ThermoFisher Scientific Orion Pro Star PD216 Operation Manual
ThermoFisher Scientific Orion Pro Star PD216 Operation Manual

ThermoFisher Scientific Orion Pro Star PD216 Operation Manual

Electrochemistry meters, ph/ion and dissolved oxygen bench meter
Table of Contents

Advertisement

Quick Links

Operations Manual
Electrochemistry Meters
Update Meter Software to
Ensure Optimum Performance
www.thermofisher.com/orionsoftware
Orion Pro Star PD216 pH/Ion
and Dissolved Oxygen Bench
Meter User Manual
68X002730 | Revision 01 | December 2024

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Orion Pro Star PD216 and is the answer not in the manual?

Questions and answers

Summary of Contents for ThermoFisher Scientific Orion Pro Star PD216

  • Page 1 Operations Manual Electrochemistry Meters Update Meter Software to Ensure Optimum Performance www.thermofisher.com/orionsoftware Orion Pro Star PD216 pH/Ion and Dissolved Oxygen Bench Meter User Manual 68X002730 | Revision 01 | December 2024...
  • Page 3: Table Of Contents

    Table of Contents Chapter 1 ........................ 8 Introduction ....................8 1.1 Intended Use ..................9 1.1.1 Intended Use ....................9 1.1.2 Non-Intended Use ..................9 1.1.3 Safety ......................9 1.1.4 Power Connection ..................10 1.2 Cleaning Instructions (Meter Only) ............ 10 Chapter 2 ......................
  • Page 4 3.3 Calibrate ..................36 3.4 Data Logs ..................37 3.4.1 Filter the Data Log ..................38 3.4.2 View Individual Data Logs ................38 3.4.3 Export the Data Log ..................39 3.5 Calibration Logs ................40 3.5.1 Filter the Calibration Log ................40 3.5.2 View Individual Calibration Logs ..............41 3.5.3 Export the Calibration Log ................42 3.6 Methods ..................
  • Page 5 3.10 Information & Help ................. 57 Chapter 4 ......................59 Calibration ....................59 4.1 pH Calibration ................. 59 4.1.1 pH Calibration Overview .................59 4.1.1.1 Determined pH Buffer Value ..............59 4.1.2 pH Calibration Procedure ................60 4.1.3 pH Value vs. Temperature ................61 4.2 ORP Calibration................
  • Page 6 Chapter 6 ......................80 Data Exporting and Importing ..............80 6.1 Data Transfer .................. 80 6.1.1 Data Transfer Settings ...................80 6.1.2 Connect the External Device ................81 6.1.3 Verify the Data Destination ................82 6.2 Remote Commands ................ 82 6.3 Meter Software Update Procedure ........... 84 Chapter 7 ......................
  • Page 7 9.5 WEEE Compliance ................98 9.6 Proper Product Disposal ..............99 Chapter 10 ......................100 Ordering Info .................... 100 10.1 Meter Ordering Information ............100 10.2 Electrode and Solution Ordering Info ..........101...
  • Page 8: Chapter 1

    ™ Pro Star PD216 pH/ion and dissolved oxygen bench meter. Orion Pro Star PD216 pH/ion and DO bench meters are designed to provide accurate and reliable testing in laboratories that need a modern, easy-to-operate meter with advanced performance and reliable data reporting.
  • Page 9: Intended Use

    Secure a wide range of electrodes using the various electrode holder options. Neatly organize electrode cables in the built-in side channels of the stand. All Orion Pro Star PD216 meters include an electrode stand, USB flash drive, computer cable and universal 100-240V 50/60 Hz power adapter with wall plugs for US/Japan, Euro, UK/Singapore, Australia/New Zealand and China.
  • Page 10: Power Connection

    Note: This symbol, in the context of a Note, indicates to follow the manufacturer specified instructions, notes and requirements set out in the manuals. CAUTION: Before using or maintaining this product, please be sure to read the manual carefully. Failure to follow these instructions may cause the product to malfunction. CAUTION: Use this product only in the way described in the product literature and in this manual.
  • Page 11: Chapter 2

