2>. <Table 1> Hazardous Substances Lead Mercury Cadmium Hexavalent Polybrominated Polybromodiphenyl Part Name Chromium...
  • Page 7 “ ” indicates that the level of the specified chemical substance exceeds the threshold level specified in the standards of SJ/T-11363-2006 and EU 2005/618/EC. Chroma is not fully transitioned to lead-free solder assembly at this moment; however, most of the components used are RoHS compliant.
  • Page 9: Safety Summary

    Failure to comply with these precautions or specific WARNINGS given elsewhere in this manual will violate safety standards of design, manufacture, and intended use of the instrument. Chroma assumes no liability for the customer’s failure to comply with these requirements.
  • Page 10: Safety Symbols

    Safety Symbols DANGER – High voltage. Explanation: To avoid injury, death of personnel, or damage to the instrument, the operator must refer to an explanation in the instruction manual. High temperature: This symbol indicates the temperature is now higher than the acceptable range of human. Do not touch it to avoid any personal injury.
  • Page 11: Revision History

    Revision History The following lists the additions, deletions and modifications in this manual at each revision. Date Version Revised Sections Oct. 2008 Complete this manual. Mar. 2009 Add the contents about “RS-232 Communication Interface”. Jul. 2013 Add “Material Contents Declaration”, “CE Declaration” and “Table 1-1 Standard Accessory”.
  • Page 13: Table Of Contents

    LCR Meter 11022 User’s Manual Table of Contents Introduction........................1-1 An Overview of Product ...................1-1 Brief Specifications ..................1-1 Incoming Inspection..................1-2 Initial Inspection ....................1-3 Common Environment Conditions ..............1-3 Maintenance & Cleaning..................1-3 Specifications (15°C~35°C, RH ≤ 75%) ..............2-1 Measurement Functions ..................2-1 Test Signals .....................2-1 Measurement Range ..................2-2...
  • Page 14 LCR Meter 11022 User’s Manual 5.3.2 RS-232 Interface Connector ..............5-4 5.3.3 RS-232 Signal Cable and Pin Table............5-4 Command Description ..................5-5 5.4.1 Command Configuration ................5-5 5.4.2 Description for Command Configuration..........5-6 5.4.3 Command Syntax ..................5-7 5.4.4 Common Command................5-8 5.4.5 Command Description ................5-9 5.4.6...
  • Page 15: Introduction

    Introduction Introduction An Overview of Product The 11022 LCR Meter is an automatic instrument used for testing and analyzing components. The unit was designed to solve the problems of low labor efficiency and low product quality that have occurred since the electronics sector began to flourish. The testing functions included in this unit are: L, C, R, and Z, which supply the perfect functions on the production line and in quality control.
  • Page 16: Incoming Inspection

    LCR Meter 11022 User’s Manual • Interfaces : GPIB, Handler and RS-232 interface Incoming Inspection Upon receipt of this instrument, please check the items as following: (1) Any damages or scratches on the surface of the product. (2) Listed in Table 1-1 and Table 1-2 are accessories for this instrument.
  • Page 17: Initial Inspection

    Remove all connected wires and cables on the instrument before cleaning. Use a brush to clean the dust on it. For internal cleaning, use a low-pressure air gun to vacuum the dust inside or send it back to the distributors or agents of Chroma for cleaning.
  • Page 19: Specifications (15°C~35°C, Rh ≤ 75%)

    Specifications Specifications (15°C~35°C, RH ≤ 75%) Measurement Functions Measurement Parameters: Primary Parameter Inductance unit: uH, mH, H, kH Capacitance unit: pF, nF, uF, mF Resistance unit: mΩ, Ω, kΩ, MΩ ⎢Z ⎢ Absolute value of impedance unit: mΩ, Ω, kΩ, MΩ Secondary Parameter Quality factor Dissipation factor...
  • Page 20: Measurement Range

