Instructions for use KaVo uniQa Table of contents Table of contents 1 User instructions ....................... 1.1 Terms and conditions of warranty..................8 2 Safety..........................2.1 Explosion hazard ......................9 2.2 Infection hazard ......................9 2.3 Electrical shock......................10 2.4 Technical condition ......................10 2.5 Electromagnetic fields ....................
Page 4
Instructions for use KaVo uniQa Table of contents 4.6.1 Armrest......................47 4.6.2 Automatic positioning of patient chair ..............47 4.6.3 Positioning the dental chair manually..............49 4.7 Adjusting the headrest ....................50 4.7.1 Adjusting the double-jointed headrest pushbutton ..........50 4.7.2...
Page 5
Instructions for use KaVo uniQa Table of contents 4.16.3 Calling-up surgery mode ...................123 4.16.4 Mounting or pulling off the handpiece or contra-angle handpiece ......124 4.16.5 Starting-up the motor ..................125 4.16.6 Using the surgical motor with programme steps ..........125 4.16.7 Using the surgical motor with "Free application" activity ........128 4.16.8 Setting the operating light (LUX)................130...
Instructions for use KaVo uniQa 1 User instructions | EAC conformity mark (Eurasian Conformity) Transportation and storage conditions (temperature range) Transportation and storage conditions (air pressure) Transportation and storage conditions (Humidity) Protect from moisture (Keep dry) Protect from impact Transport upright...
Instructions for use KaVo uniQa 1 User instructions | 1.1 Terms and conditions of warranty Please refer to the serial number of the product in all inquiries! For further in- formation, please visit: http://www.kavo.com 1.1 Terms and conditions of warranty...
Instructions for use KaVo uniQa 2 Safety | 2.1 Explosion hazard 2 Safety The instructions for use are a component of the product and must be read care- fully prior to use and must be accessible at all times. The device may only be used in accordance with the intended use, any other type of use is not permitted.
Instructions for use KaVo uniQa 2 Safety | 2.3 Electrical shock 2.3 Electrical shock Improper connection of a non-medical system to the USB interfaces of the device can lead to electrical shock and injury to patient, user and third parties.
Instructions for use KaVo uniQa 2 Safety | 2.6 Use 2.6 Use Sitting down on a dental chair that is in horizontal orientation may lead to in- jury. 4 Do not sit down or support yourself on the head or foot end or edges of the upholstery of the patient chair when it is in a horizontal position.
Instructions for use KaVo uniQa 2 Safety | 2.8 Qualification of personnel The use of non-approved accessories as well as non-approved modifications made to the medical device cause the conformity of the medical device accord- ing to MDR 2017/745 to be voided. Anybody using non-approved accessories or...
Instructions for use KaVo uniQa 2 Safety | 2.11 Intended use Backup copies must be stored in a safe place protected from access of third parties. The detailed procedure should be defined in a data protection concept. 2.11 Intended use Indications for use: ▪...
Instructions for use KaVo uniQa 2 Safety | 2.11 Intended use NOTE The USB interfaces of the system may only be connected to IT devices ap- proved by KaVo. NOTE Connecting IT equipment to the medical electrical system make sure to com- ply with EN 60601-1.
Instructions for use KaVo uniQa 2 Safety | 2.11 Intended use NOTE The MULTIflex couplings, the current K/KL motors, and the ultrasonic scaler hoses of KaVo are equipped as standard with a protective device to prevent treatment water from being drawn back into the treatment centre via the handpieces.
Page 16
Instructions for use KaVo uniQa 2 Safety | 2.11 Intended use Currently applicable packaging law NOTE Only valid for the Federal Republic of Germany. Dispose of and recycle the packaging appropriately in accordance with current packaging law, employing waste management or recycling companies. Comply with the comprehensive return system.
Instructions for use KaVo uniQa 3 Description of the product | 3.1 Treatment centre 3 Description of the product NOTE The configuration can be changed as needed and does not have to be as shown in the figure. 3.1 Treatment centre 3.1.1 KaVo uniQa T...
Instructions for use KaVo uniQa 3 Description of the product | 3.1 Treatment centre 3.1.2 KaVo uniQa TM ① Patient element ⑦ Bench ② Assistant element ⑧ Armrest ③ Headrest ⑨ Base plate ④ Backrest ⑩ Foot control ⑤ Operating light ⑪...
Instructions for use KaVo uniQa 3 Description of the product | 3.1 Treatment centre 3.1.3 KaVo uniQa S ① Patient element ⑦ Bench ② Assistant element ⑧ Armrest ③ Headrest ⑨ Base plate ④ Backrest ⑩ Foot control ⑤ Operating light ⑪...
Instructions for use KaVo uniQa 3 Description of the product | 3.2 Versions of the dentist element 3.1.4 KaVo uniQa Cart ① Patient element ⑦ Bench ② Assistant element ⑧ Armrest ③ Headrest ⑨ Base plate ④ Backrest ⑩ Foot control ⑤...
Instructions for use KaVo uniQa 3 Description of the product | 3.2 Versions of the dentist element 3.2.1 T-Table ① Tray holder ⑦ ERGOcam One ② Handle ⑧ Locking brake ③ Three function handpiece or multi- ⑨ Tray support functional handpiece ④...
