Storage, transport
16
5
Storage, transport
5.1
Storage temperature
• Permitted storage temperature: –40 to +80 °C (–40 to +176 °F)
• Use original packaging.
• Option for FMP51 and FMP54: –50 to +80 °C (–58 to +176 °F)
This range applies if the option JN "Transmitter ambient temperature" –50 °C (–58 °F)
was selected in order code 580 "Test, Certificate". If the temperature is permanently
below –40 °C (–40 °F), higher failure rates can be expected.
5.2
Transporting to the measuring point
WARNING
L
Housing or probe may become damaged or break off.
Risk of injury!
‣
Transport the measuring instrument to the measuring point in its original packaging or
by the process connection.
‣
Always secure lifting equipment (slings, eyes, etc.) at the process connection and never
lift the device by the electronic housing or probe. Pay attention to the center of gravity
of the device so that it does not tilt or slip unintentionally.
‣
Follow the safety instructions and transport conditions for devices weighing more than
18 kg (39.6 lbs) (IEC 61010).
Levelflex FMP51 Modbus