    LCR Meter 11022 User’s Manual Measurement Range Parameter Range 0.001uH ~ 99.999kH 0.001pF ~ 1.9999F 0.01mΩ ~ 99.99MΩ | Z | 0.01mΩ ~ 99.99MΩ 0.0001 ~ 9999 0.0001 ~ 9999 θ -180.00° ~ +180.00° Accuracy Within 1 year of factory calibration.
  • Page 21 Specifications 100mΩ | Z | accuracy (Ze [%]) = A + B × | Z | | Z | 180° θ accuracy (θe) = × Ze π Frequency 50Hz 60Hz 100Hz 120Hz 1kHz 10kHz 20kHz 40kHz 50kHz 100kHz Parameter 0.15 0.14 0.12 0.12...
  • Page 22 LCR Meter 11022 User’s Manual 3. D, Q Accuracy For all D value tanθe × (1 + D 2 ) Accuracy of D = ± 1 – D × tanθe For Q ≥ 10 tanθe × (1 + Q 2 ) Accuracy of Q = ±...
  • Page 23 Specifications Figure2-1 LC ⎯ |Z| conversion chart conversion chart conversion chart...
  • Page 24: Zero

    LCR Meter 11022 User’s Manual Zero Zero open: Eliminate measurement errors due to parasitic stray impedances of the test fixture. Zero short: Eliminate measurement errors due to parasitic residual impedances of the test fixture. Measurement Time Start from measurement, analog sampling calculation to Binning or Compare signal output measuring time.
  • Page 25: Installation

    Installation Installation Ambient Environment (1) Do not use the meter in a dusty or vibrating Keep from following cases: location. Do not expose it to sunlight or Direct sunlight corrosive gas. Be sure that the ambient temperature is 10 ~ 40°C and that the Dust relative humidity is below 90%.
  • Page 26: Fuse

    LCR Meter 11022 User’s Manual Fuse There is one fuse installed in the rear panel. When replacing the fuse, please turn off the power and pull the plug from the power supply. Fuse spec. AC 100V ~ 120V → T1.0A 250V AC 220V ~ 240V →...
  • Page 27: Panel Description

    Panel Description Panel Description Front Panel COPYRIGHT © JAN. 2002 VER: X.XX DATE: XX/XX/XXXX CPLD: 11022 X.XX (1) LCD Display The resolution of this instrument display is 64x240 Graphic mode LCD, so all the measurements and setting values can be shown clearly. (2) POWER switch On-off switch (3) Unknown measurement connectors...
  • Page 28 LCR Meter 11022 User’s Manual read. For example, test voltage, parameter, speed, series or parallel…etc. (6) Main index key Pressing this key allows entry to the main index screen. In this screen you may select what you want to test, for example, the analysis of unknown test result, open test, short test, compare function…etc.
  • Page 29: Rear Panel

    Panel Description Rear Panel (1) AC Line Socket This socket is an approved (by the International Electrotechnial Commission) three-line socket 320. Please use the correct wire, such as Belden SPH-386 or similar (W12 010130). (2) Fuse 1.0A or 0.5A slow flow fuse to prevent the instrument being affected by excess current in 90 ~ 125V or 190 ~ 250V.
  • Page 30 LCR Meter 11022 User’s Manual panel. • Notice the marked polarity on the panel. (7) Guard Terminal This terminal connects the cover of the instrument directly, then connect to the ground to avoid exterior obstruction that makes instrument connection incomplete and can compromise safety.
  • Page 31: Description Of System Parameters Setting

    The socket is for connecting RS-232 interface. The interface of GPIB and RS-232 can’t be used simultaneously. Description of System Parameters Setting 4.3.1 System Setup 1. The display shows Chroma’s name, model number and programming version after the main power is switched on. Chroma ATE INC. 11022 LCR Meter COPYRIGHT ©...
  • Page 32 LCR Meter 11022 User’s Manual < SYSTEM SETUP > CALIBRATION To start the system calibration MEM MANAGE To start the memory management To start the system parameter setting SYSTEM CONFIG screen 4. Press [F3] (System Configuration) key to enter system parameter setting screen as follows: <SYSTEM CONFIG>...
  • Page 33 Panel Description Note C.C: Constant Current AVERAGE TIME: Average time selection range is 1 ∼ 256. The factory default value is set as “1.” CABLE LENGTH: There are 4 kinds of cable length available 0M, 1M, 2M, and 4M. Default is 0M. After changing the cable length please run the open/short calibration CAUTION to get a more accurate results.
  • Page 34: Memory Management Instruction

