SIGLENT TECHNOLOGIES SSA5000A Series User Manual

Spectrum analyzer
Table of Contents

Advertisement

Quick Links

SSA5085A
SSA5083A
Spectrum Analyzer
User Manual
EN02A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SSA5000A Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SIGLENT TECHNOLOGIES SSA5000A Series

  • Page 1 SSA5085A SSA5083A Spectrum Analyzer User Manual EN02A...
  • Page 3: Table Of Contents

    SSA5000A Series User Manual Contents Copyright and Declaration .................... 1 Important Safety Information ..................2 General Safety Summary ..................2 Safety Terms and Symbols ..................4 Working Environment....................5 Cooling Requirements .................... 6 Power and Grounding Requirements ..............6 Cleaning ........................7 Abnormal Conditions ....................
  • Page 4 SSA5000A Series User Manual User Interface ....................... 22 Touch screen and mouse operation ..............23 Keyboard operation ....................23 Using Built-in Help ....................23 Firmware Operation ....................24 3.9.1 Check System Information ................24 3.9.2 Load Option ....................24 3.9.3 Firmware Upgrade ..................
  • Page 5 SSA5000A Series User Manual 5.3.5 Sweep Time Estimate ................... 39 5.3.6 Sweep/Measure .................... 40 Amplitude ......................41 5.4.1 Ref Level ...................... 41 5.4.2 Attenuator ..................... 41 5.4.3 RF Preamp ....................42 5.4.4 Y Axis Unit ....................42 5.4.5 Y Axis Scale Type ..................43 5.4.6...
  • Page 6 SSA5000A Series User Manual 5.7.1 Select Marker ....................62 5.7.2 Select Trace ....................62 5.7.3 Marker Type ....................62 5.7.4 Marker Postion ..................... 64 5.7.5 Relative To ....................65 5.7.6 Readout Type ....................65 5.7.7 Marker Couple ....................66 5.7.8 Marker Lines ....................
  • Page 7 SSA5000A Series User Manual 5.9.10 IQ Acquisition .................... 106 5.9.11 CCDF ......................112 Real-Time Spectrum Analyzer Mode ................ 117 Frequency & SPAN ..................... 117 6.1.1 Frequency & Span ..................117 6.1.2 Freq Offset ....................118 6.1.3 Freq Step ....................118 BW ........................
  • Page 8 SSA5000A Series User Manual 6.7.2 Select Trace ....................135 6.7.3 Marker Type ....................135 6.7.4 Relative To ....................137 6.7.5 Peak->CF ....................137 6.7.6 Update Peak/Maximum Peak ..............137 6.7.7 Minimum Peak .................... 137 6.7.8 Left Peak ....................138 6.7.9 Right Peak ....................
  • Page 9 SSA5000A Series User Manual Freq ........................163 7.3.1 Freq & Span ....................163 7.3.2 CF Step....................... 163 BW ........................164 7.4.1 EQBW ......................164 7.4.2 Window ....................... 164 Sweep ......................... 165 7.5.1 Measure/Sweep Control (Single/Continue/Restart) ........165 Trigger ......................... 166 7.6.1...
  • Page 10 SSA5000A Series User Manual 7.8.7 Data ......................175 7.8.8 Format ......................176 7.8.9 Eye Length ....................177 7.8.10 Symbol Table .................... 177 Marker ......................... 178 7.9.1 Select Marker & Select Trace ..............178 7.9.2 Marker Type ....................178 7.9.3 Marker X ..................... 179 7.9.4...
  • Page 11 SSA5000A Series User Manual 8.4.2 Attenuator ....................186 8.4.3 RF Preamp ....................187 8.4.4 Y Axis Unit ....................187 8.4.5 Y Axis Scale Type ..................188 8.4.6 Ref Offset ....................189 Trigger ......................... 190 8.5.1 Trigger Source .................... 190 8.5.2 Trigger Level ....................
  • Page 12 SSA5000A Series User Manual 8.8.5 Limit Edit ..................... 208 8.8.6 Limit Test ..................... 212 8.8.7 Setup ......................212 Meas & Meas Setup ................... 214 8.9.1 Average Type....................214 8.9.2 Meas Control ....................215 8.9.3 Scan ......................216 8.9.4 Measure ...................... 216 8.9.5...
  • Page 13 SSA5000A Series User Manual Trigger ......................... 232 9.6.1 Trigger Source .................... 232 9.6.2 Trigger Level ....................233 9.6.3 Trigger Slope ....................233 9.6.4 Trigger Delay ....................233 9.6.5 Zero Span Trigger Delay Compensation (external trigger only) ....234 9.6.6 Period (periodic trigger only) ..............234 9.6.7...
  • Page 14 SSA5000A Series User Manual 10.1 Frequency & Span ....................262 10.1.1 Carrier Frequency..................262 10.1.2 Start Offset....................262 10.1.3 Stop Offset ....................263 10.1.4 Auto Tune ....................263 10.1.5 Singal Tracking ..................263 10.2 Sweep ......................... 265 10.2.1 Sweep Points .................... 265 10.2.2...
  • Page 15 SSA5000A Series User Manual Noise Figure Measurement Mode ................282 11.1 Frequency & Span ....................284 11.1.1 Freq Context ..................... 284 11.1.2 Freq Mode....................284 11.2 BW ........................287 11.2.1 Resolution Bandwidth ................287 11.3 Sweep ......................... 287 11.3.1 Avg Time/Pt ....................287 11.3.2...
  • Page 16 SSA5000A Series User Manual 11.7.5 ENR Setup ....................300 11.7.6 Calibration ....................300 Bluetooth Measurement Mode .................. 302 12.1 Freq/Channel ...................... 302 12.1.1 Channel ....................302 12.1.2 Center Frequency ..................302 12.1.3 Center Frequency Step ................302 12.2 Amplitude ......................303 12.2.1...
  • Page 17 SSA5000A Series User Manual 12.7.2 Trigger Level ..................... 317 12.7.3 Trigger Slope .................... 317 12.7.4 Trigger Delay .................... 317 12.7.5 Trigger Delay .................... 318 12.7.6 Zero Span Trigger Delay Compensation (external trigger only) ....318 12.7.7 Period (periodic trigger only) ..............319 12.7.8...
  • Page 18 SSA5000A Series User Manual 14.2.3 Save User Config ..................332 14.2.4 Power On ....................333 14.2.5 Factory Reset ................... 333 14.2.6 Reset&Clear ..................... 333 14.3 Alignments ......................334 14.4 File ........................335 14.5 Display ........................ 336 14.6 Power ........................337 14.7 Self Test ......................
  • Page 19 SSA5000A Series User Manual 16.3 Parameter Type ....................349 16.4 Command Abbreviation ..................350 16.5 IEEE Common Commands................. 351 16.6 SCPI LIST ......................353 16.6.1 GPSA ......................353 16.6.2 RTSA ......................359 16.6.3 MA ......................362 16.6.4 EMI ......................364 16.6.5...
  • Page 21: Copyright And Declaration

    SSA5000A Series User Manual Copyright and Declaration Copyright SIGLENT TECHNOLOGIES CO., LTD All Rights Reserved. Trademark Information is the registered trademark of SIGLENT TECHNOLOGIES CO., LTD SIGLENT Declaration SIGLENT products are protected by patent law worldwide SIGLENT reserves the right to modify or change parts of or all the specifications or pricing policies at company’s sole decision.
  • Page 22: Important Safety Information

    SSA5000A Series User Manual Important Safety Information This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe condition. General Safety Summary Carefully read the following safety precautions to avoid personal injury and prevent damage to the instrument and any products connected to it.
  • Page 23 SSA5000A Series User Manual Identification of Normal State of Equipment. After the equipment is started, there will be no alarm information and error information at the interface under normal conditions. The curve of the interface will scan from left to right freely; if there is a button in the scanning process or there is an alarm or error prompt, the device may be in an abnormal state.
  • Page 24: Safety Terms And Symbols

    SSA5000A Series User Manual Safety Terms and Symbols When the following symbols or terms appear on the front or rear panel of the instrument or in this manual, they indicate special care in terms of safety. This symbol is used where caution is required. Refer to the accompanying information or documents to protect against personal injury or damage to the instrument.
  • Page 25: Working Environment

    SSA5000A Series User Manual Working Environment The design of the instrument has been verified to conform to EN 61010-1 safety standard per the following limits: Environment The instrument is used indoors and should be operated in a clean and dry environment with an ambient temperature range.
  • Page 26: Cooling Requirements

    SSA5000A Series User Manual equipment connected to the AC line (AC power). Degree of Pollution The analyzers may be operated in environments of Pollution Degree II. Note: Degree of Pollution II refers to a working environment that is dry and non-conductive pollution occurs.
  • Page 27: Cleaning

    SSA5000A Series User Manual The instrument includes a grounded cord set containing a molded three-terminal polarized plug and a standard IEC320 (Type C13) connector for making line voltage and safety ground connection. The AC inlet ground terminal is connected directly to the frame of the instrument. For adequate protection against electrical shock hazards, the power cord plug must be inserted into a mating AC outlet containing a safety ground contact.
  • Page 28: Safety Compliance

    SSA5000A Series User Manual Proper use of the instrument depends on careful reading of all instructions and labels. Warning: Any use of the analyzer in a manner not specified by the manufacturer may impair the instrument’s safety protection. This instrument should not be directly connected to human subjects or used for patient monitoring.
  • Page 29: Informations

    SSA5000A Series User Manual Informations essentielles sur la sécurité Ce manuel contient des informations et des avertissements que les utilisateurs doivent suivre pour assurer la sécurité des opérations et maintenir les produits en sécurité. Exigence de Sécurité Lisez attentivement les précautions de sé curité ci - après afin d 'éviter les dommages corporels et de prévenir les dommages aux instruments et aux produits associé...
  • Page 30 SSA5000A Series User Manual l'entretien de l'équipement n'est pas terminé et que l'entretien n'est pas confirmé. Identification de l'état normal de l'é quipement. Après le dé marrage de l'équipement, dans des conditions normales, il n'y aura pas d'information d'alarme et d'erreur au bas de l'interface, et la courbe de l'interface sera balayée librement de gauche à...
  • Page 31: Termes Et Symboles De Sé Curité

    SSA5000A Series User Manual Termes et symboles de sé curité Lorsque les symboles ou termes suivants apparaissent sur le panneau avant ou arrière de l'instrument ou dans ce manuel, ils indiquent un soin particulier en termes de sécurité. Ce symbole est utilisé lorsque la prudence est requise. Reportez-vous aux informations ou documents joints afin de vous protéger contre les blessures...
  • Page 32: Environnement De Travail

    SSA5000A Series User Manual Environnement de travail La conception de l'instrument a été certifiée conforme à la norme EN 61010-1, sur la base des valeurs limites suivantes: Environnement L'instrument doit être utilisé à l'intérieur dans un environnement propre et sec dans la plage de température ambiante.
  • Page 33 SSA5000A Series User Manual d' équipement reliés au circuit source.Dans ces bornes, des mesures préventives sont prises pour limiter la tension transitoire à un niveau inférieur correspondant. La caté gorie II d'installation (surtension) désigne le niveau local de distribution d 'énergie d' un équipement conç...
  • Page 34: Exigences De Refroidissement

    SSA5000A Series User Manual Exigences de refroidissement Cet instrument repose sur un refroidissement à air forcé avec des ventilateurs internes et des ouvertures de ventilation. Des pré cautions doivent être prises pour éviter de restreindre le flux d'air autour des ouvertures (trous de ventilateur) de chaque côté de la lunette. Pour assurer une ventilation adéquate, il est nécessaire de laisser un espace minimum de 15 cm (6 pouces) sur les...
  • Page 35: Nettoyage

    SSA5000A Series User Manual Avertissement: risque de choc électrique! Toute interruption du conducteur de terre de protection à l'intérieur ou à l'extérieur de la portée ou la déconnexion de la borne de terre de sécurité crée une situation dangereuse. L'interruption intentionnelle est interdite.
  • Page 36: Conformité En Matière De Sécurité

    SSA5000A Series User Manual Avertissement: Toute utilisation de l'instruments d'une manière non spécifié e par le fabricant peut compromettre la protection de sécurité de l'instrument. Cet instrument ne doit pas être directement connecté à des sujets humains ni utilisé pour la surveillance des patients.
  • Page 37: Quick Start

    SSA5000A Series User Manual Quick Start This chapter guides users to quickly get familiar with the appearance, panel and the user interface, as well as announcements during the first use of the analyzer. General Inspection Inspect the shipping container Keep the damaged shipping container or cushioning material until the contents of the shipment have been completely checked and the instrument has passed both electrical and mechanical tests.
  • Page 38: Preparing For Use

    SSA5000A Series User Manual Preparing for Use 3.2.1 Appearance and Dimension Adjust the supporting legs properly to use them as stands to tilt the analyzer upwards for stable placement as well as easier operation and observation of the instrument display.
  • Page 39: Front Panel

    SSA5000A Series User Manual Front Panel Figure 3-3 the Front Panel Table 3-1 Front Panel Description Name Description LCD Screen 12.1 inch multi-touch screen, resolution 1280*800 Power Switch Stand by status: Orange Power on status: White Short press: To Stand by status with current state saved...
  • Page 40: Rear Panel

    SSA5000A Series User Manual Rear Panel 9 10 11 12 Figure 3-4 Rear Panel Table 3-2 Rear Panel Name Description Handle Pull up the handle vertically for easy carrying of the instrument. When you do not need the handle, press it down.
  • Page 41 SSA5000A Series User Manual HDMI Output the LCD screen display content of the analyzer and connect it to the external monitor through HDMI cable for synchronous display. Audio 3.5 mm audio cable jack, which can be externally connected with headphones or speakers to monitor the audio demodulation output of modulated signals.
  • Page 42: User Interface

    SSA5000A Series User Manual User Interface Figure 3-5 User Interface of spectrum analyzer mode Table 3-3 Spectrum Analyzer Mode User Interface Name Description SIGLENT SIGLENT logo Mode/Measure Indicate the current working mode and measurement function of the analyzer, and click to switch, such as spectrum analysis mode, real-time spectrum mode, etc.
  • Page 43: Touch Screen And Mouse Operation

    SSA5000A Series User Manual Menu Switch Control the opening and closing of the menu area. After closing, a larger measurement result area can be displayed. Interface Status Indicates and controls the connection status of LAN, USB memory and other devices.
  • Page 44: Firmware Operation

    SSA5000A Series User Manual  Click on the items in the contents tree on the left to navigate to any topics interested.  Click the green back or forward arrow to go back or forward to the contents just read.
  • Page 45: 3.10 Remote Control

    SSA5000A Series User Manual 3.10 Remote Control The analyzer supports communication with computers via USB, LAN, and GPIB-USB interfaces. By using these interfaces, in combination with programming languages and/or NI-VISA software, users can remotely control the analyzer based on a SCPI (Standard Commands for Programmable Instruments) compliant command set, LabView and IVI (Interchangeable Virtual Instrument), to interoperate with other programmable instruments.
  • Page 46: Mode & Measurement

    SSA5000A Series User Manual Mode & Measurement The analyzer operates in multiple operating modes, each of which contains several measurements:  Spectrum Analyzer Mode (GPSA):  Swept Spectrum Analyzer (Swept SA)  Channel Power (CH Power)  Adjacent Channel Power Ratio (ACPR) ...
  • Page 47 SSA5000A Series User Manual Figure 4-1 Mode/measurement selection The analyzer supports the creation of multiple independent modes / measurements, but only one of them can be activated at the same time. At this time, other measurements enter the background. Wait until the next activation to restore the state before switching...
  • Page 48 SSA5000A Series User Manual Return Enumeration Example :INSTrument MA Command Format :INSTrument:MEASure :INSTrument:MEASure? Instruction Get/Set the measurement mode. Parameter Type Enumeration Parameter Range SA: Spectrum Analysis ACPR: Adjacent Channel Power Ratio CHPower: Channel Power OBW: Occupy Bandwidth TPOWer: Time-Domain Power...
  • Page 49: Spectrum Analyzer Mode

    SSA5000A Series User Manual Spectrum Analyzer Mode The ‘Spectrum Analyzer’ mode is the default mode of the analyzer. Frequency & Span 5.1.1 Frequency & Span Set the frequency-related parameters and functions of the analyzer. The sweep will restart every time the frequency parameters are modified.
  • Page 50 SSA5000A Series User Manual Command Format [:SENSe]:FREQuency:STOP [:SENSe]:FREQuency:STOP? Instruction Sets the stop frequency of the spectrum analyzer. Gets the stop frequency. Parameter Type Float, unit: Hz, kHz, MHz, GHz Parameter 100 Hz ~ 28 GHz Range Zero Span: 0 ~ 28 GHz...
  • Page 51: Axis Scale

    SSA5000A Series User Manual 5.1.2 X Axis Scale Set the scale type of X-axis to Linear (Lin) or Logarithmic (Log) scale. In Log scale type, the frequency scale of X-axis is displayed in the logarithmic form. Command Format :DISPlay:WINDow:TRACe:X[:SCALe]:SPACing :DISPlay:WINDow:TRACe:X[:SCALe]:SPACing?
  • Page 52: Signal Track

    SSA5000A Series User Manual Instruction Specifies the center frequency step size. Gets the center frequency step. Parameter Type Float, unit: Hz, kHz, MHz, GHz Parameter Range 1 Hz ~ 28 GHz Return Float, unit: Hz Example :FREQuency:CENTer:STEP 2 MHz Command Format...
  • Page 53: Auto Tune

    SSA5000A Series User Manual  When Marker1 is on, turn on signal tracking, a point whose amplitude does not change more than 3 dB near Marker1 will be searched and marked, and the frequency at that point will be set to center frequency.
  • Page 54: Resolution Bandwidth

    SSA5000A Series User Manual The bandwidth menu contains the RBW (Resolution Bandwidth), VBW (Video Bandwidth), average type and filter type. Filter type includes the EMI filter type that enables EMI measurement controls. 5.2.1 Resolution Bandwidth Set the resolution bandwidth in order to distinguish between signals which have frequency components that are near one another.
  • Page 55: Video Bandwidth

    SSA5000A Series User Manual 5.2.2 Video Bandwidth Set the video bandwidth in order to filter out the noise outside the video band.  Reducing the VBW will smooth the trace and helps to highlight small signals from noise, but it will also increase the sweep time (Sweep Time is affected by a combination of RBW and VBW when it is in Auto mode).
  • Page 56: Filter Type

    SSA5000A Series User Manual Instruction Specifies the ratio of the video bandwidth to the resolution bandwidth. Gets the ratio of the video bandwidth to the resolution bandwidth. Parameter Type Discrete, Float Parameter Range 0.001, 0.003, 0.01, 0.03, 0.1, 0.3, 1.0, 3.0, 10.0, 30.0, 100.0, 300.0, 1000.0...
  • Page 57: Sweep

    SSA5000A Series User Manual Sweep 5.3.1 Sweep Type There are two sweep types, Swept and FFT. Swept can only support a minimum of 3kHz RBW, and conversely FFT can only support a maximum of 10kHz RBW. 5.3.1.1 Auto Sweep Type When automatic sweep type is enabled, the analyzer automatically switches sweep types based on RBW and sweep type rules.
  • Page 58: Sweep Points

    SSA5000A Series User Manual 5.3.3 Sweep Points The number of sweep points represents the number of sweep and trace displayed points (201~10001). More sweep points will improve the resolution of waveform, but also affect the minimum sweep time, and increase the time of data processing and remote access to data, and reduce the response rate.
  • Page 59: Sweep Time Estimate

    SSA5000A Series User Manual satisfied: When Span>0: When Span=0: In general, the manual sweep time should not be longer than the automatic sweep time in this condition. Otherwise, unforeseeable anomalies may be caused and may be marked (UNCAL). It should be pointed out in particular that under the influence of the scheme, when the sweep type is FFT, the sweep time can only be calculated by the instrument itself, and any modification related to the sweep time cannot take effect.
  • Page 60: Sweep/Measure

    SSA5000A Series User Manual 5.3.6 Sweep/Measure Sweep/Measure Single/Continue controls analyzer to perform single or continuous sweep/measure. Restart Restart the current sweep or measure. In particular, if the sweep parameters are modified, a restart will be performed. Command format :INITiate:CONTinuous OFF|ON|0|1...
  • Page 61: Amplitude

    SSA5000A Series User Manual Amplitude Set the amplitude parameters of the analyzer. Through modifying these parameters, signals under measurement can be displayed in a proper mode for easier observation and minimum error. Any change of Ref Level, Attenuator Value, Preamp mode and Ref Offset will restart sweep.
  • Page 62: Rf Preamp

    SSA5000A Series User Manual Command format [:SENSe]:POWer[:RF]:ATTenuation [:SENSe]:POWer[:RF]:ATTenuation? Instructions Sets/gets the attenuation value Parameter Type Integer Parameter Range 0 dB ~ 50 dB (Even gears only) Return Integer, unit dB Example :POWer:ATTenuation 10 Command format [:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1 [:SENSe]:POWer[:RF]:ATTenuation:AUTO? Instructions Sets/gets the auto attenuation value switch...
  • Page 63: Y Axis Scale Type

    SSA5000A Series User Manual 5.4.5 Y Axis Scale Type Set the scale type of the Y-axis to Lin or Log. The default is Log.  In Lin mode, the vertical Scale value cannot be changed. The Display area is set for reference level of 0%.
  • Page 64: Ref Offset

    SSA5000A Series User Manual Command format :DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing LINear|LOGarithmic :DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing? Instructions Sets/Gets the scale display type Parameter Type Enumeration: LINear |LOGarithmic Return Enumeration: LIN|LOG Example :DISPlay:WINDow:TRACe:Y:SPACing LINear 5.4.6 Ref Offset Assign an offset to the reference level to compensate for gains or losses generated between the device under measurement and the analyzer.
  • Page 65: Trigger

