NeurOptics NPi-300 Cleaning Instructions

Pupillometer
Hide thumbs Also See for NPi-300:

Advertisement

Quick Links

NPi
Pupillometer
Cleaning Instructions
Always handle the NPi-300 Pupillometer and NPi-300 Charging Station with care because
sensitive metal, glass, plastic and electronic components are contained inside. The NPi-300
and Charging Station can be damaged if dropped or by prolonged exposure to liquid or high
humidity environments.
If the NPi-300 and Charging Station are not working properly, or are believed to have been
damaged, immediately contact NeurOptics Customer Service at
866.99.PUPIL (866-997-8745), international: +1-949- 250-9792, or email: Info@NeurOptics.com.
For additional information regarding our product, please refer to the Instructions for Use which
can be found on our website: NeurOptics.com.
Cleaning the NPi-300 Pupillometer and NPi-300 Charging Station
Isopropyl alcohol (IPA)-based cleaning solutions, in formula concentrations up to 70% IPA, are
recommended for use in cleaning the NPi-300 and Charging Station. Do not use chemicals that
can damage the NPi-300 and Charging Station surface. Some chemicals can weaken or damage
plastic parts and may cause instruments to not operate as intended. Use all cleaning products
per manufacturer's instructions, being careful to squeeze out excess liquid prior to wiping the
NPi-300 and Charging Station and do not use an oversaturated cloth.
Wipe all exposed surfaces. Follow the cleaner's manufacturer instructions as to the time required
to leave the solution on the device surface.
• DO NOT
use an oversaturated cloth. Be sure to squeeze out excess liquid prior to wiping the
NPi-300 or the Charging Station.
• DO NOT
allow the cleaner to collect on the instrument.
• DO NOT
use any hard, abrasive or pointed objects to clean any part of the NPi-300 or
Charging Station.
• DO NOT
immerse the NPi-300 or the Charging Station in liquid, or attempt to sterilize the
product, as damage to the electronic and optical componentry could occur.
Drying and Inspection Following Cleaning
Confirm the NPi-300 and Charging Station are thoroughly dry before placing the NPi-300 back
into the Charging Station.
- 300
®
Toll Free North America:

Advertisement

loading
Need help?

Need help?

Do you have a question about the NPi-300 and is the answer not in the manual?

Questions and answers

Summary of Contents for NeurOptics NPi-300

  • Page 1 Charging Station can be damaged if dropped or by prolonged exposure to liquid or high humidity environments. If the NPi-300 and Charging Station are not working properly, or are believed to have been Toll Free North America: damaged, immediately contact NeurOptics Customer Service at 866.99.PUPIL (866-997-8745), international: +1-949- 250-9792, or email: Info@NeurOptics.com.
  • Page 2 (bleach) when cleaning equipment. If products containing sodium hypochlorite (bleach) are used to clean the LCD of the NPi-300, the cleaning process should be followed by a second cleaning using a clean, soft, lint-free cloth and up to 70% IPA to ensure that all bleach residue is completely removed from the LCD.