Care instructions KaVo SYSTEM 1056 S/T 1 User notes 1.1 User guidelines 1 User notes 1.1 User guidelines Requirement Read these instructions before the initial startup to prevent misuse and damage. 1.1.1 Abbreviations Short Explanation form Instructions for use Care instructions...
Care instructions KaVo SYSTEM 1056 S/T 2 Preparation methods 2.1 Cleaning 2 Preparation methods Damage to product as a result of improper cleaning. Malfunctions, unclean surfaces and risk of infection for operator and patients. ▶ When using detergents, adhere to the manufacturer's instructions.
Page 6
Care instructions KaVo SYSTEM 1056 S/T 2 Preparation methods 2.1 Cleaning ▶ Clean all exterior surfaces with a soft cloth and a mild, non-abrasive cleaning fluid. Unit body ▶ Clean all exterior surfaces with a soft cloth and a mild, non-abrasive cleaning fluid.
Page 7
Care instructions KaVo SYSTEM 1056 S/T 2 Preparation methods 2.1 Cleaning ▶ Repeat procedure for the saliva ejector and spray mist aspiration hose. DEKASEPTOL gel basic set DEKASEPTOL gel refill canister Mat. no. 1.000.7204 Mat. no. 1.000.7205 Spittoon ▶ Add one dose of DEKASEPTOL gel to the spittoon.
Care instructions KaVo SYSTEM 1056 S/T 2 Preparation methods 2.1 Cleaning ▶ Disperse the DEKASEPTOL gel using a brush and leave to take effect. ▶ When the requisite action time has passed, press the spittoon button. The spittoon and drain are thoroughly rinsed.
Care instructions KaVo SYSTEM 1056 S/T 2 Preparation methods 2.3 Sterilisation 2.3 Sterilisation Product damage Contact corrosion ▶ Remove the sterilised item from the autoclave immediately after sterilising. CAUTION Improper sterilisation Damage to the sterilised product. Risk of infection for operator and patients.
Page 11
Care instructions KaVo SYSTEM 1056 S/T 2 Preparation methods 2.3 Sterilisation ▶ Detach cannula from 3F handpiece at an angle. ▶ Unscrew the 3F handpiece cannula in an anti-clockwise direction and in a straight line. ▶ Sterilising the probes in an autoclave:...
Page 12
Care instructions KaVo SYSTEM 1056 S/T 2 Preparation methods 2.3 Sterilisation Method Duration/Temperature Autoclave three times with an initial va‐ at least 4 minutes/134℃ ±1 cuum for Autoclave using the gravitation method at least 10 minutes/134℃ ±1 Autoclave using the gravitation method at least 60 minutes/121℃ ±1 ▶...
Page 13
Care instructions KaVo SYSTEM 1056 S/T 2 Preparation methods 2.3 Sterilisation Aspiration hose connection pieces ▶ Remove connection pieces from aspiration hoses. Note Sterilise the aspiration cannulae in accordance with the instructions supplied. ▶ To clean the connection pieces, hold under running water and clean with a brush.
Care instructions KaVo SYSTEM 1056 S/T 2 Preparation methods 2.4 Filling the water bottle 2.4 Filling the water bottle ▶ Remove water bottle by slowly rotating to the left. Note The excess pressure in the water bottle can slowly dissipate now.
Page 15
Care instructions KaVo SYSTEM 1056 S/T 2 Preparation methods 2.4 Filling the water bottle ▶ Rotate opening in lid to the 1.5 litre fill level The lid locks into the fill position. ▶ Place Oxygenal bottle on a flat surface.
Page 16
Care instructions KaVo SYSTEM 1056 S/T 2 Preparation methods 2.4 Filling the water bottle ▶ Hold Oxygenal bottle in an almost vertical position and fill the water bottle. Note If sediment can be seen in the dosing cap, this is not contamination - it is an in‐...