Operations Manual Electrochemistry Meters Update Meter Software to Ensure Optimum Performance www.thermofisher.com/orionsoftware Orion Pro Star EC212 Conductivity Bench Meter User Manual 68X002726 | Revision 01 | December 2024...
Orion ™ Pro Star EC212 conductivity bench meter. Orion Pro Star EC212 bench meters are designed to provide accurate and reliable testing in laboratories that need a modern, easy-to-operate meter with advanced performance and reliable data reporting. Orion Pro Star EC212 bench meters offer the following measurements: •...
Secure a wide range of electrodes using the various electrode holder options. Neatly organize electrode cables in the built-in side channels of the stand. All Orion Pro Star EC212 meters include an electrode stand, USB flash drive, computer cable and universal 100-240V 50/60 Hz power adapter with wall plugs for US/Japan, Euro, UK/Singapore, Australia/New Zealand and China.
Wipe the display panel with a clean microfiber cloth, taking care to wipe lightly to avoid scratching the meter display. The meter housing can also be cleaned using a clean microfiber cloth dampened with IPA (isopropyl alcohol) for disinfection purposes. Orion Pro Star EC212 Conductivity Bench Meter...
Chapter 2 Getting Started Refer to the 3.9 Information & Help section and Meter Tours to access onscreen tours that explain how to use important meter functions including meter connections and electrode stand assembly. 2.1 Universal Power Supply Assembly A universal power adapter (catalog number PSTAR-PWR) with US/Japan, Euro, UK/Singapore, Australia/New Zealand and China wall plug plates is included with the meter.
Page 11
You are now ready to turn on the meter. CAUTION: Use of a surge protector or uninterrupted power supply (UPS) is recommended, as an unintended power surge of electricity to the meter may damage the meter and void the warranty. Orion Pro Star EC212 Conductivity Bench Meter...
2.2 Electrode Stand Installation The included electrode stand (catalog number PSTAR-ARM) facilitates secure, smooth movement of electrodes into and out of solutions. A weighted base provides flexible placement of the stand around the meter. Several electrode holder and clip options accommodate a wide range of electrode types. Built-in side channels allow electrode cables to be neatly organized inside the stand.
Page 13
15mm diameter. e. The 20° angled holder secures gas-sensing ion selective electrodes (ISEs) including the ammonia electrode, catalog number 9512BNWP and high performance ammonia electrode, catalog number 9512HPBNWP. Orion Pro Star EC212 Conductivity Bench Meter...
Page 14
The RDO clip holds optical dissolved oxygen electrodes. Attach the RDO clip to the holder head by sliding the clip down on top of the tab on the front of the holder head. Use the side channels built into the electrode arm to organize the electrode cables. Press the electrode cables into the 4 grips inside the arm channels to secure the cables in place.
• Connect 1 USB flash drive with a USB-A plug to the meter for data transfer. • Connect 1 USB accessory (barcode scanner, mouse or keypad) with USB-A plug to the meter for alphanumeric character entry into meter fields including sample ID. Orion Pro Star EC212 Conductivity Bench Meter...
2.3.3 Back Panel Meter Inputs Most meter inputs are located on the back panel of the meter. Each input has a protective cap that protects the meter from dust and water intrusion. For maximum protection, keep the protective caps securely closed unless the input is in use. Power Input: Connect the power adapter, with twist-and-lock plug, to the meter.
The first time the meter is powered on, it launches the meter setup wizard. Follow the displayed steps to set critical meter and measurement settings. View detailed information on additional settings in the 3.2 Settings section. Orion Pro Star EC212 Conductivity Bench Meter...
2.5 Touchscreen Functions Orion Pro Star meters offer a 7” capacitive touchscreen interface that is compatible with most nitrile laboratory gloves. • Key interaction points on the screen are blue for easy navigation. • For screens with extended data, easily scroll by touching and sliding the display up and down. Optionally, a USB mouse can be connected to the meter and used to operate the touchscreen interface if direct contact with the screen is not desirable.
Graph: Tap the blue Graph text to view the current measurement as a graph with temperature. • Select the Read Type to change how the graph tracks measurements over time. • Tap the Display Temperature button to show or hide temperature measurements. Orion Pro Star EC212 Conductivity Bench Meter...