    LCR Meter 11022 User’s Manual Select GPIB interface position. The factory default value is set as “17”, with a range of 00 ∼ 30. TRIGGER DELAY: When using with auto machinery, external trigger EXT timing will need to be adjusted because measuring speed and auto machinery ON speed must match.
  • Page 35: Operation Instruction

    Panel Description Operation Instruction 4.4.1 Open Correction 1. After turning the instrument on and ensuring every event is OK, press [Main Index] to enter main index as following. < MAIN INDEX > Binning setting BINNING Compare setting COMPARE Zeroing correction. CORRECTION To the second page NEXT PAGE 1/2...
  • Page 36: Short Correction

    LCR Meter 11022 User’s Manual [PASS]. Press any key to escape to another index as following. < MAIN INDEX-CORRECT OPEN > SINGLE Single frequency FREQ: 100KHz RANGE: 4 CALIB: 2 Multi frequency MULTI 0.000pF MEASURE PASS… 100% PRESS ANY KEY TO ESCAPE...
  • Page 37: Setup Of Lcrz Parameter

    Panel Description The test cable is connected as the above figure shows, connect H to L Press Trig key to process short test. The display shows FAIL as short test error, it indicates short test action with problem. Check if the test cable is broken or the clip with bad contact, re-execute short test after inspecting.
  • Page 38 LCR Meter 11022 User’s Manual < MEAS. DISPLAY > Control bias source ON/OFF BIAS I : OFF Cp : 0.02 pF Control bias source value setting BIAS I 000.000A D : 0.0001 Enter to the first page NEXT PAGE 4/4 2.
  • Page 39 Please see 4.3.1 System Setup. BIAS I function can be used only when make a connection of 50 Pin Handler interface and 1320 Bias Current Source 0-10A or 1320 Bias Current Source 0-20A by using Chroma 1320 control line in Table 1-2 Optional Accessory. The connection is shown as below. To avoid control signal interferes each other, please be careful that 24 Pin Handler connector (M) and 9 Pin D-SUB connector (F) of Chroma 1320 control line can’t be connected...
  • Page 40: Binning Set

    LCR Meter 11022 User’s Manual 11022 LCR METER BIAS I 000.000A: It is the function for setting the value of 1320 Bias Current Source 0-10A or 1320 Bias Current Source 0-20A. Press [Δ], [∇], [<] and [>] to adjust the current. The factory default value is 000.000A. Please notice that the maximum current value only can be set to 10A as connecting a set of 1320 Bias Current Source 0-10A.
  • Page 41 Panel Description < MAIN INDEX-BINNING > BIN SET Bin setting. PARAMETER : Cp – D STORE Store value setting NOMINAL : 00000.0000pF AUX_HI : +99999.9999- RECALL Recall value setting AUX_LO : +00000.0000- BIN COUNT Bin test count Press [F1] then enter Bin Set screen as follows: <...
  • Page 42 LCR Meter 11022 User’s Manual BIN COUNT BIN COUNT SPEED : M Measurement speed TRIG. : INT Trigger mode Range Auto/Hold setting RANGE: AUTO 1736 1736 RESET Clear count value TOTAL: Press [MEAS DISPLAY] key to enter main measurement screen. Then press [F4] to enter the third page for BINNING function enabling, the screen will show the current binning result as below.
  • Page 43: Description Of Compare Setting Operation

    Panel Description 4.4.5 Description of Compare Setting Operation After turning on main unit power display and ensuring all is normal, press [Main Index] key to enter main function list as the following: < MAIN INDEX > Binning test setting. BINNING Compare test setting.
  • Page 44: Bias Compare