Instructions for use KaVo uniQa 3 Description of the product | 3.2 Versions of the dentist element 3.2.2 TM table ① Handle ⑥ ERGOcam One ② Three function handpiece or multi- ⑦ Tray support functional handpiece ③ Air instrument (MULTIflex coup- ⑧...
Instructions for use KaVo uniQa 3 Description of the product | 3.2 Versions of the dentist element 3.2.3 S-Table ① Handle ⑤ ultrasonic scaler PiezoLED ② Three function handpiece or multi- ⑥ USB interface functional handpiece ③ Air instrument (MULTIflex coup- ⑦...
Instructions for use KaVo uniQa 3 Description of the product | 3.3 Assistant element 3.2.4 Cart ① Handle ⑥ Three function handpiece or multi- functional handpiece ② Touchscreen for display and oper- ⑦ ERGOcam One ation ③ INTRA LUX KL 703 Motor ⑧...
Instructions for use KaVo uniQa 3 Description of the product | 3.4 Three function and multifunctional handpiece (3F and MF handpiece) ① Control element ④ Satelec Mini LED ② Three function handpiece or multi- ⑤ 2nd Saliva ejector (optional ac-...
Instructions for use KaVo uniQa 3 Description of the product | 3.5 Controls 3.5 Controls 3.5.1 Dentist element T-Table/TM-Table/Cart Touchscreen ① Tab "Treatment" ④ Tab "Cleaning" ② Tab "Timer" ⑤ Tab "Settings" ③ Tab "Patient communication" (op- tional) Tab "Treatment"...
Instructions for use KaVo uniQa 3 Description of the product | 3.5 Controls 3.5.2 Dentist element S-table Tab "Treatment" ① Direct keys "Chair functions" ⑥ "Automatic positions" keys ② "Most recent position" and "Rinse ⑦ "Collapsed position" key position" keys ③...
Instructions for use KaVo uniQa 3 Description of the product | 3.5 Controls 3.5.3 Assistant element ① Direct keys "Chair functions" ⑥ "Bowl rinsing" key ② "Most recent position" and "Rinse ⑦ "Tumbler filler" key position" keys ③ "Lock screen" key ⑧...
Instructions for use KaVo uniQa 3 Description of the product | 3.6 Foot control Name "Backrest down" key Direct keys for operating lights (selectable in "Settings" tab on the dentist element) Description Control element "Operating light on/off" key Dentist element and assistant ele- ment "Operating light dimming"...
Instructions for use KaVo uniQa 3 Description of the product | 3.7 Signs on the product Item Name Function with hand- Function with hand- piece mounted piece taken out ⑤ Foot pedal "Select Selection of treatment Starts the motor/hand- treatment mode"...
Page 31
Instructions for use KaVo uniQa 3 Description of the product | 3.7 Signs on the product Manufacturing date CE mark for medical devices Comply with all safety-related information in the accompanying docu- ments, such as warning notes and precautionary measures.
Page 32
Instructions for use KaVo uniQa 3 Description of the product | 3.7 Signs on the product Rating plate on the base of the unit Mounting site for the rating pate on the device base Control element rating plate Mounting site of rating plate and applied part mark on wireless foot control, dentist and assistant element and service table 1568.
Instructions for use KaVo uniQa 3 Description of the product | 3.8 Technical Specifications Labelling and marking of the three-function handpiece and multifunctional handpiece 3.8 Technical Specifications Electrical system Maximum power consumption (at 100 70 to 700 VA V to 240 V)
Instructions for use KaVo uniQa 3 Description of the product | 3.8 Technical Specifications Operating time (as per charge cycle) At least 1 month – The indicated oper- ating time assumes normal handling of the treatment unit and wireless foot control.
Page 35
Instructions for use KaVo uniQa 3 Description of the product | 3.8 Technical Specifications NOTE The "Water block, compact" assembly kit does not include a separation between the treatment water and water supplied by the local mains. The op- erator must observe and comply with relevant national directives concerning the prevention of backflow.
Page 36
Instructions for use KaVo uniQa 3 Description of the product | 3.8 Technical Specifications Water inflow > 4 l/min Diameter of the drain connection 40 mm Above-floor drain connection 20 mm Outflow quantity max. 4 l/min Slope of water drain pipe downstream from device: at least 10...
Instructions for use KaVo uniQa 3 Description of the product | 3.8 Technical Specifications Operating environment WARNING Inappropriate operating conditions. Impairment of the electrical safety of the device. 4 It is essential to comply with the operating conditions specified in the "Tech- nical Specifications"...
Instructions for use KaVo uniQa 3 Description of the product | 3.8 Technical Specifications Pressure dew point < -30 ºC (compressor with dry air system) Oil content < 0.1 mg/m (oil-free compressor) Contamination < 100 particles/m at particle sizes of 1 to 5 µm...
Instructions for use KaVo uniQa 3 Description of the product | 3.9 KaVo Service table 1568 (optional accessory) 3.9 KaVo Service table 1568 (optional accessory) ① Swivel arm ⑥ Non-slip mat ② Spring arm ⑦ Cup holder ③ Rotary knob (brake) ⑧...
Instructions for use KaVo uniQa 4 Operation | 4.1 Turn the device on 4 Operation Please note also the corresponding instruction and ergonomics videos on the KaVo website. Please use the QR codes or follow the corresponding link: Instruction video: https://www.kavo.com/de/einweisung Ergonomics video: https://www.kavo.com/de-de/rueckenfreundliches-arbeiten...
