Instructions for use EXPERTmatic E10 C – 1.007.5560 Table of contents Table of contents 1 User instructions....................... 2 Safety ..........................2.1 Infection hazard......................7 2.2 Improper use ....................... 7 2.3 Technical condition......................7 2.4 Accessories and combination with other equipment ............8 2.5 Qualification of personnel....................
Page 5
Instructions for use EXPERTmatic E10 C – 1.007.5560 1 User instructions Serial number UDI symbol Manufacturer Attention: Please consult the accompanying documents Follow the electronic instructions for use HIBC Code CE mark for medical devices EAC conformity mark (Eurasian Conformity)
Page 6
Instructions for use EXPERTmatic E10 C – 1.007.5560 1 User instructions CAUTION In cases which – if not prevented – can lead to minor or moderate injury. NOTICE In cases which – if not prevented – can lead to property damage.
Instructions for use EXPERTmatic E10 C – 1.007.5560 2 Safety | 2.1 Infection hazard 2 Safety NOTE All serious events occurring in relation to the product must be reported to the manufacturer and the competent authority of the member state, in which the user and/or patient resides.
Instructions for use EXPERTmatic E10 C – 1.007.5560 2 Safety | 2.4 Accessories and combination with other equipment ▪ Excessive vibration ▪ Overheating ▪ Bur is not seated firmly in the handpiece To ensure optimum function and to prevent property damage, please comply...
Page 9
Instructions for use EXPERTmatic E10 C – 1.007.5560 2 Safety | 2.6 Service and repair As a result of the use of NON-KaVo original spare parts during the repair, parts such as covers may become undone and injure the patient, user or third par- ties.
Instructions for use EXPERTmatic E10 C – 1.007.5560 3 Product description | 3.1 Intended use 3 Product description EXPERTmatic Handpiece E10 C (Mat. No. 1.007.5560) The EXPERTmatic electrical-driven handpieces are dental handpieces in accor- dance with 21 CFR § 872.4200 (dental handpieces and accessories) for the use by a trained professional in the field of general dentistry.
Instructions for use EXPERTmatic E10 C – 1.007.5560 3 Product description | 3.2 Technical Specifications E10 C (without light) ▪ Comply with the specified intended use. ▪ Protect himself or herself, the patient and third parties from hazards. ▪ Prevent contamination by the product.
Instructions for use EXPERTmatic E10 C – 1.007.5560 4 Startup and shut-down | 4.1 Checking the water quantity 4 Startup and shut-down WARNING Hazard from contaminated products. Patients, users or third parties can be infected by contaminated medical de- vices.
Page 13
Instructions for use EXPERTmatic E10 C – 1.007.5560 4 Startup and shut-down | 4.1 Checking the water quantity 13 / 28...
Instructions for use EXPERTmatic E10 C – 1.007.5560 5 Operation | 5.1 Attaching the medical device 5 Operation WARNING Rotating bur. Cuts, infection and burn injury. 4 Do not touch the bur while it is rotating. 4 Remove the bur from the handpiece after treatment to avoid injury and in- fection during storage.
Instructions for use EXPERTmatic E10 C – 1.007.5560 5 Operation | 5.3 Inserting the bur 5.3 Inserting the bur NOTE Only use carbide burs or diamond grinders that comply with ISO 1797 type 1 and type 2, are made of steel or carbide and meet the following criteria: - Shaft diameter: 2.334 to 2.350 mm...
Instructions for use EXPERTmatic E10 C – 1.007.5560 5 Operation | 5.4 Removing the bur NOTICE Bur with damaged, worn or deformed shafts. Property damage to the chuck system, bur is difficult or impossible to remove from the chuck system.
Instructions for use EXPERTmatic E10 C – 1.007.5560 6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions 6 Checking for malfunctions and troubleshooting 6.1 Checking for malfunctions CAUTION Overheating of the device. Burn injury or product damage due to over-heating.
Instructions for use EXPERTmatic E10 C – 1.007.5560 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting 4 Spray a lint-free cloth with KaVo Spray, apply it to the new O-rings and in- sert them in the recesses. 6.2.2 Cleaning the spray nozzle CAUTION Hazard from insufficient spray water flow.
Instructions for use EXPERTmatic E10 C – 1.007.5560 7 Processing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.1 Preparations at the site of use 7 Processing steps in accordance with ISO 17664-1 / ISO 17664-2 7.1 Preparations at the site of use WARNING Hazard from contaminated products.
Instructions for use EXPERTmatic E10 C – 1.007.5560 7 Processing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.2 Manual preprocessing 7.2.2 Manual internal cleaning This product is not designed for manual internal cleaning. For effective processing, the inside of the device must be subjected to auto- mated cleaning in a cleaning and disinfection unit in accordance with ISO 15883-1.
Instructions for use EXPERTmatic E10 C – 1.007.5560 7 Processing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.3 Automated preprocessing 7.3 Automated preprocessing WARNING Sharp bur in the medical device. Injury hazard from sharp and/or pointed bur.
Instructions for use EXPERTmatic E10 C – 1.007.5560 7 Processing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.4 Care products and systems - Servicing CAUTION Improper service and care. Risk of injury. 4 Perform regular proper care and servicing.
Instructions for use EXPERTmatic E10 C – 1.007.5560 7 Processing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.5 Packaging 4 Press the handpiece together with the guide bush of the chucking system to be serviced against the tip of the chucking system service coupling.
Instructions for use EXPERTmatic E10 C – 1.007.5560 7 Processing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.7 Storage – 10 minutes at 135 °C (275 °F) Drying time: 30 min. – 15 minutes at 132 C (270 Drying time: 15-30 min.
Instructions for use EXPERTmatic E10 C – 1.007.5560 8 Optional aids and consumables 8 Optional aids and consumables Available from dental suppliers. Material summary Mat. no. INTRA Instrument stand 3.005.5204 Cleanpac 10 units 0.411.9691 Cellulose pad 100 units 0.411.9862 Nozzle needle 0.410.0921...
Instructions for use EXPERTmatic E10 C – 1.007.5560 9 Terms and conditions of warranty 9 Terms and conditions of warranty This KaVo medical device is subject to the following warranty conditions: KaVo grants the end customer a warranty of proper function and guarantees...