    Slide the appropriate plug plate into the groove on the back of the power adapter. Connect the assembled power adapter to the meter input. a. Locate the 2 slots on the meter power input. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 12 b. Align the 2 tabs on the power adapter plug with the 2 indents on the meter power input. c. Insert the power adapter plug into the meter power input and twist clockwise to lock the connection. Connect the assembled power adapter to a power outlet. You are now ready to turn on the meter.
  • Page 13: Electrode Stand Installation

    The amount of up and down resistance can be adjusted using the upper dial. Dialing left decreases the resistance, making the arm easier to move up and down. Dialing right increases the resistance, making the arm harder to move up and down. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 14 b. The amount of front and back resistance can be adjusted using the lower dial. Dialing left decreases the resistance, making the arm easier to move forward and backward. Dialing right increases the resistance, making the arm harder to move forward and backward.
  • Page 15 Use the side channels built into the electrode arm to organize the electrode cables. Press the electrode cables into the 4 grips inside the arm channels to secure the cables in place. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 16: Meter Inputs

    2.3 Meter Inputs 2.3.1 Power Button The meter power button is located on the right side of the meter. Press and release the power button to power the meter on or off. 2.3.2 USB Inputs The USB meter inputs are located on the left side of the meter. Each input has a protective cap that protects the meter from dust and water intrusion.
  • Page 17: Back Panel Meter Inputs

    Stirrer Inputs: Connect the Orion stirrer probe (catalog number 096019-WA), with 3.5 mm phono jack plug, to the meter. • Lift the protective cap labeled Stirrer 1 or Stirrer 2. • Insert the Orion stirrer probe plug fully into the meter input. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 18: Meter Startup Sequence

    • 1 or 2 Orion stirrer probes can connect with the stirrer inputs on the meter; the user can assign the Stirrer 1 or Stirrer 2 input to each measurement channel in the measurement settings. • Orion stirrer probes are meter-powered and meter-controlled for easy sample stirring and rinsing. Dissolved Oxygen Input: Connect an Orion dissolved oxygen (DO) electrode, with 9-pin MiniDIN plug, to the meter.
  • Page 19: Touchscreen Functions

    2.5.1 Main Measure Mode Note: User-selectable information (i.e. instrument name) will only be shown on the measure mode display if it is entered in the settings menus. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 20 Connectivity Icons: Indicate connectivity status of USB devices and export device with the meter. Indicates a USB flash drive is connected and communicating with USB Flash Drive: the meter. Indicates a USB accessory (USB mouse, USB keyboard or USB USB Accessory: barcode scanner) is connected and communicating with the meter.
  • Page 21: Single Measure Mode View

    Graph (Single Channel View Only): Tap the blue Graph text to view the current measurement as a graph with temperature. • Select the Read Type to change how the graph tracks measurements over time. • Tap the Display Temperature button to show or hide temperature measurements. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 22: Basic Measure Mode View

    2.5.1.2 Basic Measure Mode View Refer to the 3.2.1 Instrument Settings section and Display settings and select Basic as the Custom Measure Layout to view a simplified version of the measure mode. The default meter setting is Advanced as the Custom Measure Layout and all user-selectable information (i.e. instrument name, sample ID, electrode ID, method) will only be shown when it is entered in the settings menus.
  • Page 23: Electrode Information

    2.6 Electrode Information The Orion Pro Star PD216 meter is compatible with a pH electrode, ORP (redox) electrode or ion selective electrode (ISE) with BNC connector; ATC temperature probe with 8-pin MiniDIN connector; reference half-cell electrode with pin-tip connector and DO electrode (polarographic or optical) with 9- pin MiniDIN connector.
  • Page 24: Chapter 3

    Chapter 3 Main Menu 3.1 Main Menu Tap the Main Menu icon in the measure mode to access the Main Menu. • Settings: View and customize instrument and measurement settings. • Calibrate: Perform all calibration functions, including temperature and barometer calibrations. •...
  • Page 25: Settings

    • DO Settings: View and customize measurement settings for the dissolved oxygen channel, including temperature, measure mode, sample ID, alarms, resolution, barometric pressure, salinity correction, stability criteria, stirrer and electrode ID. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 26: Instrument Settings