    SSA5000A Series User Manual Trigger The analyzer provides a variety of trigger functions, users can choose from the trigger menu. 5.5.1 Trigger Source The analyzer provides a variety of trigger sources to suit different trigger requirements. Free Run Free trigger is the default mode of the analyzer, in which the spectrum analyzer sweeps circularly and continuously.
  • Page 66: Trigger Level

    SSA5000A Series User Manual 5.5.2 Trigger Level Sets the amplitude level for the video trigger (absolute level only supported). When the video signal crosses the voltage level with the selected slope, it is triggered. Command format :TRIGger[:SEQuence]: {type}:LEVel :TRIGger[:SEQuence]: {type}:LEVel? Instructions sets the trigger level.
  • Page 67: Zero Span Trigger Delay Compensation (External Trigger Only)

    SSA5000A Series User Manual Parameter Type Float Return Float Example :TRIGger:EXTernal:DELay 5e-3 :TRIGger:FRAMe:DELay? Command format :TRIGger[:SEQuence]:{type}:DELay:STATe :TRIGger[:SEQuence]:{type}:DELay:STATe? Instructions sets the trigger delay state. gets the trigger delay state. {type}:" VIDeo ", " EXTernal " ,"FRAMe" Parameter Type Boolean Parameter Range...
  • Page 68: Offset Time

    SSA5000A Series User Manual 5.5.7 Offset Time Adjust the cumulative offset between the periodic trigger clock and trigger events. The periodic trigger clock cannot be viewed on the software, only the trigger event can be seen. Therefore, in order to adjust the trigger event time, only the offset between the periodic triggering clock and the triggering event can be adjusted.
  • Page 69: Sync Source (Periodic Trigger Only)

    SSA5000A Series User Manual 5.5.9 Sync Source (periodic trigger only) Figure 5-2 Trigger of synchronization source Command format :TRIGger[:SEQuence]:FRAMe:SYNC :TRIGger[:SEQuence]:FRAMe:SYNC? Instructions Set/Query the type of periodic synchronization Parameter Type Enumeration Parameter Range "OFF", "EXT" Return "OFF", "EXT" Example :TRIGger:FRAMe:SYNC EXT 5.5.10...
  • Page 70: Gate

    SSA5000A Series User Manual Set the trigger source of the gate to a periodic source. Similar to trigger, you can configure trigger period, trigger offset, and trigger period synchronization sources when configuring periodic sources. The configuration of the gating source affects the configuration of the trigger source.
  • Page 71 SSA5000A Series User Manual gating source first. After the trigger of the gate source, the gate signal is determined according to the gate delay and gate width. When the gate signal is high, sweep starts, data is collected and FFT transformation is carried out.
  • Page 72 SSA5000A Series User Manual 30Hz 86968 100Hz 27807 300Hz 10323 1000Hz 5447 3000Hz 2333 10000Hz 1117 Command format [:SENSe]:SWEep:EGATe:LENGth [:SENSe]:SWEep:EGATe:LENGth? Instructions Set/Query the gate length Parameter Type Float Parameter Range 2.106us~5s Return Float Example :SWEep:EGATe:LENGth 1s 5.5.11.4 Gate Delay Configure the delay between gate trigger and gate on.
  • Page 73 SSA5000A Series User Manual Gate View Start Time Set the start time on the left of the gate view, that is, set a delay. Command format [:SENSe]:SWEep:EGATe:VIEW [:SENSe]:SWEep:EGATe:VIEW? Instructions Set or query the gate view switch Parameter Type Boolean Parameter Range...
  • Page 74: Trace

    SSA5000A Series User Manual Trace Figure 5-3 trace select The sweep signal is displayed as a trace on the screen. Command format :TRACe[1]|2|3|4|5|6 [:DATA]? Instructions Get trace data Return String Example :TRACe:DATA? Command format :FORMat[:TRACe][:DATA]ASCii|REAL32| REAL :FORMat[:TRACe][:DATA]? Instructions Sets/gets the format of trace data read...
  • Page 75: Trace Type

    SSA5000A Series User Manual Command format TRACe:SELEct TRACe:SELEct? Instructions Sets/gets the current trace Parameter Type EnumerationTRACE1-6 Return Enumeration: TRACE1-6 Example TRACe:SELEct TRACE3 5.6.2 Trace Type Set the type of the current trace or disable it. The system calculates the sampled data using a specific operation method according to the trace type selected and displays the result.
  • Page 76: Trace State

    SSA5000A Series User Manual MINHold: Displays the minimum value of trace AVERage: Displays the average value of trace Return Enumeration: WRITE|MAXH|MINH| AVER Example :TRAC1:TYPE MINH 5.6.3 Trace State There are four trace states: active, view, blank, and background. Different trace states indicate the...
  • Page 77 SSA5000A Series User Manual The available types are Pos Peak, Neg Peak, Sample, Normal, Average and Quasi Peak. The default is Pos peak. Positive Peak For each trace point, Positive Peak detector displays the maximum value of data sampled within the corresponding time interval.
  • Page 78: Math

    SSA5000A Series User Manual Parameter Range Return Example :DETector:TRACe3:AUTO 1 :DETector:TRACe:AUTO:ALL 5.6.5 Math Set the computational method of the math trace. Output Z The Math result is denoted by the Z variable and can be displayed by trace. Input X, Y Input X, Y can be applied to trace.
  • Page 79: Normalize

    SSA5000A Series User Manual Command format :CALCulate[:SELected]:MATH:FUNCtion :CALCulate[:SELected]:MATH:FUNCtion? Instructions Sets/gets the trace calculation type Parameter Type Enumeration Parameter Range PDIF: Power subtracting PSUM: Power up LOFF: Logarithmic deviation LDIF: Logarithmic subtraction Return Enumeration Example :CALCulate:MATH:FUNCtion PDIF Command format :TRACe[1]|2|3|4|5|6:MATH:OFFSet :TRACe[1]|2|3|4|5|6:MATH:OFFSet?
  • Page 80: Trace Function

    SSA5000A Series User Manual Command format :DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel :DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel? Instructions Sets/gets the normalized reference level Parameter Type Float, unitdB Parameter Range -200 dB ~ 200 dB Return Float, unitdB Example :DISPlay:WINDow:TRACe:Y:NRLevel 10 Command format :DISPlay:WINDow:TRACe:Y[:SCALe]:NRPosition :DISPlay:WINDow:TRACe:Y[:SCALe]:NRPosition? Instructions Sets/reads the normalized reference location...
  • Page 81 SSA5000A Series User Manual Command format :TRACe:COPY Instructions Copy the trace Example :TRACe:COPY 1,2 Command format :TRACe:EXCHange Instructions Exchange of trace Example :TRACe:EXCHange 1,2 Command format :TRACe:PRESet:ALL Instructions Reset all traces Example :TRACe:PRESet:ALL Command format :TRACe:CLEar:ALL Instructions Clear all traces...
  • Page 82: Marker & Peak

    SSA5000A Series User Manual Marker & Peak The marker appears as a rhombic sign (as shown below) for identifying points on a trace. You can easily read the amplitude, frequency and sweep time of the marked point on the trace.
  • Page 83 SSA5000A Series User Manual frequency of the current trace.  You can use the numeric keys, knob or direction keys to move the marker. The readouts of the marker will be displayed at the upper right corner of the screen.
  • Page 84: Marker Postion

    SSA5000A Series User Manual Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MODE POSition|DELTa|FIXed|OFF :CALCulate:MARKer[1]|2|3|4:MODE? Instructions Sets/Gets the marker mode Parameter Type Enumeration Parameter Range POSition DELTa FIXed Return Enumeration: POS|DELT|FIX|OFF Example :CALCulate:MARK1:MODE POSition Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:STATe OFF|ON|0|1 :CALCulate:MARKer[1]|2|3|4|5|6|7|8:STATe? Instructions Sets/gets the marker switch status Parameter Type...
  • Page 85: Relative To

    SSA5000A Series User Manual Example :CALCulate:MARKer4:X 0.4 GHz :CALCulate:MARKer4:X 200 ms :CALCulate:MARKer4:X? Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:Y :CALCulate:MARKer[1]|2|3|4|5|6|7|8:Y? Instructions Read the value of the Y-axis of the marker point, which can also be used to read the marker noise in the marker function.
  • Page 86: Marker Couple

    SSA5000A Series User Manual Period In this type, the Normal marker shows the reciprocal of frequency; while Delta marker and Delta Pair marker show the reciprocal of frequency difference. When the frequency difference is zero, the reciprocal is infinite and 100 Ts is displayed.
  • Page 87: Marker Lines

    SSA5000A Series User Manual Instructions Set/query the marker coupling switch Parameter Type Boolean Parameter Range ON|OFF|0|1 Return Example :CALCulate:MARKer:COUPle 1 :CALCulate:MARKer:COUPle? 5.7.8 Marker Lines Mark the marker with the intersection of horizontal and vertical lines, which is more convenient to query the marker position in the waveform area.
  • Page 88 SSA5000A Series User Manual  If the Delta or Delta Pair marker is selected, the center frequency will be set to the frequency of the Delta Marker.  The function is invalid in Zero span. Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8[:SET]:CENTer Instructions Sets/gets the value of the marker X axis to the center frequency .If the marker is not open, this command automatically turns the marker on the center frequency.
  • Page 89 SSA5000A Series User Manual  If the Delta or Delta Pair marker is selected, the stop frequency will be set to the frequency of the Delta Marker.  The function is invalid in Zero span. Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8[:SET]:STOP Instructions Sets/gets the value of the marker X axis to terminate frequency If the corresponding marker is not open, sending this command will automatically open the marker at the end frequency.
  • Page 90: Marker Fn

    SSA5000A Series User Manual Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:DELTa[:SET]:CENTer Instructions Sets/gets the difference between the marker and the X axis to the center frequency This command takes effect only when the marker mode is DELTa :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MODE Example :CALCulate:MARKer3:DELTa:CENTer 5.7.11 Marker Fn Special marker functions including Noise Marker, N dB BW and Freq Counter.
  • Page 91 SSA5000A Series User Manual mkr_ndb_bw Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:BANDwidth:NDB :CALCulate:MARKer[1]|2|3|4|5|6|7|8:BANDwidth:NDB? Instructions Set/Obtain the N dB bandwidth reference value Parameter Type Float Parameter Range -100 dB ~ 100 dB Return Float Example :CALCulate:MARK1:BANDwidth:NDB 10 DB Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:BANDwidth:RESult? Instructions Result of setting/reading N dB bandwidth...
  • Page 92: Peak Search

    SSA5000A Series User Manual made based on the detection and trace types. The measurement will be more precise if RMS Avg or Sample detection type is used.  This function can be used for measuring the C/N ratio. 5.7.11.4 Turn off the noise marker, N dB BW measurement or Frequency Counter, but not the marker itself.
  • Page 93 SSA5000A Series User Manual Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum Instructions Marker searches for peaks and marks them with the specified marker (If peak-to-peak value is on, peak-to-peak value search is carried out; otherwise, single peak value search is carried out, refer to the...
  • Page 94 SSA5000A Series User Manual Peak Threshold Assign a minimum for the peak amplitude. Peaks whose amplitudes are greater than the specified peak threshold are treated as real peaks.The actual minimal peak threshold is -200dBm when shut down the Peak Threshold.
  • Page 95 SSA5000A Series User Manual 5.7.12.3 Count Peak Enable or disable continuous peak search. The default is Off. When enabled, the system will always execute a peak search automatically after each sweep in order to track the signal under measurement. Command format...
  • Page 96 SSA5000A Series User Manual Command format :CALCulate:MARKer:PEAK:SORT:ORDEr :CALCulate:MARKer:PEAK:SORT:ORDEr? Instructions Sets/Gets the peak sort type Parameter Type Enumeration Parameter Range ASC|DEC Return ASC|DEC Example :CALCulate:MARKer:PEAK:SORT:ORDEr DEC :CALCulate:MARKer:PEAK:SORT:ORDEr? Command format :CALCulate:MARKer:PEAK:TABLe:DTLimit :CALCulate:MARKer:PEAK:TABLe:DTLimit? Instructions Peak sorting based on limit selection Parameter Type Integer...
  • Page 97: Limit

    SSA5000A Series User Manual Limit The analyzer supports Pass/Fail test function. In this function, the measured curve will be compared with the pre-edited curve. If the related rules are met, the result is “Pass”, else the result is “Fail”. 5.8.1 Slect Limit Activate/Select a limit.
  • Page 98: Limit Type

    SSA5000A Series User Manual 5.8.4 Limit Type Set the limit type as upper / lower . The limit 1,3,5 is default as the lower and 2,4,6 as the lower. Command format :CALCulate:LLINe[1]|2|3|4|5|6:TYPE UPPer|LOWer :CALCulate:LLINe[1]|2|3|4|5|6:TYPE? Instructions Sets/Gets the restriction type Parameter Type...
  • Page 99 SSA5000A Series User Manual Amplitude Edit the amplitude of the current point or line. Del Point Delete the point whose number is selected in Mode. Del All Delete all the points. Save/Load Save or load the limit file. X Offset Set offsets of X axis.
  • Page 100 SSA5000A Series User Manual Parameter Range val1 : related with Span val2 : -400 dBm~330 dBm Example :CALCulate:LLINe1:ADD 100,-20 :CALCulate:LLINe2:POINt:DELete 2 Command format :CALCulate:LLINe[1]|2|3|4|5|6:DELete :CALCulate:LLINe:ALL:DELete Instructions Delete specified restrictions Delete all restrictions Example :CALCulate:LLINe1:DELete :CALCulate:LLINe:ALL:DELete Command format :CALCulate:LLINe[1]|2|3|4|5|6:TRACe :CALCulate:LLINe[1]|2|3|4|5|6:TRACe? Instructions...
  • Page 101: Limit Test

    SSA5000A Series User Manual Instructions Set or query the amplitude reference type Parameter Type Enumeration Parameter Range FIXed|RELAtive Return FIXed|RELAtive Example :CALCulate:LLINe2: AMPLitude:CMODe FIX Command format :CALCulate:LLINe[1]|2|3|4|5|6:COPY :CALCulate:LLINe[1]|2|3|4|5|6:COPY? Instructions Copy the limit Parameter Type Integer Parameter Range Return Example :CALCulate:LLINe2:COPY 5...
  • Page 102: Setup

    SSA5000A Series User Manual Command format :CALCulate:LLINe:TEST :CALCulate:LLINe:TEST? Instructions Sets/gets the status of the test switch Parameter Type Boolean Parameter Range OFF|ON|0|1 Return Example :CALCulate:LLINe:TEST 1 Command format :CALCulate:LLINe[1]|2|3|4|5|6:FAIL? Instructions Query limit test results. Return Example :CALCulate:LLINe2:FAIL 1 5.8.7 Setup Fail to stop Turn on or off the Fail to stop function.
  • Page 103: Meas & Meas Setup

    SSA5000A Series User Manual Meas & Meas Setup Provide optional measurement functions. When activated, the screen will be divided into two parts. The above part is the measure screen which displays traces, and the other part is used to display measurement results.
  • Page 104 SSA5000A Series User Manual Command format [:SENSe]:AVERage:TYPE LOGPower|POWer|VOLTage [:SENSe]:AVERage:TYPE? Instructions Set/Query the average type Parameter type Enumeration Parameter Range LOGPower POWer VOLTage Return Enumeration: LOGP|POW|VOLT Example AVERage:TYPE VOLTage Command format [:SENSe]:AVERage:TYPE:AUTO 0|1|ON|OFF [:SENSe]:AVERage:TYPE:AUTO? Instructions Set/query the average type automatically Parameter type...
  • Page 105 SSA5000A Series User Manual This machine is equipped with a headphone jack, through which demodulation signal can be output in audio mode. Audio frequency represents the frequency of the modulated signal, and the intensity of audio indicates the amplitude of the modulated signal.
  • Page 106 SSA5000A Series User Manual Instructions Set the demodulation time Get demodulation time Parameter type Float , unitms , us , s Parameter Range 5 ms ~1000 s Return Float , units Example DEMod:TIME 5 ms 5.9.1.4 Automatic coupling Auto coupling instantly sets all auto/manual functions to auto. The automatic coupling action is limited to the current measurement.
  • Page 107 SSA5000A Series User Manual Command format :DISPlay:WINDow:TRACe:X:FLINe :DISPlay:WINDow:TRACe:X:FLINe? Instructions Set/Obtain the frequency display line Parameter type Float Return Float Example :DISPlay:WINDow:TRACe:X:FLINe 100e6 Command format :DISPlay:WINDow:TRACe:X:FLINe:STATe :DISPlay:WINDow:TRACe:X:FLINe:STATe? Instructions Set/get frequency display line automatically Parameter type Boolean Parameter Range OFF|ON|0|1 Return Example :DISPlay:WINDow:TRACe:X:FLINe:STATe 1 5.9.1.6...
  • Page 108: Channel Power

    SSA5000A Series User Manual 5.9.2 Channel Power Measure the power and power density within the specified channel bandwidth. When this function is enabled, the span and resolution bandwidth are automatically adjusted to smaller values. Select Channel Power and press Meas Setup to set the corresponding parameters.
  • Page 109 SSA5000A Series User Manual Exponen tialr The measure average operation continuously calculates a weighted average for each index using a specified number of averages. Repeat: Each time the specified average is reached, the measurement resets the average counter. Command format...
  • Page 110: Acpr

    SSA5000A Series User Manual 5.9.3 ACPR Measure the power of the main channel and adjacent channels as well as the power difference between the main channel and each of the adjacent channels. When this function is enabled, the span and resolution bandwidth of the analyzer are adjusted to smaller values automatically.
  • Page 111 SSA5000A Series User Manual Command format :MEASure:ACPRatio:LOWer? :MEASure:ACPRatio:UPPer? Instructions Obtain the low frequency/high frequency adjacent channel power ratio Return Float , unitdBm Example :MEASure:ACPRatio:LOWer? 5.9.3.1 Average Mode Select the type of control used for the averaging function. This determines the average operation after the specified number of data acquisitions (average count) has been reached.
  • Page 112 SSA5000A Series User Manual Command format [:SENSe]:ACPRatio:OFFSet:BWIDth[:INTegration] [:SENSe]:ACPRatio:OFFSet:BWIDth[:INTegration]? Instructions Set the main channel bandwidth Get the main channel bandwidth Parameter type Float , unitHz , kHz , MHz , GHz Parameter Range related with Span Return Float , unitHz Example :ACPRatio:OFFSet:BWIDth 20 MHz 5.9.3.4...
  • Page 113: Obw

    SSA5000A Series User Manual 5.9.4 Integrates the power within the whole span and calculates the bandwidth occupied by this power according to the specified power ratio. The OBW function also indicates the difference (namely “Transmit Freq Error”) between the center frequency of the channel under measurement and the center frequency of the analyzer.
  • Page 114 SSA5000A Series User Manual Return Float , unitHz Example :MEASure:OBWidth:OBWidth:FERRor? 5.9.4.1 Average Mode Select the type of control used for the averaging function. This determines the average operation after the specified number of data acquisitions (average count) has been reached. Options include:...
  • Page 115 SSA5000A Series User Manual Parameter type Enumeration TPOW|OBWPower Return Enumeration Example :OBWidth:PREFerence TPOW 5.9.4.4 x dB Sets the x dB value used for the "x dB Bandwidth" result, which measures the bandwidth between two points on the signal that are x dB below the highest signal point in the OBW range.
  • Page 116: T-Power

    SSA5000A Series User Manual 5.9.5 T-Power The system enters Zero span and calculates the power within the time domain. The types of powers available include Peak, Average and RMS. Select T-Power and press Meas Setup to set the corresponding parameters.
  • Page 117 SSA5000A Series User Manual Repeat Each time the specified average is reached, the measurement resets the average counter. Command format [:SENSe]: TPOWer:AVERage:TCONtrol [:SENSe]: TPOWer:AVERage:TCONtrol? Instructions Sets/gets average mode Parameter Type Enumeration Parameter Range EXPOnentialr|REPEat Return EXPOnentialr|REPEat Example :TPOWer:AVERage:TCONtrol REPE 5.9.5.2...
  • Page 118: Toi

    SSA5000A Series User Manual Instructions Set/Query the time domain power measurement stop line Parameter type Float , unit :s Parameter Range 0 ~ 1000 s Return Float , time unit: s Example :TPOWer:RLIMit 50 s 5.9.6 Automatic measurement of IP3 (Third order Intercept Point), including the power of fundamental wave and the Third order in the power, and calculate the adjustable Intercept Point.
  • Page 119 SSA5000A Series User Manual Return Float Example :MEASure:TOI:IP3? 5.9.6.1 Average Mode Select the type of control used for the averaging function. This determines the average operation after the specified number of data acquisitions (average count) has been reached. Options include:...
  • Page 120: Spectrum Monitor

    SSA5000A Series User Manual 5.9.7 Spectrum Monitor Display the power of the swept spectrum as an intensity color map commonly referred to as a waterfall chart. Select Spectrum Monitor and press Meas Setup to set the corresponding parameters. Figure 5-13 Spectrum Monitor Display the power of spectrum of successive scans as a color map.
  • Page 121 SSA5000A Series User Manual Parameter Range Return Example [:SENSe]:SPECtrogram:RESTart 5.9.7.1 Average Mode Select the type of control used for the averaging function. This determines the average operation after the specified number of data acquisitions (average count) has been reached. Options include:...
  • Page 122: Cnr

    SSA5000A Series User Manual 5.9.8 Measure the power of the carrier and noise of the specified bandwidth and their ratio. Select CNR and press Meas Setup to set the corresponding parameters. Figure 5-14 CNR Measurement Results: C/N, Carrier Power, Noise Power.
  • Page 123 SSA5000A Series User Manual The measure average operation continuously calculates a weighted average for each index using a specified number of averages. Repeat: Each time the specified average is reached, the measurement resets the average counter. Command format [:SENSe]:CNRatio:AVERage:TCONtrol [:SENSe]:CNRatio:AVERage:TCONtrol?
  • Page 124: Harmonics