Instructions for use KaVo uniQa 4 Operation | 4.2 Moving the dentist element 4.2 Moving the dentist element CAUTION Risk of injury when the dentist element or assistant element is moved. The patient or office staff may be injured or bruised.
Instructions for use KaVo uniQa 4 Operation | 4.2 Moving the dentist element TM table S-Table 4.2.2 Move the cart CAUTION Moving and overloading the cart. Danger of tipping and damaging the cart. 4 Only use the cart on a continuously smooth floor.
Instructions for use KaVo uniQa 4 Operation | 4.3 Moving the tray NOTE The range in which the cart can be moved is restricted by the length of the lines and hoses that connect the cart to the base of the device. Only move the cart within this range.
Instructions for use KaVo uniQa 4 Operation | 4.4 Moving the assistant element 4.4 Moving the assistant element CAUTION Damage from overloading the assistant element Exceeding the maximum weight of more than 1 kg by adding handpieces, ac- cessories, etc., can cause damage.
Instructions for use KaVo uniQa 4 Operation | 4.5 Moving the patient element (optional) 4 To release the lock, the assistant element must be lifted all the way up. 4.4.1 Attaching the tray holder (optional assembly kit) 4 Mount the tray holder on the assistant element.
Instructions for use KaVo uniQa 4 Operation | 4.6 Adjusting the dental chair 4.6 Adjusting the dental chair Patient chair CAUTION Danger of injury from automatic chair adjustment The automatic adjustment of the chair position can lead to injury. 4 Only use the automatic functions under the supervision of the user!
Instructions for use KaVo uniQa 4 Operation | 4.6 Adjusting the dental chair with protective film/protective covers before use. Select the protective films/ protective covers in compliance with national regulations for medical products. Please comply with the instructions of the manufacturer of the cover.
Page 48
Instructions for use KaVo uniQa 4 Operation | 4.6 Adjusting the dental chair Feature Move to automatic position 2. Move to the collapsed position. 4 Briefly press the respective button. ð Chair automatically moves to the saved position. ð The key is activated as soon as the saved position is reached.
Instructions for use KaVo uniQa 4 Operation | 4.6 Adjusting the dental chair Saving an automatic position with the foot control ① Spray preselection/AP footswitch ③ Foot pedal ② Blown air/AP footswitch Chair positions can be saved on two foot switches; the standard setting is as follows: Foot switch "Spray preselection": Automatic position "LP"...
Instructions for use KaVo uniQa 4 Operation | 4.7 Adjusting the headrest Prerequisites ü All instruments are in their holders. If an instrument is taken out, briefly activate the stirrup switch. 4 Chair up: slide cross switch on foot control towards 1.
Instructions for use KaVo uniQa 4 Operation | 4.8 Safety shutdown 4.7.2 Adjusting the double-jointed headrest knob 4 Push in or pull out the headrest to fit the size of the patient. 4 To swivel the headrest, rotate the knob anticlockwise to the stop. After ad- justment, rotate the knob clockwise to fix it again.
Page 52
Instructions for use KaVo uniQa 4 Operation | 4.8 Safety shutdown Safety shutdowns ① Patient element (optional) pivoted ④ Kickplate/VACUstopp (2 switches) over dental chair (2 switches, right/left) ② Backrest ⑤ Stirrup on foot control ③ Support cover (2 switches) ⑥...
Page 53
Instructions for use KaVo uniQa 4 Operation | 4.8 Safety shutdown Display LED on the Safety shutdown actuated assistant element ③ Support cover ④ VACUstopp kickplate ⑤ Stirrup on foot control ⑥ Bench NOTE The chair's position cannot be changed with the key wheels when a safety shutdown is activated.
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen 4 Move the chair using the buttons of the wheel buttons of the chair. ð A beep is issued each second while the chairs moving. CAUTION Once the safety shutdown is disabled, motors move without monitor- ing.
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Menu Description Settings ▪ Users ▪ Global settings Navigation Icon Feature Description Press re- to open functions or make settings. spective button Selection list Click the "List" key to select an option from a list.
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen 4 Tap the arrow to display status messages. Status messages 4 Tap the status message and follow the request. ð As soon as the error is remedied, this is recognised and the status message disappears from the display.
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Support logic All handpieces on the dentist's side are secured against simultaneous use by holder logic. If an handpiece has been removed when the unit is switched on, the relevant holder will not be activated until the relevant handpiece has been replaced once.
Page 58
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Function Spray air cooling status Spray water cooling NaCl cooling status (optional accessory) "Cooling status spray water" (short press) "Cooling status NaCl" (long press) 4 After you set a single value or all values, tap the "Save" key to save the val- ues.
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Setting the cooling status Function No cooling Spray air cooling status Spray water cooling NaCl cooling status (optional accessory) "Cooling status spray water" (short press) "Cooling status NaCl" (long press) 4 After you set a single value or all values, tap the "Save"...
Page 60
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Define operating mode NOTE The selection of the mode depends on the treatment method and the tip used. For information about the selection of an operating mode, please refer to the "Operating modes P1 / P2 / P3 and ENDO"...
Page 61
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Settings for the multifunctional handpiece The following settings can be changed in the Treatment menu on the touch- screen: ▪ Air/water heating 4 Take the multifunctional handpiece off the holder.