    3.2.1 Instrument Settings Tap the Instrument Settings option in the Settings list to access the instrument settings. • Read Type: Set the Read Type as Continuous, Auto-Read or Timed to define how measurements are performed, saved and exported. If Timed is selected, set the time interval from 3 seconds (0:00:03) to 2 hours (2:00:00).
  • Page 27 Select All Data when exporting data to LIMS software. exported from the data log. Basic: Only critical data is exported from the data log including date, time,  primary reading and temperature. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 28 Intermediate: Moderate data is exported from the data log including basic data  plus log #, username, measure mode, secondary reading (if applicable) and short calibration log summary. Advanced: Significant data is exported from the data log including intermediate  data plus read type, read time, sample ID, method, electrode ID, full calibration log summary and instrument ID.
  • Page 29: Ph/Ion Settings

    Alarm Settings: Set alarm settings for high and/or low temperature values and calibration due reminder as needed. o Temperature Input: Select the source of temperature readings from the list. Automatic: When ATC sensors are connected to the meter, all available sensors  Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 30 are shown in the list. The meter will default to ATC sensor connected to the active channel and the ATC sensor from the other channel can be selected. Ambient: The meter’s built-in sensor is used to measure ambient air  temperature.
  • Page 31 Note: During an ion calibration, values can also be manually apart (i.e. 1.000 and 10.000). entered for each point. o 1.000, 10.00, 100.0, 1000, 10000 o Custom: Up to 5 custom standard values Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 32 • Stability Criteria: Select the Measurement Stability and Measurement Averaging functions. o Measurement Stability: Select Automatic, Fast, Medium or Slow Stability for measurements. The meter will use specific resolution criteria to determine the wait time before indicating measurements are stable. o Measurement Averaging: Select Automatic or Off for averaging measurements.
  • Page 33: Do Settings

    ATC sensors from other channels can be selected. Ambient: The meter’s built-in sensor is used to measure ambient air  temperature. While not as accurate as an ATC sensor in sample solution, it is Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 34 more accurate than not reporting temperature. Manual: When no ATC temperature sensor is connected to the meter, the meter  will default to MAN as the temperature input. Use the manual temperature input field to enter the sample temperature value, from -30.0°C to 130.0°C. •...
  • Page 35 IDs; electrode ID is shown on the display, data log, calibration log and exported records. Note: Tap the Save button once all changes are made to preserve the changes or tap the Create method button to save the current channel parameters as a new method. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 36: Calibrate

    3.3 Calibrate Tap the Calibrate option in the Main Menu to access the complete list of calibrations available. The Orion Pro Star PD216 pH/ion and DO bench meter is capable of performing: • 1 to 5 point pH calibration • 1 point ORP calibration •...
  • Page 37: Data Logs

    Note: Once data logs are deleted, the action cannot be undone. Create backup records of the data logs as needed before deleting data logs. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 38: Filter The Data Log

    3.4.1 Filter the Data Log Tap the Filter By field to select and set parameters for Username (Secure Mode only), Date & Time and Log Number to filter the data log list. Secure Mode Example Open Mode Example 3.4.2 View Individual Data Logs Tap a log on the list to view the individual data log details.
  • Page 39: Export The Data Log

    When in Secure Mode, if an Administrator or Level 1 user account is active, that account will have the option to delete the data log after it is exported. When the data log becomes full, it must be exported and then deleted before new data logs can be saved. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 40: Calibration Logs

    3.5 Calibration Logs Tap the Calibration logs option in the Main Menu to access the calibration logs. The meter will store up to 20 calibration logs per parameter with time and date stamps. Each calibration log includes calibration data with calculated results. The parameter list for calibration logs is grouped by channel.
  • Page 41: View Individual Calibration Logs

    Tap a log on the list to view the individual calibration log details. You may need to scroll down to see all included information. Tap the blue arrows to scroll up or down through the individual calibration logs. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 42: Export The Calibration Log

    Tap the Export button to export the individual calibration log. 3.5.3 Export the Calibration Log Tap the Export button to export the calibration log to a USB flash drive, computer or printer. • Use the Instrument Settings / Data Transfer Settings to view and modify the Export Destination, Output Format, Baud Rate (when applicable) and Custom Calibration log Export settings.
  • Page 43: Methods

    Use the alphanumeric keypad to enter the method name using up to 20 characters and then tap the Save button. The new method will be added to the list of methods for the selected channel. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 44: View And Load An Existing Method

