AVF250 Operation Manual OP250.1.1.7Z.R8ENG A L I A V A ® Metal Tube Variable Area Flowmeter Model AVF250 Series Operation Manual URL : http://www.alia-inc.com ALIA TECHNOLOGY LLC 633 W. 5th Street, 26th Floor, Los Angeles, CA 90071, USA e-mail : alia@alia-inc.com ■...
A. Check the overwrap first. If any damages occur, please contact ALIA Customer Service Center at once. B. Confirm whether the instrument has been damaged or its spares are lost. C. It’s recommended that you read the manual thoroughly, once you have any quires, please email or fax to ALIA technical department.
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 3. Features 5-digit flow rate&8-digit totalizer Local indication without auxiliary power 4-20 mA and scale pulse output Low pressure loss in gas and steam application Damper for gas/steam application Constant overall length ...
AVF250 Operation Manual OP250.1.1.7Z.R8ENG Dimensions and Pressure Loss m3 Hr AVF250 m3 Hr m3 Hr AVF250 Water AVF250 1000Kg/m3 eg.C 3 5Kg /cm2 Wa te r 00Kg/m3 35 Deg 3 5 Kg /cm2 Wat er 1000 Kg/ m3 35 Deg. 3 5Kg /cm2 170 5mm...
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 6.2 Straight Pipe Requirement Proper place is vital to flowmeter installation; otherwise flowmeter’s accuracy will be influenced or even get damaged. m3 Hr AVF250 m3 Hr Wat er 10 00 Kg/m 3 35 De g. 3 5Kg /c m2 AVF250 Water...
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 7. Size Calculation 7.1 Method 1) Qs=K×Q Among: Qs-flow rate of calibrated medium (water or air) under standard condition (20 C,1atm) Q-flow rate of user’s medium K-correction factor 2) Based on the Qs, check flow table (page 2 in datasheet) to choose float and measuring tube’s size. The flow data in table are both under standard condition of air or water.
Page 9
AVF250 Operation Manual OP250.1.1.7Z.R8ENG For steam medium: If AVF250 is to measure saturated steam, formula will be: 29 56 ρ Among: Qs: water’s flow rate ρ: steam’s density under operation condition (kg/m3) M: steam’s flow rate (kg/h) Choose flowmeter from Qs (water’s flow rate) after calculating as formula above. This formula is derived by taking float’s density as 7900kg/m3.
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 8. Installation Considerations and Maintenance (1) Before installation, please sweep the process pipe first so as to avoid any magnetic particles remained inside pipe as it will influence flowmeter’s performance or cause damage to flowmeter due to adhesion onto flowmeter. If it’s unavoidable, please install filter before flowmeter’s inlet.
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 9. Wiring Diagram This flowmeter adopts 4-20mA output signal of 2-wire to other external equipment of device with power supply 12~36 VDC. The maximum load resistance to output circuit is 600Ω (including resistance of cable). Generally, it’s recommended to use PVC shielded cable that resists 600V. If the environment is easily to be interfered by noises of equipments, please use 2-core shielding cable (RWP2×0.5mm) with one of its terminals connecting to grounding screw of flowmeter’s housing and another to the ground.
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 10. Keypad Key Name Button Basic Function Setting Function Enter parameter setting Save data Data setting interface Increase number & Save data Indicate data setting interface Digit shift & Decimal point shift Display Interface 10.1 TOTRL 000.00 000.00 000.00...
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 11. Operational Flowchart m3/h ∑ 02000 02008 3018 5018 SET T CUT OFF ZERO ADCAL DAMP BAUD SPAN SU-H AD 00 SET TOT SH-d FILTER AD 10 ADCAL CLR TOT SU-L Density DAO 4 AD 20 SL-d Density DAF 20...
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 12. Operation and Settings 12.1 General Settings 4-20mA output can only be set by keys. m3/h ∑ 2000 Input Password LOCK Press to change value and press to next setting. Password: 2000 SET TOT Preset Totalizer Preset totalizer as required.
Page 18
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 00000 The Corresponding Value of 4mA Output PRO-LO Press to change value. This value corresponds to 4mA output. Press to next setting. 01.500 The Corresponding Value of 20mA Output PRO-HI Press to change value. This value corresponds to 20mA output. Press to next setting.
Page 19
AVF250 Operation Manual OP250.1.1.7Z.R8ENG Alarm Low Limit Setting SU - L Press to set lower limit value (range: 0%~110%), and then press to next setting. When setting value is 0, the low limit terminal is the pulse output terminal. Hysteresis Value of Alarm Low Limit Press to change hysteresis value of alarm low limit (setting range: 0%~30%), and then press SL –...
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 12.2 Calibration Parameter Setting m3/h ∑ 2008 Input Password LOCK Press to change value and press to next setting. Password: 2008. 00000 Zero Point Adjustment ZERO Default value is 0. It’s used to calibrate lower range value. Please do not modify randomly. Press to change value and press to next setting.
Page 21
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 22733 50% Flowrate Calibration (Please do NOT adjust this parameter randomly) Adjust pointer to the location of 50%, wait until value is stable and then press to next parameter. Press to next parameter directly. Press to switch point. 25176 60% Flowrate Calibration (Please do NOT adjust this parameter randomly) Adjust pointer to the location of 60%, wait until value is stable and then press...
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 12.3 Other Parameter Settings m3/h ∑ 3018 Input Password LOCK Press to change value and press to next setting. Password: 3018 Low Flow Cutoff (Setting range: 0 –30) Press to change value and press to next setting. 8 mean 8%.
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 12.4 Communication Address and Size Setting m3/h ∑ 5018 Input Password LOCK Press to change value and press to next setting. Password: 5018 Device Address Press to change value. HART address(00~15), Modbus address(1~99) And press to next setting. 09600 Baud Rate Press...
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 13. Troubleshooting 1. Shake (1) Slight shake: it’s usually caused by medium fluctuation. By adding a buffer device or extending straight pipe will solve this problem. (2) Medium shake: it’s usually caused by medium flowing state. For gas, it’s caused by unstable operating pressure. By using a voltage-stabilizing/fluid-stabilizing device or increasing damping will solve this problem.
Page 25
AVF250 Operation Manual OP250.1.1.7Z.R8ENG 9. Wrong totalizer pulse output (1)Check whether the alarm of totalizer pulse output is correct. (2)Check whether the external power is between 5 and 24VDC and connect 1-5K resistance in series. (3) Circuit board is malfunctioned. It needs to be replaced.
AVF250 Operation Manual OP250.1.1.7Z.R8ENG Appendix Density of Common Gas and Liquid Pressure: 0.1013Mpa Temperature: 20℃ Unit: kg/m3 Density of Common Gas Name Density Name Density Name Density Name Density 1.2041 Propane 1.8332 Benzene 3.2476 Fluorine 1.5798 Nitrogen 1.1646 n-Butane 2.4163 Carbon monoxide 1.1650 Chlorine...