RF Statement
EN
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the
FR
EMC information provided in the accompanying documents.
DE
Portable and mobile RF communications equipment can affect medical electrical equipment.
SV
NL
Guidance and manufacturer's declaration-electromagnetic emissions
FI
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the The device should ensure that it is used in such an
DA
environment.
IT
ES
Emissions Test
Compliance
Electromagnetic environment – guidance
CS
CE emissions CISPR11
Group 1, Class B
The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause
PL
any interference in nearby electronic equipment.
PT
RE emissions CISPR11
Group 1, Class B
RO
HU
Harmonic emissions
Not applicable
The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public
IEC 61000-3-2
low-voltage power supply network that supplies buildings used for domestic purposes.
SK
Voltage fluctuations/Flicker
Not applicable
emissions
IEC 61000-3-3
Safety and performance
Declaration: electromagnetic emissions and immunity for equipment and systems that are not life-supporting and are specified for use
only in a shielded location.
Declaration – electromagnetic immunity
Withings Body Scan is intended for use in the electromagnetic environment specified below. The customer or the user of the Withings Body Scan should ensure that it is used in
such an environment.
Immunity test
IEC 60601 Test Level
Electrostatic discharge (ESD)
Contact: ±8 kV
IEC 61000-4-2
Air: ±2 kV, ±4 kV, ±8 kV, ±15 kV
Radiated RF EM fields
10 V/m
IEC 61000-4-3
80 wMHz to 2.7 GHz+
Proximity fields from RF wireless
communications equipment
Electrical Fast Transient/Burst
N/A
IEC 61000-4-4:2012 ED 3.0
Surge IEC 61000-4-5
N/A
Conducted RF IEC 61000-4-6
3 Vrms
24 MHz to 80MHz and 6 Vrms in ISM
and amateur radio bands between
24 Mhz and 80 MHz
Power frequency (50/ 60Hz)
30 A/m
magnetic field IEC 61000-4-8
50 Hz and 60 Hz
Voltage dips, short interruptions
N/A
and voltage variations on power
supply input lines IEC 61000-4-11
Common RF emitters
As defined per AIM Standard 7351731,
(RFID, Wireless Power Transfer,
and IEC 61000-4-39
5G Cellular)
Electromagnetic environment – guidance
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material
the relative humidity should be at least 30%.
Portable and mobile RF communications equipment should be used no closer to any part of the
equipment or system including cables than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter. Interference may occur in the vicinity
of equipment marked with the following symbol:
.
-
-
-
The power frequency magnetic field should be at levels characteristic of a typical location in
a typical commercial or hospital environment.
The main power quality should be of the kind used in a typical commercial or hospital environment.
If the user of the equipment or system requires continued operation during power main
interruptions, it is recommended that the equipment or system be powered from an uninterruptible
power supply or a battery.
-
15
EN
FR
DE
SV
NL
FI
DA
IT
ES
CS
PL
PT
RO
HU
SK

Advertisement

loading
Need help?

Need help?

Do you have a question about the Scan Monitor 2.0 and is the answer not in the manual?

Subscribe to Our Youtube Channel

Related Products for Withings Scan Monitor 2.0

This manual is also suitable for:

Wbssm2