This operating instructions contains safety information that if ignored can endanger life or result in serious injury. Read these instructions carefully before use and keep them for future reference. The original instruction is in Italian. All non-Italian instructions are translations of the original instruction. Information and specifications on this manual could be uncorrect or could have printing errors.
PURPOSE OF USE AND METERING PUMP IS INTENDED FOR CHEMICAL DOSING AND DRINKING WATER SAFETY TREATMENT. Do not use in explosive area (EX). Do not use with flammable chemicals. Do not use with radioactive chemicals. Use after a proper installation. Use the pump in accordance with the data and specifications printed on the label.
ENVIRONMENTAL Work area SAFETY Always keep the pump area clean to avoid and/or discover emissions. Recycling guidelines EWC code: 16 02 14 Always recycle according to these guidelines: 1. If the unit or parts are accepted by an authorized recycling company, then follow local recycling laws and regulations.
1. Introduction Introduction “F” and “FA” Series metering pumps are the ideal solution for low / middle dosing of chemicals. All control and setup parameters are available using accurate control knobs and a visual system (led). “F” and “FA” Series metering pumps have digital On/Off switch to ensure dosing activities (available on some models only). Pump capacity: Flow rate is determined by the stroke speed (frequency) adjustment. The stroke speed is adjustable from 0 to 100% using the adjustment knob. However dosing accuracy is guarantee within an adjustment range from 30% to 100%. The led on the panel shows the status activity of the pump. Legend: Alternating Current; Protective Earth; Standby; Warning - Note for models without main plug only: The unit shall be provided with a UL listed disconnect device, rated min. 250V, xxA Feeder should be interlocked with a no-flow protection device to automatically shut-off the pumps when there is no flow! Adequate measures shall be taken to prevent cross connection of chemicals! Chemical feeding must be stopped during backwash cycles and periods of noflow as these...
2. Unpacking Included into package: Dibbles ø6 Self tapping screws 4,5 x 40 Delayed fuse 5 X 20 Foot filter with valve Injection valve n. 1 Level probe (not included in FCO and FPDR model) Delivery pipe* (opaque PE) Suction pipe * (transparent PVC) Discharge pipe (transparent PVC 4x6) This installation manual If hose is 6x8 then there is only a 4mt long hose. Cut to obtain suction and delivery hoses. PLEASE DO NOT TRASH PACKAGING. IT CAN BE USED TO RETURN THE PUMP .
3. General description On/Off button Stroke speed knob On/Off Led - Pump’s pulses Delivery valve Discharge knob Discharge hose Suction valve Level Probe Input Power supply External Signal Input Note: Shown image does not represent a specific pump’s model. It is only for general purpose. WARNING IF EQUIPMENT IS SUPPLIED WITH A PLUG: If an appliance coupler or separable plug is used as the disconnecting device, it shall be readily identifiable and easily reached by the operator.
4. Before to Install warnings Pump’s installation and operativity is performed in 4 main steps: Pump’s installation Hydraulic Installation (hoses, level probe, injection valve) Electrical Installation (main power connection, priming) Programming the pump. Before to start, please read carefully the following safety information. IProtective clothes Wear always protective clothes as masks, gloves, safety glasses and further security devices during ALL installation procedure and while handling chemicals.
5. Installation draw Pump must be installed in a stable support at a maximum height (from tank’s bottom) of 1,5 meters. 1. Injection valve 2. Delivery hose 3. Safety breaker 4. Control panel 5. Pump head 6. Power cable 7. Outgassing hose 8. Suction hose 9. Foot filter...
