Summary of Contents for Alia AMC3200E Series

  • Page 1 ® ALIA TECHNOLOGY LLC El ectromagnetic Flowmeter Operation Manual AMC3200E Series...
  • Page 2: Table Of Contents

    Index 1. Sensor ----------------------------------------------------------------------------------------------------------------------------------2 1.1 AMF900 Flange Type-------------------------------------------------------------------------------------------------------2 1.2 AMF500 Wafer Type --------------------------------------------------------------------------------------------------------3 1.3 AMF301 Installation ---------------------------------------------------------------------------------------------------------3 1.4 AMF500 Installation ---------------------------------------------------------------------------------------------------------3 1.5 AMF601 Installation ---------------------------------------------------------------------------------------------------------4 1.6 AMF900 Installation ---------------------------------------------------------------------------------------------------------4 2. Installation Considerations---------------------------------------------------------------------------------------------------------5 2.1 Installation location ----------------------------------------------------------------------------------------------------------5 2.2 Avoid Magnetic Field interference ---------------------------------------------------------------------------------------5 2.3 Straight Pipe Distance ------------------------------------------------------------------------------------------------------5 2.4 Installation method ----------------------------------------------------------------------------------------------------------5 3.
  • Page 3: Sensor

    1. Sensor EM Flowmeter is composed by sensor and converter, be compact version and separate version, there are several specifications of sensor: 1.1 AMF900 Flange Type Size: 10Amm ~ 2000mm (3/8” A~ 80”) Liner: Neoprene Polyurethane PTFE Protection: IP68 Max. Temp: 180 C...
  • Page 4: Amf500 Wafer Type

    1.2 AMF500 Wafer Type Size: 25mm ~ 200mm (1” ~ 8”) Liner: FEP Protection: IP68 Max. Temp: 180 Installation: Wafer Size Dimensions (mm) Inch 1" 1-1/4" 1-1/2" 2" 2-1/2" 3" 4" 5" 6" 8" 1.3 AMF301 Installation 1.4 AMF500 Installation Tie-rod Packing Flange...
  • Page 5: Amf601 Installation

    1.5 AMF601 Installation 1.6 AMF900 Installation Note: 1) Pipe flange should be welded before flowmeter’s installation. Welding after flwometer’s installation is prohibited. And the welding part of pipe flange should be flat, having no sharp residue. Otherwise liner will be damaged. After flowmeter is installed, if other place in pipe needs to be welded, flowmetert’s power must be shut down.
  • Page 6: Installation Considerations

    Installation Considerations Before tube design, please take the following situations into consideration: 2.1 Installation location Please avoid sunlight when installing flowmeter; ambient temperature should be between -25 and 60 Deg.C. 2.2 Avoid Magnetic Field interference Please DO NOT install flowmeter near electric appliances such as motor-driven machine, transformer, and frequency transformer for that will cause magnetic field interference.
  • Page 7 d. Electrode position should parallel with ground The electrode position (A.B) of EM Flowmeter which is horizontally-installed or slant-installed installed should match the 2 sides (right/left) of tube, and converter (wiring box) should be on the top of tube. Exhaust valve >...
  • Page 8: Amc3200E Operation

    3. AMC3200E Operation 3.1 Wiring Diagram of Power and Signal Output Connect terminals one by one when back cover is opened. 90 260VAC 50 60Hz 4 20mA output 24VDC 4 20mA output Pulse output RS485 output Pulse output +. RS485 output To+ To- R e tu rn te mp e rtu re Su p p l y te mp e rtu re...
  • Page 9: Wiring Diagram For Separate Type

    3.2 Wiring Diagram for Separate type Converter AMC3200E Series A B C Electorde Coil Sensor AMF301 AMF500 AMF601 AMF900 A B C AMF100 Series 3.3 AMC3200E Panel & Dimensions 190mm 05 16 2017 08 08 08 +5 928 Flow 26mm 40mm 0 498 ∑...
  • Page 10: Measurement Mode

