Keep new and used batteries away from children. See the User's Manual for additional information.
NOTE
* The target range is represented by the green region of the target range indicator. The arrow will flash if your test result falls above or below this range.
*See the User's Manual for more information on the target range.
Note
Always follow your healthcare professional's instructions. For example, if your healthcare professional has advised you to immediately treat a low blood glucose result (such as by eating something), do that first.
The error code and Err will alternate on the display for all coded error messages (E-1 through E 14).
If any of the error messages below appear on the display at any time, refer immediately to the Error Messages section in the chapter Meter Maintenance and Troubleshooting of the User's Manual.
Error Code | Description |
E-1 | The test strip may be damaged, not properly inserted, or was previously used. Remove and reinsert the test strip or replace it if damaged or previously used. |
E-3 | A meter or test strip error has occurred. This error message could appear if the cap on the test strip container was not closed tightly. The test strips may have been damaged due to improper storage or handling. Never make therapy decisions based on an error message. Repeat the blood glucose test. If a second E-3 error message appears, perform a control test with the control solution and a new test strip. See the Performing a Control Test section in the chapter Control Tests of the User's Manual. If you continue to receive an E-3 error message, use an alternate method for testing your blood glucose, such as a back-up meter and test strip. If the alternate method gives an extremely high blood glucose result, or if an alternate method is not available, contact your healthcare professional immediately. In rare cases, an E-3 error message may indicate that your blood glucose is extremely high and above the system's measuring range. See the Unusual Blood Glucose Results section in the chapter Blood Glucose Tests of the User's Manual for other possible causes of the error message. |
E-4 | Not enough blood or control solution was drawn into the test strip for measurement or was applied after the test had started. Discard the test strip and repeat the blood glucose or control test. |
E-6 | Blood or control solution was applied to the test strip before the flashing drop symbol appeared on the display. Discard the test strip and repeat the blood glucose or control test. |
E-7 | An electronic error occurred. Remove the batteries, press and hold either meter button for at least 2 seconds, and reinsert the batteries. Perform a blood glucose or control test. |
E-8 | The temperature is above or below the proper range for the system. Refer to the test strip package insert for system operating conditions. Move to an area with the appropriate conditions and repeat the blood glucose or control test. Do not artificially heat or cool the meter. |
E-9 | The batteries may be out of power. Turn the meter back on. If you are in a cold environment, move to a location with a more moderate temperature and retest. If the message continues to appear after several attempts, replace the batteries. If the message reappears after the batteries have been replaced, remove the batteries, press and hold either meter button for at least 2 seconds, then reinsert the batteries. |
E 10 | The time and date setting may be incorrect. A flashing wireless symbol appears while the meter attempts to synchronize the time and date with the paired device. When the wireless symbol no longer appears, press the Up button to turn the meter off. Press the Up button a second time until the flashing test strip symbol appears. If time and date sync was not successful the meter prompts you to set them the next time you turn it on. See the Setting the Time and Date section in the chapter Your New System of the User's Manual. |
E 11 | The test strip may be damaged. Repeat the blood glucose or control test with a new test strip. |
E 12 | Your blood sample may contain a high level of ascorbate. Contact your healthcare professional. |
E 13 | Fluid or foreign material may be present in the test strip slot. Remove and reinsert the test strip or repeat the blood glucose or control test with a new test strip. If the error persists, contact Customer Service at 1-800-819-1097. |
E 14 | An electronic error has occurred. Contact Customer Service at 1-800-819-1097. |
Note
Contact Customer Service at 1-800-819-1097 if the error message continues.
For questions, contact Customer Service at 1-800-819-1097. Phone support is available from 8:00 AM - 8:00 PM ET, Monday - Friday.
For questions at all other times, or in case of emergency, contact your healthcare professional or emergency medical response.