    SSA5000A Series User Manual Instructions Sets/gets frequency offset Parameter Type Float , Unit: Hz , kHz , MHz , GHz Parameter Range 100 Hz~700 MHz Return Float , Unit: Hz Example :ACPRatio:OFFSet 20 MHz 5.9.9 Harmonics The harmonic power and total harmonic distortion of carrier signal are measured. The maximum measurable harmonic is 10th harmonic.
  • Page 125 SSA5000A Series User Manual Command format [:SENSe]:HARMonics:FREQuency:FUNDamental [:SENSe]:HARMonics:FREQuency:FUNDamental? Instructions Set the fundamental frequency Get the fundamental frequency Parameter Type Float , unit: Hz , kHz , MHz , GHz Parameter Range 1 Hz~28 GHz Return Float , unitHz Example HARMonics:FREQuency:FUNDamental 20 MHz...
  • Page 126: Iq Acquisition

    SSA5000A Series User Manual Instructions Set harmonic number Querying harmonic number Parameter Type Integer Parameter Range 2 ~ 10 Return Integer Example :HARMonics:NUMBer 5 5.9.9.5 Select Harmonic When “All” is selected, the trace shows the fundamental wave and all harmonics in the sweep bandwidth.
  • Page 127 SSA5000A Series User Manual Figure 5-16 IQ Acquisition 5.9.10.1 Amplitude Configuring the amplitude parameters of the analyzer makes the display of the measured signal visually appealing and minimizes measurement errors. By adjusting these parameters, the signal can be displayed in the current window in an easily observable format. Once the amplitude parameters have been modified, the measurement process will commence anew.
  • Page 128 SSA5000A Series User Manual Avg|Hold Number Averaging|Hold count N represents the counter when the trace type is set to ‘Average’. In a single measurement (Single) and any valid trace type set to ‘Average’, the scan will stop when the counter reaches N.
  • Page 129 SSA5000A Series User Manual Command format [:SENSe]:WAVeform:SRATe [:SENSe]:WAVeform:SRATe? Instructions Set Sampling rate Get Sampling rate Parameter Type Float , unit Hz , kHz , MHz , GHz Parameter Range 1kHz ~ 20MHz Return Float , UNIT Hz Example :WAVeform:SRATe 200kHz 5.9.10.5...
  • Page 130 SSA5000A Series User Manual (2) Click FREQ , configure the center frequency corresponding to the IQ signal. (3) Click AMPTD, Adjust the reference level, the scale and measurement time to better display the I/Q waveform on the screen. int.siglent.com...
  • Page 131 SSA5000A Series User Manual (4) Click Save/Recall , save IQ data as .txt file. int.siglent.com...
  • Page 132: Ccdf

    SSA5000A Series User Manual 5.9.11 CCDF CCDF Instructions The complementary cumulative distribution function (CCDF) reflects the probability of the peak power of the measured signal exceeding a certain threshold, which is the peak power greater than the average power. It can also be said that CCDF can describe the probability of the measured signal greater than a certain peak to average power ratio.
  • Page 133 SSA5000A Series User Manual Peak value: delta between maximum power and average power Maximum power Command format :FETCh:PSTatistic? :MEASure:PSTatistic? :READ:PSTatistic? Instructions Get the CCDF measurement result Parameter Type string Return Average input power Probability at the average input power level...
  • Page 134 SSA5000A Series User Manual Instructions Set test cycle Get test cycle Parameter Type Integer Parameter Range 1~3.2M Return Integer Example :PSTatistic:SWEep:CYCLes 500 :PSTatistic:SWEep:CYCLes? 5.9.11.3 Meas Interval Meas Interval refers to the length of time for a measurement period. Command format...
  • Page 135 SSA5000A Series User Manual 5.9.11.5 Ref/guass trace Figure 5-18 trace As shown in the figure above, blue is the saved reference trace; Purple is Gaussian trace; Yellow is the current trace. Reference trace You need to save the reference trace first. After saving the reference trace, you can view the reference trace through the View/Close button.
  • Page 136 SSA5000A Series User Manual Command format :DISPlay:PSTatistic:GAUSsian[:STATe] :DISPlay:PSTatistic:GAUSsian[:STATe]? Instructions Set the Gaussian trace switch Get the Gaussian trace switch Parameter Type Boolean Parameter Range Return Example :DISPlay:PSTatistic:GAUSsian 1 5.9.11.6 X Scale/Div Set the scale on the x axis Command format...
  • Page 137: Real-Time Spectrum Analyzer Mode

    SSA5000A Series User Manual Real-Time Spectrum Analyzer Mode This chapter introduces the function keys and menu functions of the front panel under the Real-Time Spectrum Analyzer Mode. Frequency & SPAN 6.1.1 Frequency & Span Set the frequency-related parameters and functions of the analyzer. The sweep will restart every time the frequency parameters are modified.
  • Page 138: Freq Offset

    SSA5000A Series User Manual Command Format [:SENSe]:FREQuency:SPAN [:SENSe]:FREQuency:SPAN? Instruction Sets the frequency span. Setting the span to 0 Hz puts the analyzer into zero span. Gets span value. Parameter Type Float, unit: Hz, kHz, MHz, GHz Parameter Range 5kHz~40MHz Return...
  • Page 139 SSA5000A Series User Manual measurement channels rapidly and continuously.  There are two kinds of frequency step modes: Auto and Manual. In Auto mode, the Freq step is 1/10 of the span in Non-zero span or equals the RBW while in Zero Span. In Manual mode, you can set the step using the numeric keys.
  • Page 140 SSA5000A Series User Manual Set the RBW (Resolution Bandwidth) and filter type. Resolution Bandwidth. Set the resolution bandwidth in order to distinguish between signals which have frequency components that are near one another.  Reducing the RBW will increase the frequency resolution.
  • Page 141 SSA5000A Series User Manual Command Format [:SENSe]:FILTer:TYPE KAISer | HANNing | FLATtop | GAUSsian | BHARris | RECTangular [:SENSe]:FILTer:TYPE? Instruction Sets FFT window function. Gets FFT window function. Parameter Type Enumeration Return KAIS HANN FLAT GAUS BHAR RECT Example :FILT:TYPE KAIS...
  • Page 142: Sweep

    SSA5000A Series User Manual Sweep Sets parameters about the Sweep functions, including acquisition time, sweep mode, sweep times, etc. 6.3.1 Acquisition Time Set the acquisition time of real-time spectrum analyzer within the real-time analysis span. The acquisition time can be set in “Auto” or “Manual” mode and the default is “Auto”.
  • Page 143 SSA5000A Series User Manual the system will enter continuous sweep mode and measure continuously if the trigger conditions are satisfied.  In continuous sweep mode, the system will send a trigger initialization signal automatically and enter the trigger condition judgment directly after each sweep.
  • Page 144: Amplitude

    SSA5000A Series User Manual Amplitude Set the amplitude parameters of the analyzer. Through modifying these parameters, signals under measurement can be displayed in a proper mode for easier observation and minimum error. 6.4.1 Ref Level Set the maximum power or voltage that can be currently displayed in the trace window. The value is displayed at the upper left corner of the screen grid.
  • Page 145: Rf Preamp

    SSA5000A Series User Manual  The maximum input attenuation can be set to 31 dB. When the set parameters do not meet the above formula, you can adjust the reference level by yourselve. Command format [:SENSe]:POWer[:RF]:ATTenuation [:SENSe]:POWer[:RF]:ATTenuation? [:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1 [:SENSe]:POWer[:RF]:ATTenuation:AUTO?
  • Page 146: Unit

    SSA5000A Series User Manual Parameter Range 1 dB ~ 20 dB Return Float , unitdB Example :DISPlay:WINDow:TRACe:Y:PDIVision 10 dB 6.4.5 Unit The unit can be dBm, dBmV, dBuV, Volts and Watts. The default unit is dBm. The conversion between units is as follows:...
  • Page 147: Trigger

    SSA5000A Series User Manual Trigger The trigger type can be PvT, frequency template trigger (FMT), Free Run and External. Command format :TRIGger[:SEQuence]:SOURce IMMediate|PVT|EXTernal :TRIGger[:SEQuence]:SOURce? Instructions Setting the Trigger Type Get trigger type Parameter type Enumeration Parameter Range IMMediate: Free to trigger...
  • Page 148: External

    SSA5000A Series User Manual Parameter Range 0~500ms Example :TRIGger[:SEQuence]:LEVel: DELay 20ms 6.5.3 External In this mode, an external signal (TTL signal) is input from the [TRIGGER IN] connector at the rear panel and trigger signals are generated when this signal fulfills the specified trigger edge condition.
  • Page 149: Fmt

    SSA5000A Series User Manual 6.5.4 Real-time frequency template limiting allows users to limit acquisition based on specific events in the frequency domain. The user can customize the template shape and select the frequency template mask type (greater than, less than, within and outside the template) according to the actual needs, or set the template action (normal, beep and stop), and the defined frequency template can also be saved as LIM file.
  • Page 150 SSA5000A Series User Manual Command format :TRIGger[:SEQuence]:FMT:ACTion NORMal|BEEPer|STOP :TRIGger[:SEQuence]:FMT:ACTion? Instructions Set the template trigger action | chirp | normal stop Gets the template trigger action Parameter type Enumeration Parameter Range NORMal BEEPer STOP Return Enumeration: NORM | BEEP| STOP Example :TRIGger[:SEQuence]:FMT:ACTion STOP int.siglent.com...
  • Page 151: Trace

    SSA5000A Series User Manual Trace The sweep signal is displayed as a trace on the screen. Command format :TRACe[1]|2|3 [:DATA]? Instructions Get trace data Return String Example :TRACe:DATA? Command format :TRACe[:DATA]:SPECtrum? Instructions Obtain SPECtrum data Return String Example :TRACe: SPECtrum?
  • Page 152: Trace State

    SSA5000A Series User Manual new minimum is generated in successive sweeps. Command format :TRACe[1]|2|3:TYPE WRITe|MAXHold|MINHold| AVERage :TRACe[1]|2|3:TYPE? Instructions Sets/gets the display type of trace Parameter type Enumeration Parameter Range WRITe: The trace is in normal mode, and the data is updated...
  • Page 153 SSA5000A Series User Manual Positive Peak For each trace point, Positive Peak detector displays the maximum value of data sampled within the corresponding time interval. Negative Peak For each trace point, Negative Peak detector displays the minimum value of data sampled within the corresponding time interval.
  • Page 154: Marker & Peak

    SSA5000A Series User Manual Marker & Peak The marker appears as a rhombic sign (as shown below) for identifying points on a trace. You can easily read the amplitude, frequency and sweep time of the marked point on the trace.
  • Page 155: Select Trace

    SSA5000A Series User Manual 6.7.2 Select Trace Select the trace to be marked by the current marker. Valid selections include A, B, C. Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:TRACe 1|2|3 :CALCulate:MARKer[1]|2|3|4|5|6|7|8:TRACe? Instructions Sets/Gets the marker trace Parameter type Enumeration Parameter Range 1|2|3|4|5|6 Return...
  • Page 156 SSA5000A Series User Manual 6.7.3.3 Fixed One of the marker types. When “Fixed” is selected, the X-axis and Y-axis of the marker will not change by the trace and can only be changed through the menu. The fixed marker is marked with "+".
  • Page 157: Relative To

    SSA5000A Series User Manual 6.7.4 Relative To “Relative to” is used to measure the delta values of X (Frequency or Time) and Y (Amplitude) between two markers which can mark on different traces. After the marker selects “Delta”, the original marker will become the delta measurement marker, and the related marker of the incrementing sequence number will become the reference “fixed”...
  • Page 158: Left Peak

    SSA5000A Series User Manual 6.7.8 Left Peak Search for and mark the nearest peak which is located at the left side of the current peak and meets the peak search condition. Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum:LEFT Instructions Marker searches for left peaks and marks with the specified marker.
  • Page 159: Marker To

    SSA5000A Series User Manual 6.7.12 Marker to 6.7.12.1 M->CF Set the center frequency of the analyzer to the frequency of the current marker.  If the Normal marker is selected, the center frequency will be set to the frequency of the current marker.
  • Page 160: Meas & Meas Setup

    SSA5000A Series User Manual Meas & Meas setup The real-time spectrum analyzer Mode provides several observation window combinations, including Density, spectrum+spectrogram, spectrogram, PvT, 3D+spectrogram. Command format :DISPlay:VIEW[:SELect] {type} :DISPlay:VIEW[:SELect]? Instructions Sets the current display view. Query the current view. Parameter type...
  • Page 161: Spectrogram

    SSA5000A Series User Manual The “Density” view uses colour of bitmaps to represent the signal density. Density is defined as the number of probability that frequency and amplitude points are hit during the capture interval (Acq Time). In this view, the X axis represents frequency, the Y axis represents amplitude, and the colour represents signal density.
  • Page 162: Spectrum + Spectrogram

    SSA5000A Series User Manual while D1 and D2 are the latest trace by default. Reading makers on traces D1 and D2, observe the position (time of occurrence) of the corresponding traces in the historical data, as well as the frequency and amplitude of signals.
  • Page 163: Pvt

    SSA5000A Series User Manual 6.8.4 Figure 6-4 PvT In the time domain, after the input data (IQ data) of FFT is detected, the corresponding PvT data can be obtained. The detection period is also the corresponding acquisition time. The Aladdin RTSA supports up to 50000 PVT traces for cyclic storage, and each PvT trace corresponds to a spectrogram trace.
  • Page 164 SSA5000A Series User Manual Instructions Sets/gets the average number of traces Parameter type Integer Parameter Range 1 ~ 999 Return Integer Example :AVERage:TRACe1:COUNt 10 Command format [:SENSe]:AVERage:TRACe[1]|2|3:CLEar Instructions Restart the average. Example :AVERage:TRACe2:CLEar 6.8.5.2 Persistence Sets the time when a frequency / amplitude display point's brightness fades in the persistence bitmap.
  • Page 165 SSA5000A Series User Manual size of the view frame, the start position and end position are linked. For Spectral Mode: End Position – Start Position = 600. For Spectrum + Spectrum Mode: End Position – Start Position = 298. For PvT Mode: End Position – Start Position = 298.
  • Page 166 SSA5000A Series User Manual Instructions Sets/Gets the color display scheme. Parameter type Enumeration Parameter Range WARM | COLD | GRAY Return Enumeration Example :DISPlay:VIEW:THEMe WARM 6.8.5.5 Global center frequency Set whether the current CF acts on other scan modes at the same time.
  • Page 167 SSA5000A Series User Manual between the template points, save and load the template points. Command format :CALCulate:LLINe[1]|2|3|4|5|6:TYPE UPPer|LOWer :CALCulate:LLINe[1]|2|3|4|5|6:TYPE? Instructions Sets/Gets the restriction type Parameter type Enumeration Parameter Range UPPer|LOWer Return Enumeration Example :CALCulate:LLINe1:TYPE LOWer Command format :CALCulate:LLINe[1]|2|3|4|5|6:Offset:X :CALCulate:LLINe[1]|2|3|4|5|6:Offset:X? Instructions...
  • Page 168 SSA5000A Series User Manual Parameter Range val1 : related with Span val2 : -400 dBm~330 dBm Example :CALCulate:LLINe1:ADD 100,-20 :CALCulate:LLINe2:POINt:DELete 2 Command format :CALCulate:LLINe[1]|2|3|4|5|6:DELete :CALCulate:LLINe:ALL:DELete Instructions Delete specified restrictions Delete all restrictions Example :CALCulate:LLINe1:DELete :CALCulate:LLINe:ALL:DELete Command format :CALCulate:LLINe[1]|2|3|4|5|6:TRACe :CALCulate:LLINe[1]|2|3|4|5|6:TRACe? Instructions...
  • Page 169 SSA5000A Series User Manual Instructions Set or query the amplitude reference type Parameter type Enumeration Parameter Range FIXed|RELAtive Return FIXed|RELAtive Example :CALCulate:LLINe2: AMPLitude:CMODe FIX Command format :CALCulate:LLINe[1]|2|3|4|5|6:COPY :CALCulate:LLINe[1]|2|3|4|5|6:COPY? Instructions Copy the limit Parameter type Integer Parameter Range Return Example :CALCulate:LLINe2:COPY 5...
  • Page 170: Modulation Analyzer Mode

    SSA5000A Series User Manual Modulation Analyzer Mode MA mode includes digital modulation analysis (DMA) and analog modulation analysis (AMA). Press Mode key to select and add the expected mode to window management. Press Meas Setup key to enter the corresponding measurement parameter configuration menu.
  • Page 171 SSA5000A Series User Manual real-time measurement value. Statistics are disabled by default. After performing a remeasurement, the statistical results will be cleared and the statistics will be restarted. If the averaging function is turned on, the average calculation of the measurement results will also be cleared and restarted.
  • Page 172: Demod

    SSA5000A Series User Manual 7.1.2 Demod 7.1.2.1 Demod Type & Format Modulation types are as follows: QAM modulation : 8QAM , 16QAM , 32QAM , 64QAM , 128QAM , 256QAM PSK modulation : BPSK , QPSK , 8PSK , DBPSK , DQPSK , D8PSK , pi/4 DQPSK ,...
  • Page 173 SSA5000A Series User Manual Command format :READ:DDEMod? Instructions Obtain digital demodulation results if demod type is ASK it will return: ASK err rms (% rms) ASK err peak (% pk) symbol position of ASK err peak carrier power carrier offset...
  • Page 174 SSA5000A Series User Manual The sign point is 8 , Maximum sign rate is 18.75e6 The sign point is 10 , Maximum sign rate is 15e6 The sign point is 12 , Maximum sign rate is 12.5e6 The sign point is 14 , Maximum sign rate is 10.714285e6 The sign point is 16 , Maximum sign rate is 9.375e6...
  • Page 175: Filter

    SSA5000A Series User Manual 7.1.3 Filter 7.1.3.1 Meas Filter Enable and select Meas Filter. Meas Filters that can be set include:  Root Raised Cosine  Raised Cosine  Gauss  Half Sine Command format [:SENSe]:DDEMod:FILTer[:MEASurement] [:SENSe]:DDEMod:FILTer[:MEASurement]? Instructions Set up digital demodulation measurement filter...
  • Page 176: Burst/Sync Search

    SSA5000A Series User Manual 7.1.3.3 Filter Length Sets the filter length used by the analyzer. This feature applies to MeasRef Filter and Ref Filter. Command format [:SENSe]:DDEMod:FILTer:RLENgth [:SENSe]:DDEMod:FILTer:RLENgth? Instructions Set the filter length Gets the filter length Parameter type Float...
  • Page 177 SSA5000A Series User Manual 7.1.4.2 Burst Search Burst search measures the burst power (pulses) in a signal and uses this to segment and isolate the signal for subsequent display and analysis. Using burst search, you can avoid the interference of invalid signals to the analysis process.
  • Page 178 SSA5000A Series User Manual symbol in order to make sure that the burst search finds bursts that are very close to each other. However, in case the "capture buffer" does not contain very close bursts, it is recommended that you increase the value.
  • Page 179: Bert

    SSA5000A Series User Manual Parameter type Integer Parameter Range -10000~10000 Return Integer Example :DDEMod:SYNC:SWORd:OFFSet 2 7.1.4.8 Sync Pattern Edit and display sync codewords. Command format [:SENSe]:DDEMod:SEGMent:BER:PATTern [:SENSe]:DDEMod:SEGMent:BER:PATTern? Instructions Set the BERT symbol Get BERT symbol Parameter type String Parameter Range...
  • Page 180: Analog Modulation Analysis (Ama)

    SSA5000A Series User Manual Analog Modulation Analysis (AMA) Used for modulation analysis of analog signals, a series of indicators such as carrier power, modulation rate, THD can be obtained through analysis. Analog modulation analysis includes AM, FM, PM modulation. The modulation mode, IF bandwidth, and equivalent filter can be selected, and the data can be averaged.
  • Page 181: Demod Type

    SSA5000A Series User Manual 7.2.1 Demod Type Analog modulation analysis includes AM, FM, PM modulation. Command format [:SENSe]:ADEMod:STYLe [:SENSe]:ADEMod:STYLe? Instructions Sets the analog demodulation type (cannot be set when not analog demodulation) Get analog demodulation type (cannot be queried when not analog demodulation)
  • Page 182: Average

    SSA5000A Series User Manual frequency. Command format :CALCulate:EQLPf:INDEx :CALCulate:EQLPf:INDEx? Instructions Set up analog demodulation equalization filter Obtain analog demodulation equalization filter Parameter type Enumeration Parameter Range Return Example :CALCulate:EQLPf:INDEx 2 7.2.4 Average Turn on or off the averaging option for measurement results, and you can set the number of items involved in the averaging calculation.
  • Page 183: Freq

    SSA5000A Series User Manual Freq 7.3.1 Freq & Span In MA mode, only the center frequency can be configured. Span cannot be set and is only used to show the equivalent channel bandwidth under the current configuration. Command Format [:SENSe]:FREQuency:CENTer...
  • Page 184: Eqbw

    SSA5000A Series User Manual 7.4.1 EQBW The MA mode does not support configuring the resolution bandwidth, and only displays the equivalent resolution bandwidth. Command Format [:SENSe]:BWIDth[:RESolution]? Instruction Querys equalization BW. Parameter Type None Parameter Range None Return Float, unit: Hz...
  • Page 185: Sweep

    SSA5000A Series User Manual Sweep 7.5.1 Measure/Sweep Control (Single/Continue/Restart) Sweep/Measure: Single/Continue controls the analyzer to perform a single sweep/measurement or continuous sweep/measurement. Restart: Restart the current sweep or measurement. In particular, in the continue mode, modifying some parameters will equivalently perform a restart sweep or measurement.
  • Page 186: Trigger

    SSA5000A Series User Manual Trigger 7.6.1 Trigger Source The analyzer provides a variety of trigger sources to suit different triggering needs. Free Run Free trigger is the default use mode of the analyzer, at this time, the spectrum analyzer loops and continuously scans.
  • Page 187: Trigger Level

    SSA5000A Series User Manual 7.6.2 Trigger Level Sets the amplitude level of the video trigger (only absolute levels are supported). A trigger occurs when the slope of the video signal crosses this level. When the selected trigger source is video trigger, the trigger level will be displayed as an orange line,...
  • Page 188: Zero Span Trigger Delay Compensation (External Trigger )