    3.4 Measurement Mode Converter will be in normal display after power on. There are 4 interfaces. Press to switch. 05/16/2017 10:20:20 05/16/2017 10:20:20 05/16/2017 10:20:20 05/16/2017 10:20:20  Heat FAULT Flow 0.000 m3/h 0.000 Kw Supply T 0.00 ˚С  +3.989m3 ...
  • Page 11: Auto Zero Trim

    3.7 Auto Zero Trim When pipe is full and fluid in pipe is stagnant, flowmeter may appear value due to bad grounding; you can adjust zero point to zero with auto zero trim. 05/16/2017 10:20:20 05/16/2017 10:20:20 05/16/2017 10:20:20 05/16/2017 10:20:20 ...
  • Page 12: Operation Flowchart Of Measurement Mode

    3.9 Operation Flowchart of Measurement Mode 05/16/2017 10:20:20 Flow 0.000 m3/h + 3.989m3 Basic  System  Power Query Basic System Power Language Password Password 0200 0030 PV Units New Password New Password Flow decimals Diff Temp Q max. (m3/h) Signal Total Units Power Opt.
  • Page 13 05/16/2017 10:20:20 Flow 0.000 m3/h + 3.989m3 Query  Test  Calibration Detail Test Calibration Query 0004 Zero Trim RT-Data Password Loop Trim  New password 4mA Trim Flow 20mA Trim  Loop Test Heat Empty Trim Extremum Pulse Test Tube Trim Full Trim Shut...
  • Page 14: Operation Flowchart Of Calibration Mode

    3.10 Operation Flowchart of Calibration Mode 05/16/2017 10:20:20 Flow 0.000 m3/h + 3.989m3 1111 Password Size (mm) New Password Sensor K Manual Adjust Zero K (me) Actual Zero (mv) Flow Set Coil Freq (Hz) Empty Freq (Hz) Coil Current Full Freq (Hz) Direction Indication Serial No.
  • Page 15: Operating Instructions

    3.11 Operating Instructions Parameter settings in Basic menu 05/16/2017 10:20:20 In any interface of measurement mode, press to enter “menu-selection”. Press to choose “Basic”. Flow 0.000 m3/h + 3.989m3 Main-menu Basic Choose ”Basic” and press to enter “parameter setting”.  System ...
  • Page 16 PV Units Basic Press to choose “Totalizer decimals” and press to enter. PV Accuracy Total Units Totalizer decimals Totalizer decimals Totalizer decimals Press to choose Flow totalizer decimals.  1 digit 2 digit 3 digit  Press to confirm and press to cancel.
  • Page 17 Total Decimal Basic Press to choose “T Damping(s)” and press to enter. Power Units Damping(s) T Damping(s) T Damping T Damping(s) Press to set T damping and press to move digit place. max: 99 Setting range: 0.0~ 99. You can set damping here when energy value has a large fluctuation, the ...
  • Page 18: Password Input

    Parameter Settings in System Menu 1. Signal Parameter settings Basic Password input Press to choose “System” and press to enter parameter settings in System menu.  System  Power Query Password Password input Press to move cursor and press to change value. ...
  • Page 19 2. Output parameter settings New Password System ” and Press to choose “Output press to enter. Signal Output Date (MM/DD/YY) Loop Out Output ” and press Pulse Out Press to choose “Loop Out to enter. Communication  Loop Output Press to enter corresponding output option of 4-20mA.
  • Page 20 Pulse Out Pulse – Flow Press to choose “Freq Max (Hz)” and press to enter. Freq Max(Hz) Liter/Pulse Pulse – Flow Freq max. (Hz) Freq Max(Hz) 2000.0 Press to change value. Max:5000.0 Frequency range: 100Hz~5000Hz This value corresponds to Max.Flow. Min:100.0 Press to confirm and press...
  • Page 21 Protocol Communication Baud rate Press to choose “ ” and press to enter. Baud rate Data Bit Parity Baud rate Baud rate Press to choose “Baud rate”. 9600 1200 , 2400 , 4800 , 9600 , 19200 , 38400 Default value:9600 Press to confirm and press to cancel.
  • Page 22: Date Settings

