WHALETEQ MECG 2.0 User Manual

Multichannel ecg test system
Hide thumbs Also See for MECG 2.0:
Table of Contents

Advertisement

Quick Links

MECG 2.0 | User Manual
WHALETEQ
Multichannel ECG Test System
(MECG 2.0)
User Manual
Version 2024-10-15
PC Software Version 2.0.12.1
0

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MECG 2.0 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for WHALETEQ MECG 2.0

  • Page 1 MECG 2.0 | User Manual WHALETEQ Multichannel ECG Test System (MECG 2.0) User Manual Version 2024-10-15 PC Software Version 2.0.12.1...
  • Page 2 This document could contain technical inaccuracies or typographical errors. Changes are periodically made to the information herein; these changes will be incorporated in future revisions of this document. WhaleTeq Co. LTD. is under no obligation to notify any person of the changes.
  • Page 3: Table Of Contents

    3.1 Software Installation ..............10 3.1.1 System Requirements ..................10 3.1.2 MECG 2.0 Software ..................11 3.1.3 First Time Using WhaleTeq Product - USB Driver Installation ......11 3.1.4 First Time using WhaleTeq Product – Microsoft .Net Framework 4.0 Installation ......................... 12 3.2 Set up ....................
  • Page 4 MECG 2.0 | User Manual 7.7 Graph View ..................33 Troubleshooting ................33 8.1 Long-term tests (continuous) streaming........34 Caution ....................35 10 Ordering Information ............... 35 10.1 Standard Package ................35 10.2 Optional Software, Accessories, and Services ....... 36 11 Version Information .................
  • Page 5 Table 4: Supported Command Scripts ..........30 Table 5: Demo for Command Script ..........32 Table 6: Trouble Shooting ..............33 Table 7: MECG 2.0 Standard Package ..........35 Table 8: Optional Software Add-on Pack......... 36 Table 9: Optional Accessory ............36 Table 10: Optional Calibration Service and Warranty Extension ..
  • Page 6 MECG 2.0 | User Manual List of Figure Figure 1: System Description ............. 6 Figure 2: Set up a noise-free test environment ....... 13 Figure 3: Update Application/ Firmware ......... 14 Figure 4: Firmware Update Window ..........15 Figure 5: ‘’About’’ Button ..............15 Figure 6: “Update F/W”...
  • Page 7: Introduction

    MECG 2.0 | User Manual Introduction The WhaleTeq Multichannel ECG Test System (MECG 2.0) provides a full 12 lead waveforms for testing diagnostic, ambulatory or monitoring ECG, for testing to IEC particular standards. Version 1.0 was designed according to the circuit in IEC 60601-2-51, modified by using precision, low offset op-amps (<0.5µV) and 0.1%...
  • Page 8: System Description

    (MECG 2.0) (DUT) Figure 1: System Description The MECG 2.0 software currently allows the user to select the waveform from one of three sources: CTS database (as per IEC 60601-2-25:2011) This includes the ANE, CAL and biological waveforms as detailed in the standard.
  • Page 9 MECG 2.0 | User Manual reference only, but could be used for some “single channel” performance tests. Biological waveforms from the “PhysioNet” website This allows the loading of files based on Format 16 or 212 format (other formats may be considered on request).
  • Page 10: Application Revision Update

    Equipment with serial numbers earlier than 2012-008 will continue to operate with the software as before. As MECG 2.0 is equipped with a 12-bit DAC, it is must to use the DAC compensation file. For users purchasing MECG 2.0 in 2020 (serial number as ME2001-XXXXXX), if the testing requires, please refer to 3.5.2...
  • Page 11 ±5μV for ±25µV. values under 100μV. 2.4µV for MECG Output voltage MECG 2.0 uses 12-bit DAC and MECG 2.0 and 0.15µV for resolution 2.0 (2020) uses 16-bit DAC. MECG 2.0 (2020) The output voltage on most of the Output voltage ±5mV...
  • Page 12: Start To Use Mecg 2.0

    Software Installation 3.1.1 System Requirements Users need a PC to control the multichannel ECG system (MECG 2.0) through its USB port. The PC should meet the following requirements Windows PC (XP or later, suggest to use the genuine version) Microsoft .NET 4.0 or higher...
  • Page 13: Mecg 2.0 Software