    SSA5000A Series User Manual {type}:" VIDeo ", " EXTernal " ,"FRAMe" Parameter type Float Parameter Range -500ms-500ms Return Float Example :TRIGger:EXTernal:DELay 5e-3 :TRIGger:FRAMe:DELay? 7.6.5 Zero Span Trigger Delay Compensation (External Trigger ) Under normal circumstances, after the trigger is generated, the displayed data and the data at the same time as the trigger are displayed, but because the processing time of the trigger path and the data path are different, the data displayed at the trigger time is the previous data.
  • Page 189: Reset Offset Display (Period)

    SSA5000A Series User Manual 7.6.8 Reset Offset Display (Period) The reset period triggers the time offset display. Modifying this parameter does not modify the absolute value of the internal offset. Command format :TRIGger:FRAMe:OFFSet:DISPlay:RESet Instructions Reset Period trigger offset return to zero...
  • Page 190: Auto Trigger

    SSA5000A Series User Manual 7.6.10 Auto Trigger Automatic triggering is an auxiliary triggering method used in non-Free run mode. When the user needs continuous triggering but the triggering conditions of the selected triggering type are not satisfied, automatic triggering can be used. After using the automatic trigger, the count starts after a measurement is completed.
  • Page 191 SSA5000A Series User Manual the trigger level (trigger threshold) for a single accumulation time exceeding the specified time before triggering. For the periodic trigger, the trigger moment is high level, the duration is a time period, and the other time is low level, which can be inferred according to other triggers.
  • Page 192: Ampt

    SSA5000A Series User Manual Ampt 7.7.1 Attenuator & RF Preamp According to the amplitude of the input signal, the user can set the corresponding RF front-end attenuator and amplifier, the purpose of which is to avoid display distortion when inputting large signals and reduce noise when inputting small signals.
  • Page 193 SSA5000A Series User Manual If the display type is I-Eye: -1000 ~ 1000 If the display type is Q-Eye: -1000 ~ 1000 If the display type is Wrap Phase: -1000 ~ 1000 If the display type is Unwrap Phase: -1000 ~ 1000...
  • Page 194: Trace

    SSA5000A Series User Manual Trace 7.8.1 Select Trace Select the current trace you want to select. After a trace is selected, parameters such as the reference level of the trace can be adjusted. You can also touch the screen and click the window where the trace is located to select the trace.
  • Page 195: Trace Display And Layout(Ama

    SSA5000A Series User Manual 7.8.4 Trace Display and Layout(AMA) Display up to 3 windows, each window displays different traces, respectively:  Time domain waveform;  Frequency domain waveform;  Demodulation result parameters. 7.8.5 Select Trace Select the trace in order to set the corresponding trace parameters. You can also select the trace by clicking on the trace mark displayed in the left status bar of the screen.
  • Page 196: Format

    SSA5000A Series User Manual 7.8.8 Format Select the displayed format of the trace. Command Format :TRACe[1]|2|3|4:FORMat[:Y] :TRACe[1]|2|3|4:FORMat[:Y]? Instruction Sets trace format Gets trace format Parameter Type Enumeration Parameter Range MLOG: Log Mag MLINear: Lin Mag REAL: Real IMAGinary: Imag IQ: I-Q...
  • Page 197: Eye Length

    SSA5000A Series User Manual 7.8.9 Eye Length Set the length of the Eye diagram. Command Format :TRACe:DEMod:EYE:LENGth :TRACe:DEMod:EYE:LENGth? Instruction Sets eye length. Gets eye length. Parameter Type Integer Parameter Range 2 ~ 40 Return Integer Example :TRACe:DEMod:EYE:LENGth 4 7.8.10 Symbol Table Display the demodulation digital symbols (binary or hex).
  • Page 198: Marker

    SSA5000A Series User Manual Marker 7.9.1 Select Marker & Select Trace To operate a marker, it must first be selected as the marker for the current operation. When there are multiple active marker, the marker currently operating in the waveform area will be displayed at the front, while other marker will be hollowed out (filled in black).
  • Page 199 SSA5000A Series User Manual When closing a marker, the marker type of other relative markre that it is a relative marker will automatically change to normal. Command Format :TRACe[1]|2|3|4:MARKer[1]|2|3|4:ENABle OFF|ON|0|1 :TRACe[1]|2|3|4:MARKer[1]|2|3|4:ENABle? Instruction Sets marker state. Gets marker state. Parameter Type...
  • Page 200: Reset Delta

    SSA5000A Series User Manual 7.9.4 Reset Delta Pressing this control is exactly the same as pressing the “Delta” selection on the Marker Mode radio button. The selected marker becomes a Delta Marker . If the selected marker is already a Delta marker, the reference marker is moved to the current position of the selected marker, thus resetting the Delta to zero.
  • Page 201: Emi Measurement Mode

    SSA5000A Series User Manual EMI Measurement Mode The user interface for EMI Measurement mode has three display regions showing information regarding different setting menus as shown in the figure below. Figure 8-1 EMI Measurement User Interface  Region 1#: Spectrum and setting information of scanning.
  • Page 202 SSA5000A Series User Manual search is based on the defined peak excursion and peak threshold under menu Search. A cross mark is added onto the trace for each peak signal found. Figure 8-2 CISPR-recommended EMI test flow Figure 8-3 Meas Menu For each of the peak signals found, the instrument is tuned to the signal frequency in zero span and dwells for the specified dwell time under menu Meas.
  • Page 203: Frequency & Span

    SSA5000A Series User Manual Frequency & Span 8.1.1 Meter Frequency Set the frequency of the meter measurement. The meter frequency value is displayed at the bottom of the meter window.  The frequency of the meters can be at any position, even outside the span displayed on the screen.
  • Page 204: Axis Scale

    SSA5000A Series User Manual  The span can be set down to 100 Hz and up to the full span described in Specifications. When the span is set to the maximum, the analyzer enters full span mode.  Modifying the span may cause an automatic change in RBW if they are in Auto mode.
  • Page 205: Sweep

    SSA5000A Series User Manual Sweep 8.3.1 Sweep Config 8.3.1.1 RBW / Step The parameter RBW/Step specifies the scanning step and the number of scanning points under the specified span. 8.3.1.2 Auto RDS RDS means RBW divided by Step. Auto RDS specifies RBW/Step value by span and rbw value.
  • Page 206: Amplitude

    SSA5000A Series User Manual Amplitude Set the amplitude parameters of the analyzer. Through modifying these parameters, signals under measurement can be displayed in a proper mode for easier observation and minimum error. Any change of Ref Level, Attenuator Value, Preamp mode and Ref Offset will restart sweep.
  • Page 207: Rf Preamp

    SSA5000A Series User Manual Command format [:SENSe]:POWer[:RF]:ATTenuation [:SENSe]:POWer[:RF]:ATTenuation? Instructions Sets/gets the attenuation value Parameter Type Integer Parameter Range 0 dB ~ 50 dB (Even gears only) Return Integer , unit dB Example :POWer:ATTenuation 10 Command format [:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1 [:SENSe]:POWer[:RF]:ATTenuation:AUTO? Instructions...
  • Page 208: Y Axis Scale Type

    SSA5000A Series User Manual 8.4.5 Y Axis Scale Type Set the scale type of the Y-axis to Lin or Log. The default is Log.  In Lin mode, the vertical Scale value cannot be changed. The Display area is set for reference level of 0%.
  • Page 209: Ref Offset

    SSA5000A Series User Manual Command format :DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing LINear|LOGarithmic :DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing? Instructions Sets/Gets the scale display type Parameter Type Enumeration: LINear |LOGarithmic Return Enumeration: LIN|LOG Example :DISPlay:WINDow:TRACe:Y:SPACing LINear 8.4.6 Ref Offset Assign an offset to the reference level to compensate for gains or losses generated between the device under measurement and the analyzer.
  • Page 210: Trigger

    SSA5000A Series User Manual Trigger The analyzer provides a variety of trigger functions, users can choose from the trigger menu. 8.5.1 Trigger Source The analyzer provides a variety of trigger sources to suit different trigger requirements. Free Run Free trigger is the default mode of the analyzer, in which the spectrum analyzer sweeps circularly and continuously.
  • Page 211: Trigger Slope

    SSA5000A Series User Manual 8.5.3 Trigger Slope Set the trigger polarity for external trigger, video trigger. The options are rising edge trigger and falling edge trigger. The same trigger source uses the same trigger edge for both gating and triggering.
  • Page 212: Trace

    SSA5000A Series User Manual Trace The scan signal is displayed as a trace on the screen. Command format :TRACe[1]|2|3| [:DATA]? Instructions Get trace data Return String Example :TRACe:DATA? Command format :FORMat[:TRACe][:DATA]ASCii|REAL32| REAL :FORMat[:TRACe][:DATA]? Instructions Sets/gets the format of trace data read...
  • Page 213: Trace State

    SSA5000A Series User Manual Retain the maximum level for each point of the selected trace. Update the data if a new maximum level is detected in successive sweeps. Max Hold is very effective when measuring events that may take successive scans to measure accurately. Some common applications include FM Deviation, AM NRSC, and frequency hopping or drift.
  • Page 214: Detect

    SSA5000A Series User Manual Refreshed trace data but no display. Command format :TRACe[1]|2|3:DISPlay[:STATe] :TRACe[1]|2|3:DISPlay[:STATe]? Instructions Sets/gets the display status of the trace Parameter Type Enumeration: ACTI|VIEW|BLAN| BACK Parameter Range ACTIve: Trace is in normal mode. Update data VIEW: Stops updating trace to display current trace data...
  • Page 215 SSA5000A Series User Manual range of the signal is clearly shown. Average For each trace point, Average detector displays the average value of data sampled within the corresponding time interval. Command format [:SENSe]:DETector:TRACe[1]|2|3|[:FUNCtion] [:SENSe]:DETector:TRACe[1]|2|3[:FUNCtion]? Instructions Sets/Gets the trace detection type...
  • Page 216: Marker & Peak

    SSA5000A Series User Manual Marker & Peak The marker appears as a rhombic sign (as shown below) for identifying points on a trace. You can easily read the amplitude, frequency and sweep time of the marked point on the trace.
  • Page 217 SSA5000A Series User Manual  You can use the numeric keys, knob or direction keys to move the marker. The readouts of the marker will be displayed at the upper right corner of the screen.  The readout resolution of the X-axis (frequency or time) is related to the span. For higher readout resolution, reduce the span.
  • Page 218: Marker Postion

    SSA5000A Series User Manual Instructions Sets/Gets the marker mode Parameter Type Enumeration Parameter Range POSition DELTa FIXed Return Enumeration: POS|DELT|FIX|OFF Example :CALCulate:MARK1:MODE POSition Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:STATe OFF|ON|0|1 :CALCulate:MARKer[1]|2|3|4|5|6|7|8:STATe? Instructions Sets/gets the marker switch status Parameter Type Boolean Parameter Range OFF|ON|0|1...
  • Page 219: Relative To

    SSA5000A Series User Manual Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:Y :CALCulate:MARKer[1]|2|3|4|5|6|7|8:Y? Instructions Read the value of the Y-axis of the marker point, which can also be used to read the marker noise in the marker function. To execute this command, ensure that the marker is in the onstate.:CALCulate:MARKer[1]|2|3|4|5|6|7|8:STATe...
  • Page 220: Marker Lines

    SSA5000A Series User Manual reciprocal is infinite and 100 Ts is displayed. Note: This type is invalid in Zero span. Time In this type, the Normal marker shows the time difference between the marker and the start of the sweep; while Delta marker and Delta Pair marker show the sweep time difference between the delta marker and reference marker.
  • Page 221: Marker

    SSA5000A Series User Manual Return Example :CALCulate:MARKer2:X:LINE:STATe 1 :CALCulate:MARKer2:X:LINE:STATe? 8.7.8 Marker -> 8.7.8.1 M->CF Set the center frequency of the analyzer to the frequency of the current marker.  If the Normal marker is selected, the center frequency will be set to the frequency of the current marker.
  • Page 222 SSA5000A Series User Manual Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8[:SET]:START Instructions Sets/gets the value of the marker X axis to the starting frequency, valid when the marker is on Example :CALCulate:MARKer1:START 8.7.8.4 M -> Stop Freq Set the stop frequency of the analyzer to the frequency of the current marker.
  • Page 223: Peak Search

    SSA5000A Series User Manual :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MODE Example :CALCulate:MARKer2:DELTa:SPAN ΔM->CF 8.7.8.7 Set the center frequency of the analyzer to the frequency difference between the two markers in Delta marker type.  If the Normal marker is selected, this function is invalid. ...
  • Page 224 SSA5000A Series User Manual markers. Wherein, the result of peak search is marked with the delta marker and the result of minimum search is marked with the reference marker. Peak -> CF Execute peak search and set the center frequency of the analyzer to the frequency of the peak. The function is invalid in Zero Span.
  • Page 225 SSA5000A Series User Manual 8.7.9.2 Peak Config Define the conditions of peak search for various peak searches. A real peak should meet the requirements of both the “Peak Threshold” and “Peak Excursion”. Peak Threshold Assign a minimum for the peak amplitude. Peaks whose amplitudes are greater than the specified peak threshold are treated as real peaks.
  • Page 226 SSA5000A Series User Manual 8.7.9.3 Count Peak Enable or disable continuous peak search. The default is Off. When enabled, the system will always execute a peak search automatically after each sweep in order to track the signal under measurement. Command format...
  • Page 227: Limits

    SSA5000A Series User Manual Limits The analyzer supports Pass/Fail test function. In this function, the measured curve will be compared with the pre-edited curve. If the related rules are met, the result is “Pass”, else the result is “Fail”. 8.8.1 Slect Limit Activate/Select a limit.
  • Page 228: Limit Type

    SSA5000A Series User Manual 8.8.4 Limit Type Set the limit type as upper / lower . The limit 1,3,5 is default as the lower and 2,4,6 as the lower Command format :CALCulate:LLINe[1]|2|3|4|5|6:TYPE UPPer|LOWer :CALCulate:LLINe[1]|2|3|4|5|6:TYPE? Instructions Sets/Gets the restriction type Parameter Type...
  • Page 229 SSA5000A Series User Manual X-axis Edit the X-axis value (frequency or time) of the current point. Amplitude Edit the amplitude of the current point or line. Del Point Delete the point whose number is selected in Mode. Del All Delete all the points.
  • Page 230 SSA5000A Series User Manual Parameter Type val1:frequency : Float, val2: Ampl: Float Parameter Range val1: related with Span val2: -400 dBm~330 dBm Example :CALCulate:LLINe1:ADD 100,-20 :CALCulate:LLINe2:POINt:DELete 2 Command format :CALCulate:LLINe[1]|2|3|4|5|6:DELete :CALCulate:LLINe:ALL:DELete Instructions Delete specified restrictions Delete all restrictions Example :CALCulate:LLINe1:DELete...
  • Page 231 SSA5000A Series User Manual Command format :CALCulate:LLINe[1]|2|3|4|5|6: AMPLitude:CMODe :CALCulate:LLINe[1]|2|3|4|5|6: AMPLitude:CMODe? Instructions Set or query the amplitude reference type Parameter Type Enumeration Parameter Range FIXed|RELAtive Return FIXed|RELAtive Example :CALCulate:LLINe2: AMPLitude:CMODe FIX Command format :CALCulate:LLINe[1]|2|3|4|5|6:COPY :CALCulate:LLINe[1]|2|3|4|5|6:COPY? Instructions Copy the limit Parameter Type...
  • Page 232: Limit Test

    SSA5000A Series User Manual 8.8.6 Limit Test Enable or disable the limit test function. Figure 8-5 test results Command format :CALCulate:LLINe:TEST :CALCulate:LLINe:TEST? Instructions Sets/gets the status of the test switch Parameter Type Boolean Parameter Range OFF|ON|0|1 Return Example :CALCulate:LLINe:TEST 1...
  • Page 233 SSA5000A Series User Manual Turn on or off the buzzer. When the buzzer is on, it beeps when the test result is “Fail”. Command format :CALCulate:LLINe:CONTrol:BEEP :CALCulate:LLINe:CONTrol:BEEP? Instructions Sets/gets the restricted buzzer Parameter Type Boolean Parameter Range OFF|ON|0|1 Return Example...
  • Page 234: Meas & Meas Setup

    SSA5000A Series User Manual Meas & Meas Setup 8.9.1 Average Type Choose one of the following averaging types: log power (video), power (RMS), or voltage averaging. When trace average is on, the average type is shown on the left side of the display.
  • Page 235: Meas Control

    SSA5000A Series User Manual Command format [:SENSe]:AVERage:TYPE:AUTO 0|1|ON|OFF [:SENSe]:AVERage:TYPE:AUTO? Instructions Set/query the average type automatically Parameter type Boolean Parameter Range OFF|ON|0|1 Return Example :AVERage:TYPE:AUTO 1 8.9.2 Meas Control 8.9.2.1 Sequence Display the menu controls that enable you to configure the measurement sequence. You need to press Restart Seq to start the selected measurement sequence.
  • Page 236: Scan

    SSA5000A Series User Manual 8.9.2.4 Restart Meter Start or restart the measure meter. 8.9.3 Scan A group of menus let you configure for scan quickly and easily. 8.9.4 Measure 8.9.4.1 Signal mode Set the remeasure type, either on a current signal, all signals, or the marked signal in the signal list.
  • Page 237: List Operation

    SSA5000A Series User Manual 8.9.5 List Operation 8.9.5.1 Select Signal Select one of the signals as current signal in the signal list. When a signal is selected, you can mark, unmark or delete it from the signal list. It will be relative to 1.4.5.1 Meas Signal.
  • Page 238: Meter

    SSA5000A Series User Manual 8.9.6 Meter A group of menus let you configure for meter measurement quickly and easily 8.9.6.1 Reset Peak Hold Reset the maximum hold value of the meter and start accumulating again. 8.9.6.2 Select Meter Select one of the three meters. When a meter is selected, you can set its detector and limit configure.
  • Page 239: Pulse Measurement Mode

    SSA5000A Series User Manual Pulse Measurement Mode The PULSE mode function is mainly used to measure the width, amplitude, and time parameters of pulses. Command format INSTrument:MEASure INSTrument:MEASure? Instruction Sets/Gets the Measurement mode Parameter Type Enumeration Parameter Range PULSE Return...
  • Page 240: Frequency

    SSA5000A Series User Manual Frequency 9.1.1 Frequency Set the frequency related parameters and functions of the spectrum analyzer. Command Format [:SENSe]:FREQuency:CENTer [:SENSe]:FREQuency:CENTer? Instruction Sets/Gets the center frequency. Parameter Type Float, unit: Hz, kHz, MHz, GHz Parameter Range Zero Span:0 ~ 7.5 GHz...
  • Page 241 SSA5000A Series User Manual Command Format [:SENSe]:FREQuency:CENTer:STEP[:INCRement] [:SENSe]:FREQuency:CENTer:STEP[:INCRement]? Instruction Specifies the center frequency step size. Gets the center frequency step. Parameter Type Float, unit: Hz, kHz, MHz, GHz Parameter Range 1 Hz ~ 28 GHz Return Float, unit: Hz Example...
  • Page 242: Resolution Bandwidth

    SSA5000A Series User Manual 9.2.1 Resolution Bandwidth Set Resolution BandWidth (RBW) to distinguish two signals with similar frequencies. Pay attention to the following points during use:  Reducing RBW can achieve higher frequency resolution, but it will lead to longer scanning time;...
  • Page 243: V/R Ratio

    SSA5000A Series User Manual Command Format [:SENSe]:BWIDth:VIDeo:AUTO OFF|ON|0|1 [:SENSe]:BWIDth:VIDeo:AUTO? Instruction This command turns on/off auto video bandwidth state. Gets the video bandwidth state. Parameter Type Boolean Parameter Range OFF|ON|0|1 Return Example BWIDth:VIDeo:AUTO OFF 9.2.3 V/R Ratio Set the ratio of VBW to RBW. This value is different while measuring different kinds of signals: ...
  • Page 244: Sweep

    SSA5000A Series User Manual Sweep 9.3.1 Sweep Points The number of sweep points represents the number of sweep and trace displayed points (201~10001). More sweep points will improve the resolution of waveform, but also affect the minimum sweep time, and increase the time of data processing and remote access to data, and reduce the response rate.
  • Page 245: Sweep/Measure

    SSA5000A Series User Manual Parameter Type Boolean Parameter Range OFF|ON|0|1 Return Example :SWEep:TIME:AUTO ON Command Format [:SENSe]:SWEep:SPEed NORMal|ACCUracy [:SENSe]:SWEep:SPEed? Instruction Toggles the sweep speed between normal and accuracy. Parameter Type Enumeration Parameter Range ACCUracy|NORMal Example :SWEep: SPEed NORMal 9.3.3 Sweep/Measure...
  • Page 246: Amplitude

    SSA5000A Series User Manual Amplitude Set the amplitude parameters of the analyzer. Through modifying these parameters, signals under measurement can be displayed in a proper mode for easier observation and minimum error. Any change of Ref Level, Attenuator Value, Preamp mode and Ref Offset will restart sweep.
  • Page 247: Ref Offset

    SSA5000A Series User Manual Return Example :POWer[:RF]:ATTenuation:AUTO ON Command format [:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1 [:SENSe]:POWer[:RF]:GAIN[:STATe]? Instructions Sets/gets the preset amplifier inside the switch spectrometer Parameter Type Boolean Parameter Range OFF|ON|0|1 Return Example :POWer:GAIN ON 9.4.2 Ref Offset Assign an offset to the reference level to compensate for gains or losses generated between the device under measurement and the analyzer.
  • Page 248 SSA5000A Series User Manual Set the unit of the Y-axis to dBm, dBmV, dBuV, dBuA, Volts (RMS) or Watts. Default is dBm. The conversion relationships between units are as follows. Where, R denotes the reference impedance. The default value is 50Ω and can be adjusted by pressing “Correction ->...
  • Page 249 SSA5000A Series User Manual Volts dBμV = 20lg ( 1μV Volts dBmV = 20lg ( Volts dBμA = 10 lg ( × ) − 10 lg ( R ) +10lg10 Volts Watts = The impedance selection here only represents numerical calculation and does not represent the switching of actual impedance.
  • Page 250: Correction

