SIGLENT TECHNOLOGIES CP5000 Series Instruction Manual

Current probe

Advertisement

Quick Links

User Manual
Current Probe Instruction Manual
2015 SIGLENT TECHNOLOGIES CO.,LTD

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CP5000 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SIGLENT TECHNOLOGIES CP5000 Series

  • Page 1 User Manual Current Probe Instruction Manual 2015 SIGLENT TECHNOLOGIES CO.,LTD...
  • Page 3: Safety Notices

    SIGLENT CP5000 Series Current Probe Safety Notices Note A Note notice denotes important instructions. CAUTION A CAUTION notice denotes a hazard. It calls attention to an operation procedure ,practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data.
  • Page 4: Notes On Safety

    SIGLENT Notes on Safety This device is designed to comply with IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the device. Be certain that you understand the instructions and precautions in the manual before use .We disclaim any responsibility for accidents or injuries not resulting directly from device defects.
  • Page 5 SIGLENT CAUTION To avoid damage to the device, protect it from vibration or shock during transport and handling, and be especially careful to avoid dropping. Do not store or use the device where it could be exposed to direct sunlight, high temperature, humidity, or condensation.
  • Page 6: Table Of Contents

    SIGLENT 目 录 CP5000 Series Current Probe ........... I CP5000 Series Summary......................I Notes on Safety ......................... II Summary ........................... 1 Key features ..........................1 Application ..........................1 Description of Probe Parts ......................2 Normal and Typical Characteristics ................... 4 Mechanical Characteristics......................
  • Page 7: Summary

    SIGLENT Summary The CP8000 series uses a combination of Hall effect and transformer technology which enables measurements to be made on DC, AC and impulse currents. It’s key features include highly accurate current measurements , wide bandwidth , easy current measurements , DC/AC measure , over-current protected and indication(buzzer and LED indicator) , double ranges selection , low current measurements , degauss and auto zero function.
  • Page 8: Description Of Probe Parts

    SIGLENT Description of Probe Parts 1)CP5030 CP5030A 2) CP5150 CP5500 1 . Sensor head 2 Current Probe...
  • Page 9 SIGLENT This clamps the conductor being measured, and carries out the actual current measurement. It is a precision assembly including a molded component, a ferritecore,and a Hall effect element. It may be damaged if subjected to suddenchanges in ambient temperature, or mechanical strain of shock, and therefore great care should be exercised in handling it.
  • Page 10: Normal And Typical Characteristics

    SIGLENT Accessories BNC Cable :100cm,MALE X MALE (CK-310) Power Adapter (12V/1.2A) (CK-612) Normal and Typical Characteristics Model CP5030 CP5030A CP5150 CP5500 BW(-3dB) DC-50MHz DC-100MHz DC-12MHz DC-5MHz Rise time ≤7ns ≤3.5ns ≤29ns ≤70ns Continuous maximum 30Arms 30Arms 150Arms 500Arms input range Max peak 300A 750A...
  • Page 11 SIGLENT Figure 1-1 CP5030 Frequency response Figure 1-2 CP5030A Frequency response Figure 2-1 CP5030 Continuous maximum Figure 2-2 CP5030A Continuous maximum input rating (Frequency derating) input rating(Frequency derating) Figure 3-1 CP5030 Input impedance VS Frequency Figure 3-2 CP5030A Input impedance VS Frequency Current Probe 5...
  • Page 12 SIGLENT Figure 4 CP5150 Frequency response Figure 5 CP5150 Continuous maximum input rating (Frequency derating) Figure 6 CP5150 Input impedance VS Frequency 6 Current Probe...
  • Page 13 SIGLENT Figure 7 CP5500 Frequency response Figure 8 CP5500 Continuous maximum input rating (Frequency derating) Figure 9 CP5500 Input impedance VS Frequency Current Probe 7...
  • Page 14: Mechanical Characteristics

    SIGLENT Mechanical Characteristics Model CP5030 CP5030A CP5150 CP5500 Measurement Conductor 20mm Diameter Max. Cable Length 1.5m BNC Length 100cm Power Adapter Dimensions 72*62*31mm; Cable length 1.5m Current Clamp Dimensions 75*40*18mm 175*68*29mm (L*W*H) Termination Unit(L*W*H) 119*49*28mm Probe Weight 240g 500g 510g Environmental Characteristics Model CP5030...
  • Page 15 SIGLENT The reading may be affected by the position within the clamp aperture of the conductor being measured. The conductor should be in the center of the clamp aperture. When carrying out a measurement, press the opening lever until the “UNLOCK” indication disappears and check that the sensor head is properly closed.
  • Page 16: Preparations For Measurement