6. Hydraulic Installation Hydraulic connections are: Suction Hose with level probe and foot filter Delivery Hose with injection valve Discharge Hose Suction Hose. Completely unscrew tightening nut from pump’s head and remove assembling components: tightening nut, holding ring and pipe holder. Assembly as shown in fig. (A). Insert hose into pipe holder until it reaches the bottom. Lock hose on pump’s head by screwing down the tightening nut. Use only hands to do it! Connect other side of the hose to the foot filter using the same procedure. Suction Hose Tightening Nut Holding Ring Pipe Holder O-ring Valve fig. (A)
Page 11
6. Hydraulic Installation Assembling foot filter with level probe. Level probe must be assembled with foot filter using the provided kit. Foot valve is made to be installed into tank’s bottom without sediments priming problem. Completely unscrew level probe’s nut and assembly as described in fig. (B) paying attention to the black ring: it must be inserted from floater side. Lock nut on the opposite side of the floater using hands only. to BNC connector to suction hose (pump’s level input) level probe black ring fixing nut foot filter fig. (B) Connect BNC from level probe into pump’s level input (front side of the pump). Put level probe assembled with foot filter into tank’s bottom. Warning: If there is a mixer installed into tank, install a suction lance instead of level probe / foot filter. Delivery Hose. Completely unscrew tightening nut from pump’s head and remove assembling components: tightening nut, holding ring and pipe holder. Assembly as shown in fig. (A). Insert hose into pipe holder until it reaches the bottom. Lock hose on pump’s head by screwing down the tightening nut. Use only hands to do it! Connect other side of the hose to the injection valve using the same procedure.
Page 12
6. Hydraulic installation Injection Valve. Injection valve must be installed on plant from water’s input. Injection valve will open at pressure greater than 0,3bar. Discharge hose. Insert one side of discharge hose into discharge connector as shown in fig (C). Insert other side of discharge hose into product’s tank. During priming procedure product exceeding will flow into tank. To Delivery hose Discharge Knob Discharge hose To Suction hose To delivery hose fig (C) Discharge Knob Discharge hose To suction hose fig. (C) For priming procedure see the related chapter.
Page 13
6. Hydraulic installation Self-venting pump head. to bleed hose to delivery hose to suction hose To bleed hose To delivery hose To suction hose Self-venting pump head must be used when using chemicals that produce gas (i.e. hydrogen peroxide, ammonium, sodium hypoclorite at particular conditions). Hoses assembling procedure (including purge hose) is described in fig. (A). Notes: - suction, delivery and purge valves are DIFFERENT! Do not exchange them! - delivery and purge hoses are made of same material! - it’s allowed to lightly bend discharge hose! - during calibration procedure (“TEST”) insert discharge hose into BECKER test-tube!
7. Electrical installation All electrical connections must be performed by AUTHORIZED AND QUALIFIED personnel only. Before to proceed, please, verify the following steps: - verify that pump’s label values are compatible with main power supply. - pump must be connected to a plant with a differential switch (0,03A sensitivity) if there isn’t a good ground. - to avoid damages to the pump do not install it in parallel with heavy inductance load (for example: engines). A relay switch must be used. See below picture. P - Dosing Pump R - Relay I - Switch or safety device E - Electrovalve or inductance load A - Main Power - On pump’s mother board there is a further protection against over voltages (275V - 150V) and distribution line noises (4KV for max 50µsec) as shown:...
Page 15
7. Electrical Installation Once verified previous steps proceed as follows: - check that “BNC” of level probe has been connected as described in “Hydraulic Installation” chapter. - connect “BNC” and external signal to pump’s “INPUT” connectors. Level Probe Input (bottom view) External Signal Input Level Probe Input available on: FTE, FPDR, FIC, FIS, FPV, FPVM, FCL External Signal Input available on: FPDR (optional), FIC, FIS, FPV, FPVM, FTE...
8. Models LEVEL ALARM CL, IS, IC, PV, TE and PVM type pump are provided with a liquid level alarm to indicate product tank is empty. The level probe is connected to the right BNC plug on pump’s bottom panel. The level probe is made of a N.O. reed contact (10VA, 1A max., 230Vac max.) closed by a floating magnet housed in a (PP) plastic box. When the product level goes below the minimum the magnet closes the reed contact. The pump stops and the red LED on pump’s front panel indicates the alarm status. PUMP TYPES Pumps mod. “FCL ” (12-24 Vac/Vdc), “FIC”, “FIS”, “FPV” and “FPVM” are equipped with a bicolour led. Led on, red colour: low level product alarm. Check product’s tank and restore the level. Led on, blinking green colour: pump normal operating mode.