    LCR Meter 11022 User’s Manual < MEAS. DISPLAY – ΔABS> BIAS V: OFF External DCV setting Cp : 0.0001nF BINNING: OFF Binning setting : 0.0001 Compare setting COMPARE: ON-Δ : HI : HI NEXT PAGE 3/4 To the 4 page Press [F3], meanwhile compare setting function will show difference value percentage of primary parameter and COMPARE nominal value.
  • Page 45 To avoid control signal interferes each other, please be careful that 24 Pin Handler connector (M) and 9 Pin D-SUB connector (F) of Chroma 1320 control line can’t be connected simultaneously. Meanwhile, 24 Pin Handler connector (M) should be connected and power off the two units and then perform connection.
  • Page 46 LCR Meter 11022 User’s Manual <MAIN INDEX-BIAS COMP. TEST > SPEED: M Measurement speed setting LsO : 0.6450mH TRIG.: EXT Trigger mode setting 87.6 FAIL LsB : 0.6040mH R1:A 10Ω The primary measurement range setting 86.5 R2: A 10Ω The secondary measurement range setting...
  • Page 47 The connection is shown as below. To avoid control signal interferes each other, please be careful that 24 Pin Handler connector (M) and 9 Pin D-SUB connector (F) of Chroma 1320 control line can’t be connected simultaneously. Meanwhile, 9 Pin D-SUB connector (F) should be connected and power off the two units and then perform connection.
  • Page 48 LCR Meter 11022 User’s Manual <MAIN INDEX-BIAS COMP. SET > DIGIT UP The digit cursor on for adjusting upward Ls O_NOMINAL: 600.000uH DIGIT DOWN The digit cursor on for adjusting downward Hi: 630.000uH Lo: 570.000uH QO_Lo_LIMIT: 070.000 DIGIT Move the cursor to right for one digit Ls B_Hi : 550.000uH...
  • Page 49 Panel Description <SYSTEM CONFIG> CONTINUE Test continuously BIAS C. MODE : MASTER TEST FAIL STOP STOP Fail and stop test BAUDRATE : NONE <MAIN INDEX-BIAS COMP. TEST > SPEED: M Measurement speed setting LsO : 0.6450mH TRIG.: MAN Trigger mode setting 87.6 LsB : R1: A...
  • Page 50: Description Of Loading Correction Function

    5. The 24 Pin Handler connector (M) is able to input control signal or output judgment signal as using MASTER MODE with Chroma 1320 control line for performing BIAS COMPARE measurement. It is for the use of automatic equipment, the pins are shown below.
  • Page 51: Operation Of Bin Sorting 99 Groups

    Panel Description Enable LOAD function < MAIN INDEX-CORRECT LOAD> LOAD: OFF Disable LOAD function FREQUENCY: 1.0KHz PARAMETER: Rs-Xs MEAS. LOAD Measure parameter actual value REF. A VAL: +00000.0000-Ω Ω MEAS A VAL: REF. B VAL: +00000.0000-Ω Ω MEAS B VAL: Set the frequency and parameter requires along with parameter reference value desire to set then turn LOAD function on.
  • Page 52 LCR Meter 11022 User’s Manual BIN SET Bin setting < MAIN INDEX-BINNING > PARAMETER: Cp – D STORE Store setting value NOMINAL : 00000.0000pF RECALL Recall setting value AUX_HI : +99999.9999- AUX_LO : +00000.0000 Press [F1] then enter Bin Setting screen is as below: The digit which cursor on for adjusting upward.
  • Page 53: Dual Frequency Function (Option)

    Panel Description 4.4.9 Dual Frequency Function (Option) 1. Dual frequency function is mainly operating DUT capacitor parameter of multi-bandwidth. For example, Multi-layer Ceramic Capacitor (MLCC) applies to integrated circuit and filtering circuit its operation bandwidth is extra widely. The capacitance value should be measured in lower operation frequency, whereas ESR value should be noticed in higher operation frequency.
  • Page 54 LCR Meter 11022 User’s Manual <SYSTEM CONFIG> 25 OHM 25 OHM mode VM/IM DISPLAY : OFF CONST RSOURCE : 100/25 OHM AVERAGE TIME : 001 CABLE LENGTH : 0M BEEPER : LOW KEY LOCK : OFF Next. P 2/2 To the 1 page The calculation of trigger delay time is decided by DUT capacity.
  • Page 55 Panel Description < MAIN INDEX-DUAL FREQ. > SETTING Dual frequency function setting STORE Store setting 1ST PARA: Cp-D FREQ: 1.0KHz NOMINAL: +00000.0000-F RECALL Recall setting 2ND PARA : D FREQ: 1.0KHz TEST Dual frequency function test Press [F1] to go to modification setting screen is shown as below. <MAIN INDEX-DUAL FREQ.
  • Page 57: Description Of Gpib/Rs-232 Interface

