Warnings and safety notes Before first use Always read the full instructions for use before taking your first blood glucose measure- ment. Check the measuring unit in accordance with the instructions for use. Risk of infection All components of the monitor, its accessories and/or its replacement parts may come into contact with human blood and are therefore a possible source of infection.
Page 3
® Transfer measured values using Bluetooth ® • Activate Bluetooth in your smartphone’s settings. • Download the free “beurer HealthManager Pro” from the Apple App Store or Google Play. • Start the app and follow the instructions. • Select GL60 in the app and follow the instructions. ®...
Making and changing basic settings The abbreviations shown in the display vary depending on the selected language. Detailed explana- tions can be found in the following table: Full German (DE) Full English (EN) Sì Sí Nein Setzen Rég Beep Beep Beep Açk Ketone...
Page 5
Full German (DE) Full English (EN) Bluetooth Bluetooth Zielwert Target Value Verbinden Pairing Nicht OK not ok Bşz Löschen delete Unterzucker Hypoglycaemia Niedrig Hoch high...
Preparing the monitor and lancing device Inserting the lancet needle Turn o cap from the lancing device. Insert lancet needle in the lancing device. Hold the shaft of the lancet needle and remove protective disc by turning. Retain the protective disc. Reapply the cap to the lancing device. Turn the moving upper part of the cap until the desired penetration depth is displayed.
Taking a blood sample Wash your hands with soap and warm water before taking a blood sample. Pricking your finger Firmly position the lancing device slightly to the side of the centre of the fingertip. Press the trigger. Please discard the first drop of blood. The blood sample taken must: •...
Page 8
Apply blood to the test strip Hold the blood absorption gap of the test strip against the blood drops until the gap is fully filled, and the measurement process starts on the meter display, followed by displaying the measurement result Normal value: 2 hours after a meal: 70 –...
Page 9
Evaluating critical measured values Measurement result is outside the detec- Immediate medical treatment table range is required. below 10 mg/dL (below 0.56 mmol/L) Have a suitable snack. Low blood glucose level Follow the instructions from below 70 mg/dL (below 3.9 mmol/L) mg/dL mmol/L your physician.
Page 10
Evaluating critical measured values High blood glucose level, Conduct ketone test Consult 16 7 possibly ketones above 250 mg/dL (13.9 your physician on how to mg/dL mmol/L mmol/L) do this. Take another measurement using a new test strip. If the Very high blood glucose level result is the same as before, above 630 mg/dL (35.0 mmol/L)
Information on the display Flat batteries. Replace all batteries. Temperature of the measuring en- Repeat the test using a new test strip as soon vironment, blood glucose monitor as the measuring environment, monitor and test strip have reached the permissible opera- or test strip above the permitted range.
Page 12
Information on the display Repeat the measurement using a new test Err002 Insu cient blood on the test strip strip. Remove and reinsert batteries. If the problem Err004 System error persists, contact Customer Services. Remove and reinsert batteries. If the problem Unknown error messages persists, contact Customer Services.