    ․ Download the MECG 2.0 software from WhaleTeq’s official website. ․ Unzip the file and double click the installation file ․ When the installation process is completed, click the MECG 2.0 icon on the desktop to execute or select MECG 2.0 software in the start menu 3.1.3 First Time Using WhaleTeq Product - USB Driver Installation If Windows device manager can’t recognize WhaleTeq product, please...
  • Page 14: First Time Using Whaleteq Product - Microsoft .Net Framework 4.0

    Note: if the socket is changed, it may take the PC a short amount of time to recognize and connect to the system. Run the MECG 2.0 software. If the USB module is not recognized, a message will be displayed. In this case, repeat the process, ensuring sufficient time for the PC to recognize the USB module prior to starting the MECG 2.0 software.
  • Page 15: Environment, Noise Reduction

    A noise free environment is necessary for testing ECG equipment. This can be achieved relatively easily by using a metal bench or metal sheet underneath the ECG device under test, the MECG 2.0, and also connecting together the ground as shown:...
  • Page 16: Live Update Application/ Firmware

    How to check whether your MECG 2.0 supports firmware update or not? Answer: Connect the MECG 2.0 to a PC and go to the ‘’About’’ dialog to check if ‘’F/W Version’’ and ‘’H/W Version’’ show. Please see ‘’3.5.1 How to...
  • Page 17: How To Update Firmware

    MECG 2.0 | User Manual Not Support Firmware Update Support Firmware Update Figure 4: Firmware Update Window Caution: There are risks of losing data if improper options are performed during the Firmware Update period. 3.5.1 How to Update Firmware If the device supports “Firmware Update” feature, here are the steps to update the device firmware: Step 1.
  • Page 18: Figure 6: "Update F/W" Window

    MECG 2.0 | User Manual Step 2. Click the “Update F/W” button and select the firmware image file Figure 6: “Update F/W” Window Step 3. The application will show an information dialog. After pressing “OK”, the updating operation cannot be terminated.
  • Page 19: How To Update Dac Compensation File

    To update the file, please follow the steps below: 1. Copy the file to the “C:\Whaleteq\” folder. 2. When the MECG 2.0 software is running, it reads the serial number of the connected MECG 2.0, and looks for the corresponding DAC compensation file.
  • Page 20: Software Development Kit (Sdk)

    10µV, so that DAC compensation is irrelevant. DAC compensation data will be kept by WhaleTeq, so that any time if the data is lost (e.g.: transferring to a new PC), a copy can be requested.
  • Page 21: Self-Calibration

    ±1%. Since no minimum limit is stated (and it is impossible to apply ±1% for very small voltages), WhaleTeq has applied a rule of 20% of the EUT limit. This is effectively ±5µV for up to 500µV, ±8µV from 500 to 800µV and ±1% for values above 800µV.
  • Page 22: Figure 10: Self-Calibration Settings

    “Play” as the figure below: Figure 10: Self-calibration settings 2. Connect a multimeter to the RA and LA jacks of the MECG 2.0 to measure the DC mV. It should be in ±1% of 5 mV as the figure below. The MECG 2.0 output amplitude gives alternating 5s (0.1 Hz square wave)
  • Page 23: Figure 11: Connection To A Multimeter

    MECG 2.0 | User Manual Figure 11: Connection to a multimeter 3. Connect the multimeter to the RA and V1 jacks of the MECG 2.0 to measure the DC mV. It should be in ± 1% of 8.33 mV [V1 = V1 + (RA+LA+LL)/3 = 5 + (0+5+5)/3 = 8.33 mV] as the figure below.
  • Page 24 4. Repeat step 3 to connect the multimeter to the RA and V2, and then the V3 to V6 of the MECG 2.0 to measure the DC mV. All the V2 to V6 should be in ±1% of 8.33 mV.
  • Page 25: Standalone Behavior

    MECG 2.0 | User Manual Standalone Behavior To support subclause 202.6.2.1.10 and 202.6.2.6 in IEC 60601-2-25. The MECG 2.0 (with firmware version 2.3 or above) is kept outputting the CAL20110 signal after power-on, and stopped outputting the signal after connecting to MECG 2.0 software.
  • Page 26: Cts/Cse Database

