Continuously adjustable digital microliter 8-channel pipette (16 pages)
Summary of Contents for Rainin EDP-10
Page 1
All manuals and user guides at all-guides.com sales@artisantg.com artisantg.com (217) 352-9330 | Click HERE Find the Mettler Toledo / Rainin EP-10ML at our website:...
Page 2
All manuals and user guides at all-guides.com ™ Electronic Pipette Six pipettes for volume ranges from 0.5 µL to 2.5 mL 100 µL model shown...
20.0 WARRANTY INFORMATION ..........39 Prices and specifications are subject to change without notice. “EDP”, “RAININ Certified” and “Clean-Pak” are trademarks of Rainin Instrument, LLC. EDP Pipettes are manufac- tured under U.S. Patent Nos. 4,671,123. Other U.S. and national patents issued and...
All manuals and user guides at all-guides.com DESCRIPTION RAININ EDP Electronic Pipettes measure liquid volumes from one to 2500 microliters Six EDP models have nominal volume ranges of 10, 25, 100, 250, 1000, and 2500 accurately and precisely. Volume measurements may occur in any of four different microliters, respectively.
Improper connection could result in damage to either the Power Pack or the EDP pipette. ately with the shipping carrier. Rainin Instrument LLC can assist you in filing a claim. When the Power Pack is connected, the instrument should purr, move, and execute the...
All manuals and user guides at all-guides.com LIQUID MEASUREMENT MODES QUICK REFERENCE: GENERAL KEYBOARD DISABLED The EDP Electronic Pipette can operate in four distinct liquid measurement modes, DISPENSE described in Table 4.1. When power is first applied, or after changing the liquid end, the EDP pipette defaults to the P mode.
All manuals and user guides at all-guides.com QUICK REFERENCE: PIPETTE MODE (#1) OPERATION: READY PICKUP DISPENSE BLOWOUT READY PISTON TO SET MODE POSITION: Press TO SET VOLUME Press 1 5 . 0 E (Desired volume) HOME DOWN TO CLEAR, EMPTY, OR RETURN HOME Pull Pull...
All manuals and user guides at all-guides.com QUICK REFERENCE: MULTIDISPENSE MODE (#2) OPERATION: PICKUP DISPENSE DISCARD REMAINDER READY BLOWOUT READY PISTON TO SET MODE POSITION: Press TO SET NEW VOLUMES V1: Press 1 0 0 0 Repeat (Desired pickup volume) HOME V2: Press 2 1 0 DOWN...
All manuals and user guides at all-guides.com QUICK REFERENCE: TITRATE MODE (#3) OPERATION: PICKUP DISPENSE DISCARD V2 + REMAINDER READY BLOWOUT READY PISTON TO SET MODE POSITION: Press TO SET NEW VOLUMES V1: Press 1 0 0 . 0 (Desired pickup volume) HOME V2: Press 1 5 .
All manuals and user guides at all-guides.com QUICK REFERENCE: DILUTE MODE (#4) OPERATION: PICKUP PICKUP DILUENT SAMPLE PICKUP READY DISPENSE BLOWOUT READY PISTON TO SET MODE POSITION: Press TO SET NEW VOLUMES V1: Press 9 0 . 0 (Desired diluent volume) HOME V2: Press 1 0 .
All manuals and user guides at all-guides.com QUESTIONS AND ANSWERS Q #1: Q #4: Is it necessary to enter the decimal point when setting a volume? What happens if no encoder plug is present? No. If the volume is a whole microliter, the decimal point and zeros following it need not If the keyboard is not locked the numeric display will show “–...
Page 12
All manuals and user guides at all-guides.com Q #6: WARNING: What does it mean when the volume indicator on the display is flashing? In the M and T modes, a possible trap exists if V1 has been inadver- ULTIDISPENSE ITRATE tently set to a very small value, especially if the user cannot remember what value was entered for V1.
Page 13
All manuals and user guides at all-guides.com At all times during the dispense cycle, the volume display records the total volume The keyboard lock is for protection. When the EDP pipette is used to repeat the same program over a long period of time, locking the keyboard eliminates the possibility of dispensed.
All manuals and user guides at all-guides.com CHARGING THE POWER PACK Q #17: Operating the EDP Electronic Pipette with the Portable Power Pack installed provides Can the EDP pipette be used to measure the volume of liquid in a container? maximum freedom of movement and ease of use.
WALL TRANSFORMER Performance to specification by EDP Pipettes is guaran- external contacts on the sides of the drive module. teed only when RAININ Certified Tips are used as recom- While the instrument is being charged on the Stand, Seal mended. Rainin Instrument LLC accepts no responsibility the symbol “FC”...
All manuals and user guides at all-guides.com Tips for use with EDP are: HINTS FOR PROPER PIPETTING EDP Electronic Pipettes are calibrated to deliver accurate volumes under the following conditions: Model EDP-10 EDP-25 EDP-100 EDP-250 EDP-1000 EDP-2500 Maximum 1. The liquid pipetted is water or any liquid with a density not significantly different from, Volume 10 µL...
All manuals and user guides at all-guides.com CHANGING LIQUID ENDS Any of the six available liquid ends can be rapidly mounted on the EDP Drive Module. Liquids of viscosity moderately higher than water, such as serum and plasma, can be measured accurately by using the reverse-mode pipetting strategy to aspirate a larger To remove the liquid end, first move the EDP pipette to the home position by pressing volume than is dispensed.
All manuals and user guides at all-guides.com REMOVING THE TIP EJECTOR CAUTION: Except for the 2500 microliter model, each EDP liquid end has an integral tip ejector, a When attaching a new liquid end, always install movable cylinder surrounding the liquid end and extending up through the retainer ring. the liquid end on the drive module before inserting The tip ejector is operated by pressing the tip ejector button on the head of the drive the encoder plug.
Periodically, the liquid end should be disassembled for inspection: First remove the liquid end from the For a proper seal between the liquid end and the tip, it is important to use only RAININ drive module. (It is not necessary to remove the Certified™...
Rainin Instrument LLC maintains a complete serv- Reinsertion is most easily done with the shaft ice department to correct the problem. Please call RAININ for instructions regarding inverted. After tightening with the wrench, push the service for your EDP instrument.
ACCURACY means the closeness with which the dispensed volume approximates the volume setting of the instrument. Accuracy is specified as “Mean Error,” the maximum By phone 8am–8pm EST: 800-4-RAININ (800-472-4646) amount by which the mean value of a large number of replicate measurements of the By fax: 510-564-1617 or 781-938-1152 same volume will deviate from the set volume.
All EDP Electronic Pipettes are warranted for one year against defects in materials and Liquid Ends: workmanship. For warranty information regarding the warranty or repair of any EDP 6100-065 6100-461 Liquid End, 10 µL instrument, please call Rainin Instrument, LLC at 800-472-4646. 6100-066 same Liquid End, 25 µL 6100-067 same Liquid End, 100 µL...
Page 23
All manuals and user guides at all-guides.com Rainin Instrument, LLC 7500 Edgewater Drive, Box 2160 Oakland, CA 94621-0060 www.rainin.com PI-17 R1 WEB...
Page 24
All manuals and user guides at all-guides.com...