    SIGLENT Continuous input of current exceeding the rated maximum or repeated activation of the safety function may result in damage to the unit. The probe is rated for maximum input under two conditions in addition to the input maximums shown in the “Characteristics ". Such as CP8030A,these are (1) 30 Apeak for continuous input and (2) 50 Apeak for pulse widths 10 μs.
  • Page 17: Service Strategy

    SIGLENT Service Strategy Problem Possible cause Processing methods Power off Power on Can’t measuring DC Oscilloscope set AC mode Oscilloscope set AC mode Sensor is not locked Lock the sensor Demagnetization and Zero Sensor with magnetization Adjustment Can’t adjust to zero Turn the coarse adjustment Zero Adjustment out of range trimmer...
  • Page 18: Cp4000 Series Current Probe

    SIGLENT CP4000 Series Current Probe CP4000 Series Summary Continuous Maximum Model Peak Current BW(-3dB) Range switch Input Range 100mV/A 300kHz CP4070A 200A 70Arms 10mV /A 50mV/A CP4070 200A 70Arms 150kHz 5mV /A 500mV/A CP4050 140A 50Arms 1MHz 50mV /A 50mV/A CP4020 20Arms 100kHz...
  • Page 19: General Safety Instructions

    SIGLENT General Safety Instructions Read the following safety instructions to avoid injury and prevent damage to this product or any products connected to it. Use this product only as specified.  Only qualified personnel should perform service procedures.  To avoid fire or Personal injury. ...
  • Page 20: Getting Started

    SIGLENT Getting Started The CP4000 series current probe enables a general purpose oscilloscope to display AC and DC current signals up to 100 amps Peak (70A RMS). The current probe can also make AC and DC measurements with a multimeter by using the recommended accessory MT-246(BNC-to- banana) plug adapter.
  • Page 21: Basic Operation

    SIGLENT Basic Operation Before using the probe, the batteries or specified power adaptor must be installed. WARNING! DO not clamp the probe onto circuits with voltages greater than 600 VAC. Personal Injury or damage to the probe may result. Always connect the CP4000 current probe output to the instrument before clamping onto the circuit under test.
  • Page 22: Specifications

    SIGLENT Congratulations on your purchase of CP4000 series, a multifunctional current probe. When connecting to a digital meter, use the recommended MT-246N (BNC to banana adapter). Connect the black lead to the meter COM, and the red lead to the VΩ input . To measure only AC current, set the meter to measure AC volts.
  • Page 23: Mechanical Specifications

    SIGLENT Power Supply 9V battery Maximum working Voltage 300V CAT III 600V CAT II Maximum working Voltage Mechanical Specifications Model CP4020 CP4070 CP4050 CP4070A Dimensions 231*67*36mm 262*81*36mm 280*70*32mm Maximum Conductor size 10.3mm 10.3mm 11mm Cable Length 200cm 100cm 100cm Weight 310g 310g 260g...
  • Page 24 SIGLENT CP4050 Maximum current versus frequency CP4070A Maximum current versus frequency CP4050 DC signal linearity in the 50mV/A range CP4070A DC signal linearity in the 10mV/A range CP4050 Phase versus frequency at 1A peak CP4070A Phase versus frequency at 1A peak 18 Current Probe...
  • Page 25 SIGLENT Certifications and compliances Compliance was demonstrated to the following specification as listed in the Official Journal of the European Union: Low Voltage Directive 73/23/EEC,as amended by 93/68/EEC EN 61010-1/A2:1995 EC Declaration of Safety requirements for electrical equipment for Conformity Low Voltage measurement, control, and laboratory use.
  • Page 26: Care And Maintenance

    4)Please put the probe in the package provided to prevent shock. 5)Do not forcefully pull the input and output lead to prevent bending, twisted and folding. Contact SIGLENT SIGLENT TECHNOLOGIES CO., LTD Address: 3/F, Building 4, Antongda Industrial Zone, Liuxian Road, 68 District, Baoan District, Shenzhen, P.R. CHINA...