2.5.1.2 Basic Measure Mode View Refer to the 3.2.1 Instrument Settings section and Display settings and select Basic as the Custom Measure Layout to view a simplified version of the measure mode. The default meter setting is Advanced as the Custom Measure Layout and all user-selectable information (i.e. instrument name, sample ID, electrode ID, method) will only be shown when it is entered in the settings menus.
2.6 Electrode Information The Orion Pro Star EC212 meter is compatible with a conductivity electrode with 8-pin MiniDIN connector. The Orion Pro Star EC212 meter is available in a meter only configuration (purchase electrode separately) or in meter kits with electrodes and solutions included for different applications.
Chapter 3 Main Menu 3.1 Main Menu Tap the Main Menu icon in the measure mode to access the Main Menu. • Settings: View and customize instrument and measurement settings. • Calibrate: Perform all calibration functions, including temperature and barometer calibrations. •...
• Conductivity Settings: View and customize measurement settings for conductivity, TDS, salinity or resistivity, including temperature, measure mode, sample ID, alarms, nominal cell constant, reference temperature, temperature compensation, calibration points, stability criteria, stirrer, electrode ID and electrode type. Orion Pro Star EC212 Conductivity Bench Meter...
3.2.1 Instrument Settings Tap the Instrument Settings option in the Settings list to access the instrument settings. • Read Type: Set the Read Type as Continuous, Auto-Read or Timed to define how measurements are performed, saved and exported. If Timed is selected, set the time interval from 3 seconds (0:00:03) to 2 hours (2:00:00).
Page 25
Custom Data Log Export: Select an option to customize the amount of information Select All Data when exporting data to LIMS software. exported from the data log. Basic: Only critical data is exported from the data log including date, time, primary reading and temperature. Orion Pro Star EC212 Conductivity Bench Meter...
Page 26
Intermediate: Moderate data is exported from the data log including basic data plus log #, username, measure mode, secondary reading (if applicable) and short calibration log summary. Advanced: Significant data is exported from the data log including intermediate data plus read type, read time, sample ID, method, electrode ID, full calibration log summary and instrument ID.
Alarm Settings: Set alarm settings for high and/or low temperature values and calibration due reminder as needed. o Temperature Input: Select the source of temperature readings from the list. Automatic: When an ATC sensor is connected to the meter, the sensor is Orion Pro Star EC212 Conductivity Bench Meter...
Page 28
shown on the list. Ambient: The meter’s built-in sensor is used to measure ambient air temperature. While not as accurate as an ATC sensor in sample solution, it is more accurate than not reporting temperature. Manual: When no ATC temperature sensor is connected to the meter, the meter ...
Page 30
Solution (25- Coefficient Solution (25- Coefficient 50°C) (%/°C) 50°C) (%/°C) 1.90 o nLFn: For non-linear natural waters, meter applies a non-constant correction factor to natural water samples that have a varying response to changes in temperature. Recommended for low conductivity waters in equilibrium with air and natural waters like natural ground, well or surface waters.
Page 31
Cat. No. 013016MD). Note: Tap the Save button once all changes are made to preserve the changes or tap the Create method button to save the current parameters as a new method. Orion Pro Star EC212 Conductivity Bench Meter...
3.3 Calibrate Tap the Calibrate option in the Main Menu to access the complete list of calibrations available. The Orion Pro Star EC212 Conductivity bench meter is capable of performing: • 1 to 5 point conductivity calibration • 1 to 5 point TDS calibration •...
Note: Once data logs are deleted, the action cannot be undone. Create backup records of the data logs as needed before deleting data logs. Orion Pro Star EC212 Conductivity Bench Meter...
3.4.1 Filter the Data Log Tap the Filter By field to select and set parameters for Username (Secure Mode only), Date & Time and Log Number to filter the data log list. Secure Mode Example Open Mode Example 3.4.2 View Individual Data Logs Tap a log on the list to view the individual data log details.
When in Secure Mode, if an Administrator or Level 1 user account is active, that account will have the option to delete the data log after it is exported. When the data log becomes full, it must be exported and then deleted before new data logs can be saved. Orion Pro Star EC212 Conductivity Bench Meter...
3.5 Calibration Logs Tap the Calibration logs option in the Main Menu to access the calibration logs. The meter will store up to 20 calibration logs per parameter with time and date stamps. Each calibration log includes calibration data with calculated results. Tap to select a parameter to view the calibration log list for that parameter.