    SSA5000A Series User Manual Correction Measured value can be corrected in specific x-axis and y-axis. Now, there are eight corrections, witch enter into force at the same time. Select Correction Select a correction (1-8) to operating. Correct Switch If the selected switch enter into force.
  • Page 251 SSA5000A Series User Manual Menu Input/Output>Correction>Edit Correction example :CORRection:CSET1:POINt:DELete 0 Close All the Correction All the correction is not effective. Delet One/All the Correction Empty all the points of a specail/all correction(s). Command Format [:SENSe]:CORRection:CSET[1]2|3|...|8:DELete Illustration Delete all points of a special correction.
  • Page 252: Trigger

    SSA5000A Series User Manual Trigger The analyzer provides a variety of trigger functions, users can choose from the trigger menu. 9.6.1 Trigger Source The analyzer provides a variety of trigger sources to suit different trigger requirements. Free Run Free trigger is the default mode of the analyzer, in which the spectrum analyzer sweeps circularly and continuously.
  • Page 253: Trigger Level

    SSA5000A Series User Manual 9.6.2 Trigger Level Sets the amplitude level for the video trigger (absolute level only supported). When the video signal crosses the voltage level with the selected slope, it is triggered. Command format :TRIGger[:SEQuence]: {type}:LEVel :TRIGger[:SEQuence]: {type}:LEVel? Instructions sets the trigger level.
  • Page 254: Zero Span Trigger Delay Compensation (External Trigger Only)

    SSA5000A Series User Manual Parameter Type Float Return Float Example :TRIGger:EXTernal:DELay 5e-3 :TRIGger:FRAMe:DELay? Command format :TRIGger[:SEQuence]:{type}:DELay:STATe :TRIGger[:SEQuence]:{type}:DELay:STATe? Instructions sets the trigger delay state. gets the trigger delay state. {type}:" VIDeo ", " EXTernal " ,"FRAMe" Parameter Type Boolean Parameter Range...
  • Page 255: Offset Time

    SSA5000A Series User Manual 9.6.7 Offset Time Adjust the cumulative offset between the periodic trigger clock and trigger events. The periodic trigger clock cannot be viewed on the software, only the trigger event can be seen. Therefore, in order to adjust the trigger event time, only the offset between the periodic triggering clock and the triggering event can be adjusted.
  • Page 256: Sync Source (Periodic Trigger Only)

    SSA5000A Series User Manual 9.6.9 Sync Source (periodic trigger only) Figure 9-1 Trigger of synchronization source Command format :TRIGger[:SEQuence]:FRAMe:SYNC :TRIGger[:SEQuence]:FRAMe:SYNC? Instructions Set/Query the type of periodic synchronization Parameter Type Enumeration Parameter Range "OFF", "EXT" Return "OFF", "EXT" Example :TRIGger:FRAMe:SYNC EXT 9.6.10...
  • Page 257 SSA5000A Series User Manual Set the trigger source of the gate to a periodic source. Similar to trigger, you can configure trigger period, trigger offset, and trigger period synchronization sources when configuring periodic sources. The configuration of the gating source affects the configuration of the trigger source.
  • Page 258: Trace

    SSA5000A Series User Manual Trace The waveform trace is obtained by connecting a set of discrete points to represent the scanned signal. The connection method is shown in the following figure: Figure 9-2 Selection of trace lines The x value of the trace point represents frequency or time (in the case of zero sweep width), and the y value represents the signal amplitude of the current frequency (or time).
  • Page 259: Trace State

    SSA5000A Series User Manual Max Hold Retain the maximum level for each point of the selected trace. Update the data if a new maximum level is detected in successive sweeps. Max Hold is very effective when measuring events that may take successive scans to measure accurately.
  • Page 260: Detect

    SSA5000A Series User Manual View The trace data will not be refreshed, and the current latest trace will be displayed in a fixed frame. Blank Trace data is no longer refreshed or displayed Background Refreshed trace data but no display.
  • Page 261 SSA5000A Series User Manual value of the sample data segment in turn: Odd-numbered data points display the maximum value and even-numbered data points display the minimum value. In this way, the amplitude variation range of the signal is clearly shown.
  • Page 262: Marker & Peak

    SSA5000A Series User Manual Marker & Peak The marker appears as a rhombic sign (as shown below) for identifying points on a trace. You can easily read the amplitude, frequency and sweep time of the marked point on the trace.
  • Page 263 SSA5000A Series User Manual frequency of the current trace.  You can use the numeric keys, knob or direction keys to move the marker. The readouts of the marker will be displayed at the upper right corner of the screen.
  • Page 264: Marker Postion

    SSA5000A Series User Manual Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MODE POSition|DELTa|FIXed|OFF :CALCulate:MARKer[1]|2|3|4:MODE? Instructions Sets/Gets the marker mode Parameter Type Enumeration Parameter Range POSition DELTa FIXed Return Enumeration: POS|DELT|FIX|OFF Example :CALCulate:MARK1:MODE POSition Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:STATe OFF|ON|0|1 :CALCulate:MARKer[1]|2|3|4|5|6|7|8:STATe? Instructions Sets/gets the marker switch status Parameter Type...
  • Page 265: Readout Type

    SSA5000A Series User Manual Example :CALCulate:MARKer4:X 0.4 GHz :CALCulate:MARKer4:X 200 ms :CALCulate:MARKer4:X? Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:Y :CALCulate:MARKer[1]|2|3|4|5|6|7|8:Y? Instructions Read the value of the Y-axis of the marker point, which can also be used to read the marker noise in the marker function.
  • Page 266: Marker Couple

    SSA5000A Series User Manual Inverse Time = 1 / Δ Time. Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:X:READout FREQuency | TIME | PERiod | INTIme :CALCulate:MARKer[1]|2|3|4|5|6|7|8:X:READout? Instructions Sets/gets the marker in X-axis reading mode Parameter Type Enumeration Parameter Range FREQuency: frequency TIME PERiod INVERSE_TIME...
  • Page 267: Marker Table

    SSA5000A Series User Manual Parameter Range ON|OFF|0|1 Return Example :CALCulate:MARKer2:X:LINE:STATe 1 :CALCulate:MARKer2:X:LINE:STATe? 9.8.8 Marker Table Enable or disable the Marker Table. Display all the markers enabled on the lower portion of the screen, including marker number, trace number, marker readout type, X-axis readout and amplitude. Through this table you can view the measurement values of multiple points.
  • Page 268 SSA5000A Series User Manual between the two points in the active function area. "----" would be displayed if the search fails. The parameters in the figure are shown as: Figure 9-3 N dB parameter Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:BANDwidth:NDB :CALCulate:MARKer[1]|2|3|4|5|6|7|8:BANDwidth:NDB? Instructions Set/Obtain the N dB bandwidth reference value...
  • Page 269: Peak Search

    SSA5000A Series User Manual Instructions Sets/gets the marker frequency counter state Parameter Type Boolean Parameter Range ON|OFF|0|1 Return Example :CALCulate:MARK1:FCOunt 1 Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:FCOunt:X? Instructions Read the marker frequency counter reading Return Float Example :CALCulate:MARK1:FCOunt:X? 9.8.9.3 Noise Marker Execute the Noise marker function for the selected marker and read the normalized noise power spectral density.
  • Page 270 SSA5000A Series User Manual Search for and mark the nearest peak which is located at the left side of the current peak and meets the peak search condition. Next Right Peak Search for and mark the nearest peak which is located at the right side of the current peak and meets the peak search condition.
  • Page 271 SSA5000A Series User Manual :CALCulate:MARKer:PEAK:SEARch:MODE :CALCulate:MARKer:PEAK:THReshold :CALCulate:MARKer:PEAK:EXCursion) Example :CALCulate:MARKer1:MAXimum:LEFT Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:PTPeak Instructions Performs a peak-to-peak search, marking with the specified marker Example :CALCulate:MARKer1:PTPeak Command format :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum[:SET]:CENTer Instructions Execute peak search and set the center frequency of the analyzer to the frequency of...
  • Page 272 SSA5000A Series User Manual Command format :CALCulate:MARKer:PEAK:EXCursion :CALCulate:MARKer:PEAK:EXCursion? Instructions Sets/gets a relative threshold for the peak search criteria Parameter Type Float , unitdB Parameter Range 0 ~ 200.0 dB Return Float , unitdB Example :CALCulate:MARKer:PEAK:EXCursion 10 Command format :CALCulate:MARKer:PEAK:EXCursion:STATe :CALCulate:MARKer:PEAK:EXCursion:STATe?
  • Page 273 SSA5000A Series User Manual Command format :CALCulate:PEAK:TABLe? Instructions Get peak table data Return String Example :CALCulate:PEAK:TABLe? 9.8.10.5 Sort Order Sort all signals in the peak table by the specific order. Sort order based on the frequency / time(zero span) , ampt or delta to limit.
  • Page 274: Meas & Meas Setup

    SSA5000A Series User Manual Meas & Meas Setup Provide optional measurement functions. When activated, the screen will be divided into two parts. The above part is the measure screen which displays traces, and the other part is used to display measurement results.
  • Page 275 SSA5000A Series User Manual Command format [:SENSe]:AVERage:TYPE LOGPower|POWer|VOLTage [:SENSe]:AVERage:TYPE? Instructions Set/Query the average type Parameter type Enumeration Parameter Range LOGPower POWer VOLTage Return Enumeration: LOGP|POW|VOLT Example AVERage:TYPE VOLTage Command format [:SENSe]:AVERage:TYPE:AUTO 0|1|ON|OFF [:SENSe]:AVERage:TYPE:AUTO? Instructions Set/query the average type automatically Parameter type...
  • Page 276 SSA5000A Series User Manual frequency, etc. Trace average The sum of all amplitudes of the entire trace divided by the number of scan points (dBm). Determine the top (TOP) and bottom (BASE) of the pulse, positive peak (PosPeak), negative peak...
  • Page 277 SSA5000A Series User Manual Pulse duration and pulse center The pulse duration is the time difference between a positive pulse (negative pulse) passing through the pulse duration reference (RefDuration, 50%) on the rising edge (falling edge) and a falling edge (rising edge) passing through the pulse duration reference (RefDuration, 50%).
  • Page 278 SSA5000A Series User Manual Instructions Set/Get Amplitude Line Switch Parameter Type Boolean Parameter Range OFF|ON|0|1 Return Example :DISPlay:WINDow:TRACe:Y:DLINe:STATe 1 Command format [:SENSe] :DISPlay:WINDow:TRACe:Y:DLINe [:SENSe] :DISPlay:WINDow:TRACe:Y:DLINe? Instructions Setting/Get amplitude line values Parameter Type Enumeration Parameter Range Return Example :DISPlay:WINDow:TRACe:Y:DLIN 20 Command format...
  • Page 279 SSA5000A Series User Manual Command format [:SENSe]:PULSe:LABel:LEVel:REFerence:DURation [:SENSe]:PULSe:LABel:LEVel:REFerence:DURation? Instructions Set/Get Duration Ref line Switch Parameter Type Boolean Parameter Range OFF|ON|0|1 Return Example :PULSe:LABel:LEVel:REFerence:DURation 1 Command format [:SENSe]:PULSe:REFerence:DURation [:SENSe]:PULSe:REFerence:DURation? Instructions Sets/gets Duration Ref line Position percentage Parameter Type Boolean Parameter Range...
  • Page 280 SSA5000A Series User Manual Command format [:SENSe] :PULSe:LEVel:USER:BOTTom [:SENSe] :PULSe:LEVel:USER:BOTTom? Instructions Sets/getsPulse Base value Parameter Type Boolean Parameter Range Return Example :PULSe:LEVel:USER:BOTTom 12 Command format [:SENSe]:PULSe:LABel:INSTant:CENTer [:SENSe]:PULSe:LABel:INSTant:CENTer? Instructions Sets/gets Center line Switch Parameter Type Boolean Parameter Range OFF|ON|0|1 Return Example...
  • Page 281 SSA5000A Series User Manual Command format [:SENSe]:PULSe:LABel:INSTant:REFerence:LOW [:SENSe]:PULSe:LABel:INSTant:REFerence:LOW? Instructions Sets/gets Low Instant Intersection switch Parameter Type Boolean Parameter Range OFF|ON|0|1 Return Example :AVERage:TYPE:AUTO 1 Command format [:SENSe]:PULSe:LABel:LEVel:REFerence:LOW [:SENSe]:PULSe:LABel:LEVel:REFerence:LOW? Instructions Sets/getsLow Ref line Switch Parameter Type Boolean Parameter Range OFF|ON|0|1 Return...
  • Page 282: Phase Noise Measurement Mode

    SSA5000A Series User Manual Phase Noise Measurement Mode The phase noise mode is mainly used for measuring single sideband phase noise testing analysis, jitter analysis, etc. 10.1 Frequency & Span 10.1.1 Carrier Frequency Specify the center frequency of the carrier wave.
  • Page 283: Stop Offset

    SSA5000A Series User Manual Command format [:SENSe]:LPLot:FREQuency:OFFSet:STARt [:SENSe]:LPLot:FREQuency:OFFSet:STARt? Instruction Sets/gets Start offset of scanned upper sideband Parameter Type Float, unit:Hz , kHz , MHz , GHz Parameter Range 1Hz~2.8GHz Return Float, unit:Hz Example :LPLot:FREQuency:OFFSet:STARt 100 Hz :SENSe:LPLot:FREQuency:OFFSet:STARt 0.2 GHz 10.1.3...
  • Page 284 SSA5000A Series User Manual carrier signal during each acquisition and returns a new carrier frequency if necessary. Command format [:SENSe]:FREQuency:CARRier:TRACk[:STATe] [:SENSe]:FREQuency:CARRier:TRACk[:STATe]? Instruction Sets/gets Signal tracking status Parameter Type bool Parameter Range 0|1|OFF|ON Return Example :FREQuency:CARRier:TRACk ON :SENSe:FREQuency:CARRier:TRACk:STATe OFF Command format...
  • Page 285: 10.2 Sweep

    SSA5000A Series User Manual 10.2 Sweep 10.2.1 Sweep Points Sweep Points characterized the number of scanning and trace display points (601~10001). More scanning points will improve the resolution of the waveform, while also affecting the minimum scanning time, increasing data processing time and remote access time, and reducing response rate.
  • Page 286: 10.3 Amplitude

    SSA5000A Series User Manual 10.3 Amplitude 10.3.1 Ref Level Set absolute power reference value. The position of the reference level can be set to top, middle, or bottom. Command format :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel? Instruction Sets/gets reference level. Parameter Type Float, unitdBc...
  • Page 287: 10.4 Marker

    SSA5000A Series User Manual 10.4 Marker 10.4.1 Select Marker Select one of the eight markers. The default is Marker1. When a marker is selected, you can set its type, trace to be marked, readout type and other related parameters. The enabled marker will appear on the trace selected through the Select Trace option and the readouts of this marker are also displayed in the active function area and at the upper right corner of the screen.
  • Page 288 SSA5000A Series User Manual  You can use the numeric keys, knob or direction keys to move the marker. The readouts of the marker will be displayed at the upper right corner of the screen.  The readout resolution of the X-axis (frequency or time) is related to the span. For higher readout resolution, reduce the span.
  • Page 289: Marker Postion

    SSA5000A Series User Manual Command format :CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:STATe OFF|ON|0|1 :CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:STATe? Instruction Sets/gets the marker switch status Parameter Type Boolean Parameter Range OFF|ON|0|1 Return Example :CALCulate:MARK1:STATe ON Command format :CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:REFerence 1|2|3|4 :CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:REFerence? Instruction Sets/Gets the marker relative to Parameter Type Enumeration Parameter Range...
  • Page 290 SSA5000A Series User Manual When the cursor function is turned on, if the current cursor is not open, the cursor type of the current cursor will be automatically changed to [Normal]. When the cursor is turned off, its cursor function will automatically turn off.
  • Page 291: Spurious Search

    SSA5000A Series User Manual Command format :CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:RMSNoise: RESult? Instruction Set/Get the measurement results of integrated noise Parameter Type Float, unit depending on the set result type Return Float, unit depending on the set result type Example :CALCulate:LPLot:MARKer1:RMSNoise:RESult? 10.4.6 Spurious search Search for stray signals on the original trace.
  • Page 292: 10.5 Trace

    SSA5000A Series User Manual 10.5 Trace 10.5.1 Select Trace The analyzer allows for up to four traces to be displayed at the same time. Each trace has its own color (Trace A - Yellow, Trace B - Purple, Trace C - Light blue).
  • Page 293: Trace Function

    SSA5000A Series User Manual BLANK Turn off the display of the trace and all measurement functions based on that trace. The trace in the closed state is not displayed, and the data is in the previous state. Command format :TRACe[1]|2|3:LPLot:MODE RAW|SMOothed|VIEW|BLANk...
  • Page 294 SSA5000A Series User Manual Command format :TRACe:PRESet:ALL Instruction Reset all traces Example :TRACe:PRESet:ALL Clear All Traces Clear all traces without affecting the state of any functions or variables in the instrument. Command format :TRACe:CLEar:ALL Instruction Change all trace data to default minimum values...
  • Page 295: 10.6 Meas

    SSA5000A Series User Manual 10.6 Meas 10.6.1 Average Command format [:SENSe]:LPLot:AVERage:STATe 0|1|OFF|ON [:SENSe]:LPLot:AVERage:STATe? Instruction Set/Get the average type Parameter Type bool Parameter Range 0|1|OFF|ON Return bool Example :LPLot:AVERage:STATe ON 10.6.1.1 Average/Hold Number Sets the terminal count number N for Average, Max Hold and Min Hold trace types. This number is an integral part of how the average trace is calculated.
  • Page 296: View

    SSA5000A Series User Manual 10.6.2 View The view menu allows for the selection of display views: Normal, decade, spurious. Command format :DISPlay:LPLot:VIEW NORMal|DECade|SPURious :DISPlay:LPLot:VIEW? Instruction Sets/gets the view mode Parameter Type Enumeration Parameter Range NORMal|DECade|SPURious Return EnumerationNORMal|DECade|SPURious Example :DISP:LPLot:VIEW NORMal...
  • Page 297: Carrier Amplitude Limit

    SSA5000A Series User Manual of data, while average refers to the average amplitude of the same frequency point in different frames. The smooth window setting requires an average number of points. For a trace, the amplitude value of each frequency point on the trace is centered on the current frequency point, and the number is the number of scan points multiplied by the smoothing window/100, which is the average amplitude value of the original data of these points.
  • Page 298 SSA5000A Series User Manual X Offset Set offsets of X axis. Y Offset Set offsets of Y axis X-scale Match the X-axis with the selected constraints as much as possible adjustment For linear frequency scaling, the range between the starting and ending frequencies is 12.5% higher than the range between the minimum and...
  • Page 299 SSA5000A Series User Manual Instructions Add limit point Delete limit points Parameter Type val1: frequency : Float, val2: Ampl: Float Parameter Range val1: related with Span val2: -400 dBm~330 dBm Example :CALCulate:LLINe1:ADD 100,-20 :CALCulate:LLINe2:POINt:DELete 2 Command format :CALCulate:LLINe[1]|2|3|4|5|6:DELete :CALCulate:LLINe:ALL:DELete Instructions...
  • Page 300 SSA5000A Series User Manual Command format :CALCulate:LLINe[1]|2|3|4|5|6: AMPLitude:CMODe :CALCulate:LLINe[1]|2|3|4|5|6: AMPLitude:CMODe? Instructions Set or query the amplitude reference type Parameter Type Enumeration Parameter Range FIXed|RELAtive Return FIXed|RELAtive Example :CALCulate:LLINe2: AMPLitude:CMODe FIX Command format :CALCulate:LLINe[1]|2|3|4|5|6:COPY :CALCulate:LLINe[1]|2|3|4|5|6:COPY? Instructions Copy the limit Parameter Type...
  • Page 301 SSA5000A Series User Manual Command format :CALCulate:LLINe:FAIL:STOP :CALCulate:LLINe:FAIL:STOP? Instructions The set/query limit test stops if it fails Parameter Type Boolean Parameter Range OFF|ON|0|1 Return Example :CALCulate:LLINe:FAIL:STOP OFF int.siglent.com...
  • Page 302: Noise Figure Measurement Mode

    SSA5000A Series User Manual Noise Figure Measurement Mode Noise Figure Mode is used to measure Noise Figure, Thermal Noise, Equivalent Noise Temperature, etc. Different measurement results can be selected to display under Trace menu. Calibration is required before measurement. Calibration requires entering the ENR provided by the noise source provider.
  • Page 303 SSA5000A Series User Manual -- USB Type-B: Power and control signal input for connecting to the spectrum analyzer. -- BNC: + 28V pulse power supply output for connecting the noise source to provide input bias. Connection: The NSD28 noise source driver is connected to the USB interface of the spectrum analyzer using the attached USB connection line.
  • Page 304: 11.1 Frequency & Span

    SSA5000A Series User Manual 11.1 Frequency & Span 11.1.1 Freq Context It is necessary to set or view the RF/IF/LO frequencies when measuring a frequency converting device. Freq Context defines the way frequencies are shown in the result window and the menu.
  • Page 305 SSA5000A Series User Manual 11.1.2.1 Swept Set the Center Frequency, Span, Start Frequency, Stop Frequency and Points. Measurement frequencies are linearly distributed Start Frequency and Stop Frequency. In this mode, Fixed Freq and Edit Freq List are invisible. When Layout is Auto, changes to Graph layout.
  • Page 306 SSA5000A Series User Manual 11.1.2.2 Fixed Set the Fixed Frequency that will be measured. In this mode, Center Freq, Span, Start Freq, Stop Freq and Edit Freq List are invisible. Points are fixed at 1 and cannot be edited (be grey out).
  • Page 307: Resolution Bandwidth