    3.6.2 View and Load an Existing Method From the methods list, tap the load (triangle) icon or view (eye) icon to view the method summary information for an existing method on the list. From the method summary screen, perform one of the following actions: a.
  • Page 45: Incremental Techniques

    When double known addition or double known subtraction is performed, the slope is also determined with the sample concentration. For certain samples, incremental techniques for ion concentration analysis may offer significant benefits Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 46 for accuracy, precision, compensation for matrix effects and preparation simplicity. There is no need to prepare multiple calibration standards, create a calibration curve or perform periodic recalibration. Available incremental techniques are single known addition, double known addition, single known subtraction, double known subtraction, single analyte addition and single analyte subtraction. •...
  • Page 47: Incremental Technique Tips

    9 mV or more for a divalent ion selective electrode. The sample and standard must be volumetrically measured with pipets and graduated cylinders. The total volume of additions made should not exceed 10% of the starting sample volume. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 48: Create A New Incremental Technique

    Any complexing agent must be present in excess (at least 50 to 100 times) or not at all. The ratio of free to complexed ions must remain constant over the analysis. An indication that a complexing agent is present, but not in sufficient quantity to maintain a constant free ion to complexed ion ratio, is an abnormally high or low slope.
  • Page 49 2 silver ions need 1 sulfide ion to • subtract them and the reaction ratio is 2/1 = 2. Name: Unique name to identify the programmed settings • • Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 50: View And Run An Existing Incremental Technique

    3.7.3 View and Run an Existing Incremental Technique From the incremental technique list, tap the run (triangle) icon or view (eye) icon to view the incremental technique summary information for an existing incremental technique on the list. From the incremental technique summary screen, perform one of the following actions: a.
  • Page 51: Delete An Existing Incremental Technique

    Confirm you want to delete the incremental technique by tapping the Delete button again. The meter will return to the incremental techniques list and the selected incremental technique will be removed. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 52: Secure Mode

    3.8 Secure Mode Tap the Secure Mode option in the Main Menu to access the secure mode function. Select the Secure Mode option and tap the Activate button to initiate the Secure Mode. Once the Secure Mode is activated, create up to 3 administrator and 30 user accounts, each with unique username, password and customized meter access.
  • Page 53: Create New Account (Admin Only)

    • • admin set Ready Type • • • admin set Mode • • • admin set Sample ID • • • admin set Obtain Measurements • • • • • Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 54: Edit Existing Account (Admin Only)

    Record the username and temporary password. a. The first time the user logs into their account, they will need to enter the temporary password and then create their own password. b. The first time a new admin logs into their account, they will also need to create 3 security questions.
  • Page 55: Deactivate Secure Mode (Admin Only)

    Tap the Logout button. Confirm you want to logout and tap the Logout button. When no account is logged into the meter, the login screen will be displayed. To login to the meter, select your username and enter your password. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 56: Notifications

    3.9 Notifications Tap the Notifications option in the Main Menu to view a list of the latest warnings and errors that have been shown on the meter. Onscreen notifications include info, success, warning and error notifications. Icon Notification Description Info notifications are triggered when there is a recommendation for the user to perform an action.
  • Page 57: Information & Help

    DO calibration, measurements, meter connections, electrode stand assembly, pH electrode preparation, pH electrode storage and pH electrode cleaning. The following are example screens from the pH calibration tour: Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 58 • Meter Tests & Troubleshooting Guides: Perform a meter self test, electrode stability test or pH electrode check for troubleshooting purposes. Select the type of test and follow all onscreen prompts to perform the selected test. o Meter Self Test: Meter will perform a series of internal verification tests to ensure the meter is functioning according to specifications.
  • Page 59: Chapter 4

    • USA: 1.68, 4.01, 7.00, 10.01, 12.46 • Pure Water: 4.10, 6.97, 9.15 • DIN: 1.68, 4.01, 6.86, 9.18, 12.46 • Custom: Up to 5 custom buffer values • STD: 1.00, 3.00, 6.00, 8.00, 10.00, 13.00 Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 60: Ph Calibration Procedure