From the parameter-specific calibration log list, filter the list (Secure Mode only), view individual calibration logs and export the calibration log. Secure Mode Example Open Mode Example Tap the Filter By field to set parameters for Username (Secure Mode only) to filter the calibration log list. Orion Pro Star EC212 Conductivity Bench Meter...
3.5.2 View Individual Calibration Logs Tap a log on the list to view the individual calibration log details. • You may need to scroll down to see all included information. • Tap the blue arrows to scroll up or down through the individual calibration logs. •...
When in Secure Mode, if an Administrator or Level 1 user account is active, that account will have the option to delete the calibration log after it is exported. When the calibration log becomes full, it must be exported and then deleted before new calibration logs can be saved. Orion Pro Star EC212 Conductivity Bench Meter...
3.6 Methods Tap the Methods option in the Main Menu to access the methods list. Methods offer the option to save measurement settings to standardize analyses and SOPs. From the methods list, create new methods; load, view and edit existing methods; export existing methods to a USB flash drive and import methods from a USB flash drive.
Tap the Delete button to delete the selected method. Confirm you want to delete the method by tapping the Delete button again. The meter will return to the methods list and the selected method will be removed. Orion Pro Star EC212 Conductivity Bench Meter...
3.7 Secure Mode Tap the Secure Mode option in the Main Menu to access the secure mode function. Select the Secure Mode option and tap the Activate button to initiate the Secure Mode. Once the Secure Mode is activated, create up to 3 administrator and 30 user accounts, each with unique username, password and customized meter access.
• admin set Obtain Measurements • • • • • Record the username and temporary password. a. The first time the user logs into their account, they will need to enter the temporary Orion Pro Star EC212 Conductivity Bench Meter...
password and then create their own password. b. The first time a new admin logs into their account, they will also need to create 3 security questions. Tap the Save button. 3.7.1.3 Edit Existing Account (Admin Only) Admins can modify the first name, last name, email and access level of active accounts as well as reset the password.
Tap the Logout button. Confirm you want to logout and tap the Logout button. When no account is logged into the meter, the login screen will be displayed. To login to the meter, select your username and enter your password. Orion Pro Star EC212 Conductivity Bench Meter...
3.8 Notifications Tap the Notifications option in the Main Menu to view a list of the latest warnings and errors that have been shown on the meter. Onscreen notifications include info, success, warning and error notifications. Icon Notification Description Info notifications are triggered when there is a recommendation for the user to perform an action.
• Meter Tours: View onscreen meter tours to learn how to use important meter functions. Tours include conductivity calibration, measurements, meter connections and electrode stand assembly. The following are example screens from the conductivity calibration tour: Orion Pro Star EC212 Conductivity Bench Meter...
Page 48
• Meter Tests & Troubleshooting Guides: Perform a meter self test or electrode stability test for troubleshooting purposes. Select the type of test and follow all onscreen prompts to perform the selected test. o Meter Self Test: Meter will perform a series of internal verification tests to ensure the meter is functioning according to specifications.
The non-volatile meter memory retains all data when powered off and in the event of power failure. 4.1.1.1 Determined Conductivity Standard Value During the conductivity calibration, the meter will attempt to automatically recognize each conductivity Orion Pro Star EC212 Conductivity Bench Meter...
standard value using the selected Nominal Cell Constant, Conductivity Standard Set and the signal measured by the conductivity electrode. Once the reading is stable and the conductivity standard is recognized, the meter will automatically display the conductivity value at 25°C. •...
4.1.4 TDS Calibration Overview The TDS measure mode is automatically calibrated when a conductivity calibration is performed and the TDS factor is set. This is typically the most accurate way to calibrate for TDS measurements. Alternatively, the meter can be calibrated in TDS mode using TDS standards and the TDS calibration will be applied to TDS measurements only.
Page 53
(4, 3, 2 or 1), only the Next button will appear for each calibration point until the set number of points is completed, at which point the Done button will appear. Orion Pro Star EC212 Conductivity Bench Meter...
4.1.6 Salinity Calibration Overview The salinity measure mode is automatically calibrated when a conductivity calibration is performed. This is typically the most accurate way to calibrate for salinity measurements. Alternatively, the meter can be calibrated in salinity as % NaCl mode using a salinity as % NaCl standard and the salinity calibration will be applied to salinity as % NaCl measurements only.
