UNPACKING THE SHIPPING BOX Shipping Box Contents ........................x A. Overview A.1 The AQLite-Standard Air Monitor A.1.1 Ozone Measurements Using the 2B Technologies Model 108-L ..........1 A.1.2 Sensor Package: CO, CO , PM , and PM Measurements ........... 3 A.1.3 Viewing and Acquiring Data with the AQLite-Standard ............
Page 3
Theory of Operation ....................... 41 Adaptive Filter ........................43 Instrument Default Settings ....................44 Labeled Photo of the Ozone Monitor in the AQLite ............... 44 SPECIFICATIONS: Model 108-L Ozone Monitor in the AQLite Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 4
ZERO AND SPAN CHECKS – OZONE MONITOR APPENDIX 2: The Sensor Package of the AQLite OVERVIEW SENSOR PACKAGE TECHNICAL INFORMATION Sensor Specifications ......................61 Power and Battery ......................... 61 Printed Circuit Board ......................61 GPS Module .......................... 61 Model AQLite-Standard Air Monitor Manual Rev. B-2...
IDENTIFICATION RECORDS Record the following information for future reference: Unit serial number: ______________________________________ Warranty start date: _______________________________________ (date of receipt) Model AQLite-Standard Air Monitor Manual Rev. B-2...
PRINTING HISTORY This manual covers the AQLite-Standard Air Monitor used for measurement of ozone ), carbon monoxide (CO), carbon dioxide (CO ) and particulate matter (PM ) in air, with options for measuring nitrogen dioxide (NO ) and sulfur dioxide (SO New editions of this manual are complete revisions that reflect updates to the instrument itself, as well as clarifications, additions and other modifications of the text.
2B Technologies warrants its products against defects in materials and workmanship. 2B Technologies will, at its option, repair or replace products that prove to be defective. The warranty set forth is exclusive and no other warranty, whether written or oral, is expressed or implied.
Limitation of Remedies and Liability The remedies provided herein are the Customer's sole and exclusive remedies. In no event shall 2B Technologies be liable for direct, indirect, special, incidental or consequential damages (including loss of profits) whether based on contract, tort or any other legal theory.
WARNING: ATENCION: If this instrument is used in a manner not specified by 2B Technologies, Si este instrumento se usa de una forma no especificada por 2B USA, the protection provided by the instrument may be impaired.
Shipping Box Contents Open the shipping box and verify that it contains all of the items on the shipping list. If anything is missing or obviously damaged, contact 2B Technologies immediately by email at techsupport@twobtech.com or by phone at +1(303)273-0559.
A. Overview of the AQLite-Standard A. Overview The AQLite-Standard Air Monitor The AQLite-Standard Air Monitor is a compact and versatile package that combines an Environmental Protection Agency (EPA)-Federal Equivalent Method (FEM) instrument for measuring ozone (O ) with multiple sensors for high-quality yet affordable...
Page 12
Indicator lights Sample air - inlet Optical bench (absorption cell) Solenoid valve Inlet filter Cell outlet Bulkheads (inlet- bottom; exhaust- top) Power cable feedthrough Monitor Figure A.1. Model 108-L Ozone in the AQLite Model AQLite-Standard Air Monitor Manual Rev. B-2...
, PM , and PM Measurements The AQLite-Standard uses sensors to measure air pollutants. The sensor package is mounted on the inside of the door of the AQLite Air Monitor, as shown in the photo at right. The sensors and other components of the AQLite-Standard are shown in the photo below.
There, the data can be accessed for analysis using the Data Portal on top menu bar of the 2B Technologies website: https://twobtech.com. This integrated web- based data stream is the primary data acquisition mode for use of the AQLite-Standard. averaging...
Users who wish to know more detail about the Ozone Monitor and the sensor package, will find information in the Appendices: Appendix 1: The Model 108-L Ozone Monitor Appendix 2: The AQLite-Standard Sensor Package and Options Model AQLite-Standard Air Monitor Manual Rev. B-2...
The choice among these methods must be specified when the AQLite is ordered so that the proper converter is installed (DC to DC, or AC to DC). If you later wish to change the configuration of your AQLite, please contact 2B Technologies. B.1.2 Power Cable Feedthrough To create the correct air flow in the AQLite, you must properly seal the hole where the DC or AC power wire enters the enclosure.
