Shoulder
Reference Guide
1. Pass suture through soft tissue in desired stitch configuration. Feed sutures
through the anchor eyelet using the suture threader.
3. Remove the drill and drill guide. Advance the loaded suture anchor into the
operative cannula. Remove excess suture slack, approximate the anchor to the
prepared bone site, and ensure that the sutures are not twisted around the anchor.
5. While maintaining slight downward pressure on the inserter, rotate the knob
clockwise to deploy the anchor until a minimum of five (5) clicks are heard. Do not
rotate the knob counterclockwise during and after anchor deployment.
Turn until 5 audible
clicks are heard
Supporting healthcare professionals
2. Insert the appropriate drill guide and obturator. Remove the obturator. While
firmly holding the guide in place on the bone at the desired implantation site, use
the recommended drills to prepare the insertion site. For dense bone, it may be
required to cycle the drill in and out of the implantation site. Ensure drill is always
powered during removal.
4. Establish and maintain axial alignment of the suture anchor with prepared
insertion site. Place tip of anchor into the prepared hole. Apply tension to suture
limbs to reduce labrum to desired location. Use a mallet to tap the inserter handle
until the laser mark is flush with cortical bone.
6. Slowly remove the inserter by pulling straight back or back tapping the distal
end of the handle with a mallet. Do not wiggle the handle to disengage the
anchor. Discard the insertion device and cut the excess suture.
For use in Australia only

Advertisement

loading
Need help?

Need help?

Do you have a question about the MICRORAPTOR REGENESORB and is the answer not in the manual?

Questions and answers

Summary of Contents for Smith & Nephew MICRORAPTOR REGENESORB

  • Page 1 Shoulder Reference Guide Supporting healthcare professionals 1. Pass suture through soft tissue in desired stitch configuration. Feed sutures 2. Insert the appropriate drill guide and obturator. Remove the obturator. While through the anchor eyelet using the suture threader. firmly holding the guide in place on the bone at the desired implantation site, use the recommended drills to prepare the insertion site.
  • Page 2 MICRORAPTOR™ REGENESORB Suture Anchor and Guide System Reference # Description MICRORAPTOR REGENESORB Suture Anchors 72205021 MICRORAPTOR Knotless Peek Suture Anchor MICRORAPTOR REGENESORB Drill Guides, Drills and Obturators 72205022 MICRORAPTOR Drill 72205169 MICRORAPTOR Drill Guide, Crown Tip 72204991 MICRORAPTOR Drill Guide, Crown Tip...

This manual is also suitable for:

72205021