To complete the calibration, tap the Done button. The meter will automatically save the calibration information to the calibration log and display a summary of the calibration. Tap the Measure button to proceed to the main measure mode. Orion Pro Star EC212 Conductivity Bench Meter...
Meter Tours to access onscreen tours that explain how to use important meter functions including measurements. The Orion Pro Star EC212 meter is capable of measuring pH, mV or ORP with temperature Refer to the 2.5 Touchscreen Functions section for details on the display components and icons.
• Tap the Measure button to start a measurement. • When the measurement is stable, the stable icon is shown and the measurement is locked on the display until the Measure button is tapped again. Orion Pro Star EC212 Conductivity Bench Meter...
Page 58
• The stable measurement is automatically saved to the data log and exported to an external device according to the Data Transfer Settings in the 3.2.1 Instrument Settings section. • Tap the Measure button to start a new measurement. • The Orion stirrer probe will turn on automatically when the Measure button is tapped. When on, tap the Stirrer button to manually turn off the stirrer probe.
Timed Read Type: Sample ID entry field will appear when the Start button is tapped and each time interval is reached (3 seconds to 2 hours). Select the Apply ID to Multiple Orion Pro Star EC212 Conductivity Bench Meter...
Readings option to use the same sample ID for subsequent readings and stop the Sample ID entry field from appearing. • Off: No sample ID is shown for measurements, logs or exported records. In the measure mode, when sample ID is set to auto-incremental or manual, the sample ID can also be entered by tapping the blue Sample ID text.
Record the displayed measurements as required. Remove the electrode from the sample. Repeat steps 1-5 for all samples. When all samples have been measured, store the electrode according to the instructions in the electrode manual. Orion Pro Star EC212 Conductivity Bench Meter...
5.3 Data Viewing Data Log When saving measurements to the meter data log, refer to the 3.4 Data Logs section for detailed information and instructions on viewing and exporting the meter’s data log. Save up to 10000 data logs in the meter’s non-volatile memory. Each data log is saved with its associated date and time stamp.
Select the “All Data” option when exporting data to LIMS software. • Custom Calibration Log Export: Select the amount of information to be exported from the calibration log. Select the “All Data” option when exporting data to LIMS software. Orion Pro Star EC212 Conductivity Bench Meter...
6.1.2 Connect the External Device USB Flash Drive For data transfer using a USB flash drive: • Power on the meter. • Insert a USB flash drive with USB-A plug into the USB meter input. Computer For data transfer using a computer: •...
• Empty commands (i.e. just a <CR>) will be ignored and a new prompt will be issued. • <CR> (Carriage Return, ASCII 13) is used to terminate a command. Whenever this character is received, the internal buffer will be processed. • Remote commands are not case sensitive. Orion Pro Star EC212 Conductivity Bench Meter...
Page 66
Remote Commands List Remote Command Action Example Command PGETMEAS <CR> Prints the current measurement immediately Prints the current measurement for a set number of PGETMEAS Data times PGETMEAS 2 Count <CR> Data Count = 1, 2, 3, etc. Prints all current calibration data GETCAL <CR>...
Wait while the meter software is updating. Do not remove the USB flash drive while the meter software is updating. Once the meter software update is complete, follow any onscreen prompts. The meter will restart and then the software update process is complete. Orion Pro Star EC212 Conductivity Bench Meter...
Chapter 7 Troubleshooting 7.1 Meter Troubleshooting Meter display not powering on: • Use a surge protector or uninterrupted power supply (UPS) so an unintended power surge of electricity does not damage the meter or meter power supply. • Check that the wall outlet is functional. •...
Conductivity Calibration: • Clean the conductivity electrode, 0.1M HCl is recommended and adding pepsin is helpful if protein build-up occurs. See the electrode user manual for detailed instructions. Orion Pro Star EC212 Conductivity Bench Meter...
• Discard the current standards, check the expiration date on the standard bottle or pouch, pour a fresh portion of the standard into a clean, dry beaker and repeat the calibration. • Do not allow conductivity standards to remain exposed to air for extended periods of time. •...