Page 17
AQLite enclosure. Place the rubber sleeve with slit around the wire. 5. From the outside of the enclosure, feed the cable through the threaded fitting. Note orientation of threaded fitting Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 18
8. On the outside of the enclosure, be sure that the split rubber sleeve is around the power cable. Gently push the sleeve into the fluted end of the fitting. Push the black rubber sleeve into the fitting Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 19
• If the direct AC configuration is being used your instrument must have been ordered with an AC-DC converter rather than a DC-DC converter. Install the wiring as shown: Position 1: CASE GROUND Position 2: LINE Position 3: NEUTRAL Model AQLite-Standard Air Monitor Manual Rev. B-2...
(charging takes about 4 hours). • Solid red LED indicates that proper battery charging is occurring. • If blinking red, battery is not properly charging. • Light goes off when battery is fully charged. Model AQLite-Standard Air Monitor Manual Rev. B-2...
The back of the enclosure has three holes at the top and three holes at the bottom for mounting screws. The holes are 1/4” in diameter and should be used with stainless steel bolts and washers. The dimensional drawing is shown on the next page. Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 22
B. Installation and Startup Model AQLite-Standard Air Monitor Manual Rev. B-2...
The sensors require a warmup period of about an hour. ****** The steps in this Section B have ensured that your AQLite is functioning. The next section of the manual describes how to make measurements with the AQLite. Model AQLite-Standard Air Monitor Manual Rev. B-2...
When power is supplied to the AQLite-Standard, the sensor package and the Ozone Monitor begin making measurements. These measurements are integrated into a data stream that is uploaded via cellular to a 2B Technologies website and stored on a micro- SD card in the AQLite.
Page 25
After a few moments, you will see the dashboard, which gives a display of your devices (identified by their serial numbers) and their locations on a map: The AQLite serial number is located on the back of the instrument case. Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 26
(note, though, that if your customized AQLite-Standard does not have one or more of the sensors, its indicator may appear yellow even if your AQLite is fully operational).
Page 27
(regardless of whether the AQLite was turned on and off at different times during the time window). The file column headings in the csv are largely self-explanatory. A few clarifications are given below. Model AQLite-Standard Air Monitor Manual Rev. B-2...
4. Type “W” to see the files that are on the SD card. An example of a list of files is shown below: 0: 1080_20210708_163131.txt 1: 1080_20210708_163516.txt 2: 1080_20210708_171250.txt 3: 1080_20210708_184952.txt 4: 1080_20210708_185151.txt 5: 1080_20210708_185818.txt End of files on Sd. Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 29
9. Select the lines you wish to analyze, and then copy and paste them into a spreadsheet such as Excel. Note that your data line may look slightly different than the above example, depending on the sensors you have selected for your AQLite. Model AQLite-Standard Air Monitor Manual Rev. B-2...
6. Type “?” to see the list of available serial commands. Type “x” to exit the menu and resume measurements. The list of commands is shown in the next section. 7. Type “x” to exit the menu and resume measurements. Model AQLite-Standard Air Monitor Manual Rev. B-2...
Adjust NO2 slope Adjust NO2 zero Get build version Enable NO2 Disable NO2 Output the PMIC system configuration Increase the charge current by 64 mA Increase the current input limit by 100 mA Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 32
Enable/Disable Sensible-iot data push. If enabled, time zone will be ignored - UTC will be used. Output this menu Exits this menu and resumes measurements Commands listed in parentheses (k, l, u) are not currently operative. Model AQLite-Standard Air Monitor Manual Rev. B-2...
Exit menu and return to measuring. Default settings: Avg=10 s, offset=0, slope=1, adaptive filter difference=0, adaptive filter percent=0, adaptive filter long average length=25, adaptive filter short average length=10, T in °C, P in mbar, O in ppb. Model AQLite-Standard Air Monitor Manual Rev. B-2...
The steps to change it are below. But again: we strongly recommend using your AQLite at the factory settings of 5 minutes for the upload frequency. Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 35
D. Accessing the Serial Menu ***************** We recommend consulting with 2B Technologies before making any of the changes mentioned in (a)-(f) below. a) Connect the AQLite’s USB cable to the USB of your computer and enter the serial menu as described previously in Section D.1.
Though many other serial commands are given in Section D.2, they will not commonly be used during normal operation. Calibration settings, for example, are only needed if the user elects to perform calibrations. Contact 2B Technologies with inquiries about any of the commands not described in this manual.
