5
The shutter valve must be in open position, in order to allow
the patient to breath normally.
Once the patient has reached the position of Total Pulmonary
Capacity, move the cursor to the closed position and perform the
maneuver.
The Icons give access to:
Escapes this screen and goes back to the previous one
Enables to modify the Patient data
Starts the maneuver
/
Changes the test, from MEP to MIP or vice versa
Performs the Report (If there are performed maneuvers)
6
Once the maneuver is finished, repeat step 3 in order to
perform more maneuvers.
It is advisable a minimum of 3 maneuvers. The best 3 maneuvers
should not differ more than 5 % between them, and the last
maneuver should not be the best.
MIP-MEP module userÕs manual
10
511-BI0-MU2
¥
REV. 2.02