    SSA5000A Series User Manual 11.2 BW 11.2.1 Resolution Bandwidth Set the resolution bandwidth in order to distinguish between signals which have frequency components that are near one another. Command Format [:SENSe]:NFIGure:BWIDth|BANDwidth[:RESolution] [:SENSe]:NFIGure:BWIDth|BANDwidth[:RESolution]? Instruction Specifies the resolution bandwidth. For numeric entries, all RBW types choose the nearest (arithmetically, on a linear scale, rounding up) available RBW to the value entered.
  • Page 308: Points

    SSA5000A Series User Manual Parameter Range OFF|ON|0|1 Return Example :NFIGure:SWEep:TIME:AUTO 1 11.3.2 Points Command Format [:SENSe]:NFIGure:SWEep:POINts [:SENSe]:NFIGure:SWEep:POINts? Instruction Sets/Gets sweep points. Parameter Type Integer Parameter Range 2~10001 Return Integer Example :NFIGure:SWEep:POINts 2500 11.3.3 Sweep/Measure Sweep/Measure Single/Continue controls analyzer to perform single or continuous sweep/measure, Restart Restart the current sweep or measure.
  • Page 309: 11.4 Amplitude

    SSA5000A Series User Manual 11.4 Amplitude 11.4.1 Attenuator & Preamp & Ref Level Set the amplitude parameters of the analyzer. Through modifying these parameters, signals under measurement can be displayed in a proper mode for easier observation and minimum error. Any change of Ref Level, Attenuator Value, and Preamp mode will restart sweep.
  • Page 310: Y Axis Scale

    SSA5000A Series User Manual Instructions Sets/gets the auto attenuation value switch Parameter Type Boolean Parameter Range OFF|ON|0|1 Return Example :POWer:ATTenuation:AUTO ON Command format [:SENSe]:NFIGure:POWer[:RF]:GAIN[:STATe] [:SENSe]:NFIGure:POWer[:RF]:GAIN[:STATe]? Instructions Sets/gets the preamplifier inside the switch spectrometer Parameter Type Boolean Parameter Range OFF|ON|0|1 Return...
  • Page 311: Auto Scale

    SSA5000A Series User Manual Gain Y Factor T effective K (Kelvin) P hot dB or dBm/Hz P cold dB or dBm/Hz 11.4.4 Auto Scale Auto Scale can be used to adjust the scale of the current and all windows adaptively to the waveform data.
  • Page 312: Trace Type

    SSA5000A Series User Manual frequency point. More results can be displayed in a one-row table when using Result Table menu to turn on other results. Table: Displays discrete results for each frequency point in a tabular form. Command Format :DISPlay:NFIGure:FORMat...
  • Page 313: Trace State

    SSA5000A Series User Manual 11.5.3 Trace State Command format :TRACe[1]|2:NFIGure:DISPlay[:STATe] :TRACe[1]|2:NFIGure:DISPlay[:STATe]? Instructions Sets/gets the display status of the trace in Graph layout Parameter Type Enumeration Parameter Range ACTive: Trace is in normal mode. Update data VIEW: Stops updating trace to display current trace data...
  • Page 314: 11.6 Marker

    SSA5000A Series User Manual Instruction Sets/Gets display result in Meter layout. Parameter Type Enumeration Parameter Range NFIGure | NFACtor | GAIN | YFACtor | TEEFective | PHOT | PCOLd Return NFIGure | NFACtor | GAIN | YFACtor | TEEFective | PHOT | PCOLd...
  • Page 315: Marker All Off

    SSA5000A Series User Manual 11.6.3 Marker All Off Command Format :CALCulate:NFIGure:MARKer:AOFF Instruction Sets all markers turn to Off. Example :CALCulate:NFIGure:MARKer:AOFF 11.7 Meas Setup 11.7.1 Settings Sets the terminal count number N for Average, Max Hold and Min Hold trace types. This number is an integral part of how the average trace is calculated.
  • Page 316 SSA5000A Series User Manual Parameter Range AMPLifier: amplifier DOWNconv: downconverter UPConv: upconverter Return AMPL | DOWN | UPC Example :NFIGure:MODE:DUT DOWNconv :NFIGure:MODE:DUT? 11.7.3.2 Sideband Sets the LO offset of the DUT. When DSB is selected, only the frequencies of the LSB is displayed in the menu and diagram.
  • Page 317: Ext Lo Setup

    SSA5000A Series User Manual 11.7.3.5 Ext LO Power Command Format [:SENSe]:NFIGure:MODE:LOSCillator:DUT:PARameter:POWer[:LEVel] [:SENSe]:NFIGure:MODE:LOSCillator:DUT:PARameter:POWer[:LEVel]? Instruction Sets/Gets the fixed power of external LO source. Parameter Type Float Parameter Range -100 dBm ~ 100 dBm Return Float. Unit: dBm Example :NFIGure:MODE:LOSCillator:DUT:PARameter:POWer -50 11.7.4 Ext LO Setup 11.7.4.1...
  • Page 318 SSA5000A Series User Manual 11.7.4.3 Min Freq / Max Freq Specify the frequency range of the external LO source. Attempting to set the external LO source to a value less than Min Freq will clip to Min Freq, and attempting to set a value greater than Max Freq will clip to Max Freq.
  • Page 319 SSA5000A Series User Manual Denominator to 1e9, the command sent will be FREQ 20GHZ The auxiliary command is used to send commands between each setting of power and frequency in the following order: auxiliary command > power command > frequency command > auxiliary command >...
  • Page 320: Enr Setup

    SSA5000A Series User Manual 11.7.4.5 Add Source to List You can search, identify, connect, and control external LO devices to provide LO signal output through USB or LAN. One or more successfully identified devices will appear in the list of Available Source List, showing their manufacturer, model, and serial number information.
  • Page 321 SSA5000A Series User Manual re-calibrated frequency point is inconsistent with the original calibration data frequency point, the original calibration data will not be deleted. If it has not been calibrated, it cannot be measured. It will display on the window.
  • Page 322: Bluetooth Measurement Mode

    SSA5000A Series User Manual Bluetooth Measurement Mode The Bluetooth measurement function provides RF emission headphone testing that complies with the Bluetooth Core Specification standards to assist users in testing, evaluating, and verifying Bluetooth devices. It supports testing of low-power Bluetooth, basic speed, and enhanced data rate Bluetooth.
  • Page 323: 12.2 Amplitude

    SSA5000A Series User Manual 12.2 Amplitude 12.2.1 Scale Set the vertical division value of the Cartesian coordinate display format, which is only valid for RF envelope, Demod waveform, and RF spectrum waveform Command format :DISPlay:WINDow#:TRACe[1]|2|3|4:Y[:SCALe]:PDIVision :DISPlay:WINDow#:TRACe[1]|2|3|4:Y[:SCALe]:PDIVision? Instructions Set/get scale Parameter Type...
  • Page 324: 12.3 Marker/Peak

    SSA5000A Series User Manual 12.3 Marker/Peak 12.3.1 Select Marker Select one of the four markers. The default is Marker1. When a marker is selected, you can set its type, trace to be marked, and other related parameters. The enabled marker will appear on the trace selected through the Select Trace option and the readouts of this marker are also displayed in the active function area and at the upper right corner of the screen.
  • Page 325 SSA5000A Series User Manual 12.3.3.2 Delta One of the marker types. It is used to measure the delta values of X (Frequency or Time) and Y (Amplitude) between the reference point and a certain point on the trace. When selected, a pair of markers appears on the trace: Fixed Related Marker (marked by a combination of the marker number and letter “...
  • Page 326: Marker X

    SSA5000A Series User Manual Return Enumeration: POS|DELT|OFF Example :TRACe:MARKer:TYPE POSition Command format :CALCulate:MARKer#[:SET]:RESEt:DELTa Instruction Difference marker resets to 0 Only valid when the current cursor is a differential marker Example :CALCulate:MARKer2:RESEt:DELTa Command format :CALCulate:MARKer:AOFF Instruction Close all markers Example :CALCulate:MARKer:AOFF 12.3.4...
  • Page 327: Peak

    SSA5000A Series User Manual Parameter Type Enumeration Parameter Range 1|2|3|4|5|6|7|8 Return Enumeration Example :CALCulate:MARKer1:REFerence 3 12.3.6 Peak Set the current cursor position to the position of the peak with the maximum or minimum amplitude of the trace.  Peak: Set the current marker position to the position with the maximum amplitude on the trace.
  • Page 328: 12.4 Trace

    SSA5000A Series User Manual 12.4 Trace 12.4.1 Trace Num & Layout The screen can display up to 4 windows and arrange them according to the layout settings. The layout methods include:  Single  Stacked 2  Grid 1,2 ...
  • Page 329 SSA5000A Series User Manual 2.Demod Waveform Displays Demodulation Results. 3.RF spectrum int.siglent.com...
  • Page 330 SSA5000A Series User Manual By performing FFT on the collected data, it displays the frequency domain characteristics of the signal. 4.Result Sum Summary of measurement results that comply with Bluetooth core specifications. int.siglent.com...
  • Page 331 SSA5000A Series User Manual 5. Measured Polar Constellation diagram reflecting signal modulation performance (only valid when the result is Enhanced data rate type) Command Format :TRACe#:FORMat[:Y] :TRACe#:FORMat[:Y]? Instruction Set/get the format of the select trace Parameter Type Enumeration Parameter Range...
  • Page 332: 12.5 Sweep

    SSA5000A Series User Manual 12.5 Sweep Set the time for the instrument to capture a single data packet. Command Format [:SENSe]:SWEep:TIME [:SENSe]:SWEep:TIME? Instruction Set/get the sweep time Parameter Type Float with unit:ks,s,ms,us Parameter Range 1us ~ 1s Return Float, Unit s...
  • Page 333: 12.6 Measurement

    SSA5000A Series User Manual 12.6 Measurement Bluetooth Measurement provides support for Bluetooth measurement of three specifications: Basic Rate Enhanced Data Rate Low Energy According to the Bluetooth protocol, BR/EDR are both classic Bluetooth, and are used in the same measurement that can automatically recognizes signal. Compared to Classical Bluetooth, Low Energy Bluetooth has differences in channel usage, data rate, and data transmission methods.
  • Page 334 SSA5000A Series User Manual Δf1 Average Δf2 Average Δf1 Max Δf2 Max Δf2>185KHz Freq Drift Max Drift Rate Freq Offset Initial Freq Offset 12.6.1.1 Basic Rate & Enhanced Data Rate The main difference between Basic Rate and Enhanced Data Rate lies in transmission speed and transmission quality, while Enhanced Rate is an improvement on Basic Rate Bluetooth.
  • Page 335: Average

    SSA5000A Series User Manual Freq Drift Max Drift Rate ICFT Other signals only support demodulation and power calculation. 12.6.1.2 Enhanced Data Rate Enhanced data rate measurement supports the following types of signal demodulation: ESCO_2-EV3, ESCO_3-EV3, ESCO_2-EV5, ESCO_3-EV5 ACL_2-DH1, ACL_3-DH1, ACL_AUX, ACL_2-DH3, ACL_3-DH3, ACL_2-DH5, ACL_3-DH5...
  • Page 336: 12.7 Trigger

    SSA5000A Series User Manual 12.7 Trigger The analyzer provides a variety of trigger functions, users can choose from the trigger menu. 12.7.1 Trigger Source The analyzer provides a variety of trigger sources to suit different trigger requirements. Free Run Free trigger is the default mode of the analyzer, in which the spectrum analyzer sweeps circularly and continuously.
  • Page 337: Trigger Level

    SSA5000A Series User Manual 12.7.2 Trigger Level Sets the amplitude level for the video trigger (absolute level only supported). When the video signal crosses the voltage level with the selected slope, it is triggered. Command format :TRIGger[:SEQuence]: {type}:LEVel :TRIGger[:SEQuence]: {type}:LEVel? Instructions sets the trigger level.
  • Page 338: Trigger Delay

    SSA5000A Series User Manual Parameter Type Float Return Float Example :TRIGger:EXTernal:DELay 5e-3 :TRIGger:FRAMe:DELay? Command format :TRIGger[:SEQuence]:{type}:DELay:STATe :TRIGger[:SEQuence]:{type}:DELay:STATe? Instructions sets the trigger delay state. gets the trigger delay state. {type}:" VIDeo ", " EXTernal " ,"FRAMe" Parameter Type Boolean Parameter Range...
  • Page 339: Period (Periodic Trigger Only)

    SSA5000A Series User Manual same time. However, the processing time of the trigger path and the data path is different. As a result, the data displayed at the trigger time is the previous data. This does not affect the integrity of the data and does not cause data loss at the trigger point.
  • Page 340: Reset Time Offset Display

    SSA5000A Series User Manual 12.7.9 Reset Time Offset Display Reset the display of cycle trigger time offset. Command format :TRIGger[:SEQuence]:FRAMe:OFFSet:DISPlay:RESet Instructions Reset Period trigger offset return to zero Example :TRIGger:FRAMe:OFFSet:DISPlay:RESet 12.7.10 Sync Source (periodic trigger only) Figure 12-1 Trigger of synchronization source...
  • Page 341: Input And Output

    SSA5000A Series User Manual Input and Output 13.1 Freq Ref Input Frequency Reference Source include inside reference source, external reference source and auto-selection reference source. When external reference source is not connected, the button of external reference is not available. Defualt auto-selection reference source.
  • Page 342 SSA5000A Series User Manual Command Format [:SENSe]:CORRection:CSET#[:STATe] [:SENSe]:CORRection:CSET#[:STATe]? Illustration Set Correction Switch Status Parameter Type Boolean Menu Input/Output>Correction Return Value example :CORRection:CSET1 0 :CORRection:CSET2 1 Edit Correction Editing, preserving, loading the selected correction. Command Format [:SENSe]:CORRection:CSET[1]2|3|...|8:DATA {x1,y1 } [:SENSe]:CORRection:CSET#:DATA? Illustration...
  • Page 343: If Output

    SSA5000A Series User Manual Return Value example CORRection:CSET1:DELete Command Format [:SENSe]:CORRection:CSET:ALL:DELete Illustration Delete all the corrections Return Value example :CORRection:CSET:ALL:DELete IF Output 13.4 The intermediate frequency (IF) output signal can be used for external signal processing and can be obtained via the ‘IF Out’ output connector on the instrument's rear panel. Please note that the IF output is only useful in Zero Span.
  • Page 344 SSA5000A Series User Manual Use another spectrum analyzer to observe the IF signal. Configure the center frequency to 120MHz and the span to 50MHz. IF Output when input an 1GHz RF signal int.siglent.com...
  • Page 345 SSA5000A Series User Manual IF Output when input a 3.5 GHz NR signal int.siglent.com...
  • Page 346: System

    SSA5000A Series User Manual System 14.1 System 14.1.1 About Figure 14-1 about Area 1 Displays product information, including product name, serial number, Host ID, and IP address. Area 2 Displays firmware information including software version, FPGA version, CPLD version, and hardware version.
  • Page 347: Hardware

    SSA5000A Series User Manual 14.1.2 Hardware Figure 14-2 Hardware Area 1 displays the current operating status of the analyzer, including the RF board temperature, CPU temperature, and fan speed. Area 2 displays automatic alignment (temperature drift compensation) information, including calibration status, time of last calibration, temperature, and temperature difference.
  • Page 348: Connect Setting

    SSA5000A Series User Manual Press this key to select the desired display language. Command format :SYSTem:LANGuage CHINESE|ENGLISH :SYSTem:LANGuage? Instructions Setting languages Acquire languages Parameter type Enumeration Parameter Range CHINESE ENGLISH Return Enumeration: CHINESE|ENGLISH Example :SYSTem:LANGuage CHINESE :SYSTem:LANGuage? 14.1.5 Connect Setting 14.1.5.1...
  • Page 349 SSA5000A Series User Manual Return Gateway string :SYSTem:COMMunicate:LAN:GATeway “192.168.1.1” Example :SYSTem:COMMunicate:LAN:GATeway? :SYSTem:COMMunicate:LAN:SMASk {“xxx.xxx.xxx.xxx”} Command format :SYSTem:COMMunicate:LAN:SMASk? Instructions Set the subnet mask according to your computer network Settings Obtain the subnet mask Parameter type String Parameter Range The IP address must comply with the nic specifications (0-255:0-255:0-255:0-255)
  • Page 350: Date And Time

    SSA5000A Series User Manual 14.1.6 Date and Time Switch the time and date display status in the upper right corner of the screen. Modify the system time display format, including YMD, MDY, and Dmy. Command format :SYSTem:TIME :SYSTem:TIME? Instructions Setting the System Time...
  • Page 351 SSA5000A Series User Manual command information. To use the Help document, select a menu button, then press the Help button on the keyboard or touch the “Help on this setting”.At this point, the analyzer will jump to the relevant chapter of the corresponding button in the manual.
  • Page 352: 14.2 Reset

    SSA5000A Series User Manual 14.2 Reset 14.2.1 Preset Perform a reset operation based on the reset type currently set. Command format :SYSTem:PRESet Note Reset the device parameters based on the reset type Example :SYSTem:PRESet 14.2.2 Preset Type Perform the reset operation based on the current reset type. Select a configuration type for the spectrum analyzer to reset and load.Reset Settings Preset types include default, last time, and User.
  • Page 353: Power On

    SSA5000A Series User Manual 14.2.4 Power On Select the configuration to be loaded during power-on. Default configuration, last configuration, and User configuration are available Command format :SYSTem:PON:TYPE DFT|LAST|USER :SYSTem:PON:TYPE? Note Set the startup load configuration type Get the startup load configuration type...
  • Page 354: 14.3 Alignments

    SSA5000A Series User Manual 14.3 Alignments The alignment function is used to calibrate errors caused by temperature drift. Automatic alignment: After automatic alignment is turned on, the analyzer determines and triggers temperature error calibration logic based on temperature changes. Calibrate now: Perform a temperature error calibration immediately.
  • Page 355: 14.4 File

    SSA5000A Series User Manual 14.4 File Activate the analyzer file operation dialog box to perform file-related operations. Include: File Browser, File Browser, and Recall File. :MMEMory:STORe STA|TRC|COR|CSV|LIM|JPG|BMP|PNG, ”{file}” Command format Note Different modes support different types of file formats Parameter type characterstring :MMEMory:STORe STA, ”ABC.sta”...
  • Page 356: 14.5 Display

    SSA5000A Series User Manual 14.5 Display Set screen brightness. Set grid brightness in waveform area. Set buzzer state Set buzzer volume Command format :DISPlay:WINDow:TRACe:GRATicule:GRID:BRIGhtness :DISPlay:WINDow:TRACe:GRATicule:GRID:BRIGhtness? Instructions Set the grid brightness of the waveform area Get waveform area grid brightness Parameter type...
  • Page 357: 14.6 Power

    SSA5000A Series User Manual 14.6 Power This menu provides shutdown and restart operation buttons. Analysis Provides energy saving options. You can set the analyzer to automatically turn off the display if no operation is performed within a specified period of time.
  • Page 358: Remote Control

    SSA5000A Series User Manual Remote Control The analyzer features LAN, USB Device, and GPIB_USB module interfaces. By using a computer with these interfaces, and a suitable programming language (and/or NI-VISA software), users can remotely control the analyzer based on SCPI (Standard Commands for Programmable Instruments) command set, LabView and IVI (Interchangeable Virtual Instrument), to interoperate with other programmable instruments.
  • Page 359: Lan Port

    SSA5000A Series User Manual 15.1.2 LAN port Figure 15-1 LAN config Refer to the following steps to finish the connection via LAN: Install NI-VISA on your PC for VXI driver. Or without NI-VISA, using socket or telnet in your PC’s Operating System.
  • Page 360: Gpib-Usb Host Port

    SSA5000A Series User Manual 15.1.3 GPIB-USB Host port Refer to the following steps to finish the connection via USB: Install NI-VISA on your PC for GPIB driver. Connect the analyzer USB Host port to a PC’s GPIB card port, with SIGLENT USB-GPIB adaptor.
  • Page 361: 15.2 Build Communication

    SSA5000A Series User Manual 15.2 Build Communication 15.2.1 VISA NI-VISA includes a Run-Time Engine version and a Full version. The Run-Time Engine version provides NI device drivers such as USB-TMC, VXI, GPIB, etc. The full version includes the Run-Time Engine and a software tool named NI MAX that provides a user interface to control the device.
  • Page 362 SSA5000A Series User Manual Figure 15-5 Set the install path, default path is “C:\Program Files\National Instruments\”, you can change it. Click Next, dialog shown as above. Figure 15-6 Click Next twice, in the License Agreement dialog, select the “I accept the above 2 License Agreement(s).”, and click Next, dialog shown as below:...
  • Page 363: Sockets/Telnet

    SSA5000A Series User Manual Click Next to run installation. Figure 15-8 Now the installation is complete, reboot your PC. 15.2.2 Sockets/Telnet Through the LAN interface, VXI-11, Sockets and Telnet protocols can be used to communicate with the analyzer. VXI-11 is provided in NI-VISA, while Sockets and Telnet are commonly included in PC’s OS initially.
  • Page 364: 15.3 Remote Control Capabilities

    SSA5000A Series User Manual 15.3 Remote Control Capabilities 15.3.1 User-defined Programming Users can use SCPI commands to program and control the analyzer. For details, refer to the introductions in “Programming Examples”. 15.3.2 NI MAX Users can control the analyzer remotely by sending SCPI commands via NI-MAX software. NI_MAX is National Instruments Measurement and Automation Explorer.
  • Page 365 SSA5000A Series User Manual Note: The “*IDN?” command (known as the Identification Query) returns the instrument manufacturer, instrument model, serial number, and other identification information. 15.3.2.2 Using LAN Select “Add Network Device”, and select “VISA TCP/IP Resource” as shown: Run NI MAX software.
  • Page 366 SSA5000A Series User Manual Figure 15-13 After a brief scan, the connection should be shown under Network Devices: Figure 15-14 Right-click on the product and select Open NI-VISA Test Panel: Figure 15-15 Click “Input/Output” option button and click “Query” option button. If everything is OK, you will see the Read operation information returned as shown below.
  • Page 367: Web Browser

