DEAR CUSTOMER, THANK YOU FOR CHOOSING A GLOBUS PRODUCT. WE REMAIN AT YOUR DISPOSAL FOR ANY ASSISTANCE OR ADVICE YOU MAY NEED. DOMINO s.r.l. via Vittorio Veneto 52 31013 - Codognè - TV - Italy Tel. (+39) 0438.7933 Fax. (+39) 0438.793363 e-mail: info@globusitalia.com...
Index TECHNICAL FEATURES ................................ 4 INTENDED USE of the device ............................6 DEVICE CLASSIFICATION ..............................6 SAFETY PRECAUTIONS ............................... 7 Warnings before the use ............................7 Warnings during the use ............................8 Personal protection ..............................8 Absolute contraindications ............................8 Relative contraindications ............................
160 x 99 x 35.4mm Weight 440 g Certificates CE MDD certificate Laser class (CEI EN 60825-1) Laser handpiece for Physiolaser 1000 “Handpiece Green 808 nm” Diode type GaAlAs Laser class (CEI EN 60825-1) Max RMS power (CEI EN 60825-1) 1 W ±20% Optics spot dimension 11 mm –...
Page 5
Power supply Producer GLOBTEK Model GTM41060-2512 100-240V 50-60Hz 600mA max 2.08A Battery Battery pack Ni-MH 7.2V 1.8Ah Conditions of use and storage of the device Temperature: from 0°C to 35°C Maximum relative humidity: from 15% to 93% Atmospheric pressure: from 700 hPa to 1060 hPa Storage and shipping temperature: from -10°C to 45°C Maximum relative humidity:...
INTENDED USE of the device These laser therapy devices are used in physiotherapy and podiatry, for the treatment of pain, trauma and pathologies affecting the musculoskeletal system. These devices are used in physiotherapy and podiatry, for applications in rehabilitation medicine, sports medicine, orthopedics in outpatient and hospital settings, in operational areas such as outpatient clinics, medical practices and rehabilitation centers.
SAFETY PRECAUTIONS For maximum safety, the device must be used following the rules and the limitations of the user manual. The manufacturer declines all responsibility with reference to a different use from what is indicated in this manual. The full or partial reproduction in any form and by any electronic or mechanical means of the texts and/or pictures contained in this manual without the written authorization of the manufacturer is forbidden.
- If the device is not connected to the mains, sudden shutdown shortly after switching on could indicate a low battery. Recharge as reported in the section "HOW TO CHARGE THE BATTERIES". - Please make sure the guiding light is working. If it does not work please pay attention, as the device emits even if the light is not visible.
• On the abdominal and uterine area in pregnant women. There is no scientific evidence that demonstrates potential risks to the fetus or mother, but for prudential purposes it is advisable to avoid direct irradiation on pregnant subjects. • On patients with neoplasms. It is possible that the therapeutic application of the laser can accelerate carcinogenesis in cancer patients.
EQUIPMENT Carrying case Device Laser handpiece for Physiolaser 1000 “HANDPIECE GREEN 808 nm” (REF G5796) with 11 mm optics – area 0,95 cm² Laser handpiece for Physiolaser 500 “HANDPIECE WHITE 808 nm” (REF G5767) with 11 mm optics – area 0,95 cm²...
Device Power supply by electricity mains. Physiolaser 1000 -500 can operate connected to the mains. Connect the power supply to the connector as shown in the figure (below). Handpiece and emergency button connection Insert the connector of the handpiece and the emergency button in the appropriate housings located in the upper part of the unit (see photo below).
The handpiece must be inserted in the first output of the device. The emergency button must be inserted in the third output of the device before starting a treatment with the laser. Outputs 2 and 4 are used only during service and repair processes by qualified personnel. Battery recharge The device is supplied with a set of rechargeable nickel-metal hydrate batteries (7.2V, 1.8Ah) with high performance and without memory effect.
DESCRIPTION OF THE DEVICE Below is a brief description of the keyboard and commands of the device. Buttons increase/decrease parameter value P+ Button FN button * Button LEFT Button P- Button Right Button ON/OFF/OK button ON/OFF/OK button ..Confirms the selection, starts and pauses the treatment. Pressed for 3 seconds, it allows switching on and off.
DEVICE USE General instructions Turning the device on/off To turn on or off, press and hold the ON / OFF / OK button until an acoustic signal is heard. Unlock device with PIN code When the device is not in use it must be protected against unauthorized use, for this purpose the PIN function is on.
Pulsed mode: laser emission with variable duty cycle and frequency that depends on the individual program (frequency that can be set for programs created by the user). Program type The device supports 2 types of programs: Scanning mode: the treatment is carried out uniformly over one area. Point mode: the treatment is performed by points using the handpiece statically on each point.
Page 17
Both in the programs already in the library and in free programming, the power can be adjusted up to 60 minutes of treatment. With very low powers, the treatment duration timer may be updated every 2-3 seconds instead of second by second. This is completely normal and does not indicate a malfunction.