    4.1.2 pH Calibration Procedure Ensure the active measure mode is set to pH. a. If using a method, ensure the method is loaded and active. Tap the Calibrate button to start the calibration. a. If viewing multiple measurement channels, select the pH calibration option. Rinse the pH electrode and ATC sensor with purified water and place into the pH buffer.
  • Page 61: Ph Value Vs. Temperature

    11.728 1.715 4.071 6.837 8.964 11.609 1.724 4.086 6.841 8.942 11.499 1.733 4.103 6.847 8.922 11.396 1.744 4.122 6.854 8.903 11.301 1.755 4.142 6.861 8.885 11.213 1.766 4.163 6.984 9.455 11.130 Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 62 STD: 1.00, 3.00, 6.00, 8.00, 10.00, 13.00 Temp. pH 1.00 pH 3.00 pH 6.00 pH 8.00 pH 10.00 pH 13.00 (°C) Buffer Buffer Buffer Buffer Buffer Buffer 0.97 3.02 6.04 8.19 10.27 14.02 0.98 3.02 6.03 8.15 10.22 13.81 0.98 3.02 6.02 8.11...
  • Page 63: Orp Calibration

    To accept the ORP value, no action is required. b. To edit the ORP value, tap the edit value text. Tap the ORP value and use the numeric Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 64 keypad to enter the known ORP value of the standard at the measured temperature. Tap the Save button to save the entered value. Tap the Done button to complete the calibration. The meter will automatically save the calibration information to the calibration log and display a summary of the calibration. Tap the Measure button to proceed to the main measure mode.
  • Page 65: Ion Calibration

    (ATC) probe is used with the ion selective electrode to ensure temperatures are the uniform. The non-volatile meter memory retains all data when powered off and in the event of power failure. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 66: Ion Calibration Procedure

    4.3.2 Ion Calibration Procedure Ensure the active measure mode is set to ion and the selected ISE Type matches the ion selective electrode to be calibrated. a. If using a method, ensure the method is loaded and active. Tap the Calibrate button to start the calibration. a.
  • Page 67: Dissolved Oxygen Calibration

    It is necessary to set the proper temperature, barometric pressure and salinity values prior to performing any DO calibration or measurement. The non-volatile meter memory retains all data when powered off and in the event of power failure. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 68: Saturation Do Calibration Procedure

    4.4.2 % Saturation DO Calibration Procedure Ensure the active measure mode is set to DO. a. If using a method, ensure the method is loaded and active. Tap the Calibrate button to start the calibration. a. If viewing multiple measurement channels, select the DO calibration option. Rinse the DO electrode with purified water and place into the prepared calibration sleeve, water sample, BOD bottle or airtight flask.
  • Page 69: Concentration Do Calibration Procedure

    This may be helpful if the electrode is responding more slowly than usual. Tap force stabilize text to accelerate the reading stability (may give less accurate calibration results). Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 70: Temperature Calibration

    4.5 Temperature Calibration 4.5.1 Temperature Calibration Overview The thermistor sensor used for automatic temperature compensation (ATC) and measurements is both accurate and stable, so frequent calibration is not required. If desired, perform a temperature calibration to calibrate the ATC sensor when it is connected to the meter. A 1 to 2 point temperature offset calibration can be performed, up to ±5.0°C for each calibration point.
  • Page 71: Barometer Calibration

    To complete the calibration, tap the Done button. The meter will automatically save the calibration information to the calibration log and display a summary of the calibration. Tap the Measure button to proceed to the main measure mode. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 72: Chapter 5

    3.10 Information & Help section and Meter Tours to access onscreen tours that explain how to use important meter functions including measurements. The Orion Pro Star PD216 meter offers 2 measurement channels, capable of measuring the following parameters: • Channel 1 (pH): pH, mV, ORP or ion concentration with temperature •...
  • Page 73 82.0% or greater than 114.0%. Ranges are adjustable. The ion calibration will use the units selected in Calibration the pH/Ion Channel settings. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 74: Measure Mode Read Types

