Proline Promass O 100 PROFINET
Endress+Hauser
Base accuracy
Design fundamentals → 145
Mass flow and volume flow (liquids)
• ±0.05 % o.r. (optional for mass flow: PremiumCal; order code for "Calibration flow",
option D)
• ±0.10 % o.r. (standard)
Mass flow (gases)
±0.35 % o.r.
Density (liquids)
Under reference
Standard density
conditions
[g/cm³]
±0.0005
1)
Valid range for special density calibration: 0 to 2 g/cm³, +5 to +80 °C (+41 to +176 °F)
2)
order code for "Application package", option EE "Special density" (for nominal diameter ≤ 100 DN)
3)
Valid range for extended density calibration: 0 to 2 g/cm³, +20 to +60 °C (+68 to +140 °F)
4)
order code for "Application package", option E1 "Extended density"
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
Zero point stability
DN
[mm]
80
100
150
250
Flow values
Flow values as turndown parameters depending on nominal diameter.
SI units
DN
1:1
[mm]
[kg/h]
80
180 000
100
350 000
150
800 000
Wide-range
calibration
Density
specification
[g/cm³]
[g/cm³]
±0.0005
±0.001
[in]
3
4
6
10
1:10
1:20
[kg/h]
[kg/h]
18 000
9 000
35 000
17 500
80 000
40 000
Technical data
Extended density
calibration
1) 2)
[g/cm³]
±0.0005
Zero point stability
[kg/h]
[lb/min]
9
0.330
14
0.514
32
1.17
88
3.23
1:50
1:100
[kg/h]
[kg/h]
3 600
1 800
7 000
3 500
16 000
8 000
3) 4)
1:500
[kg/h]
360
700
1 600
143
Need help?
Do you have a question about the Proline Promass O 100 and is the answer not in the manual?