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Function ▪ Laser mode on: Key is active ▪ Laser mode off: Key is inactive NOTE For more information on the individual modes, please refer to the Operating light chapter.
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen ð This causes the "Timer" menu to be displayed. 4 Tap the key next to Timer to edit a previously generated timer. 4.9.5 Patient communication menu KaVo CONNECTbase Consult instructions for use KaVo CONNECTbase.
Page 64
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Using the CONEXIO menu CONEXIO consists of 3 different levels: ▪ Level 1 - Patient list: display of all available patients ▪ Level 2 - Patient file: display of the available images of the patient ▪...
Page 65
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Item Name Pressing but- Function ① Stirrup switch long Deselect selected patients. ② "LP/preselected briefly Move cursor to previous patient spray" footswitch ③ "SP/blown air" foot...
Page 66
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Symbol Pressing but- Function briefly Return to patient file briefly Displays the selected patient briefly To image display briefly To filter selection Display Note: Filter is active...
Page 67
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen CONEXIO Filter selection Tooth filter Time filter 67 / 172...
Page 68
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Source filter Symbol Pressing but- Function briefly Back to patient file briefly Reset filter briefly Filter option: By tooth briefly Filter option: By time briefly Filter option: By type and source...
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Symbol Pressing but- Function briefly Tooth (missing - Setting via "Clinical monitoring status") Time filter Symbol Pressing but- Function briefly Show all images briefly Show images from last session...
Page 70
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Symbol Pressing but- Function briefly Back to patient file Dropdown Selection of display mode Single | Compare | Split briefly Image preview for stock images, Changing the image with arrow keys briefly Saving changes in image (e.g.
Page 71
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Symbol Pressing but- Function ▪ Kritisch: Critical Tooth is shown in red ▪ Erledigt: Done briefly Tooth selection/deselection by clicking briefly Display of clinical monitoring status ▪...
Page 72
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Symbol Pressing but- Function briefly Back to patient file Dropdown Selection of display mode Single | Compare | Split briefly The live camera image is always shown in the left frame.
Page 73
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Symbol Pressing but- Function briefly Open/close Full Screen briefly Toggle between list view and tile view Split image display In this mode, up to 6 images from different sources (cameras, microscopes, X- ray) can be displayed.
Page 74
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Symbol Pressing but- Function briefly Enlarging and reducing image detail, image navigation by moving the yellow frame briefly Setting contrast and brightness with slider. To close the menu, tap the "Contrast Brightness"...
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Item Name Pressing but- Function ④ - ⑦ "Chair position/ briefly Open/close Full Screen motor rotational direc- tion" cross switch ⑧ "Level preselection/ briefly Selection key handpieces"...
Page 76
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Function Once the rinsing position (SP) is reached, the bowl rinsing is activated, i.e. the bowl is being wetted. Leaving the rinsing position (SP), the bowl is rinsed again. Selectable for "Home screen" in "Settings" tab.
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Setting the rinsing time 4 Actuate the "Bowl rinsing time" key until the desired rinsing time is achieved and the following view is shown. 4 Confirm the bowl rinsing time by saving once the desired rinsing time is reached.
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen User settings The following items can be selected in the "User" menu: ▪ Users ▪ Types of treatment ▪ Language ▪ Light ▪ Home screen Users 4 Tap the "+"...
Page 79
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Treatment mode 4 Tap the "Treatment mode" key to edit the types of treatment. Types of treatment 4 Tap the key of the type of treatment to be changed, e.g. "Excavation", and rename the type of treatment.
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Select a language Light 4 Tap the "Light" key to edit the lamp settings. Set light Home screen 4 Tap the "Home screen" key to configure the home screen with up to six dir- ect keys.
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Select direct keys Global settings The following items can be selected in the "Global settings" menu: ▪ Display ▪ Sounds and volume ▪ Date and time ▪...
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Global settings / Display Sounds and volume 4 Tap the "Sounds and volume" key to set the key sounds and the volume. General settings / Sounds and volume Date and time 4 Tap the "Date and time"...
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen General settings / Date and time Advanced settings The following items can be set in the "Advanced settings" menu: ▪ Status after the unit is switched on ▪...
Page 84
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Instruments 4 Tap the "Instruments" key to make the following settings: - Instrument light and spray temperature - Spray water for ultrasonic scaler - Water for three function / multifunctional handpiece...
Page 85
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Advanced settings / Handpieces Tumbler & bowl 4 Tap the "Tumbler & bowl" key to make the following settings: - Automatic bowl rinsing - Tumbler filling and tumbler sensor...
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Advanced settings / Tumbler and bowl Boiler temperature 4 Tap the "Boiler temperature" key to set the boiler temperature. Advanced settings / Boiler temperature Foot control 4 Tap the "Foot control"...
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Advanced settings / Foot control Operating light 4 Tap the "Operating light" key to set the automatic activation of the operat- ing light. Advanced settings / Operating light Weekly cleaning 4 Tap the "Weekly cleaning"...
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Advanced settings / Weekly cleaning Advanced settings / Weekly cleaning Suction system 4 Tap the "Suction system" key to set VACUstopp. 88 / 172...