    Protocol Communication Press again to confirm and return to Output. Baudrate Parity Dev Address Loop Out Output Press again to confirm and return to System. Pulse Out Communication 3. Date settings System New Password Press to choose “Date” and press to enter.
  • Page 23 Parameter Settings in Power Menu Basic Main-menu  Press to choose “Power” and press to enter. System Press to confirm and press to cancel. Power Query Password Password Press to move cursor and press to change value.  Input password: 0030 0030 Press to confirm and press...
  • Page 24 New Password Power Press again to return to “main-menu”. Diff Temp Power Opt. Pressure Opt Basic Main-menu Press to choose “Query” and press to enter.  System Power Query...
  • Page 25 Parameter settings in Query menu Basic Main-menu  System Press to confirm and enter “Query”. Power Query Record Query-menu Press to choose object, press to confirm and enter “Query”. RT-Data ΣFlow Queryable item: real-time data, ΣFlow (total flow), ΣHeat (total heat), Extremum, Shut, ΣHeat Extremum Alarm.
  • Page 26 Record Query-menu Extremum Extremum” RT-Data ΣFlow Press to choose “ ” and press to enter “ ΣHeat Extremum Shut Alarm Extremum Statistics Day Black part is selected. Press to check daily/ monthly/ yearly statistics of FL+ and FL- FL+: (forward/ reverse flow rate), PwH and PwC (heat/ cold flow rate) andΔTH and ΔTC (heat/ Data not exist! cold temperature difference).
  • Page 27: Password Modification

    Parameter Settings in Test Menu System Main-menu  Press to choose “Query” and press to enter. Power Press to confirm and press to cancel. Query Test Password Password Press to change value.  Input password: 0004  0004 Press to confirm and press to cancel.
  • Page 28 Parameter settings in Calibration menu Power Main-menu  Press to choose “Calibration” and press to enter. Query  Test Calibration Zero Trim Calibration Press to choose “Zero Trim” and press to enter. Loop Trim Tube Trim Zero Trim Zero Trim Press to choose zero trim or not and press to confirm.
  • Page 29 Loop Trim  Press again to return to Calibration. 4mA Trim 20mA Trim Zero Trim Calibration Press to choose “Tube Trim” and press to enter. Loop Trim Tube Trim Tube Trim Empty Trim Full Trim Press to choose “Empty Trim” and press to enter.
  • Page 30: Operation Instruction Of Calibration Mode

    3.12 Operation Instruction of Calibration Mode In any interface of calibration mode, press to enter parameter settings. P22A1 0.000m3/h 05/16/2017 10:20:20 05/16/2017 10:20:20 05/16/2017 10:20:20  Heat 0.00 T 0.00 Flow 0.000 m3/h 0.000 Kw Supply T 0.00 ˚С  0.00m/s 614 +3.989m3 ...
  • Page 31 Sensor K Sensor K setting 01.00000 Press to change value, press to move cursor and input corresponding sensor factor. Max:99.99999 Setting range: 0.01000 ~ 99.99999 Min: 0.01000 Sensor factor should be set according to matching sensor. 01.00000 Press to confirm and press to cancel.
  • Page 32 Flow Set Coil Freq(Hz) Press to choose ”Indication” and press to enter. Coil Current Direction Indication Indication Indication Press to choose “Indication”. Forward 1. Forward 2. Reverse Default setting: Forward Press to confirm and press to cancel. Reverse Coil Freq(Hz) Flow Set Press again to return to Detail.
  • Page 33 Parameter Settings in Power Set Menu New Password Detail Press to choose “Power Set” and press to enter. Flow Set Power Set Manual Adjust Detail Medium Press to choose “Medium” and press to enter. Density (g/cm3) Heat C Temp Trim Medium Medium Press...
  • Page 34 Tin Revise Tin Revise Setting Press to change value, press to move cursor and input value. +0.0 Max: 3.0 Setting range: -3.000 ~ +3.000 Min: -3.0 Suppose Ti is 32℃, if set +1, Ti will be 31 C. If -1, Ti will be 33℃. +0.0 Press to confirm and press...
  • Page 35 Parameter Settings in Manual Adjust Menu Detail New Password Press to choose “Manual Adjust” and press to enter. Flow Set Power Set Manual Adjust Manual Adjust Actual Zero(mv) Press to choose “Actual Zero” and press to enter. Empty Freq(Hz) Full Freq(Hz) Actual Zero Setting Actual Zero(mV) Press...
  • Page 36: Clear Log