Screenshots of a running treatment program Scanning mode: Wave length Peak power Battery status Average power Remaining program time Dosage per cm² Duty Cycle Total dosage transferred Treatment area Point mode: Rev 9_19.07.2023...
Page 19
Wave length Average power Battery status Point number/total points Remaining program time Total dosage transferred Dosage per cm² Peak power By keeping the * key pressed, you go to the screen-page that usually characterizes a laser therapy protocol. The following screen-page will appear with the specific parameters of the program being executed: Name of the program Frequency...
Page 20
How to save a program to easily find it for a new use In the main menu, select PROGRAM LIST and select the desired program. Choose SAVE TO FAVORITES Start the treatment. For a new use we will find the program in the FAVORITES section (Folder) of the main menu.
Page 21
Screenshots of programs in the Programming section Scanning mode Choosing the scan mode, this screen will appear: Mode Intensity Battery status Remaining program time Optic diameter Continuous emission Dosage per cm² Emitted energy Treatment area Duty cycle Pulsed emission Rev 9_19.07.2023...
Page 22
Point mode Choosing the point mode, instead, this screen will appear: Mode Intensity Battery status Energy per point Remaining program time Continuous emission Optic diameter Emitted energy Dosage per cm² Duty cycle Point number/total points Treatment area Pulsed emission Rev 9_19.07.2023...
Page 23
Using the adjustment keys, you can set the frequency (if the pulsed emission mode has been chosen), the energy per point, the power intensity and the number of points. By pressing the * key, the parameter “D” (Power density) in J / cm² is displayed at the bottom and, with the adjustment keys, it can be changed.
Page 24
g. Laser unlock: It is possible to activate or deactivate the safety function during the “point” laser treatment. The device is supplied with the “Key on” function, i.e. during the point treatment, to start the treatment of each point it is necessary to press the “OK”...
Program List AESTHETICS - FITNESS (non-medical treatments) X = Program available in the device Beauty Physiolaser 500 Physiolaser 1000 FACE AESTHETIC Wrinkles Acne Dehydrated skin Skin rejuvenation Face skin - lifting Expression Wrinkles Neck wrinkles Rosacea Teleangectasie BODY AESTHETIC Breast stretch marks Cell.
MAINTENANCE AND CLEANING Maintenance and cleaning of the device - In case of real or alleged malfunctioning, do not tamper with the device and do not try to repair it by yourself. Do not try to repair the device and do not open it. Only specialized and authorized centers can repair it.
APPLICATION METHOD Functionalities The main goal of laser therapy treatment is to exploit the analgesic and anti-inflammatory effects that laser radiation has on human tissue. The laser light is absorbed by the different molecules normally present within the surface layers of biological tissues, called chromophores. This absorption is lower in the near infrared region, the wavelengths present in this range are therefore able to better penetrate the soft tissues through the skin and muscles;...
Page 29
the drainage of algogenic substances. It interacts with the activity of large-caliber myelinated fibers, modulating the perception of pain. • Biostimulating effect: metabolic activation of fibroblasts, increase in ATP, increase in the supply of nutrients and growth factors, necessary for tissue regeneration. Use method To activate the laser radiation, the button on the handpiece must be pressed.
Annex 1: SYMBOLS and LABELING of medical devices It refers to the manufacturer Warning Keep the device dry This symbol on your device indicates that it complies with the requirements of the directives on medical devices (93/42/EEC 47/2007EEC). The number of the notified body is 0476. It indicates that this is a class II device.
Page 31
Symbol for polythene Device GLOBTEK Model: GTM41060-2512 PRI: 100-240V 50-60Hz 0.6A max SEC: 12V 2.08A Laser type: Physiolaser 1000 Input power: 12V 760mA IP20 Input battery: 7.2V 910mA Wave output: 808nm Max Power: 1000mW (Cont. DC 100% Pulse DC 50%)
Annex 2: INFORMATION SIGNS Information signals visible on the display Correct power emission: The treatment is going on properly. Laser emission. Cable not connected: Ensure that the cable is connected to the correct output. Device in pause: Press OK to continue the treatment. The device is in stand-by mode: Press the button on the handpiece to start the emission.
Page 33
N.B. Before returning the device for repairs, we suggest reading carefully the user manual and consulting the Globus website. If you have to return the product for repair, please contact your dealer or the Globus customer service. Rev 9_19.07.2023...
Annex 3: EMC accompanying documents Essential performance PERFORMANCE CONDITION RISK ACCEPTED EVENT Laser External Display information The machine must stop the laser disturbance no longer readable. emission. (Burst). The machine must maintain the emission and accept the commands. Lack of internal Interruption of the The machine must signal the battery power supply.
Page 35
In compliance with: EN 60601-1: 2006 + A1: 2013 EN 60601-1-2: 2015 EN 60601-2-22: 2013 Warning: radiofrequency communication devices (including accessories like antennas or antenna cables) must be used at least 3 meters away from every part (including the cables and the accessories) of the device.