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Advanced settings / Suction system Information The following items can be selected in the "Information" menu: ▪ Device information ▪ System check ▪ Specialised dealers Global settings / Information Device information 4 Tap the "Device information"...
Instructions for use KaVo uniQa 4 Operation | 4.9 Using functions through the touchscreen Global settings / Device information System check 4 Tap the "System check" key to open the system check. Global settings / System check Specialised dealers 4 Tap the "Specialised dealer" key to display specialised dealers.
Instructions for use KaVo uniQa 4 Operation | 4.10 Using functions through the control panel of the assistant element Global settings / Specialised dealer 4.10 Using functions through the control panel of the assistant element 4.10.1 Using the chair functions...
Instructions for use KaVo uniQa 4 Operation | 4.10 Using functions through the control panel of the assistant element Dentist element key Assistant element key Feature The chair moves up. The chair moves down. The backrest moves up- ward. The backrest moves downward.
Instructions for use KaVo uniQa 4 Operation | 4.10 Using functions through the control panel of the assistant element Select the timer time 4 To start a timer time, e.g. Timer 1, press the "Timer 1" key. ð Time on the timer starts to run. A beep is issued after the timer time is elapsed.
Instructions for use KaVo uniQa 4 Operation | 4.10 Using functions through the control panel of the assistant element 4 Open the settings menu by pressing the symbol. ð This causes the settings menu of the multifunctional handpiece to be displayed.
Instructions for use KaVo uniQa 4 Operation | 4.11 Using the three function and multifunctional handpiece ① Chair kickplate In the delivery state, only the spray mist suction stops in the VACUstopp func- tion. 4.11 Using the three function and multifunctional...
Page 96
Instructions for use KaVo uniQa 4 Operation | 4.11 Using the three function and multifunctional handpiece CAUTION Cannula that is worn or not locked into place. Injury from swallowing the cannula. 4 Before each treatment, ensure that the cannula is locked into place and firmly seated.
Page 97
Instructions for use KaVo uniQa 4 Operation | 4.11 Using the three function and multifunctional handpiece NOTE If only the cold light is preselected (heater: off), the multifunctional handpiece shines when it is removed from the holder. 4 Remove the handpiece from the holder.
Instructions for use KaVo uniQa 4 Operation | 4.11 Using the three function and multifunctional handpiece 4 Press the "Handpieces" foot pedal. ð The light turns on. 4.11.1 Replacing the lamp (multifunctional handpiece only) CAUTION Danger of injury if the valve body is hot (heating elements, high-pres- sure lamp).
Instructions for use KaVo uniQa 4 Operation | 4.12 Using PiezoLED 4 Push the holder ③ forward and pull the defective LED lamp ② out of the mount. 4 Insert a new LED lamp (mat. no. 1.007.5372). The following may happen after you turn on the LED lamp: ▪...
Instructions for use KaVo uniQa 4 Operation | 4.13 Operating the foot switch 4.13 Operating the foot switch General functions Cable-connected foot control and wireless foot control Item Name Function with hand- Function with hand- piece mounted piece taken out ①...
Instructions for use KaVo uniQa 4 Operation | 4.13 Operating the foot switch 4 Press the "SP" foot switch. ð The chair moves into the saved position. 4 Press the "LP" foot-operated button. ð The chair moves into the saved position.
Instructions for use KaVo uniQa 4 Operation | 4.13 Operating the foot switch 4 With the handpieces in the holders, press the foot pedal. ð The treatment mode advances upon each actuation of the foot pedal. 4.13.3 Preselecting dentist 4 Hold down the foot pedal and press the stirrup switch (with handpieces stored).
Instructions for use KaVo uniQa 4 Operation | 4.13 Operating the foot switch 4 Press "Preselected spray" footswitch. ð The cooling level is switched each time the foot switch is pressed: no cooling - air - spray. ð The cooling level is displayed on the dentist and assistant element.
Instructions for use KaVo uniQa 4 Operation | 4.13 Operating the foot switch 4.13.10 Special functions of the wireless foot control CAUTION Electrical power Personal injury or damage to the wireless foot control. 4 The user must never touch the charger connector and the patient at the...
Page 105
Instructions for use KaVo uniQa 4 Operation | 4.13 Operating the foot switch Residual Foot control Status display / charge status Beep capacity status display Active actu- Flashes yellow ation (approx. 200 millisecond intervals) < 10 % Idle state Flashes yellow...
Page 106
Instructions for use KaVo uniQa 4 Operation | 4.13 Operating the foot switch 4 To couple the foot control and the treatment centre, actuate the "Couple wireless foot control" button. ð The following window appears: 4 For coupling, press and hold down the stirrup switch button, the cross switch up button and the foot pedal until you hear a beep.
Page 107
Instructions for use KaVo uniQa 4 Operation | 4.13 Operating the foot switch NOTE Only one wireless foot control per treatment centre can be registered to a RF receiver at any given time. If another wireless foot control was previously re- gistered, the last wireless foot control to be registered will be deleted with every new start of the synchronisation process.
Page 108
Instructions for use KaVo uniQa 4 Operation | 4.13 Operating the foot switch NOTE If the wireless foot control is switched off, the display of the receiver also shows "Disconnected". 4 Use the "Up" or "Down" keys to select the "Pairing" menu item, and activate it with the "OK"...
