Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 Table of contents Table of contents 1 User instructions ........................ 4 2 Safety..........................6 2.1 Infection hazard ......................6 2.2 Technical condition......................6 2.3 Accessories and combination with other equipment ............6 2.4 Qualification of personnel....................
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 1 User instructions 1 User instructions Dear User Congratulations on purchasing this KaVo quality product. By following the in- structions below you will be able to work smoothly, economically and safely.
Page 5
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 1 User instructions Note: Please note accompanying documents EAC conformity mark (Eurasian Conformity) GOST R certification Transportation and storage conditions (Temperature range) Transportation and storage conditions (Air pressure) Transportation and storage conditions...
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 2 Safety | 2.1 Infection hazard 2 Safety The instructions for use are a component of the product and must be read care- fully prior to use and be accessible at all times.
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 2 Safety | 2.4 Qualification of personnel 2.4 Qualification of personnel Application of the product by users without the appropriate medical training could injure the patients, the users or third parties.
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 3 Product description | 3.1 Intended use 3 Product description MULTIflex LUX Midwest coupling (Mat. no. 0.553.1390) 3.1 Intended use Indications for use: The MULTIflex LUX Midwest coupling is an accessory of the medical device de- signed for coupling the supply hose to dental turbines and dental instruments featuring a Midwest connector.
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 3 Product description | 3.3 Transportation and storage conditions 3.3 Transportation and storage conditions NOTICE Startup after refrigerated storage. Malfunction. ▶ Prior to startup, strongly refrigerated products must be allowed to warm up to a temperature of 20 °C to 25 °C (68 °F to 77 °F).
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 4 First use 4 First use WARNING Hazard from contaminated products. Infection hazard for care provider and patient. ▶ Prior to initial commissioning and after each use, reprocess the product and accessories.
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 5 Operation | 5.1 Connecting the coupling to the hose 5 Operation Note At the beginning of each workday, the water-conducting systems should be rinsed for at least 2 minutes (without transmission handpieces being at- tached) and if there is a risk of contamination from reflux or back suction, the system may also need to be rinsed for 20 to 30 seconds after each patient.
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 5 Operation | 5.4 Removing instruments 5.4 Removing instruments ▶ Hold the coupling tight, and pull the instrument forward while twisting slightly. 12 / 20...
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 6 Troubleshooting | 6.1 Checking for malfunctions before initial start-up 6 Troubleshooting 6.1 Checking for malfunctions before initial start-up NOTICE The coupling system is fitted with light. Defects on the light source.
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use 7 Reprocessing steps in accordance with ISO 17664 7.1 Preparations at the site of use WARNING Hazard from contaminated products.
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 7 Reprocessing steps in accordance with ISO 17664 | 7.3 Automated reprocessing KaVo recommends the following products based on material compatibility. The microbiological efficacy must be ensured by the disinfectant manufacturer.
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 7 Reprocessing steps in accordance with ISO 17664 | 7.4 Sterilisation 7.4 Sterilisation Non-sterilisable. 7.5 Storage ▶ Reprocessed products should be stored protected from dust with minimum exposure to germs in a dry, dark and cool place.
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 8 Tools 8 Tools Available from dental suppliers. Material summary Mat.No. 0.411.1563 Spare washer 0.553.4762 O-ring, large 1.004.2776 O-ring, small 1.004.2775 17 / 20...
Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390 9 Warranty terms and conditions 9 Warranty terms and conditions The following warranty conditions apply to this KaVo product: KaVo provides the end customer with a warranty of proper function and guar-...