0.001 ppt minimum; 0.01 PSU minimum; or 0.01 % NaCl minimum, auto-ranging Relative accuracy ±0.5% reading Modes Natural Seawater (ppt), Practical Salinity (PSU),or % NaCl Resistivity Range 1.0 Ω to 100.0 MΩ Resolution 0.1 Ω minimum, auto-ranging Relative accuracy ±0.5% reading Orion Pro Star EC212 Conductivity Bench Meter...
Page 72
Temperature Range -30.0 to 130.0 ºC, -22.0 to 266.0 ºF Resolution 0.1 ºC, 0.1 ºF Relative accuracy ±0.3 ºC, ±0.5 ºF Calibration points 1 or 2 points Offset adjustment Up to ±5.0 ºC Temperature Automatic with ATC sensor, ambient with built-in meter temperature sensor or source manual entry Measurement Functions...
Page 73
216.3 mm x 153.2 mm x 91.62 mm, 8.52” x 6.03” x 3.61” (L x W x H) Weight 1.00 kg, 2.20 lbs. Environmental Conditions Environmental Indoors conditions Altitude Up to 2,000 Meters Operating 5°C to 45°C temperature Orion Pro Star EC212 Conductivity Bench Meter...
Operating relative 5 to 85%, non-condensing humidity Storage -20°C to 60°C temperature Storage relative 5 to 85%, non-condensing humidity Mains fluctuation ±10% of range (100-240VAC) Installation category Pollution degree Protection class Note: Specifications subject to change without notice. All relative accuracy values are influenced by the displayed least significant digit and should include ±...
Chapter 9 Regulatory Compliance 9.1 Regional Approvals European Union: The European CE marking is applied on products that meet all the applicable requirements of the European Directives. Products not marked with a CE marking either do not operate in the 230V / 50Hz voltage range or are not intended to be sold to the EU Member States or European Economic area (EEA).
9.3.2 Canadian ISED IC Notice This ISM digital apparatus complies with Canadian ICES-001, Class A. Cet appareil ISM est conforme á la norme NMB-001 du Canada, Classe A. 9.3.3 South Korean KC Registration Orion Pro Star EC212 Conductivity Bench Meter...
[Translation of the above user notification: Equipment for business broadcasting and communication specified in Article 11, Subparagraph 1), Items (가), (바), and (사 1). This device has undergone suitability assessment for use in a business environment. If used in a residential environment, there may be concerns regarding radio interference.] 9.4 Environmental Compliance 9.4.1 REACH - Europe...
Ulteriori informazioni sulla conformità di Thermo Fisher Scientific con queste Direttive, l’elenco delle ditte di riciclaggio nel Vostro paese e informazioni sui prodotti Thermo Scientific che possono essere utili alla rilevazione di sostanze soggette alla Direttiva RoHS sono disponibili sul sito http://www.thermofisher.com. Orion Pro Star EC212 Conductivity Bench Meter...
Conformité WEEE. Ce produit doit être conforme à la directive euro-péenne (2012/19/EU) des Déchets d’Equipements Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant. Thermo Fisher Scientific s’est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de l’union européenne et ce produit devraitêtre collecté...
(PSTAR2120), 4-cell epoxy/graphite-body K=0.475 conductivity/ATC PSTAR2125 electrode (013005MD), 1413 μS/cm conductivity standard pouches Orion Pro Star EC212 conductivity bench meter pure water kit, includes meter with stand (PSTAR2120), ultra pure water 2-cell K=0.1 conductivity/ATC electrode with PSTAR2126 flow cell (013016MD), 100 μS/cm conductivity standard bottles...
10.2 Electrode and Solution Ordering Info Description Catalog Number Conductivity/ATC DuraProbe 4-cell K=0.475 epoxy/graphite body electrode 013005MD Conductivity/ATC ultra pure water 2-cell K=0.1 stainless steel body electrode 013016MD 100 µS/cm conductivity/TDS standard, 5 x 60 mL 011008 1413 µS/cm conductivity/TDS standard, 5 x 60 mL 011007 12.9 mS/cm conductivity/TDS standard, 5 x 60 mL 011006...
Page 84
Orion Pro Star EC212 Conductivity Bench Meter Scientists have trusted Thermo Scientific Orion Lab Products for over 60 years to provide them with accurate and reliable analytical testing instruments in the lab and in the field. Since 1962, Thermo Scientific Orion products have set the standard for quality in electrochemistry. We offer instruments for analytical and optical measurement that allow scientists to solve some of the most challenging issues facing society today.