probfit BA 183C/07/en/11.99 Nr. 50090675 CPA 450 Retractable Assembly for pH/Redox Measurement Operating Instructions Endress Hauser The Power of Know How...
If you have any questions, consult your has been settled. supplier or the Endress+Hauser sales agency in your area (see back cover of Check that the delivery is complete and these operating instructions for addresses).
CPA 450 1 General information PA450E01.CHP Product structure You can identify the assembly variant by the ENDRESS+HAUSER order code on the nameplate. CPA 450 Order code: CPA450-1F110 Serial no.: Spec.: PN=4bar T=-15...130°C Fig. 1.1 Nameplate of CPA 450 Retractable assembly CPA 450...
2 Safety probfit CPA 450 Safety Intended use The manually operated retractable assembly It is the operator’s responsibility to assure that CPA 450 is intended for installation of pH/ the following safety regulations are observed: redox sensors in tanks and pipelines.
CPA 450 2 Safety PA450E02.CHP Safety features Safety devices The assembly is protected against external influences and damage by means of the following design measures: • media-resistant material • stopcock Notes for installation in pressurised systems Warning: • The maximum operating pressure of the assembly must not be exceeded.
CPA 450 Installation Measuring system The complete measuring system consists of: Optional: • Retractable assembly Probfit CPA 450 • Junction box VBA for measuring cable • 1 gel-filled pH combination electrode, extension length 120 mm, with or without integrated •...
3 Installation probfit CPA 450 Assembly installation Handle Cable gland Pg 11 Sleeve Cable guide Immersion tube Setting collar Pan head screw M5x16 DIN 912 Union nut Clamping ring Flanged sleeve G ¼ rinse connection Lock ring Inner pipe Adapter...
Page 11
CPA 450 3 Installation PA450E03.CHP Instructions for assembly, electrode and measuring cable installation Loosen union nut ½ revolution. Extend immersion tube completely. Close ball valve. Unscrew lock ring with AF 55 or hook wrench. Remove assembly. Loosen cable gland.
Page 12
3 Installation probfit CPA 450 Mounting instructions (continued) Plug electrode head onto cable connector and Plug protection cage onto electrode and hand-tighten. Push electrode head into inner pipe hand-tighten. all the way to stop. Insert tube assembly in ball valve.
CPA 450 4 Maintenance PA450E04.CHP Maintenance • soiling or blocking of diaphragm → this Electrode soiling may impair measurement to such an extent that the electrode ceases to causes poor response and unstable function at all, e.g. due to: measurement.
4 Maintenance probfit CPA 450 Cleaning via the rinse chamber The integrated rinse chamber permits cleaning via a hose connection set and injector CYR 10 (see Accessories). Instructions for electrode cleaning Loosen union nut 1/2 revolution. Extend immersion tube completely.
PARAM MONTAGE5.CDR Mount electrode and assembly in reversed dircetion as at unmounting. Maintenance and replacement of wearing parts The retractable assembly Probfit CPA 450 • Grease dry O-rings (particularly the requires only little maintenance. However, O-rings on the protection cage).
5 Accessories and spare parts probfit CPA 450 Accessories and spare parts The following accessories may be ordered • Hook wrench DIN 1810 design B separately: D 58 - 62 mm Order no. 50090687 • Spray cleaning injector CYR 10 (see Techn.
CPA 450 6 Technical data PA450E06.CHP Technical data Manufacturer Endress+Hauser General specifications Designation Probfit CPA 450 Installation Immersion depth 250 mm, 700 mm No. of electrode positions in sensor holder 1 mounting position for combination gel electrode Electrode length...