Page 1
USER R MA ANUA LG-u uniO N 42 and L -uni EN 11 1 0...
Page 2
5. In the event of failure, LLG shall be under no liability for any injury, or any loss or damage as the result of the failure other than the guarantee conditions.
Page 3
We would like to take this opportunity to thank you for preferring this LLG Labware product. Please read the operating instructions carefully and keep them handy for future reference. Please detain the packing material until you see that the unit is in good condition and it is operating properly.
Page 5
USE AND FUNCTION 1.1. Both LLG-uniOVEN 42 and LLG-uniOVEN 110 are designed to heat or dry samples in laboratories. They offer excellent drying and heating conditions. They maintain temperatures between 50 °C and 250 °C and keep the temperature stable within the given tolerances.
Page 6
OPTIONAL ACCESSORIES 2.2. 6.263 671 Shelf for LLG-uniOVEN 42 6.263 672 2 x Shelf carrier for LLG-uniOVEN 42 6.263 676 Shelf for LLG-uniOVEN 110 6.263 677 2 x Shelf carrier for LLG-uniOVEN 110 Note: For each shelf two shelf carriers are necessary.
Page 7
GENE ERAL PRE ESENTAT TION 2. .3. Figure 1: In ner side of LLG-uniOV VEN 42/ LLG G-uniOVEN N 110 1. Contro l panel 8. S helf carrier holder 2. Safety thermostat 9. S helf carrier 3. On / Of ff switch 10.
Page 8
3. INSTALLATION PROCEDURE LIFTING AND TRANSPORT 3.1. Due to the heavy weight of the LLG-uniOVENs, all lifting and transport must be carried out using proper handling equipment. They must be supported from underneath and never turned over. UNPACKING 3.2. Remove the packing cardboard box and the second nylon packing around the oven.
Page 9
The instruments, which will be dried up and heated, should not be combustible and explosive, especially please kindly check that IF MENTIONED WARNINGS ARE NOT CONSIDERED, LLG WILL NOT BE RESPONSIBLE FROM THEIR RESULTS.
Page 10
4. O OPERAT TING PR RINCIPL SWITC CHING O Push on n/off switch See tha at the micro processor c control syst em activate Learn th he functıon of the cont trol and disp play panel ( Section 4.2 CONT TROL PAN...
Page 11
Temperature Display This display shows chamber temperature during “stand-by” and during the operation (thermometer position),during program preparation, the temperature setting values or the alarm setting values, Failure codes, EoF warning when power is interrupted, the set temperature values and alarm values. Time Display This display shows the elapsed time during the operation and the set value during programming.
Page 12
PREP ARATION N OF USE ER SETTI he device h as a passw word protect ted menu. T The passwo ord is set to “000” when n the evice first st tarts. Follow w the below w steps to ch hange the p password, u pdate curre...
Page 13
4.3.1. oP Operator Menu Parameters 1: Recording Period: This time is recording period of temperature and error information. 2: Lid Alarm Time: Not used for all LLG-uniOven models. 3: Lid Alarm Range: Not used for all LLG-uniOven models. 4: Timer Set Band: When the read temperature reaches the "Set Temperature - TIMER SET BAND"...
Page 14
PROG GRAMMIN NG SUMM MARY ou can set t the steam s sterilizer on stand-by. F Follow the b below steps to set and save e values. Push t he encoder r button By pus shing the en ncoder butto on select SE ET menu.
Page 15
See th e paramete er flashing o on the time display. By turn ning the enc coder button n set opera ting time value ( (01 minute t to 99 hours s 54 minutes s or Hold). Push t he encoder r button and d save set v value.
Page 16
5.1. The oven does not require any periodical maintenance which is carried out by the operator. Please contact to LLG partner for an authorised service or maintenance. CLEANING 5.2. After unplugging the equipment and the equipment is at the room temperature, wipe down the chamber to remove any undesirable effects of the operation, for example spillage.
Page 17
ASE OF A ANY ERRO OR, THE P PROGRAM İS STOP PPED AUTO OMATİCAL LLY AND İMMED DİATELY. PLEAS SE CONTA ACT TO AN AUTHORIZ ZED LLG P PARTNER T TO REQUE EST TECHN NICAL HELP IF AN ERR ROR OCCU...
Page 18
8. EL LECTRIC CAL CIR RCUIT D DIAGRA LLG-uniO OVEN 42 a and LLG-u uniOVEN N 110 Elec c trical circ rcuit diag ram...
Page 19
9. WA ARNING G LABE LLG-uniOVE VEN 110 FUSE (2×1 10A) ROUNDED PL G-uniOVEN 4 FUSE FUSE ( 2×4A) (2×3A)