    SSA5000A Series User Manual Figure 15-16 15.3.3 Web browser The analyzer can be remotely controlled through PC or web browser of mobile terminal installing any driver. It mimics the touch screen/mouse clickable display function, just like a physical instrument. The browser also supports screenshot and firmware update functions.
  • Page 368: Scpi Overview

    SSA5000A Series User Manual SCPI Overview 16.1 Command Format SCPI commands present a hierarchical tree structure containing multiple subsystems, each of the subsystems is made up of a root keyword and several subkeywords. The command string usually starts with “:”, the keywords are separated by “:” and the followed parameter settings are separated by space.
  • Page 369: 16.3 Parameter Type

    SSA5000A Series User Manual Braces { } The parameters in the braces are optional which can be ignored or set for one or more times. For example: :CALCulate:LLINe[1]|2:DATA <x-axis>,<ampl>{,<x-axis>, <ampl>}, in the command, the {,<x-axis>, <ampl>} parameters can be ignored or set for one or more times.
  • Page 370: 16.4 Command Abbreviation

    SSA5000A Series User Manual example: [:SENSe]:BWIDth:VIDeo:RATio <number> The parameter <number> could only be one of 0.001, 0.003, 0.01, 0.03, 0.1, 0.3, 1.0, 3.0, 10.0, 30.0, 100.0, 300.0, 1000.0. 16.4 Command Abbreviation All of the commands are not case sensitive, so you can use any of them. But if abbreviation is used, all the capital letters in the command must be written completely.
  • Page 371: 16.5 Ieee Common Commands

    SSA5000A Series User Manual 16.5 IEEE Common Commands *IDN *RST *CLS *ESE *ESR? *OPC *SRE *STB? *WAI *TRG *TST? Command Format *IDN? Instruction Returns an instrument identification information string. The string will contain the manufacturer, model number, serial number, software number, FPGA number and CPLD number.
  • Page 372 SSA5000A Series User Manual Command Format *ESR? Instruction Queries and clears the standard event status event register. (This is a destructive read.) The value returned reflects the current state (0/1) of all the bits in the register. Menu None Example...
  • Page 373: 16.6 Scpi List

    SSA5000A Series User Manual 16.6 SCPI LIST 16.6.1 GPSA Mode/Meas :INSTrument[:SELect] :INSTrument:MEASure Freq [:SENSe]:FREQuency:CENTer [:SENSe]:FREQuency:STARt [:SENSe]:FREQuency:STOP [:SENSe]:FREQuency:CENTer:STEP[:INCRement] [:SENSe]:FREQuency:CENTer:STEP:AUTO [:SENSe]:FREQuency:OFFSet [:SENSe]:FREQuency:SPAN [:SENSe]:FREQuency:SPAN:FULL [:SENSe]:FREQuency:SPAN:ZERO [:SENSe]:FREQuency:SPAN:PREVious [:SENSe]:FREQuency:SPAN:HALF [:SENSe]:FREQuency:SPAN:DOUBle [:SENSe]:FREQuency:TUNE:IMMediate :DISPlay:WINDow:TRACe:X[:SCALe]:SPACing :CALCulate:MARKer:TRCKing[:STATe] [:SENSe]:BWIDth[:RESolution] [:SENSe]:BWIDth[:RESolution]:AUTO [:SENSe]:BWIDth:VIDeo [:SENSe]:BWIDth:VIDeo:AUTO [:SENSe]:BWIDth:VIDeo:RATio [:SENSe]:FILTer:TYPE Sweep [:SENSe]:SWEep:TIME [:SENSe]:SWEep:TIME:AUTO :INITiate:CONTinuous :INITiate[:IMMediate] [:SENSe]:SWEep:MODE...
  • Page 374 SSA5000A Series User Manual :DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision :DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing :UNIT:POWer :DISPlay:WINDow:TRACe:Y:SCALe:RLEVel:OFFSet [:SENSe]:POWer[:RF]:ATTenuation [:SENSe]:POWer[:RF]:ATTenuation:AUTO [:SENSe]:POWer[:RF]:GAIN[:STATe] Trigger :TRIGger[:SEQuence]:SOURce :TRIGger[:SEQuence]:{type}:LEVel :TRIGger[:SEQuence]:{type}:DELay :TRIGger[:SEQuence]:{type}:SLOPe :TRIGger[:SEQuence]:EXTernal:DELay:COMPensation :TRIGger[:SEQuence]:FRAMe:PERiod :TRIGger[:SEQuence]:FRAMe:OFFSet :TRIGger[:SEQuence]:FRAMe:OFFSet:DISPlay:RESet :TRIGger[:SEQuence]:FRAMe:SYNC [:SENSe]:SWEep:EGATe:SOURce [:SENSe]:SWEep:EGATe[:STATe] [:SENSe]:SWEep:EGATe:VIEW [:SENSe]:SWEep:EGATe:DELay [:SENSe]:SWEep:EGATe:LENGth [:SENSe]:SWEep:EGATe:METHod [:SENSe]:SWEep:EGATe:VIEW:STARt Trace TRACe:SELEct :TRACe[1]|2|3|4|5|6:TYPE :TRACe[1]|2|3|4|5|6:DISPlay[:STATe] :TRACe[1]|2|3|4|5|6 [:DATA]? :FORMat[:TRACe][:DATA] [:SENSe]:DETector:TRACe[1]|2|3|4|5|6[:FUNCtion] [:SENSe]:DETector:TRACe[1]|2|3|4|5|6:AUTO [:SENSe]:DETector:TRACe:AUTO:ALL :TRACe[1]|2|3|4|5|6:MATH:X :TRACe[1]|2|3|4|5|6:MATH:Y...
  • Page 375 SSA5000A Series User Manual :CALCulate:NTData[:STATe] :DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel :DISPlay:WINDow:TRACe:Y[:SCALe]:NRPosition :DISPlay:WINDow:NTTRace[:STATe] :TRACe:COPY :TRACe:EXCHange :TRACe:PRESet:ALL :TRACe:CLEar:ALL Marker :CALCulate:MARKer:SELEct :CALCulate:MARKer[1]|2|3|4|5|6|7|8:STATe :CALCulate:MARKer[1]|2|3|4|5|6|7|8:TRACe :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MODE :CALCulate:MARKer[1]|2|3|4|5|6|7|8:X :CALCulate:MARKer[1]|2|3|4|5|6|7|8:Y? :CALCulate:MARKer[1]|2|3|4|5|6|7|8:REFerence :CALCulate:MARKer:AOFF :CALCulate:MARKer:TABLe :CALCulate[:SELected]:MARKer:COUPle :CALCulate:MARKer[1]|2|3|4|5|6|7|8:X:READout :CALCulate:MARKer[1]|2|3|4|5|6|7|8:X:READout:AUTO :CALCulate:MARKer[1]|2|3|4|5|6|7|8:X:LINE:STATe :CALCulate:MARKer[1]|2|3|4|5|6|7|8:FUNCtion :CALCulate:MARKer[1]|2|3|4|5|6|7|8:BANDwidth[1]|2|3|4|5|6|7|8:NDB? :CALCulate:MARKer[1]|2|3|4|5|6|7|8:BANDwidth:RESult? :CALCulate:MARKer[1]|2|3|4|5|6|7|8:FCOunt[:STATe] :CALCulate:MARKer[1]|2|3|4|5|6|7|8:FCOunt:X? :CALCulate:MARKer[1]|2|3|4|5|6|7|8[:SET]:CENTer :CALCulate:MARKer[1]|2|3|4|5|6|7|8[:SET]:STEP :CALCulate:MARKer[1]|2|3|4|5|6|7|8[:SET]:START :CALCulate:MARKer[1]|2|3|4|5|6|7|8[:SET]:STOP :CALCulate:MARKer[1]|2|3|4|5|6|7|8[:SET]:RLEVel :CALCulate:MARKer[1]|2|3|4|5|6|7|8:DELTa[:SET]:SPAN :CALCulate:MARKer[1]|2|3|4|5|6|7|8:DELTa[:SET]:CENTer :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MINimum...
  • Page 376 SSA5000A Series User Manual :CALCulate:MARKer[1]|2|3|4|5|6|7|8:PTPeak :CALCulate:MARKer[1]|2|3|4|5|6|7|8:CPSearch[:STATe] :CALCulate:MARKer:PEAK:THReshold :CALCulate:MARKer:PEAK:THReshold:STATe :CALCulate:MARKer:PEAK:EXCursion :CALCulate:MARKer:PEAK:EXCursion:STATe :CALCulate:MARKer:PEAK:TABLe :CALCulate:PEAK:TABLe? :CALCulate:MARKer:PEAK:SORT :CALCulate:MARKer:PEAK:SORT:ORDER :CALCulate:MARKer:PEAK:TABLe:DTLimit:STATe :CALCulate:MARKer:PEAK:TABLe:DTLimit |1|2|3|4|5|6 Limit :CALCulate:LLINe[1]|2|3|4|5|6:STATe :CALCulate:LLINe[1]|2|3|4|5|6:TYPE :CALCulate:LLINe[1]|2|3|4|5|6:MARGin :CALCulate:LLINe[1]|2|3|4|5|6:MARGin:STATe :CALCulate:LLINe[1]|2|3|4|5|6:Offset:X :CALCulate:LLINe[1]|2|3|4|5|6:Offset:Y :CALCulate:LLINe[1]|2|3|4|5|6:DATA :CALCulate:LLINe[1]|2|3|4|5|6:ADD :CALCulate:LLINe[1]|2|3|4|5|6:POINt:DELete :CALCulate:LLINe[1]|2|3|4|5|6:DELete :CALCulate:LLINe:ALL:DELete :CALCulate:LLINe[1]|2|3|4|5|6:TRACe :CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:INTerpolate:TYPE :CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:CMODe :CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:INTerpolate:TYPE :CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:CMODe :CALCulate:LLINe[1]|2|3|4|5|6:COPY :CALCulate:LLINe[1]|2|3|4|5|6:BUILd :CALCulate:LLINe:TEST :CALCulate:LLINe:CONTrol:BEEP :CALCulate:LLINe:FAIL:STOP :CALCulate:LLINe[1]|2|3|4|5|6:FAIL?
  • Page 377 SSA5000A Series User Manual [:SENSe]:AVERage:TYPE :COUPle:ALL :DISPlay:WINDow:TRACe:Y:DLINe:STATe? :DISPlay:WINDow:TRACe:Y:DLINe :DISPlay:WINDow:TRACe:X:FLINe:STATe? :DISPlay:WINDow:TRACe:X:FLINe [:SENSe]:DEMod [:SENSe]:DEMod:EPHone [:SENSe]:DEMod:VOLume [:SENSe]:DEMod:TIME :INSTrument:COUPle:FREQuency:CENTer [:SENSe]:CHPower:BWIDth:INTegration [:SENSe]:CHPower:FREQuency:SPAN:POWer :UNIT:CHPower:POWer:PSD :CHPower:MEASure:CHPower? :CHPower:MEASure:CHPower:CHPower? :CHPower:MEASure:CHPower:DENSity? [:SENSe]:CHPower:AVERage:TCONtrol [:SENSe]:CHPower:FREQuency:SPAN:POWer :UNIT:CHPower:POWer:PSD :CHPower:MEASure:CHPower? :CHPower:MEASure:CHPower:CHPower? :CHPower:MEASure:CHPower:DENSity? [:SENSe]:CHPower:AVERage:TCONtrol ACPR [:SENSe]:ACPRatio:BWIDth:INTegration [:SENSe]:ACPRatio:OFFSet:BWIDth[:INTegration] [:SENSe]:ACPRatio:OFFSet[:FREQuency] :MEASure:ACPRatio:ACPower:MAIN? :MEASure:ACPRatio:LOWer:POWer? :MEASure:ACPRatio:UPPer:POWer? :MEASure:ACPRatio:LOWer? :MEASure:ACPRatio:UPPer? [:SENSe]:ACPower:AVERage:TCONtrol [:SENSe]:OBWidth:PERCent [:SENSe]:OBWidth:XDB...
  • Page 378 SSA5000A Series User Manual :MEASure:OBWidth:CENTroid? [:SENSe]:OBWidth:PREFerence [:SENSe]:OBWidth:INTegration[:METHod] :MEASure:OBWidth:OBWidth:FERRor? [:SENSe]:OBWidth:AVERage:TCONtrol T-Power [:SENSe]:TPOWer:FREQuency:CENTer [:SENSe]:TPOWer:LLIMit [:SENSe]:TPOWer:RLIMit :MEASure:TPOWer? [:SENSe]:TPOWer:AVERage:TCONtrol :MEASure:TOI? :MEASure:TOI:IP3? [:SENSe]:TOI:AVERage:TCONtrol Sepctrum [:SENSe]:SPECtrogram:STATe Monitor [:SENSe]:SPECtrogram:RESTart [:SENSe]:SPECtrogram:AVERage:TCONtrol [:SENSe]:CNRatio:BANDwidth:INTegration [:SENSe]:CNRatio:BANDwidth:NOISe [:SENSe]:CNRatio:OFFSet :CNRatio:MEASure:CNRatio? :CNRatio:MEASure:CNRatio:CARRier? :CNRatio:MEASure:CNRatio:NOISe? [:SENSe]: CNRatio:AVERage:TCONtrol Harmonics [:SENSe]:HARMonics:FREQuency:FUNDamental [:SENSe]:HARMonics:FREQuency:FUNDamental:AUTO [:SENSe]:HARMonics:FREQuency:STEP[:INCRement] [:SENSe]:HARMonics:FREQuency:STEP[:INCRement]:AUTO [:SENSe]:HARMonics:NUMBer [:SENSe]:HARMonics:SELect int.siglent.com...
  • Page 379: Rtsa

    SSA5000A Series User Manual 16.6.2 RTSA Freq [:SENSe]:FREQuency:CENTer [:SENSe]:FREQuency:STARt [:SENSe]:FREQuency:STOP [:SENSe]:FREQuency:CENTer:STEP[:INCRement] [:SENSe]:FREQuency:CENTer:STEP:AUTO [:SENSe]:FREQuency:OFFSet [:SENSe]:FREQuency:SPAN [:SENSe]:FREQuency:SPAN:FULL [:SENSe]:FREQuency:SPAN:ZERO [:SENSe]:FREQuency:SPAN:PREVious [:SENSe]:FREQuency:SPAN:HALF [:SENSe]:FREQuency:SPAN:DOUBle [:SENSe]:BWIDth[:RESolution] [:SENSe]:BWIDth[:RESolution]:AUTO [:SENSe]:FILTer:TYPE Sweep :INITiate[:IMMediate] :INITiate:CONTinuous [:SENSe]:ACQuisition:TIME :DISPlay:PAUSe Trigger :TRIGger[:SEQuence]:SOURce :TRIGger[:SEQuence]:LEVel:LEVel :TRIGger[:SEQuence]:LEVel:DELay :TRIGger[:SEQuence]:EXTernal:DELay :TRIGger[:SEQuence]:EXTernal:SLOPe :TRIGger[:SEQuence]:FMT:STATe :TRIGger[:SEQuence]:FMT:ACTion Ampt :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel :DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision :UNIT:POWer [:SENSe]:POWer[:RF]:ATTenuation...
  • Page 380 SSA5000A Series User Manual Trace :TRACe[1]|2|3 [:DATA]? :TRACe[:DATA]:SPECtrum? :TRACe[:DATA]:PVT? :TRACe[1]|2|3:TYPE :TRACe[1]|2|3DISPlay[:STATe] [:SENSe]:DETector:TRACe[1]|2|3|4|5|6[:FUNCtion] [:SENSe]:DETector:TRACe:PVTime [:SENSe]:DETector:TRACe:SPECtrogram Marker :CALCulate:MARKer[1]|2|3|4|5|6|7|8:TRACe :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MODE :CALCulate:MARKer[1]|2|3|4|5|6|7|8:X :CALCulate:MARKer[1]|2|3|4|5|6|7|8:Y? :CALCulate:MARKer[1]|2|3|4|5|6|7|8:REFerence :CALCulate:MARKer:AOFF :CALCulate:MARKer[1]|2|3|4|5|6|7|8[:SET]:CENTer :CALCulate:MARKer[1]|2|3|4|5|6|7|8[:SET]:START :CALCulate:MARKer[1]|2|3|4|5|6|7|8[:SET]:STOP :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MINimum :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum:NEXT :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum:LEFT :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum:RIGHt :CALCulate:MARKer[1]|2|3|4|5|6|7|8:PTPeak :CALCulate:MARKer[1]|2|3|4|5|6|7|8:CPSearch[:STATe] Linit :CALCulate:LLINe[1]|2|3|4|5|6:STATe :CALCulate:LLINe[1]|2|3|4|5|6:TYPE :CALCulate:LLINe[1]|2|3|4|5|6:MARGin :CALCulate:LLINe[1]|2|3|4|5|6:MARGin:STATe :CALCulate:LLINe[1]|2|3|4|5|6:Offset:X :CALCulate:LLINe[1]|2|3|4|5|6:Offset:Y :CALCulate:LLINe[1]|2|3|4|5|6:DATA :CALCulate:LLINe[1]|2|3|4|5|6:ADD...
  • Page 381 SSA5000A Series User Manual :CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:CMODe :CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:INTerpolate:TYPE :CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:CMODe :CALCulate:LLINe[1]|2|3|4|5|6:COPY :CALCulate:LLINe[1]|2|3|4|5|6:BUILd Meas :AVERage:TRACe[1]|2|3|4|5|6:COUNt :DISPlay:VIEW[:SELect] :DISPlay:VIEW:DENSity:PERSistence :DISPlay:VIEW:DENSity:PERSistence:INFinite :DISPlay:VIEW:SPECtrogram:TRAC:STOP :DISPlay:VIEW:SPECtrogram:TRAC:OFFSet int.siglent.com...
  • Page 382 SSA5000A Series User Manual 16.6.3 [:SENSe]:AVERage[:STATe] [:SENSe]:AVERage:COUNt [:SENSe]:DDEMod:MODulation :DDEMod[:FORMat]:SRATe [:SENSe]:DDEMod[:FORMat]:SYMBol:POINts [:SENSe]:DDEMod[:FORMat]:RLENgth [:SENSe]:DDEMod:FILTer[:MEASurement] [:SENSe]:DDEMod:FILTer:REFerence [:SENSe]:STATistic:STATe :CALCulate:RESTart :READ:DDEMod? [:SENSe]:DDEMod:SYNC:BURSt[:STATe] [:SENSe]:DDEMod:SYNC:SLENgth [:SENSe]:DDEMod:SYNC:BURSt:THREshold [:SENSe]:DDEMod:SYNC:BURSt:MINLength [:SENSe]:DDEMod:SYNC:BURSt:MINGap [:SENSe]:DDEMod:SYNC:SWORd[:STATe] [:SENSe]:DDEMod:SYNC:SWORd:OFFSet [:SENSe]:DDEMod:SYNC:SWORd:PATTern [:SENSe]:DDEMod:SEGMent:BER:STATe [:SENSe]:DDEMod:SEGMent:BER:PATTern [:SENSe]:ADEMod:STYLe :CALCulate:IFBW:INDEx :CALCulate:EQLPf:INDEx :READ:ADEMod? Freq [:SENSe]:FREQuency:CENTer [:SENSe]:FREQuency:CENTer:STEP[:INCRement] [:SENSe]:FREQuency:SPAN? [:SENSe]:BWIDth[:RESolution] [:SENSe]:DDEMod:FFT:WINDow:TYPE Sweep :INITiate[:IMMediate] :INITiate:CONTinuous Trigger...
  • Page 383 SSA5000A Series User Manual :TRIGger[:SEQuence]:{type}:SLOPe :TRIGger[:SEQuence]:FRAMe:PERiod :TRIGger[:SEQuence]:FRAMe:OFFSet :TRIGger[:SEQuence]:FRAMe:OFFSet:DISPlay:RESet :TRIGger[:SEQuence]:FRAMe:SYNC :TRIGger[:SEQuence]:ATRigger:STATe :TRIGger[:SEQuence]:ATRigger :TRIGger[:SEQuence]:HOLDoff:STATe :TRIGger[:SEQuence]:HOLDoff :TRIGger[:SEQuence]:HOLDoff:TYPE Ampt [:SENSe]:POWer[:RF]:ATTenuation [:SENSe]:POWer[:RF]:ATTenuation:AUTO :TRACe1|2|3|4:Y[:SCALe]:RLEVel :TRACe1|2|3|4:Y[:SCALe]:PDIVision :TRACe1|2|3|4[:Y]:AUToscale [:SENSe]:POWer[:RF]:GAIN[:STATe] Trace :CALCulate:PARameter:COUNt :DISPlay:LAYout :TRACe[1]|2|3|4:DATA:NAME :TRACe[1]|2|3|4:FORMat[:Y] :TRACe:DEMod:EYE:LENGth :TRACe:DEMod:TABLe:FORMat Marker :TRACe[1]|2|3|4:MARKer[1]|2|3|4:ENABle :TRACe[1]|2|3|4:MARKer[1]|2|3|4:TYPE :TRACe[1]|2|3|4:MARKer[1]|2|3|4:X :TRACe[1]|2|3|4:MARKer[1]|2|3|4:Y? :TRACe[1]|2|3|4:MARKer[1]|2|3|4:REFerence :CALCulate[:SELected]:MARKer:COUPle int.siglent.com...
  • Page 384: Emi