    Description of GPIB/RS-232 Interface Description of GPIB/RS-232 Interface Overview Using the IEEE-488.2 interface, the users can operate the 11022 by remote control, transfer data and etc. The RS-232 interface is for option. This chapter is mainly for providing the description of the 11022 interface Bus (GPIB) or RS-232 command for the users’ writing program to control the 11022 in order to do test data process.
  • Page 58: Ieee-488 Interface Connector

    LCR Meter 11022 User’s Manual 5.2.4 IEEE-488 Interface Connector • Pin configuration of the meter side connector is as below shown. 1 DIO1 13 DIO5 2 DIO2 14 DIO6 3 DIO3 15 DIO7 4 DIO4 16 DIO8 5 EOI 17 REN...
  • Page 59: Response Of Interface Messages

    Description of GPIB/RS-232 Interface 5.2.6 Response of Interface Messages • The meter is capable of responding to the following messages. Interface Message Response ‧ Only addressed devices that receive this command are set to local mode. GTL (Go To Local) ‧...
  • Page 60: Rs-232 Interface Connector

    LCR Meter 11022 User’s Manual Baud Rate: 9600 / 19200 / 28800 / 38400 / 57600 Parity: NONE Flow Control: Hardware Transfer Bit: 1 start bit + 8 data bits + 1 stop bit Query command uses Q and A method.
  • Page 61: Command Description

    Description of GPIB/RS-232 Interface Command Description 5.4.1 Command Configuration Command is with tree configuration level. All of 11022 related commands can see totally from the following figure. Root ABORt CALCulate FORMat LIMit BEEPer CONDition [STATe] LOWer [DATA] STATe UPPer [DATA] MATH EXPRession CATalog NAME...
  • Page 62: Description For Command Configuration

    LCR Meter 11022 User’s Manual HANDler ALARm CONs INTEgration LOCal TRIGger DELay [IMMediate] SOURce EDGE BINning STATe UPPer LOWer RESUlt NOMInal MODE STATe UPPer LOWer Dual FETCh? TRIGger SOURce [:SENSe] FIMPedance APERture BINning MODE FIRSt BINning NOMInal UPPer LOWer CALCulate1...
  • Page 63: Command Syntax

    Description of GPIB/RS-232 Interface :CALCulate:MATH:EXPRession:NAME Then can get NAME command correctly. Moreover, if send (Example: NAME and CATalog) two commands simultaneously. Be able to use the following method make the instruction more easily and clearly. :CALCulate : MATH: EXPRession:NAME;CATalog Its’ meaning is the same as the following two instructions. Isn’t it more briefly? :CALCulate:MATH:EXPRession:NAME :CALCulate:MATH:EXPRession:CATalog The commands need using colon ( : ) to separate between level and level.
  • Page 64: Common Command

    LCR Meter 11022 User’s Manual (3) Before the first parameter in every command need a blank character. To end character To end character has three kinds: [CARRIAGE RETURN] (0Dh), [NEW LINE] (0Ah) and [CARRIAGE RETURN] (0Dh) + [NEW LINE] (0Ah) 5.4.4 Common Command...
  • Page 65: Command Description