1. Pull gently to remove silicone tubing from the barb fitting of the AQLite inlet 2. Attach the silicone tubing to the white barb fitting of the ozone scrubber/filter 3. Zeroing setup shown above Model AQLite-Standard Air Monitor Manual Rev. B-2...
7. Important: Also restore the measurement time to 5 minutes. E.1.2 Calibration of the Ozone Monitor It is recommended that the AQLite be returned to 2B Technologies annually for cleaning and calibration service. At that time, both the Ozone Monitor and the sensor package will be calibrated.
E.2.1 Zeroing and Calibration of the Sensors The zeroing, calibration, and occasional replacement of the AQLite’s sensors are done at 2B Technologies, at a frequency that depends on the environmental sampling conditions. Contact 2B Technologies to discuss recommended scheduling. E.2.2 Recommended: Frequent Checks of the Sensor Functioning It is advisable in all cases to observe the sensor readings and assess whether they make sense for the conditions.
(~20,000 hours) and solenoid valve (rarely fails). It is recommended that the instrument be returned to 2B Technologies if any of these components fail. Alternatively, the user may install these components at their own risk. In that case, please contact 2B Technologies for instructions.
Contaminated Flow Path: The instrument will typically have a large positive or negative offset and the ozone readings will be low once corrected for the measured offset. Lamp Glow Power Low Flow High Ozone Lamp Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 42
Wrong serial cable work. used. provided. Some data collection devices require a ”cross over” cable in which pins 1 and 3 are exchanged between the two ends of the cable. Use a “cross over” cable Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 43
(second light on the circuit fallen below 0.6 V. and check lamp connection to circuit board; see Figure A.1 of main board. Run Lamp Test from menu. If photodiode voltage is noisy and is Model AQLite-Standard Air Monitor Manual Rev. B-2...
• Be sure the all the cable connections to the sensor package are secure. F.2.3 Service through 2B Technologies 2B Technologies offers reasonably priced customer service for instrument repairs. The calibration service for the ozone monitor includes cleaning of the entire flow path with methanol, testing of all components for proper function, installation of a new internal ozone measurement scrubber and calibration against a NIST-traceable standard.
Package to Ozone Monitor Monitor Air Intake for Ozone Monitor (not visible) Power Cable Gland Sampling Fan Exhaust Air Intake for Sensor Package Figure G.1. Overview of the AQLite-Standard Air Monitor (inside and outside views) Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 46
Sample air - inlet Optical bench (absorption cell) Solenoid valve Inlet filter Cell outlet Bulkheads (inlet- bottom; exhaust- top) Power cable feedthrough Figure G.2. The Ozone Monitor of the AQLite Air Monitor. Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 47
Connection (mounted near top of the AQLite door) GPS Module Particulate Matter (PM) Lithium Ion Battery Sensor (located under PM sensor) (Plantower PM5003) Figure G.3. The Sensor Package of the AQLite-Standard Air Monitor Model AQLite-Standard Air Monitor Manual Rev. B-2...
Note that replacement of the solenoid valve requires a knowledge of soldering. Please see the 2B Technologies website for a full and updated list of parts and pricing: https://twobtech.com/parts-online.html. Please contact us with questions or for any items not listed below.
I. Service Log I. Service Log 2B Tech Model #__AQLite-Standard Air Monitor AQLite Serial #_________ Calibrated (note if Date/ New O Cleaned Pump Lamp Other sensors, O Hours Scrubber Monitor Monitor, or Monitor Monitor both) Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 50
I. Service Log Calibrated (note if sensor Date/ New O Cleaned Pump Lamp Other package, O Hours Scrubber Monitor Monitor, or Monitor Monitor both) Model AQLite-Standard Air Monitor Manual Rev. B-2...
APPENDIX 1: The Model 108-L Ozone Monitor 1. INTRODUCTION - OZONE MONITOR The 2B Technologies Model 108-L Ozone Monitor inside your AQLite is designed to enable accurate measurements of ozone in air over a wide dynamic range extending from a few parts-per-billion by volume (ppb) to parts-per-million (ppm) levels based on the well-established technique of absorption of ultraviolet light at 254 nm.
Page 52
The pressure and temperature within the absorption cell are measured so that the ozone concentration can be expressed as a mixing ratio in parts-per-billion by volume (ppbv). The following animated video developed by 2B Technologies provides a detailed explanation of how this and other UV-based ozone monitors measure ozone. Click on the link https://www.youtube.com/watch?v=3qBdl5qSYm4...