    Mode Example Calibration Details Notes Optical DO calibration details include the optical cap life icon, which shows: Optical DO • 3 green bars when ≥180+ days remain. Calibration • 2 orange bars when 31 to 180 days remain. 1 red bar when ≤30 days remain. 5.1.1 Measure Mode Read Types Refer to the 3.2.1 Instrument Settings...
  • Page 75 • Orion stirrer probes will turn on automatically when the Start button is tapped. When on, tap the Stirrer button to manually turn off the Orion stirrer probes. Orion stirrer probes will turn off automatically when the Stop button is tapped. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 76: Sample Id

    Timed, Waiting Timed, Reading Stabilizing Timed, Reading Stable 5.1.2 Sample ID When using Sample ID to identify samples with measurements, determine which channels will be active in the measure mode and refer to the 3.2.2 pH/Ion Settings 3.2.3 DO Settings sections and Sample ID setting to turn on sample ID, set the type of sample ID and enter the sample ID.
  • Page 77: Stirrer Button

    Press and hold the Stirrer button to access and change the stirrer control settings. Note: The Stirrer button is hidden if the stirrer function is turned off for all actively shown channels. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 78: General Measurement Procedure

    5.2 General Measurement Procedure Refer to the electrode manuals for detailed instructions on electrode preparation, calibration and usage. Make sure the electrodes are fully prepared, properly calibrated and working correctly before taking sample measurements. Refer to the 3.2.1 Instrument Settings section and Read Type settings to define how measurements are performed and when measurements are saved and exported.
  • Page 79: Data Viewing

    • When in Secure Mode, the meter will save up to 20 calibration logs and prompt the Administrator and Level 1 accounts to export and delete calibration logs when the limit is reached. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 80: Chapter 6

    Chapter 6 Data Exporting and Importing 6.1 Data Transfer Export live measurements, completed calibrations, data logs, calibration logs, methods, user lists (Secure Mode only) and audit trails (Secure Mode only) from the meter to a USB flash drive, computer or Orion ink printer (catalog number STARA-106). To successfully export data: Set all required data transfer settings in the Instrument Settings menu.
  • Page 81: Connect The External Device

    Baud Rate : 115200 bps Baud Rate : 9600 bps Data bits: 8 Data bits: 8 Parity: None Parity: None Stop bits: 1 Stop bits: 1 Flow Control: None Flow Control: None Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 82: Verify The Data Destination

    6.1.3 Verify the Data Destination USB Flash Drive For data transfer using a USB flash drive, perform a test export of data from the meter to the USB flash drive. Remove the USB flash drive from the meter, insert it in a computer and open the test export on the computer.
  • Page 83 TIME MODE = 1 (PM), 2 (AM) or 3 (24 Hour) SETMODE MODE Sets the meter measurement mode. MODE = PH or SETMODE PH <CR> GETMODE <CR> Prints the active measurement mode Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 84: Meter Software Update Procedure

    6.3 Meter Software Update Procedure Note: Back up all data before performing the meter software update procedure, as it will be erased when the meter software update is completed. Power on the meter. Download the latest Orion Pro Star meter software at www.thermofisher.com/orionsoftware. Unzip/extract the Orion Pro Star meter software file.
  • Page 85: Chapter 7

    Perform a factor reset on the meter. o Replace the electrode. • pH Electrode Icon: o Troubleshoot the pH buffers and pH electrode. o Replace the pH buffers. o Clean and condition the pH electrode. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 86: Measurement Troubleshooting

    o Replace the pH electrode as needed. 7.2 Measurement Troubleshooting Reading is unstable, slow to stabilize: • Consider changing the displayed measurement resolution. For example, some pH measurements only require 0.01 pH reporting. Changing the pH resolution from 0.001 to 0.01 will reduce the amount of time needed for the reading to stabilize.
  • Page 87 Dissolved Oxygen Calibration: • Perform maintenance on the DO electrode according to the electrode user manual. • For polarographic DO electrodes, ensure the DO electrode is fully polarized before performing a calibration. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 88: Technical Support

    • For zero percent saturation calibrations, prepare a sodium sulfite solution by dissolving about 15.0 g Na in 250 mL distilled water. Transfer the solution to a BOD bottle or airtight flask. 7.4 Technical Support Contact our Orion Technical Support team at 1-800-225-1480, +1-978-232-6000 or WLP.TechSupport@thermofisher.com.
  • Page 89: Chapter 8