    SSA5000A Series User Manual 16.6.4 Freq [:SENSe]:FREQuency:CENTer [:SENSe]:FREQuency:STARt [:SENSe]:FREQuency:STOP [:SENSe] :FSCan:RANGe [:SENSe] :FREQuency:MIDSpan [:SENSe]:FREQuency:OFFSet :DISPlay:WINDow:TRACe:X[:SCALe]:SPACing [:SENSe]:BWIDth[:RESolution] [:SENSe]:BWIDth[:RESolution]:AUTO [:SENSe]:BANDwidth[:RESolution] [:SENSe]:BANDwidth[:RESolution]:AUTO [:SENSe]:FSCan:SCAN:BWIDth[:RESolution] [:SENSe]:FSCan:SCAN:BWIDth[:RESolution]:AUTO Sweep :INITiate:FSCan:CONTinuous :INITiate:METer:CONTinuous :SENSe]:SWEep:TIME :SENSe]:SWEep:COUNt :SENSe]:QPD:DWELl:TIME :SENSe]:SWEep:POINts :SENSe]:FSCan:SCAN:PRBW :SENSe]:FSCan:SCAN:PRBW:AUTO Trigger :TRIGger[:SEQuence]:SOURce :TRIGger[:SEQuence]:VIDeo:LEVel :TRIGger[:SEQuence]:LEVel :TRIGger[:SEQuence]:EXTernal:SLOPe :TRIGger[:SEQuence]:EXTernal:DELay:STATe :TRIGger[:SEQuence]:EXTernal:DELay Ampt :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel :DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision...
  • Page 385 SSA5000A Series User Manual Trace :TRACe:SELEct :TRACe:MODE :TRACe[1]|2|3:FSCan:TYPE :TRACe[1]|2|3:TYPE [:SENSe]:AVERage:TRACe[1]|2|3:COUNt [:SENSe]:AVERage:TRACe[1]|2|3:CLEar [:SENSe]:DETector:TRACe[1]|2|3[:FUNCtion] [:SENSe]:FSCan:DETector:TRACe[1]|2|3 [:SENSe]:DETector:TRACe[1]|2|3:AUTO [:SENSe]:DETector:TRACe:AUTO:ALL [:SENSe]:AVERage:TRACe[1]|2|3? :TRACe[1]|2|3 [:DATA]? :TRACe[1]|2|3:FSCan[:DATA]? :TRACe[1]|2|3:MAXimum:X? :TRACe[1]|2|3:MAXimum:Y? :FORMat[:TRACe][:DATA] :TRACe[1]|2|3:DISPlay[:STATe] :TRACe:COPY :TRACe:EXCHange :TRACe:PRESet:ALL Marker :CALCulate:MARKer:SELEct :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MODE :CALCulate:FSCan:MARKer[1]|2|3|4|5|6|7|8:MODE :CALCulate:MARKer[1]|2|3|4|5|6|7|8:STATe :CALCulate[:SELected]:MARKer:AOFF :CALCulate:MARKer[1]|2|3|4|5|6|7|8:TRACe :CALCulate:FSCan:MARKer[1]|2|3|4|5|6|7|8:TRACe :CALCulate:MARKer[1]|2|3|4|5|6|7|8:REFerence :CALCulate:FSCan:MARKer[1]|2|3|4|5|6|7|8:REFerence :CALCulate:MARKer[1]|2|3|4|5|6|7|8:RELative:TO:MARKer :CALCulate:MARKer[1]|2|3|4|5|6|7|8:RELative:TO :CALCulate:MARKer[1]|2|3|4|5|6|7|8:X :CALCulate:FSCan:MARKer[1]|2|3|4|5|6|7|8:X...
  • Page 386 SSA5000A Series User Manual :CALCulate:MARKer[1]|2|3|4|5|6|7|8:X:DELTa :CALCulate:MARKer[1]|2|3|4|5|6|7|8:X:REFerence :CALCulate:MARKer[1]|2|3|4|5|6|7|8:Y:DELTa? :CALCulate:MARKer[1]|2|3|4|5|6|7|8:Y:REFerence? :CALCulate:MARKer[1]|2|3|4|5|6|7|8 [:SET]:CENTer :CALCulate:FSCan:MARKer[1]|2|3|4|5|6|7|8 [:SET]:SLISt :CALCulate:FSCan:MARKer[1]|2|3|4|5|6|7|8 [:SET]:METer :CALCulate:FSCan:MARKer[1]|2|3|4|5|6|7|8:TO:METer :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum[:SET]:CENTer :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum:NEXT :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum:LEFT :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum:RIGHt :CALCulate:MARKer[1]|2|3|4|5|6|7|8:PTPeak :CALCulate:MARKer[1]|2|3|4|5|6|7|8:MINimum :CALCulate:MARKer:PEAK:EXCursion :CALCulate:MARKer:PEAK:THReshold :CALCulate:MARKer:PEAK:NUMber Limit :CALCulate:LLINe:SEL :CALCulate:LLINe[1]|2|3|4|5|6:STATe :CALCulate:LLINe[1]|2|3|4|5|6:MARGin:STATe :CALCulate:LLINe[1]|2|3|4|5|6:MARGin :CALCulate:LLINe[1]|2|3|4|5|6:TYPE :CALCulate:LLINe[1]|2|3|4|5|6:OFFSet:X :CALCulate:LLINe[1]|2|3|4|5|6:OFFSet:Y :CALCulate:LLINe[1]|2|3|4|5|6:DATA :CALCulate:LLINe[1]|2|3|4|5|6:ADD :CALCulate:LLINe[1]|2|3|4|5|6:DELete :CALCulate:LLINe[1]|2|3|4|5|6:TRACe :CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:INTerpolate:TYPE :CALCulate:LLINe[1]|2|3|4|5|6:CONTrol:INTerpolate:TYPE :CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:CMODe...
  • Page 387 SSA5000A Series User Manual Meas [:SENSe]:FSCan:SEQuence [:SENSe]:FSCan:SCAN:POINts? [:SENSe]:FSCan:SCAN:TIME :DISPlay:METer[:STATe] [:SENSe]:METer:DETector:DWELl [:SENSe]:METer:DETector :CALCulate:METer:LIMit[:DATA] :CALCulate:METer:LIMit:STATe :CALCulate:METer:LIMit:ULLine :CALCulate:METer:POWer[:CURRent]? :SENSe]:METer:PHOLd:RESet :CALCulate:SLISt:MARK:SIGNal :CALCulate:SLISt:MARK:CLEar:SIGNal :CALCulate:SLISt:MARK:ALL :CALCulate:SLISt:MARK:CLEar:ALL :CALCulate:SLISt:DELete:SIGNal :CALCulate:SLISt:DELete:ALL :CALCulate:SLISt:DELete:MARKed :CALCulate:SLISt:SORT:TYPE :CALCulate:SLISt:SORT:DAMPlitude :CALCulate:SLISt:SORT:DLDelta :CALCulate:SLISt:SORT:ORDer [:SENSe]:FSCan:SEQuence:REStart [:SENSe]:FSCan:SEQuence:REMeasure [:SENSe]:FSCan:SEQuence:REMeter :CALCulate:SLISt:SET:METer :CALCulate:FSCan:SLISt:SET:METer :CALCulate:FSCan:SLISt:DATA? :DISPlay:MEASure[:STATe] [:SENSe]:FSCan:FINal:DETector#:DWELl [:SENSe]:FSCan:FINal:DETector#:LDELta int.siglent.com...
  • Page 388: Pulse

    SSA5000A Series User Manual 16.6.5 PULSE Freq [:SENSe]:FREQuency:CENTer [:SENSe]:FREQuency:CENTer:STEP[:INCRement] [:SENSe]:FREQuency:CENTer:STEP:AUTO [:SENSe]:FREQuency:OFFSet [:SENSe]:BWIDth[:RESolution] [:SENSe]:BWIDth[:RESolution]:AUTO [:SENSe]:BWIDth:VIDeo [:SENSe]:BWIDth:VIDeo:AUTO [:SENSe]:BWIDth:VIDeo:RATio [:SENSe]:BWIDth:VIDeo:RATio:CONfig? [:SENSe]:BANDwidth[:RESolution] [:SENSe]:BANDwidth:VIDeo [:SENSe]:BANDwidth[:RESolution]:AUTO [:SENSe]:BANDwidth:VIDeo:AUTO [:SENSe]:BANDwidth:VIDeo:RATio "[:SENSe]:BANDwidth:VIDeo:RATio:CONfig? Sweep :INITiate:CONTinuous [:SENSe]:SWEep:POINts Trigger :TRIGger[:SEQuence]:SOURce :TRIGger[:SEQuence]:VIDeo:SLOPe :TRIGger[:SEQuence]:EXTernal:SLOPe :TRIGger[:SEQuence]:VIDeo:LEVel :TRIGger[:SEQuence]:LEVel :TRIGger[:SEQuence]:VIDeo:DELay:STATe :TRIGger[:SEQuence]:VIDeo:DELay :TRIGger[:SEQuence]:EXTernal:DELay:STATe :TRIGger[:SEQuence]:EXTernal:DELay :TRIGger[:SEQuence]:EXTernal:DELay:COMPensation :TRIGger[:SEQuence]:FRAMe:DELay :TRIGger[:SEQuence]:FRAMe:DELay:STATe :TRIGger[:SEQuence]:FRAMe:PERiod...
  • Page 389 SSA5000A Series User Manual Ampt :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel ":DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision :SENSe]:POWer[:RF]:ATTenuation [:SENSe]:POWer[:RF]:ATTenuation:AUTO :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet [:SENSe]:POWer[:RF]:GAIN[:STATe] :DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing :UNIT:POWer :TRACe[:Y]:AUToscale Trace :TRACe:MODE :TRACe:TYPE :SENSe]:AVERage:TRACe:CLEar [:SENSe]:DETector:TRACe[:FUNCtion] [:SENSe]:DETector:TRACe:AUTO [:SENSe]:DETector:TRACe:AUTO:ALL [:SENSe]:AVERage:TRACe? :TRACe[:DATA]? :TRACe:MAXimum:X? :FORMat[:TRACe][:DATA] :TRACe:DISPlay[:STATe] :TRACe:PRESet:ALL :TRACe:CLEar:ALL int.siglent.com...
  • Page 390 SSA5000A Series User Manual 16.6.6 Freq [:SENSe]:FREQuency:CARRier [:SENSe]:LPLot:FREQuency:OFFSet:STARt [:SENSe]:LPLot:FREQuency:OFFSet:STOP [:SENSe]:FREQuency:CARRier:SEARch [:SENSe]:FREQuency:CARRier:TRACk[:STATe] [:SENSe]:FREQuency:CARRier:TRACk:SPAN [:SENSe]:FREQuency:CARRier:TRACk:SPAN:AUTO [:SENSe]:FREQuency:CARRier:TRACk:ACCuracy [:SENSe]:BWIDth[:RESolution] Sweep :INITiate:CONTinuous [:SENSe]:SWEep:POINts Ampt :DISPlay:LPLot:WINDow:TRACe:Y[:SCALe]:RLEVel :DISPlay:LPLot:WINDow:TRACe:Y[:SCALe]:PDIVision :DISPlay:LPLot:WINDow:TRACe:Y[:SCALe]:RPOSition Trace :TRACe:SELEct :TRACe:LPLot:MODE :TRACe:LPLot:TYPE :TRACe:LPLot:COPY :TRACe:LPLot:EXCHange :TRACe:PRESet:ALL :TRACe:CLEar:ALL [:SENSe]:LPLot:DETector[:FUNCtion] [:SENSe]:AVERage:TRACe? :TRACe[:DATA]? :FORMat[:TRACe][:DATA] Marker :CALCulate:MARKer:TABLe[:STATe] :CALCulate:LPLot:MARKer:SELEct :CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:MODE :CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:STATe :CALCulate:LPLot:MARKer:AOFF...
  • Page 391 SSA5000A Series User Manual :CALCulate:LPLot:MARKer:COUPle[:STATe] :CALCulate:MARKer[1]|2|3|4|5|6|7|8[:SET]:RESEt:DELTa :CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:FUNCtion :CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:RMSNoise:MODE :CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:BAND:SPAN :CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:BAND:LEFT :CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:BAND:RIGHt :CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:RMSNoise:RESult Linit :CALCulate:LLINe:SEL :CALCulate:LLINe[1]|2|3|4|5|6:STATe :CALCulate:LPLot:LLINe[1]|2|3|4|5|6:DISPlay :CALCulate:LLINe[1]|2|3|4|5|6:MARGin :CALCulate:LLINe[1]|2|3|4|5|6:MARGin:STATe :CALCulate:LLINe[1]|2|3|4|5|6:TYPE :CALCulate:LLINe[1]|2|3|4|5|6:Offset:X :CALCulate:LLINe[1]|2|3|4|5|6:Offset:Y :CALCulate:LLINe[1]|2|3|4|5|6:DELete :CALCulate:LLINe:ALL:DELete :CALCulate:LLINe[1]|2|3|4|5|6:TRACe :CALCulate:LLINe[1]|2|3|4|5|6:DATA :CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:INTerpolate:TYPE :CALCulate:LLINe[1]|2|3|4|5|6:CONTrol:INTerpolate:TYPE :CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:INTerpolate:TYPE :CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:CMODe :CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:CMODe :CALCulate:LLINe[1]|2|3|4|5|6:ADD :CALCulate:LLINe[1]|2|3|4|5|6:POINt:DELete :CALCulate:LLINe[1]|2|3|4|5|6:COPY :CALCulate:LLINe[1]|2|3|4|5|6:BUILd :CALCulate:LLINe:TEST :CALCulate:LLINe:CONTrol:BEEP :CALCulate:LLINe[1]|2|3|4|5|6:FAIL? Meas [:SENSe]:LPLot:AVERage:COUNt [:SENSe]:LPLot:AVERage:STATe...
  • Page 392 SSA5000A Series User Manual :DISPlay:LPLot:VIEW :CALCulate:LPLot:SPURious:TABLe :CALCulate:LPLot:DECade:TABLe [SENSe]:CARRier:THReshold:MINimum int.siglent.com...
  • Page 393 SSA5000A Series User Manual 16.6.7 Freq :SENSe]:NFIGure:FREQuency:MODE [:SENSe]:NFIGure:FREQuency:CENTer [:SENSe]:NFIGure:FREQuency:STARt [:SENSe]:NFIGure:FREQuency:STOP [:SENSe]:NFIGure:FREQuency:SPAN [:SENSe]:NFIGure:FREQuency:FIXed [:SENSe]:NFIGure:FREQuency:LIST:DATA:DELete [:SENSe]:NFIGure:BWIDth[:RESolution] [:SENSe]:NFIGure:BANDwidth[:RESolution] [:SENSe]:NFIGure:BWIDth[:RESolution]:AUTO [:SENSe]:NFIGure:BANDwidth[:RESolution]:AUTO Sweep [:SENSe]:NFIGure:SWEep:POINts [:SENSe]:NFIGure:SWEep:TIME [:SENSe]:NFIGure:SWEep:TIME:AUTO :INITiate:CONTinuous Ampt :DISPlay:NFIGure:TRACe:NFIGure:Y[:SCALe]:RLEVel :DISPlay:NFIGure:TRACe:NFACtor:Y[:SCALe]:RLEVel :DISPlay:NFIGure:TRACe:GAIN:Y[:SCALe]:RLEVel :DISPlay:NFIGure:TRACe:TEEFective:Y[:SCALe]:RLEVel :DISPlay:NFIGure:TRACe:PHOT:Y[:SCALe]:RLEVel :DISPlay:NFIGure:TRACe:PCOLd:Y[:SCALe]:RLEVel :DISPlay:NFIGure:TRACe:NFIGure:Y[:SCALe]:PDIVision :DISPlay:NFIGure:TRACe:NFACtor:Y[:SCALe]:PDIVision :DISPlay:NFIGure:TRACe:GAIN:Y[:SCALe]:PDIVision :DISPlay:NFIGure:TRACe:TEEFective:Y[:SCALe]:PDIVision :DISPlay:NFIGure:TRACe:PHOT:Y[:SCALe]:PDIVision :DISPlay:NFIGure:TRACe:PCOLd:Y[:SCALe]:PDIVision :DISPlay:NFIGure:TRACe:NFIGure:Y[:SCALe]:RPOSition :DISPlay:NFIGure:TRACe:GAIN:Y[:SCALe]:RPOSition :DISPlay:NFIGure:TRACe:TEEFective:Y[:SCALe]:RPOSition :DISPlay:NFIGure:TRACe:PHOT:Y[:SCALe]:RPOSition :DISPlay:NFIGure:TRACe:PCOLd:Y[:SCALe]:RPOSition...
  • Page 394 SSA5000A Series User Manual [:SENSe]:NFIGure:POWer[:RF]:ATTenuation:AUTO [:SENSe]:NFIGure:POWer[:RF]:GAIN[:STATe] Trace :DISPlay:NFIGure:FORMat :DISP:NFIG:FORM:AUTO :DISPlay:NFIGure:DATA:TRACe :DISPlay:NFIGure:TRACe:COMBined[:STATe] :TRACe:NFIGure:TYPE :TRACe:NFIGure:DISPlay[:STATe] Marker :CALCulate:NFIGure:MARKe[1]|2|3|4|5|6|7|8:X :CALCulate:NFIGure:MARKer[1]|2|3|4|5|6|7|8:Y :CALCulate:NFIGure:MARKer[1]|2|3|4|5|6|7|8:MODE :CALCulate:NFIGure:MARKer[1]|2|3|4|5|6|7|8:TRACe :CALCulate:NFIGure:MARKer:AOFF Meas [:SENSe]:NFIGure:CORRection:ENR[:MEASurement]:TABLe:DATA:DELete [:SENSe]:NFIGure:CORRection:ENR:MEASurement:TABLe:DATA [:SENSe]:NFIGure:CALibration:DELete int.siglent.com...
  • Page 395 SSA5000A Series User Manual 16.6.8 Freq [:SENSe]:FREQuency:CENTer [:SENSe]:FREQuency:CHANnel Sweep :INITiate:CONTinuous [:SENSe]:SWEep:TIME Ampt :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel :DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision Trace :CALCulate:PARameter:SELect :CALCulate:PARameter:COUNt :TRACe:FORMat[:Y] Trigger :TRIGger[:SEQuence]:SOURce :TRIGger[:SEQuence]:VIDeo:LEVel :TRIGger[:SEQuence]:LEVel :TRIGger[:SEQuence]:VIDeo:DELay:STATe :TRIGger[:SEQuence]:VIDeo:DELay :TRIGger[:SEQuence]:VIDeo:SLOPe :TRIGger[:SEQuence]:EXTernal:DELay:STATe :TRIGger[:SEQuence]:EXTernal:DELay :TRIGger[:SEQuence]:EXTernal:DELay:COMPensation :TRIGger[:SEQuence]:EXTernal:SLOPe :TRIGger[:SEQuence]:RFBurst:SLOPe :TRIGger[:SEQuence]:FRAMe:DELay:STATe :TRIGger[:SEQuence]:FRAMe:DELay :TRIGger[:SEQuence]:FRAMe:PERiod :TRIGger[:SEQuence]:FRAMe:OFFSet :TRIGger[:SEQuence]:FRAMe:OFFSet:DISPlay:RESet :TRIGger[:SEQuence]:FRAMe:SYNC :TRIGger[:SEQuence]:ATRigger:STATe :TRIGger[:SEQuence]:ATRigger :TRIGger[:SEQuence]:HOLDoff:STATe :TRIGger[:SEQuence]:HOLDoff...
  • Page 396: Other

    SSA5000A Series User Manual 16.6.9 Other [:SENSe]:ROSCillator:SOURce:TYPE [:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude] Correction [:SENSe]:CORRection:CSET#[:STATe] [:SENSe]:CORRection:CSET#:ADD [:SENSe]:CORRection:CSET:ALL:DELete [:SENSe]:CORRection:CSET#:DATA? [:SENSe]:CORRection:CSET#:DELete [:SENSe]:CORRection:CSET:ALL:DELete System :SYSTem:CONFigure:SYSTem? :SYSTem:LANGuage :SYSTem:COMMunicate:LAN:TYPE :SYSTem:COMMunicate:LAN:IPADdress :SYSTem:COMMunicate:LAN:GATeway :SYSTem:COMMunicate:LAN:SMASk :SYSTem:WEB:PSW :SYSTem:TIME :SYSTem:DATE :SYSTem:LKEY :SYSTem:COMMunicate:GPIB:ADDRess Reset :SYSTem:PRESet :SYSTem:PRESet:TYPE :SYSTem:PRESet:USER[1]|2|3|4|5|6|7:SAVE :SYSTem:PRESet:USER[1]|2|3|4|5|6|7:LOAD :SYSTem:PON:TYPE :SYSTem:FDEFault :SYSTem:CLEAr Calibration :CALibration:STATe :CALibration File :MMEMory:STORe...
  • Page 397: 17 Service And Support

    SSA5000A Series User Manual 17 Service and Support 17.1 Service Summary SIGLENT warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of three years (accessories for a period of one year) from the date of shipment from an authorized Siglent distributor.
  • Page 398: 17.2 Troubleshooting

    SSA5000A Series User Manual 17.2 Troubleshooting Before calling SIGLENT, or returning an analyzer for service, perform the quick checks listed below. This check may eliminate the problem. If the problem remains still, please contact SIGLENT and provide your device information in the back of the analyzer.
  • Page 399 SSA5000A Series User Manual To calculate the system errors and check the measurement results and precision, refer to the introductions in “Specifications”. To reach these specifications, please: (1) Check whether all the external devices are successfully connected and are working normally.
  • Page 400 SSA5000A Series User Manual Table 17-1 System Message User system message Message on screen System message description (1~199) SWT_OOR (1) Sweep time out of range RBW_OOR(2) RBW out of range SWT_CCOFM(3) Can't change the sweep time in FFT mode MRKT_UNDEF(4)
  • Page 401 SSA5000A Series User Manual QP_RBW_OOR (26) RBW out of range when do quasi peak scan LAN_PLUG_IN (150) Ethernet cable plug in LAN_PLUG_OUT (151) Ethernet cable plug out IP_CONFLICT (152) IP address conflict IP_INVALID (153) IP address invalid NETM_INVALID (154) Netmask address invalid...
  • Page 402 SSA5000A Series User Manual About SIGLENT SIGLENT is an international high-tech company, concentrating on R&D, sales, production and services of electronic test & measurement instruments. SIGLENT first began developing digital oscilloscopes independently in 2002. After more than a decade of continuous development, SIGLENT has...

This manual is also suitable for: