Operating menu
Additional information
Measured thickness upper layer
Navigation
Description
DC value
Navigation
Description
Calculated DC value
Navigation
Description
130
UP Interface thickness (= thickness of upper medium)
On the local display, the measured interface thickness is indicated together with the
manual interface thickness. By comparing these two values the device can
automatically adjust the dielectric constant of the upper medium.
Setup → Advanced setup → Interface → Meas.thick.u.lay
Displays the measured interface thickness. (Thickness UP of the upper medium).
Setup → Advanced setup → Interface → DC value
Displays relatvie dielectric constant ε
Setup → Advanced setup → Interface → Calc. DC value
Displays calculated (i.e. corrected) relative dielectric constant ε
medium.
UP
UP
of the upper medium (DC
r
Levelflex FMP51 Modbus
A0013313
) before correction.
1
(DC1) of the upper
r
Endress+Hauser