Instructions for use KaVo uniQa 4 Operation | 4.14 Using the operating light 4.13.11 Charging the wireless foot control CAUTION Risk of injury and material damage from incorrect use of the charger for the wireless foot control. Personal injury or damage to the wireless foot control or charger.
Instructions for use KaVo uniQa 4 Operation | 4.14 Using the operating light CAUTION Premature hardening of composite fillings. A light intensity that is too high can have a negative impact on the durability of the treatment. 4 Select the appropriate dimming level according to the processing time.
Instructions for use KaVo uniQa 4 Operation | 4.15 Using the KL 703 LED in ENDO mode (optional accessory) Set the brightness and type of dimming 4 Tap the "Dim mode" key to toggle between COMPOsave and Dim mode. ð The active mode is indicated by the blue background.
Page 112
Instructions for use KaVo uniQa 4 Operation | 4.15 Using the KL 703 LED in ENDO mode (optional accessory) CAUTION Use of impermissible file systems. Impermissible file systems may damage the product or cause personal injury. 4 Only use approved NiTi file systems with a conicity >2% that are suitable for rotary reprocessing.
Instructions for use KaVo uniQa 4 Operation | 4.15 Using the KL 703 LED in ENDO mode (optional accessory) NOTE The INTRAINTRA LUX KL 703 LED motor has a torque range from 0.15 to 3 Ncm. If the torque is in excess of 2.0 Ncm, KaVo recommends using a redu- cing contra-angle handpiece 3:1 or 8:1 in order to reduce the load on and heating of the motor.
Instructions for use KaVo uniQa 4 Operation | 4.15 Using the KL 703 LED in ENDO mode (optional accessory) 4 Unfold the "Type of treatment" selection list and select "Endodontics". ð The display switches to "Endodontics" as the treatment mode.
Instructions for use KaVo uniQa 4 Operation | 4.15 Using the KL 703 LED in ENDO mode (optional accessory) 4.15.4 File database Endodontics mode (optional accessories) has a file database integrated in it. It is necessary to check if the file data are up-to-date by comparing them to the respective manufacturer information.
Page 116
Instructions for use KaVo uniQa 4 Operation | 4.15 Using the KL 703 LED in ENDO mode (optional accessory) 4 Tap the "Pen" control element to edit the sequence. ð The display shows file sequence "Profile 1" in editing mode.
Page 117
Instructions for use KaVo uniQa 4 Operation | 4.15 Using the KL 703 LED in ENDO mode (optional accessory) 4 Edit "Profile 1" with the keyboard and tap the "Save" key to save the value. 4 Tap the second column called "Type" and use the arrow keys to select the file system (here: HyFlex EDM).
Page 118
Instructions for use KaVo uniQa 4 Operation | 4.15 Using the KL 703 LED in ENDO mode (optional accessory) ð Once the file system and file geometry have been selected, the columns "Torque", "Speed" and "Colour bar" (colour of the file) are assigned auto- matically except for the user-defined files.
Page 119
Instructions for use KaVo uniQa 4 Operation | 4.15 Using the KL 703 LED in ENDO mode (optional accessory) 4 In order to define user-defined files, tap on the column named "Type" and use the arrow keys to jump to "FILE" and make the selection 4 All other data must be set manually by the user according to the informa- tion provided by the file manufacturer.
Instructions for use KaVo uniQa 4 Operation | 4.15 Using the KL 703 LED in ENDO mode (optional accessory) Changing parameters short-term 4 Unfold the "Profile" list and select a profile as desired. 4 Use the "slider" to set speed and torque.
Instructions for use KaVo uniQa 4 Operation | 4.15 Using the KL 703 LED in ENDO mode (optional accessory) If the set torque differs from the recommended value of a file, a greyed-out factory symbol is shown. Use the "Save" key for saving in profiles 1 to 8. You can save after setting each parameter individually or after setting all parameters.
Instructions for use KaVo uniQa 4 Operation | 4.16 Using the SL600 surgical motor (optional accessory) Torque mode Autorev / Forward 4 Press the foot pedal. ð The motor starts by rotating clockwise (unless selected otherwise). When the set torque is reached, the motor switches to counterclockwise rotation.
Instructions for use KaVo uniQa 4 Operation | 4.16 Using the SL600 surgical motor (optional accessory) Operating mode NOTE The maximal motor load is 30 seconds operating time / 9 minutes pause (full load at maximal speed). 4.16.1 Connecting and operating the pump for physiological saline The surgical motor set includes the "Pump for physiological saline solution"...
Instructions for use KaVo uniQa 4 Operation | 4.16 Using the SL600 surgical motor (optional accessory) 4 Unfold the "Treatment mode" selection list and select "Surgery". 4.16.4 Mounting or pulling off the handpiece or contra- angle handpiece NOTE Follow the instructions for use, service instructions and installation instruc- tions in the instrument packaging.
Instructions for use KaVo uniQa 4 Operation | 4.16 Using the SL600 surgical motor (optional accessory) Remove the straight or contra-angle handpiece 4 Pull the coolant hose off the straight or contra-angle handpiece. 4 Pull the instrument off the motor in axial direction.
Instructions for use KaVo uniQa 4 Operation | 4.16 Using the SL600 surgical motor (optional accessory) device is the same as the current treatment step. Operating errors can thus be largely prevented. Default values have been set at the factory for the parameters, speeds, torques, transmission ratios and coolant flow rate for every activity according to application.
