Download Print this page
NeurOptics NPi-300 Quick Start Manual

NeurOptics NPi-300 Quick Start Manual

Pupillometer
Hide thumbs Also See for NPi-300:

Advertisement

Quick Links

NPi
Pupillometer
Quick Start Guide
Power Up
Charging the NPi-300 Pupillometer
1.
Connect the NPi-300 Power Adapter to the NPi-300 Charging Station and
plug into a power outlet. The indicator light at the base of the Charging
Station will display a white color to indicate power has been established to
the Charging Station.
2.
Place the NPi-300 into the Charging Station. The Charging Station indicator
light will turn blue, and the LCD screen will display
icon, indicating the NPi-300 is charging. The indicator light will turn
when fully charged.
NOTE: Prior to first-time use, ensure the NPi-300 is fully charged and
Date/Time are set accurately (refer to Setting Date and Time section).
3.
The NPi-300 Pupillometer goes to sleep in the Charging Station to
efficiently charge:
The NPi-300 will initially turn ON (or remain on) upon being placed in the
Charging Station.
After 5 minutes in the Charging Station, the NPi-300 will go to sleep to
efficiently charge. The screen will go dark. If any button is pressed or the
screen is touched within this 5-minute window, the time period before the
NPi-300 goes to sleep is extended by an additional 5 minutes.
To use the NPi-300 after it has gone to sleep in the Charging Station, simply
remove from the Charging Station, and it will wake up automatically.
4.
If the NPi-300 Pupillometer is not in the Charging Station, to conserve
battery life it will:
Go into sleep mode after 5 minutes. To turn ON, touch the screen or push
any button.
Power down after 20 minutes.
Turning On the NPi-300 Pupillometer
5.
If the NPi-300 is out of the Charging Station and has powered down, press
(do not hold) the On/Off button
- 300
®
on the side of the device.
within the battery
2
green
1
3
5

Advertisement

loading

Summary of Contents for NeurOptics NPi-300

  • Page 1 NPi-300 goes to sleep is extended by an additional 5 minutes. • To use the NPi-300 after it has gone to sleep in the Charging Station, simply remove from the Charging Station, and it will wake up automatically. If the NPi-300 Pupillometer is not in the Charging Station, to conserve battery life it will: •...
  • Page 2 Press the RIGHT or LEFT buttons (green circles) to return to the Home Screen. Measuring Pupils Using the NPi-300 Pupillometer The NPi-300 provides objective pupil size and reactivity data independent of examiner – removing variability and subjectivity. The NPi-300 expresses pupil reactivity numerically as the Neurological Pupil index, NPi (see the Neurological Pupil index Scale below).
  • Page 3 Patient ID with either alpha or numeric characters. Scan Barcode Using Incorporated Barcode Scanner Select Scan Code. The NPi-300 will emit a white light from the top of the device. Center the light over the barcode until you hear an audible beep.
  • Page 4 Cleaning and Maintenance Isopropyl alcohol (IPA)-based cleaning solutions, in formula concentrations up to 70% IPA, are recommended for use in cleaning the NPi-300 Pupillometer and NPi-300 Charging Station. Please refer to the Cleaning and Maintenance Instructions included in the Instructions For Use, which can be found at NeurOptics.com.