1. Product Overview 1.1 Main Structure of the Product The InSight Equine-IA Equine Immunoassay Analyser is composed of the control unit, optical unit, mechanical unit and the output unit/display unit, as shown in the figure below. Optical Unit Output Unit/Display Unit...
• Do not use the analyser in unstable environments such on unlevel or vibrating surfaces. • Do not place the analyser in a location where it is difficult to disconnect the device. • Water or debris should not enter the analyser. If this occurs, please contact Woodley Equipment Company.
Page 6
Woodley Equipment Company, other users are not permitted to remove, modify or repair the analyser. Any violation will invalidate the analyser warranty. Woodley Equipment Company will not bear any responsibility for possible personal injury, fire risk or electric shocks caused by violation of the warranty.
Page 7
Attention: • The analyser should be inspected periodically. If there are any damages, cracks or other abnormalities, please contact Woodley Equipment Company. • When cleaning the analyser, wipe using a damp, lint free cloth. The following solutions can be used for cleaning: alcohol or mild detergent.
B. Test with temporary degradation or loss of function or performance, but self-restoring. C. Test with temporary degradation or loss of function or performance, but requiring operator intervention or system reset. Table 2 Electromagnetic Emission Emission Test Compliance GB 4824 RF Emission Group 1 GB 4824...
2. Check the analyser and accessories to see if they are in good condition. NOTE: If there are any problems, please contact Woodley Equipment Company. 3.1.2 Analyser Placement 1. The analyser should be placed in a clean and ventilated room with temperature between 10°C –...
Page 10
4. Used samples and materials should be regarded as waste and disposed of in accordance with local regulations. Warnings: • If any signs of damage to the analyser function are found or an error message appears during operation, do not continue the test. Contact Woodley Equipment Company.
4. Software Interface Operation Introduction 4.1 Main Interface Press and hold the power button to turn on the analyser, and enter the software main interface (Figure 4.1.2) after passing through the software startup interface (Figure 4.1.1). The main interface contains four submenus [Test], [History Record], [General Settings] and [About]. Figure 4.1.1 Figure 4.1.2 4.2 Testing...
Page 12
Figure 4.2.1 Figure 4.2.2 After the sample test is complete, you can view the result information (Figure 4.2.3). Click the ‘Print’ button to print the report as required or click ‘Testing Finished’ to exit the test. Figure 4.2.3...
The default setting is permanent. If you wish to change this setting, please contact Woodley Equipment Company. ‘Institution’ can be used to fill in the laboratory information. Please contact Woodley Equipment Company if you wish to change this.
Page 14
Notice: Please keep ‘Scan barcode’ open. When ‘Scan barcode’ option closed,Analyzer will always test with the current item ID chip in every test. Figure 4.4.1 Figure 4.4.2 4.4.1 How to Connect to Wi-Fi 1. Turn the analyser on. On the Menu page, click ‘General Settings’ to open the settings page. Figure 4.4.3 Figure 4.4.4...
2. Click ‘Wi-Fi Settings’. Select the Wi-Fi you’d like to connect to and input the password. Figure 4.4.5 Figure 4.4.6 About and Software update Click ‘About’ to view the device model, software version and contact information. Click ‘Restore Factory Settings’ and the analyser will automatically reset (Figure 4.5.1). NOTE: After restoring the factory settings, all detection data, user data and configuration information will be cleared, please operate with caution.
Figure 4.5.1 Android System Setting On the Android system, you can set the time zone, date, display, advanced Wi-Fi settings and Android system settings etc. 4.6.1 Enter Android System • On the Main Menu (Figure 4.6.1.1), click ‘General Settings’ to open the General Settings page (Figure 4.6.1.2).
Page 17
• Click the Menu/List icon on the upper left of the page to open the Function List page (Figure 4.6.1.3). Click ‘Exit the software’. Enter the password ‘12345678’ and press ‘Confirm’ to exit software (Figure 4.6.1.4). Figure 4.6.1.3 Figure 4.6.1.4 •...
Page 18
4.6.2 Navigation Bar and Display • Navigation Bar Click ‘Settings’ on the Android apps page and press ‘Display’ (Figure 4.6.2.1). Toggle the ‘Hide Navigation Bar’ so that the button turns green and the Navigation Bar will be shown on the bottom (Figure 4.6.2.2). Note: Please hide the Navigation Bar after finishing Android system setting.
Page 19
Figure 4.6.2.3 Figure 4.6.2.4 4.6.3 Date and Time The analyser will automatically update the date/time and the time zone after connecting to the Wi-Fi. To modify it manually, perform the following steps: • Click ‘System’ on the settings page (Figure 4.6.3.1). •...
Page 21
Figure 4.6.4.4 Figure 4.6.4.5 Figure 4.6.4.6 4.6.5 USB Device Support and Software Installation USB device function is supported on the Equine-IA software page but it is closed on the Android system. To open USB device support and install the app on the Android system, perform the following steps: •...
Page 23
4.6.6 Keyboard Type and Language of the Android System Keyboard Type and Language can be set on the Android system. • Click ‘System’ on the Android system settings page (Figure 4.6.6.1). • Click ‘Language & Input’ (Figure 4.6.6.2). • Click ‘Add a Language’. English is the default language (Figure 4.6.6.3). Figure 4.6.6.1 Figure 4.6.6.2 Figure 4.6.6.3...
Page 24
Figure 4.6.6.4 Figure 4.6.6.5 Figure 4.6.6.6 On the APP page, press and hold the Globe icon (Figure 4.6.6.7) and select the new language from the list (Figure 4.6.6.8). The current language keyboard will show in the middle of the spacebar (Figure 4.6.6.9). Figure 4.6.6.7 Figure 4.6.6.8 Figure 4.6.6.9...
Page 25
• Click ‘System update’ on the app page (Figure 4.6.7.1). • Click ‘Select Firmware’, the update file name will show on the text box (Figure 4.6.7.2). • Click ‘Start upgrade’ to upgrade the Android system (Figure 4.6.7.2). Figure 4.6.7.1 Figure 4.6.7.2 Figure 4.6.7.3 •...
Power switch is not turned on Turn on the switch. The analyser won’t The power adapter is not switch on Reconnect the power adapter. connected Please contact Woodley Screen has broken Equipment Company. The screen doesn’t display Please contact Woodley Problem with operating system Equipment Company.
Page 27
Restart the analyser and Communication failure Analyser stops during retest. testing Please contact Woodley Equipment Company. Check the kit insert to make sure Repeat the test checking the correct sample type and sample requirements. volume was used Abnormal test result...
6. Storage and Transportation Conditions Storage When the analyser won’t be in use for an extended period of time, it should be stored in its original packaging in a well-ventilated and clean room. The storage temperature is -10°C to 40°C. The relative humidity should be less than 93%.
Page 30
Old Station Park Buildings, St. John Street, Horwich, Bolton, BL6 7NY, UK Tel: +44 (0) 1204 669033 Email: sales@woodleyequipment.com Web: www.woodleyequipment.com...