    Parameter Settings in Serial No. Menu Flow Set Detail Press to choose “Serial No.” and press to enter. Power Set Manual Adjust Serial No. Serial No. Serial No. Press to change value, press to set sensor number. 00000000 Press to confirm and press to cancel.
  • Page 37 Parameter Settings in Save Settings Menu Detail Serial No. Press to choose “Save Settings” and press to enter. If save settings, factory settings will Clear Log be overwritten. Clear Total Note: DO NOT set it randomly, or you can not retrieve the original parameters. Save Settings Save Settings ...
  • Page 38: Amc3200E App Software

    Read-write mode: read and write any parameters in converter (default mode). If enter the wrong Bluetooth password, you will be prompted to enter in read-only mode. Terminal mode and client mode are not accessible for the moment. Read WriteMode TerminalMode ClientMode ALIA TECHNOLOGY LLC Software version 1 2 4 : . .
  • Page 39 ⑤ If parameters are to be changed, please choose read-write mode and then choose login. Cell phone will search device automatically. See pictures below. 16 22 16 22 Searching Searching MEYERREADING MEYERREADING SEARCHING BONDED DEVICE SEARCHING BONDED DEVICE DATA DATA ALIA0000000003 Searching ALIA0000000003...
  • Page 40 ⑦ After entering system, you will see Bluetooth name in the middle of upper interface, mode in upper right interface, flowmeter version, password and nickname in bottom interface. Set Nickname: you can set a nickname for AMC3200E, and this won't change the Bluetooth name. The nickname will be displayed only after Bluetooth name.
  • Page 41 ⑨ Once Bluetooth password is conformed, system will be as below: 16 22 . . . PAUSE Real time data REAL-TIME METER TOTALIZER MORE READ DATA Flow rate 0 0 m3 h Fwd totalizer 1035 L Rev totalizer 701 L Net Totalizer 334 L Energy rate...
  • Page 42: Power Unit

    16 22 16 22 ..PAUSE PAUSE Real time data Real time data REAL-TIME REAL-TIME METER METER TOTALIZER MORE TOTALIZER MORE READ DATA READ DATA Flow rate Output current 0 0 m3 h Inlet water temperature Output frequency value 0 0 c 0 0 Hz...
  • Page 43 A. For example, if power unit KW is changed to MW: . . . Parameter REFRESH ENERGY1 ENERGY2 BASICS INTERFACE Power unit Current Output selection Energy meter Frequency Output selection Energy meter Energy selection AUTO Pressure selection 0.6 Mpa Medium selection Other Flow Decimals Re altime...
  • Page 44 C. After pressing "confirm", screen will be shown as below. The unit before arrow is current unit while after is the changed MW. Click "set" to finish settings. 16 22 16 22 ..Parameter Parameter REFRESH REFRESH ENERGY1...
  • Page 45 Fourth function: Virtual keys. 3 Virtual keys and 3 keys on converter have the same functionality. 16 22 . . . Virtual Key Left Down Right Support short press and long press . P a ra m e te r Re a ltime F lo wme te r Op e ra tio n...
  • Page 46 Sixth function: Parameter backup. Recreate a file and rename it. Press "Function" in upper part to choose: export measurement data, export config data, export measurement config data and import setting parameters. 16 22 16 22 ..Parameter Parameter FUNCTION...
  • Page 47: Common Alarm Code Indication

    Zero Zero alarm good grouding. PT_H Alarm Check PT-H connect or not PT_L Alarm Check PT-L connect or not Normal 6. APP Download Link Link 1: scan QR code to download. Link 2: enter the following link to download: http://www.alia-inc.net/download/alia/amc3200/alia_AMC3200.apk...
  • Page 48 Quality we care ! ALIAMAG ALIAPANEL ALIASONIC ALIADP ALIAPT ALIAVA ALIAVTX Tel: +1-213-533-4139 Fax: +1-213-223-2317 URL: www alia inc com Email: alia @alia inc com 633 W. 5th Street, 26th Floor, Los Angeles, CA 90071, USA...