    MECG 2.0 | User Manual 04- Output Waveform: Start and stop the waveform output 05- Other Waveforms: Select other waveforms such as Sine, triangle (press “Load” to load them into memory and screen) 06- Command Replay: Use “Command” to auto display waveform...
  • Page 27: Load Ecg Source From File

    Load ECG Source from File This function is set up to work with popular waveforms from the PhysioNet website and other databases. As the MECG 2.0 software can work directly with the website, the user does not need any knowledge about the PhysioNet site, file formats and the like.
  • Page 28: Whaleteq Format

    Do not change the keywords, including “start”, “Lead I”, “Lead II”, and “V1” to “V6”, or the data file cannot be correctly loaded. 7.3.2 EDF Format The MECG 2.0 software supports loading and playing EDF file format. To enable this function, you must purchase an additional authorization code.
  • Page 29: Display Parameters / Output Waveforms

    PhysioNet Non-Invasive Fetal ECG Database (nifecgdb) can be downloaded and played directly in the MECG 2.0 software. Figure 15: ‘’ECG File Load’’ Window Connect the MECG 2.0 to a PC and click ‘’EDF’’, then the ‘’License Activation’’ pop-up window shows. Figure 16: ‘’License Activation’’ Window To purchase this authorization code, please contact WhaleTeq and provide the ID of your MECG 2.0.
  • Page 30 MECG 2.0 | User Manual 3 – Leads I, II, III 1 – Lead I only Note: Leads III, aVR, aVL, aVF are derived from Lead I, II and in general do not use the data if supplied. Waveforms can be started or stopped at anytime.
  • Page 31: Other Waveforms

    MECG 2.0 | User Manual Other Waveforms Basic waveforms are selectable as below: Figure 18: Other Waveforms The following other functions can be selected: Sine wave with adjustable amplitude and frequency Triangle wave with adjustable amplitude and frequency Square wave with adjustable...
  • Page 32: Command Replay

    The command script could be automatically generated while the user operates the MECG 2.0. This means the user could operate the test procedure for the one time, then edit, save and replay the script. Supported Command Scripts:...
  • Page 33 MECG 2.0 | User Manual Command Parameter Description Example LoadLocal_ Waveform Load the AHA file (*.ecg) D:\whalete Waveform waveform LoadLocal_AHA path on file path (*.ecg) file q\Whaleteq your local on local PC. MECG\1001 .ecg LoadLocal_ Waveform Load the AHA AHA_TXT file (*.txt)
  • Page 34: Table 5: Demo For Command Script

    MECG 2.0 | User Manual Command Parameter Description Example before replay for continuing specified next seconds then command. continue. Invoke an executable to run. executabl Process Command Executable e file NotePad.ex Process replay would PathArg path and continue e result.txt...
  • Page 35: Graph View

    Recognition of USB devices needs to be done in order: USB module (test unit) not 1) Close the MECG 2.0 software if it’s opened recognized (USB 2) Disconnect the USB module for ~2s driver is installed 3) Reconnect the USB module...
  • Page 36: Long-Term Tests (Continuous) Streaming

    While modern PCs give the appearance of real time, the core structure does not guarantee interruption free streaming of serial data to a peripheral. Previous versions of the MECG 2.0 have incorporated a feature to detect interruptions, stop streaming and inform the user.
  • Page 37: Caution

    ➢ WhaleTeq does not recommend to connect test equipment with DUT to conduct Electrostatic Discharge (ESD) test. This may cause unexpected damages to test equipment.
  • Page 38: 10.2 Optional Software, Accessories, And Services

    USB isolator for reducing the power noise from PC. Recommended to use with SECG 100-OT00001 4.0, MECG 2.0, HRS200, HRS100+, SEEG 100 and SEEG 100E. ․ Optional Calibration Service and Warranty Extension Table 10: Optional Calibration Service and Warranty Extension Part No.
  • Page 39: 11 Version Information

    MECG 2.0 | User Manual 11 Version Information Table 11: Version Information Version Modified Content Issue Date Chap 4 Software Development Kit (SDK) 2020-12-31 Chap 9 Caution 2020-12-31 Chap 10 Ordering Information Chap 11 Version information 2021-06-28 2021-06-28 Chap 9 Caution 3.5.2 How to Update DAC Compensation...

This manual is also suitable for:

100-ec00101

Table of Contents