During operation of the adaptive filter, the firmware will automatically switch between two different filter lengths based on the conditions at hand, as determined by settings specified by the user. 2B Technologies recommends the following settings for the adaptive filter:...
Labeled Photo of the Ozone Monitor in the AQLite On the next page is a labeled photo of the Ozone Monitor showing most of the components given in the previous schematic (Appendix 1 / Section 1.1 / Figure 1.1). Model AQLite-Standard Air Monitor Manual Rev. B-2...
Page 55
Inlet filter Cell outlet Bulkheads (inlet- bottom; exhaust- top) Power cable feedthrough Appendix 1 / Figure 1.2. The Ozone Monitor in the AQLite. Compare to the schematic in Appendix 1 / Figure 1.1. Model AQLite-Standard Air Monitor Manual Rev. B-2...
User-defined scaling factor in serial menu Adaptive Filter Available; user-defined parameters Data Transfer Baud Rate 2400 Relay responds based on ozone set points (user- Relay with Two Set Points defined in serial menu) Model AQLite-Standard Air Monitor Manual Rev. B-2...
AQLite and its Ozone Monitor. Alternatively, we are investigating options for using a solar panel and/or batteries to power the AQLite’s Ozone Monitor and sensor package. Contact 2B Technologies to inquire about the status of that research.
Once the zero of the instrument is corrected, the slope may be adjusted so that the instrument readout agrees with a standard ozone source (such as the 2B Technologies Model 306 Ozone Calibration Source) or with the readout from another instrument whose calibration is considered...
#8 (ground). To change the analog output voltage scaling factor, send the character ‘f’ in the Ozone Monitor’s serial menu and enter a number between 1 and 99999. For example, if you Model AQLite-Standard Air Monitor Manual Rev. B-2...
• The bottom light indicates the ozone level is above 100 ppb. If the Monitor is sampling ambient air, personnel in the vicinity should take precautions to avoid breathing unsafe levels of ozone. Model AQLite-Standard Air Monitor Manual Rev. B-2...
(~20,000 hours) and solenoid valve (rarely fails). It is recommended that the instrument be returned to 2B Technologies if any of these components fail. Alternatively, the user may install these components at their own risk. In that case, please contact 2B Technologies for instructions.
2B Tech or if readings are noisy Troubleshooting Please refer to Section F.2.1 of the main part of this manual for troubleshooting information. Model AQLite-Standard Air Monitor Manual Rev. B-2...
This calibration may be performed by the user with the guidelines provided in this section. Alternatively, the ozone monitor may be returned to 2B Technologies for calibration service. 2B Tech’s calibration service includes cleaning of the entire flow path with methanol, testing of all components for proper function, installation of a new internal ozone scrubber and calibration against a NIST-traceable standard.
Calibration Setup Preparation As indicated in the EPA Technical Assistance Document, there are several tests that should be performed prior to calibration to ensure the accuracy of the measurements. These tests include: Model AQLite-Standard Air Monitor Manual Rev. B-2...
R = Dilution ratio = Measured concentration of = Ozone generator flow original concentration = Diluent zero air flow = Measured concentration of E = Linearity error, in percent diluted concentration Model AQLite-Standard Air Monitor Manual Rev. B-2...
1. Generate an ozone concentration slightly less than the concentration range of interest and allow the ozone generator to warm up for at least 5 minutes. The same zero air supply used for making zero air measurements must be used in the ozone generator. Model AQLite-Standard Air Monitor Manual Rev. B-2...
Z and S as determined above. If the calibration has been done in units other than ppb, the offset (Z) value must be converted to ppb for entry into the instrument software. Model AQLite-Standard Air Monitor Manual Rev. B-2...
The most likely problems in the ozone monitor are leaks, a malfunctioning ozone measurement scrubber, a contaminated valve, or contamination in the optical setup. See the “Troubleshooting” section of this manual (Section F.2.1 of the main manual). Model AQLite-Standard Air Monitor Manual Rev. B-2...
(located under PM sensor) (Plantower PM5003) The AQLite-Standard comes with the sensors shown in the photo above. In the customizable options, the CO sensor location and another location directly opposite from it on the board (unoccupied in the photo above) can be populated with user’s...
4 hours. The battery recharges in about 4 hours. Printed Circuit Board The sensor package is built on a customized printed circuit board. GPS Module The GPS module is a U-Blox CAM-M8. Model AQLite-Standard Air Monitor Manual Rev. B-2...