ENGLISH Contents Read these instructions for use carefully. Follow the warnings and safety notes. Keep these in- 1. Included in delivery ............3 structions for use for future reference. Make the 2. Signs and symbols ............3 instructions for use accessible to other users. If 3.
Page 4
Separate the product and packaging elements and dis- CAUTION pose of them in accordance with local regulations. Indicates a potentially impending danger. If it is not avoided, Device protected against foreign objects ≥12.5 mm and slight or minor injuries may result. IP22 against water dripping at an angle NOTICE...
3. INTENDED USE 4. WARNINGS AND SAFETY NOTES Purpose GENERAL NOTES The thermometer is intended exclusively for measuring human body temperature in the ear canal (ear temperature mode) or on the fore- Product information head (forehead temperature mode). The thermometer is only intend- •...
• The measurements taken by you are for your information only NOTES ON HANDLING BATTERIES – they are no substitute for a medical examination. Consult a doctor before you undertake any medical treatment. WARNING • Keep the device away from children, animals and vermin. •...
• Exposure of batteries to an environment with extremely high • Temperatures measured with di erent thermometers temperatures or an extremely low air pressure may result in ex- should never be compared with one another. plosion or leakage of flammable liquids and gases. •...
6. DEVICE DESCRIPTION 7. INITIAL USE The associated drawings are shown on page 3. Inserting the batteries 1. Loosen the screw on the battery compartment lid and pull the lid Forehead/cover cap LED (red) (fever indicator) o downwards. LED (green) 2.
8. SWITCHING ON AND SETTING THE • The ear thermometer may be used by children only un- THERMOMETER der adult supervision. Measurement is usually possible over the age of 6 months. In infants under 6 months, Press the button for 1 second to switch the thermometer on. the ear canal is still very narrow so the temperature of The device performs a brief self test: all elements of the display are the eardrum often cannot be recorded and the result...
2. Place the measuring head with the forehead/cover cap fitted on • Insert the sensor tip carefully and press the button for 1 se- the temples, hold down the button and move the thermometer cond. smoothly over the forehead to the other temple. •...
To switch back to the Ear/Forehead thermometer mode, press the Do not use the device again until it is completely dry. buttons simultaneously for 3 seconds until the sym- Storing the device bol disappears and there is a short beep. When switched on again, the object temperature mode is exited automatically.
We reserve the right to make technical changes to improve and de- „Err“ from 50 to 104 °F (10°C temperature ranges. velop the product. to 40°C). Name and model FT 65 Type FT 65 12. DISPOSAL Measurement Infrared measurement For environmental reasons, do not dispose of the device in house- method hold waste at the end of its service life.
Clinical accuracy Ear thermometer mode (at the eardrum): Acoustic signal When the device is switched on, during Clinical bias: -0.03 °C measurement, at the end of the measure- Limit of Agreement: 0.36 °C ment and if the body temperature is above Clinical repeatability: ±...
regard to electromagnetic compatibility. Please note that portable Notification of incidents and mobile HF communication systems may interfere with this de- For users/patients in the European Union and identical regulation vice. For more details, please contact our Customer Services at the systems (EU Medical Device Regulation (MDR) 2017/745), the follo- address indicated.
Page 16
UK Importer: Beurer UK Ltd., Suite 16, Stonecross Place, Stonecross Lane North, WA3 2SH Lowton, United Kingdom Beurer GmbH • Söflinger Str. 218 • 89077 Ulm, Germany • www.beurer.com www.beurer-healthguide.com • www.beurer-gesundheitsratgeber.com...
Need help?
Do you have a question about the FT 65 and is the answer not in the manual?
Questions and answers