SIGLENT CP5000 Series Current Probe Safety Notices Note A Note notice denotes important instructions. CAUTION A CAUTION notice denotes a hazard. It calls attention to an operation procedure ,practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data.
SIGLENT Notes on Safety This device is designed to comply with IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the device. Be certain that you understand the instructions and precautions in the manual before use .We disclaim any responsibility for accidents or injuries not resulting directly from device defects.
Page 5
SIGLENT CAUTION To avoid damage to the device, protect it from vibration or shock during transport and handling, and be especially careful to avoid dropping. Do not store or use the device where it could be exposed to direct sunlight, high temperature, humidity, or condensation.
SIGLENT 目 录 CP5000 Series Current Probe ........... I CP5000 Series Summary......................I Notes on Safety ......................... II Summary ........................... 1 Key features ..........................1 Application ..........................1 Description of Probe Parts ......................2 Normal and Typical Characteristics ................... 4 Mechanical Characteristics......................
SIGLENT Summary The CP8000 series uses a combination of Hall effect and transformer technology which enables measurements to be made on DC, AC and impulse currents. It’s key features include highly accurate current measurements , wide bandwidth , easy current measurements , DC/AC measure , over-current protected and indication(buzzer and LED indicator) , double ranges selection , low current measurements , degauss and auto zero function.
SIGLENT Description of Probe Parts 1)CP5030 CP5030A 2) CP5150 CP5500 1 . Sensor head 2 Current Probe...
Page 9
SIGLENT This clamps the conductor being measured, and carries out the actual current measurement. It is a precision assembly including a molded component, a ferritecore,and a Hall effect element. It may be damaged if subjected to suddenchanges in ambient temperature, or mechanical strain of shock, and therefore great care should be exercised in handling it.
SIGLENT Accessories BNC Cable :100cm,MALE X MALE (CK-310) Power Adapter (12V/1.2A) (CK-612) Normal and Typical Characteristics Model CP5030 CP5030A CP5150 CP5500 BW(-3dB) DC-50MHz DC-100MHz DC-12MHz DC-5MHz Rise time ≤7ns ≤3.5ns ≤29ns ≤70ns Continuous maximum 30Arms 30Arms 150Arms 500Arms input range Max peak 300A 750A...
Page 11
SIGLENT Figure 1-1 CP5030 Frequency response Figure 1-2 CP5030A Frequency response Figure 2-1 CP5030 Continuous maximum Figure 2-2 CP5030A Continuous maximum input rating (Frequency derating) input rating(Frequency derating) Figure 3-1 CP5030 Input impedance VS Frequency Figure 3-2 CP5030A Input impedance VS Frequency Current Probe 5...
Page 12
SIGLENT Figure 4 CP5150 Frequency response Figure 5 CP5150 Continuous maximum input rating (Frequency derating) Figure 6 CP5150 Input impedance VS Frequency 6 Current Probe...
Page 13
SIGLENT Figure 7 CP5500 Frequency response Figure 8 CP5500 Continuous maximum input rating (Frequency derating) Figure 9 CP5500 Input impedance VS Frequency Current Probe 7...
Page 15
SIGLENT The reading may be affected by the position within the clamp aperture of the conductor being measured. The conductor should be in the center of the clamp aperture. When carrying out a measurement, press the opening lever until the “UNLOCK” indication disappears and check that the sensor head is properly closed.
SIGLENT Continuous input of current exceeding the rated maximum or repeated activation of the safety function may result in damage to the unit. The probe is rated for maximum input under two conditions in addition to the input maximums shown in the “Characteristics ". Such as CP8030A,these are (1) 30 Apeak for continuous input and (2) 50 Apeak for pulse widths 10 μs.
SIGLENT Service Strategy Problem Possible cause Processing methods Power off Power on Can’t measuring DC Oscilloscope set AC mode Oscilloscope set AC mode Sensor is not locked Lock the sensor Demagnetization and Zero Sensor with magnetization Adjustment Can’t adjust to zero Turn the coarse adjustment Zero Adjustment out of range trimmer...
SIGLENT General Safety Instructions Read the following safety instructions to avoid injury and prevent damage to this product or any products connected to it. Use this product only as specified. Only qualified personnel should perform service procedures. To avoid fire or Personal injury. ...
SIGLENT Getting Started The CP4000 series current probe enables a general purpose oscilloscope to display AC and DC current signals up to 100 amps Peak (70A RMS). The current probe can also make AC and DC measurements with a multimeter by using the recommended accessory MT-246(BNC-to- banana) plug adapter.
SIGLENT Basic Operation Before using the probe, the batteries or specified power adaptor must be installed. WARNING! DO not clamp the probe onto circuits with voltages greater than 600 VAC. Personal Injury or damage to the probe may result. Always connect the CP4000 current probe output to the instrument before clamping onto the circuit under test.
SIGLENT Congratulations on your purchase of CP4000 series, a multifunctional current probe. When connecting to a digital meter, use the recommended MT-246N (BNC to banana adapter). Connect the black lead to the meter COM, and the red lead to the VΩ input . To measure only AC current, set the meter to measure AC volts.
SIGLENT Power Supply 9V battery Maximum working Voltage 300V CAT III 600V CAT II Maximum working Voltage Mechanical Specifications Model CP4020 CP4070 CP4050 CP4070A Dimensions 231*67*36mm 262*81*36mm 280*70*32mm Maximum Conductor size 10.3mm 10.3mm 11mm Cable Length 200cm 100cm 100cm Weight 310g 310g 260g...
Page 24
SIGLENT CP4050 Maximum current versus frequency CP4070A Maximum current versus frequency CP4050 DC signal linearity in the 50mV/A range CP4070A DC signal linearity in the 10mV/A range CP4050 Phase versus frequency at 1A peak CP4070A Phase versus frequency at 1A peak 18 Current Probe...
Page 25
SIGLENT Certifications and compliances Compliance was demonstrated to the following specification as listed in the Official Journal of the European Union: Low Voltage Directive 73/23/EEC,as amended by 93/68/EEC EN 61010-1/A2:1995 EC Declaration of Safety requirements for electrical equipment for Conformity Low Voltage measurement, control, and laboratory use.
4)Please put the probe in the package provided to prevent shock. 5)Do not forcefully pull the input and output lead to prevent bending, twisted and folding. Contact SIGLENT SIGLENT TECHNOLOGIES CO., LTD Address: 3/F, Building 4, Antongda Industrial Zone, Liuxian Road, 68 District, Baoan District, Shenzhen, P.R. CHINA...