Page 127
Instructions for use KaVo uniQa 4 Operation | 4.16 Using the SL600 surgical motor (optional accessory) Recommended programming when placing multiple implants consecutively NOTE Default values for the respective transmission ratios can be selected with the sliders. Activity Step Icon...
Instructions for use KaVo uniQa 4 Operation | 4.16 Using the SL600 surgical motor (optional accessory) Activity Step Icon Speed Torque Transmis- [rpm] [Ncm] sion ratio Placing im- 15 – 50 5 – 80 16:1 plant 20 (D) 25 (D)
Page 129
Instructions for use KaVo uniQa 4 Operation | 4.16 Using the SL600 surgical motor (optional accessory) ▪ Transmission ratio ▪ Direction of motor rotation ▪ Activity Changing and saving parameters 4 Use the "slider" to set speed and torque. 4 Tap the "NaCl" key to select the coolant flow rate.
Instructions for use KaVo uniQa 4 Operation | 4.16 Using the SL600 surgical motor (optional accessory) Setting activity 4 Tap the activity symbol to change the activity of the respective step (toggle function). 4.16.8 Setting the operating light (LUX) 4 Push the cross-switch to the right to activate the light (without the motor and pump running).
Instructions for use KaVo uniQa 4 Operation | 4.17 Using the pump for physiological saline (optional accessory) NOTE Auto-calibration should be carried out only with KaVo surgical handpieces with a transmission ratio of 16:1, 20:1 or 27:1. The auto-calibration cannot be run with third-party handpieces or handpieces with different transmission ratios.
Page 132
Instructions for use KaVo uniQa 4 Operation | 4.17 Using the pump for physiological saline (optional accessory) CAUTION Coolant container made of glass. Danger of injury due to falling or shattering glass coolant container. 4 Do not use glass bottles as coolant containers on the dentist part of the dental unit.
Instructions for use KaVo uniQa 4 Operation | 4.17 Using the pump for physiological saline (optional accessory) 4.17.1 Connecting the coolant NOTE The liquid-conveying parts are not sterile! They must be sterilised before the first treatment. All parts conveying liquids must be kept sterile.
Instructions for use KaVo uniQa 4 Operation | 4.17 Using the pump for physiological saline (optional accessory) Connecting the coolant container and hose set NOTE The coolant must be selected to suit the planned application. The flow rate of the coolant is dependent on the instrument used. The user must set an ad- equate flow of coolant and check this.
Instructions for use KaVo uniQa 4 Operation | 4.17 Using the pump for physiological saline (optional accessory) NOTE Make sure to place the pump hose in the pump appropriately such that the pump hose does not get clamped or pinched by the lock. Route all hoses re- laxed and without tension.
Instructions for use KaVo uniQa 4 Operation | 4.17 Using the pump for physiological saline (optional accessory) 4 Move the cross switch down for four seconds until the signal sounds to turn off the saline pump. 4.17.3 Changing the coolant container The coolant container can be changed as follows: 4 Close the hose clamp.
Instructions for use KaVo uniQa 4 Operation | 4.18 Using the USB interface 4.18 Using the USB interface CAUTION Electrical power. Electrical shock. 4 Set up the external PC outside of the patient environment keeping a min- imum distance of 1.5 m.
Instructions for use KaVo uniQa 4 Operation | 4.20 Service table 1568 (optional accessory) 4.20 Service table 1568 (optional accessory) CAUTION Over-travel beyond swivel range. Material damage. 4 Comply with rotary knob swivel range of 180°. CAUTION Exceeding the load limits.
Instructions for use KaVo uniQa 4 Operation | 4.21 Third-party device connector Dimensions and swing ranges (in mm) 4.21 Third-party device connector The third-party device connector allows dental devices from third-party manu- facturers to be used on the KaVo treatment centre.
Instructions for use KaVo uniQa 4 Operation | 4.22 Using the water bottle 4.22 Using the water bottle CAUTION Insufficient cooling due to lack of patient water. A lack of water or the water bottle not being inserted can lead to personal in- jury.
Instructions for use KaVo uniQa 5 Reprocessing methods DIN EN ISO 17664 5 Reprocessing methods DIN EN ISO 17664 NOTE The reprocessing methods are specified in the Care instructions. Please note also the corresponding hygiene video on the KaVo website. Please use the QR code or the following link: https://www.kavo.com/hygiene-easy-way...
Instructions for use KaVo uniQa 6 Additional equipment and assembly kits | 6.1 Equipment 6 Additional equipment and assembly kits General information concerning the connection of additional equipment: The following products can be combined with the KaVo treatment centre. Moreover, the KaVo treatment centres, in part, provide standardised interfaces.
Instructions for use KaVo uniQa 6 Additional equipment and assembly kits | 6.3 Dentist element Name Description Satelec Mini LED The assistant element can be equipped with Sate- lec Mini LED. 6.3 Dentist element Description Description Three function and multi-...
Page 144
Instructions for use KaVo uniQa 6 Additional equipment and assembly kits | 6.4 Patient chair NOTE Any attachment of products without the approval of KaVo or in the absence of a standardised interface is prohibited. If this is done regardless, the CE con- formity, and thus the proof of the conformity of the product, may be voided.
NOTE If several medical electrical devices (ME device) or electrical devices from sev- eral manufacturers combined into a system are connected to the KaVo dental unit, the manufacturer data contained in the instructions for use for all products subject to the safety checks must also be noted.
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.1 Introduction NOTE The following tests and measurements must be documented, for example in the medical device book. We recommend using the templates at the end of the document.
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.1 Introduction Measurements ▪ Measurement of the protective earth resistance in accordance with IEC 62353 ▪ Measurement of the equipment leakage current - Alternative measuring method in accordance with IEC 62353 ▪...
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.2 Instructions for the safety check 7.2 Instructions for the safety check 7.2.1 Preparatory measures to be undertaken on the device WARNING Electrical power. Death or injury from electric shock.
Page 149
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.2 Instructions for the safety check ▪ Stability of the device ▪ Mechanical integrity of the safety switches. ▪ Absence of damage to the cladding or casing (cracks, breakage) ▪...
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.2 Instructions for the safety check Control of the availability of the necessary documents 4 Check if the required instructions for use and care instructions are available in the surgery.
Page 151
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.2 Instructions for the safety check NOTE Connection cables such as data cables and cables for the functional earth could simulate protective conductor connections. These types of supplement- ary but unintentional protective earth connections could lead to erroneous measurements.
Page 152
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.2 Instructions for the safety check Connecting the safety tester without the KaVo measuring cable to the treatment centre. 4 Switch L + N of the on-site power supply cord to be voltage-free.
Page 153
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.2 Instructions for the safety check 4 Connect TP2 measuring socket ② to the safety tester. NOTE If a light polymerisation unit is installed on the assistant element, the light polymerisation unit must be connected to the safety tester as an applied part.
Page 154
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.2 Instructions for the safety check ACPs on the treatment centre No ACPs need to be connected to the protective conductor (PE) during the measurement on the treatment centre, as all relevant parts are connected to the protective conductor (PE) at the factory and are included in the test.
Page 155
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.2 Instructions for the safety check Scanning the treatment centre with the test tip Scan the device body and chair at least at the following spots: ① PE terminal ④...
Page 156
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.2 Instructions for the safety check ① Dentist element T: Fixing screw on ③ Dentist element Cart: table bot- the underside of the dentist ele- ment ② Dentist element S: Fixing screw ④...
Page 157
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.2 Instructions for the safety check Scanning the operating light with the test tip ① Fastening screw of the handle support when the gripping sleeve has been removed Scanning the monitor with the test tip ①...
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.2 Instructions for the safety check WARNING Danger to persons due to a lack of care exercised during the safety checks and testing. Death or injury from electric shock.
Page 159
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.2 Instructions for the safety check - Test of the limit switches ▪ Functional test ... Safety shutdowns ① Patient element (optional) pivoted ④ Kickplate/VACUstopp (2 switches) over dental chair ②...
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.2 Instructions for the safety check 7.2.5 Assessment and documentation NOTE All tests conducted must be documented comprehensively. The docu- ments must contain at least the following particulars:...
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.3 Test protocol for the safety check 7.3 Test protocol for the safety check Test Protocol - Safety Check (Test of safety and function according to §7 section 3 Medical Device Operation Ordinance)
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.4 Annex - Additional measuring points 7.4.1 Additional scan points SL X for Module ERGOcam One ① Screw in bottom part of housing NOTE The modules are not earthed with a safety conductor. In the case of excessive PE resistance, the electrical connection between the module and the dentist element must be improved.
Instructions for use KaVo uniQa 7 Safety check - Test instructions | 7.4 Annex - Additional measuring points Ceiling adapter for operating light assembly kit ① Base plate for the ceiling adapter ② Surroundings of the protective conductor connector ③ Surroundings of the protective connector terminal 7.4.2 Additional measuring points AP X for the EGA/EPA...
Instructions for use KaVo uniQa 8 Troubleshooting 8 Troubleshooting NOTE In case of malfunction, consult the separate instructions for the use and care of the respective instruments (e.g., turbine, motor, camera, etc.). Malfunction Cause Remedy Nothing works. Main switch is off.
Page 165
Instructions for use KaVo uniQa 8 Troubleshooting Malfunction Cause Remedy The suction hoses do not have Slides on the conical sec- 4 Open the slide valve. any suction. tions are closed. 4 Replace sieves. Sieves in suction con- nector are blocked.
Instructions for use KaVo uniQa 9 Information concerning the electromagnetic compatibility according to EN IEC 60601-1-2 | 9.1 Operating environ- ment and EMC warning notes 9 Information concerning the electromagnetic compatibility according to EN IEC 60601-1-2 9.1 Operating environment and EMC warning notes This product is neither life-sustaining nor coupled to the patient.
Page 167
Instructions for use KaVo uniQa 9 Information concerning the electromagnetic compatibility according to EN IEC 60601-1-2 | 9.2 Results of the electromagnetic tests Requirement Class / test level Radiated emitted interference Class B [30 MHz – 1000 MHz] DIN EN 61000-3-2 VDE 0838-2 / 03.2010...
Page 168
Instructions for use KaVo uniQa 9 Information concerning the electromagnetic compatibility according to EN IEC 60601-1-2 | 9.2 Results of the electromagnetic tests Requirement Class / test level 6 V in ISM bands 6 V in amateur radio bands DIN EN 61000-4-8 VDE 0847-4-8 / 11.2010...