    Description of GPIB/RS-232 Interface Main Board Test Key Control (15) *STB It queries the contents of the Status Bit Register. 5.4.5 Command Description ABORt Command (1) Command: ABORt Parameter: None Return: None Function: It terminates processing trigger system immediately, that is say to reset trigger system.
  • Page 66 LCR Meter 11022 User’s Manual (7) Command: BINning: NO: STATe Parameter: {ON(1) | OFF(0)} Return: {1 | 0} Function: It sets if BIN NO>9 function in SYSTEM CONFIG is enabled. (8) Command: BINning: NO Parameter: {1~99} Return: {1~99} Function: It sets BIN NO when BIN NO>9 function in SYSTEM CONFIG is enabled.
  • Page 67 Description of GPIB/RS-232 Interface (2) Command: CALCulate2:FORMat Parameter: {IMAGinary|PHASe|D|Q|REAL|RS|XS|LP|INV} Return: {PHASe|D|Q|REAL|MLINear|LP|INV} Function: It sets or queries the current measurement secondary parameter. Description: IMAGinary Impedance imaginary PHASe Phase Dissipation factor Quality factor REAL Impedance real Equivalent series resistance Imaginary series equivalent value MLINear Absolute value of impedance Parameter...
  • Page 68 LCR Meter 11022 User’s Manual Description: ON(1) Turn on comparison function OFF(0) Turn off comparison function (7) Command: CALCulate{1|2}:LIMit:UPPer[:DATA] Parameter: {The high limit value | MAXimum | MINimum Return: The high limit value, the format is<NR3> Function: It sets or queries the high limit value.
  • Page 69 Description of GPIB/RS-232 Interface Parameter: {Primary, secondary parameter the nominal value in COMPARE (reference value) (NR3)} Return: None Function: It sets primary, secondary parameter the nominal value in COMPARE mode. If the nominal value of primary parameter set on 100, command is DATA [: DATA] REF1, 100;...
  • Page 70 LCR Meter 11022 User’s Manual Equivalent series capacitance Equivalent parallel inductance Equivalent series inductance Equivalent series impedance Equivalent series resistance Equivalent parallel resistance (2) Command: DUAL:FIRSt:CALCulate2:FORMat Parameter: {IMAGinary|PHASe|D|Q|REAL|RS|XS} Return: {IMAGinary|PHASe|D|Q|REAL|RS|XS} Function: It sets or queries the 1 measurement secondary parameter in dual frequency function.
  • Page 71 Description of GPIB/RS-232 Interface Function: The 2 measurement setting or query in dual frequency function if auto range selection mode is enabled. (7) Command: DUAL: SECOnd [:SENSe]:FIMPedance:RANGe[:UPPer] Parameter: {Measurement range value |UP|DOWN} Unit: [MOHM | kOHM | OHM | MAOHM], if the unit isn’t given then it will be defined as OHM.
  • Page 72 LCR Meter 11022 User’s Manual function, i.e. the high limit value of measurement secondary parameter. (14) Command: DUAL:FIRSt:BINning:LOWer:AUX Parameter: {Low limit (NR3) of the 1st measurement secondary parameter.} Return: {Low limit (NR3) of the 1st measurement secondary parameter. Function: The AUX-LO value of 1st measurement setting Binning in dual frequency function, i.e.
  • Page 73 Description of GPIB/RS-232 Interface NOGO (19) Command: DUAL: TRIGger:SOURce Parameter: {BUS|EXTernal|INTernal|MANual} Return: {BUS|EXTernal|INTernal|MANual} Function: It sets or queries trigger mode currently in dual frequency function. Description: BUS Bus trigger EXTernal External trigger INTernal Internal trigger MANual Manual trigger (20) Command: DUAL [:SENSe]:FIMPedance:APERture Parameter: 0.025 (Fast), 0.065 (Medium), 0.500 (Slow) Return: 0.025 (Fast), 0.065 (Medium), 0.500 (Slow) Function: It sets or queries measurement speed currently in dual frequency function.
  • Page 74 LCR Meter 11022 User’s Manual FORMat Command (1) Command: FORMat[:DATA] Parameter: {ASCii} Return: ASK Function: It sets or queries the data format of transmission numeral. INITiate Command Set (1) Command: INITiate[:IMMediate] Parameter: None Return: None Function: All sequences escape from idle status and enter initial status.
  • Page 75 Description of GPIB/RS-232 Interface [:SENSe]:CORRection:DATA:SHORt:RANGe:1OHM? Parameter: None Return: SHORT correction value of RANGE 1Ω, the format is < NR3 > Function: It sets or queries SHORT correction value of RANGE 1Ω, that is R, X (7) Command: [:SENSe]:CORRection:DATA:SHORt:RANGe:100MOHM [:SENSe]:CORRection:DATA:SHORt:RANGe:100MOHM? Parameter: None Return: SHORT correction value of RANGE 100mΩ, the format is <...
  • Page 76 LCR Meter 11022 User’s Manual (3) Command: SOURce:BIAS:CURRent:CONDition Parameter: {ON(1)|OFF(0)} Return: {1|0} Function: It returns and sets the output condition of bias current source. Description: ON(1) Enable bias current source output. OFF(0) Disable bias current source output. (4) Command: SOURce:VOLTage[:LEVel][:IMMediate] :OFFSet:STATe...
  • Page 77 Description of GPIB/RS-232 Interface (3) Command: SYSTem:ERRor? Parameter: None Return: It returns error message in queue, the fixed return is 0. (4) Command: SYSTem:KLOCk Parameter: {ON(1)|OFF(0)} Return: {1|0} Function: It sets or queries whether 11022 key is locked or not. (5) Command: SYSTem:LOCal Parameter: None Return: None...
  • Page 78: Command Notices