    ORP or E method Calibration offset ±250 mV Range 0.0001 to 19999 Resolution 0.0001 minimum, 1 to 5 digits Relative accuracy ± 0.2 mV or ±0.05 % of the reading, whichever is greater Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 90 Units ppt, ppm, ppb, mg/L, M, mM, %, none or user-defined Calibration points 1 to 5 points Calibration Automatic standard recognition, manual entry option method Calibration 1.000, 10.00, 100.0, 1000; 10000 or up to 5 custom standards standards Slope type Linear or segmented Calibration Slope and E...
  • Page 91 User set for button tap, notifications and alarms, stable reading Maintenance Up to 5 electrode maintenance reminders and up to 5 solution maintenance Reminder reminders Sleep mode Selectable time interval of 5, 10, 20, 30, 60 minutes or off Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 92 Meter tests Self-test, stability test and pH electrode check Meter tours and Calibration modes, measurement mode, meter connections, electrode stand guides assembly and pH electrode maintenance Stirrer probe inputs Stirrer probe 5 selectable speeds from 250 RPM to 3600 RPM speeds Channel 1: BNC, 8-pin MiniDIN (ATC), reference pin-tip Electrode inputs...
  • Page 93: Meter Dimensions

    The warranty on your meter shall not apply to defects resulting from: • Improper or inadequate maintenance by customer • Unauthorized modification or misuse • Operation outside of the environment specifications of the products Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 94: Return Of Items

    8.3.1 Return of Items Authorization must be obtained from our Technical Support Team or authorized distributor before returning items for any reason. Please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against damage or loss.
  • Page 95: Chapter 9

    9.3 Electromagnetic Compatibility The meter requires special precautions regarding electromagnetic compatibility (EMC) and needs to be installed and put into service according to the EMC information provided in this instruction manual. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 96: Fcc Statement (Usa)

    Recommendations for actions to ensure the product remains safe regarding electromagnetic disturbances: Do not modify the product beyond the intended user-adjustable settings. It is important to use only accessories as supplied or recommended by Thermo Fisher Scientific. The equipment should be visually inspected regularly for damaged cables and connectors. Damaged cables should be replaced.
  • Page 97: Environmental Compliance

    EN 63000:2018. This confirms that any potential trace contamination levels of the substances listed above are below the maximum level set by the latest regulations or are exempt due to their application. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 98: Weee Compliance

    9.5 WEEE Compliance WEEE Compliance. This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2012/19/EU. It is marked with the following symbol. Thermo Fisher Scientific has contracted with 1 or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them.
  • Page 99: Proper Product Disposal

    Safety Data Sheets (SDS), or regulations applicable to your country and region. • This product has materials and components that may be recycled or reused according to local guidelines and regulations. • Clean up any chemical safety hazards using appropriate methods. Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 100: Chapter 10

    DO electrode maintenance kit; BOD adapter Orion Pro Star PD216 pH/ion and DO bench meter pH and BOD kit; includes meter with stand (PSTAR2160); ROSS Ultra Triode refillable epoxy-body pH/ATC electrode (8157BNUMD);...
  • Page 101: Electrode And Solution Ordering Info

    910199 and pH electrode storage solution ROSS All-in-One pH buffer kit: 475 mL each pH 4.01, 7.00 and 10.01 buffers 810199 and ROSS pH electrode storage; 60 mL electrode cleaning solution Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 102 Description Catalog Number pH 1.68 buffer, 475 mL 910168 pH 4.01 buffer, 475 mL 910104 pH 6.86 buffer, 475 mL 910686 pH 7.00 buffer, 475 mL 910107 pH 9.18 buffer, 475 mL 910918 pH 10.01 buffer, 475 mL 910110 pH 12.46 buffer, 475 mL 910112 pH 4.01 buffer, 25 individual-use pouches 910425...
  • Page 103 User Notes ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ ________________________________________________________________________________________ Orion Pro Star PD216 pH/Ion and DO Bench Meter...
  • Page 104 Orion Pro Star PD216 pH/Ion and Dissolved Oxygen Bench Meter Scientists have trusted Thermo Scientific Orion Lab Products for over 60 years to provide them with accurate and reliable analytical testing instruments in the lab and in the field. Since 1962, Thermo Scientific Orion products have set the standard for quality in electrochemistry.

Table of Contents