    LCR Meter 11022 User’s Manual Function: It sets LINK1320 ON/OFF. TRIGger Command Set (1) Command: TRIGger:DELay Parameter: Trigger delay time Unit: [MS|S] if there is no command unit the auto define S. Return: Trigger delay time, the format is <NR3>.
  • Page 79: Format Of Data Transfer

    Description of GPIB/RS-232 Interface (2) When users give TRIGger command, the sequential other setting commands (non FETCh ? commands) will be given. The program needs to write the function of time delay. This delay time is the measurement time of 11022, otherwise the part of sequential commands can’t normal function.
  • Page 80: Status Reporting Structure

    LCR Meter 11022 User’s Manual Status Reporting Structure Service Request Figure 5-1 Status Reporting Structure The 11022 LCR Meter can send an SRQ (Service Request) control signal when it requires the controller to perform a task. When the 11022 LCR Meter generates an SRQ, it also sets Bit 6 of the status Byte Register, RQS (Request Service) bit.
  • Page 81: Status Byte Register

    Description of GPIB/RS-232 Interface Status Byte Register The Status Byte Register is composed of eight bits that summarize an overlaying status data structure. The Status Byte Register can be read using either *STB? Or SPOLL, which return a decimal expression of the contents of the register (equal to the sum of the total bit weights of all the bits set to “1”).
  • Page 82: Standard Event Status Register

    LCR Meter 11022 User’s Manual Standard Event Status Register The Standard Event Status Register is frequently used and is one of the simple. Standard Event Status Register Read by *ESR? Standard Event Status Enable Register Summary Message Event Set by *ESE...
  • Page 83: Standard Operation Status Group

    Description of GPIB/RS-232 Interface Bit No Bit Weight Description Power on Bit – This bit is set when the 11022 LCR Meter has been turned OFF and then ON since the last time this register was read. Always zero (0) Command Error Bit- This bit is set if the following command errors occur.
  • Page 84 LCR Meter 11022 User’s Manual 15 14 4 3 2 1 0 - - - - Operation Status Condition Register (Reserved) Transition filter 4 3 2 1 - - - Operation Status Event Register Operation Status Enable Register 4 3 2 1 0...
  • Page 85: Description Of Handler Interface

    Description of Handler Interface Description of Handler Interface Binning, Compare and Bias Compare in 11022 as well as LINK 1320(1320 Bias Current Source 0-10A and 1320 Bias current Source 0-20A) are all connected to external unit by Handler interface. There are connectors 50 Pin (Standard) and 24 Pin (Optional) in Binning, Compare and Bias Compare, LINK 1320 only has 50 Pin connector.
  • Page 86: Pin

    LCR Meter 11022 User’s Manual 6.1.2 24 Pin Name Description /EXT External trigger 3.20 BIN 7 BIN 7, primary parameter test value within BIN 7 sets range 4.24 BIN 8 BIN 8, primary parameter test value within BIN 8 sets range Ground 9.13...
  • Page 87: Pin

    Description of Handler Interface EOT End of Test 45-46 COM COMMON 47-48 49-50 GND Connect with cover 6.2.2 24 Pin Name Description /EXT External trigger 3.20 A LO Primary parameter test value too low 4.24 A HI Primary parameter test value too high GND Ground 9.13 A NG Primary parameter test value not in specification...
  • Page 88: Description For Bias Compare Handler Interface Pin

    LCR Meter 11022 User’s Manual Description for Bias Compare Handler Interface Pin 6.3.1 50 Pin Name Description VEXT External DC voltage acceptable voltage range is 5V ~ 24V. 3-18 /EXT External trigger 20-21 VEXT External DC voltage acceptable voltage range is 5V ~ 24V.
  • Page 89: Description For Link 1320 Handler Interface Pin

    Description of Handler Interface Description for Link 1320 Handler Interface Pin 6.4.1 50 Pin Name Description External DC voltage acceptable voltage range is 5V ∼ 24V. VEXT IBCLK COMMON IBDATA XB20ARDY IBSTR IBIRDY COMMON XBC1320 VINT Internal DC voltage +5V 14-19 External DC voltage acceptable voltage range is 5V ∼...
  • Page 90: Handler Interface Description Of Dual Frequency (Option)

    LCR Meter 11022 User’s Manual The secondary parameter will over high/low limit setting value if BIN number is judged as “0”. Therefore, the binary string data is “00000000” outputted by Handler interface. The primary parameter test value isn’t in all setting specifications if BIN result is judged as “OUT”. Thus, the binary string data is “01100100”...
  • Page 91: Pin

    Description of Handler Interface 6.6.2 24 Pin Signal Name Description /EXT External trigger The secondary parameter test value with second frequency is out 3, 20 SFQ NOGO of setting range The secondary parameter test value with second frequency is in 4, 24 SFQ GO setting range...
  • Page 92 LCR Meter 11022 User’s Manual /EXT Previous BIN value Bin value currently Measurement FAST MEDIUM SLOW Speed T1 ≥ 1.25μs 1.25μs 1.25μs Measurement T2 < Time T3 < 21ms 51ms 360ms Set Trigger Edge to Rising state in system parameter operation then set INTEG. CYCLE to a cycle.
  • Page 93: Setting With Trigger Delay Time (Freq. ≥ 100Hz)

    Description of Handler Interface Measurement Speed FAST MEDIUM SLOW T1 ≥ 1.25μs 1.25μs 1.25μs Measurement T2 < Time T3 < 21ms 51ms 360ms 6.7.2 Setting with Trigger Delay Time (Freq. ≥ 100Hz) Set Trigger Edge to Falling state in system parameter operation then set INTEG. CYCLE to a cycle.
  • Page 94 LCR Meter 11022 User’s Manual /EXT Previous BIN value BIN value currently Measurement FAST MEDIUM SLOW Speed T1 ≥ 1.25μs 1.25μs 1.25μs Measurement T2 < Time T3 < (21+T ) ms (51+T ) ms (360+T ) ms DELAY DELAY DELAY T1 is /EXT external trigger signal, this external trigger time should be more than 1.25μs to...
  • Page 95: Appendix A Optional Accessory

    Optional Accessory Appendix A Optional Accessory A110211 Component Test Fixture This fixture is specially designed for testing DIP passive component impedance. It adopts 4-terminal measured method which can test resistance range extensively. DIP passive component suitable maximum pitch is evaluated as 73mm and minimum pitch is 5mm.
  • Page 96 10mm. The rest are evaluated as more suitable dimensions are 4mm, 5mm, 6.3mm and 8mm respectively. A110236 Rack Mountain Kit This kit is specially designed for Chroma LCR 11022 fixes on large-sized system rack conveniently such as Chroma 8800 ATS system.
  • Page 97 Optional Accessory A110244 High Capacitance Capacitor Test Fixture This fixture is specially designed for testing high capacitance capacitor. High capacitance capacitor junction is M5 screw hole and outer diameter over 6mm. The DUT is with very low impedance when clip the junction by using test clamp isn’t easy to be fixed.
  • Page 99 Kuei-shan Hwaya Technology Park Taoyuan County 33383, Taiwan 33383 台灣桃園縣龜山鄉 華亞科技園區華亞一路 66 號 T +886-3-327-9999 F +886-3-327-8898 Mail: info@chromaate.com http://www.chromaate.com Copyright by CHROMA ATE INC. All Rights Reserved. All other trade names referenced are the properties of their respective companies.