Page 17
8. Models FCO / FCE Constant pump with flow control (front panel knob may be set from 0% to 100% of pump nominal capacity). Flow control is electronically set and it operates on pump injections number. To avoid linearity problem do not set the knob of pump flow between 0% and 10%. FCO / FCE may operate in constant dosing mode or On-Off mode (using an external signal). Dosing example using an FCO / FCE 0505: to dose 2,5 l/h at 5bar counter-pressure rotate front panel knob to 50%. FCO / FCE has a divider (x- 0,1) to reduce by ten times the pump capacity by dividing the pump stroke speed. How to enable “divider mode”: - set the pump into STANDBY* mode;...
Page 18
8. Models Constant dosing pump with level alarm. A red led indicates that the product’s tank is empty. During this condition the pump does not dose. The pump has flow control (front panel knob may be set from 0% to 100% of pump nominal capacity). Flow control is electronically set and it operates on pump injections number. To avoid linearity problem do not set the knob of pump flow between 0% and 10%. FCL has a divider (x- 0,1) to reduce by ten times the pump capacity by dividing the pump stroke speed. How to enable “divider mode”: - set the pump into STANDBY* mode; - keeping pressed the on/off button, wait 3 flashes from the status led. The pump will start the dosing activity with the stroke speed reduced ten times than the value set on stroke lenght knob. To disable the “divider mode”, power OFF the pump. Keeping pressed the on/off button, wait 3 flashes of the status led. level alarm on/off button stroke speed knob status led STATUS LED The led on the frontal panel shows the pump’s operating status through four flashing:...
Page 19
8. Models Proportional/constant pump driven by a current signal. Setting the switch on constant position, the pump has flow control (front panel knob may be set from 0% to 100% of pump nominal capacity). Flow control is electronically set and it operates on pump injections number. To avoid linearity problem do not set the knob of pump flow between 0% and 10%. Setting the switch on the proportional position, the pump capacity is set proportionally to analog current input signal; a signal change will be followed by a linear change of pump’s capacity. The current range is 0÷20 mA (0 dosing, 20 max dosing). The maximum pump capacity for maximum input signal is set by the % marked knob. The FIC pump can be controlled by any electronic device (such as pH-meter, redox-meter, etc) that has an analog current signal output. This signal must be applied to the bipolar cable provided with the pump, using the following connections: - Red wire : positive (+) - Black wire : negative (-) proportional constant stroke speed knob on/off button...
Page 20
8. Models Proportional/constant pump driven by a digital signal. Setting the switch on constant position, the pump has flow control (front panel knob may be set from 0% to 100% of pump nominal capacity). Flow control is electronically set and it operates on pump injections number. To avoid linearity problem do not set the knob of pump flow between 0% and 10%. Setting the switch on the front panel in the proportional position, for each external “voltage free” pulse corresponds a magnet stroke. If proportional position is set, then the % marked knob does NOT affect the pump capacity. FIS proportional dosing pump can be driven by any external device (such as PCs, PLCs, etc) that has a digital signal. This digital signal (N.O. contact) must be applied to the cable provided with the pump. Connect a 0÷5 or 0÷12 Vdc input signal option using the following connections: - Red wire : positive (+) - Black wire : negative (-) proportional constant stroke speed knob on/off button...
Page 21
8. Models Proportional/constant pump driven by a water meter digital signal. Setting the switch on constant position, the pump has flow control (front panel knob may be set from 0% to 100% of pump nominal capacity). Flow control is electronically set and it operates on pump injections number. To avoid linearity problem do not set the knob of pump flow between 0% and 10%. It is furthermore possible to divide the maximum magnet strokes per minute by 1, 10 and 100 using the switch on the front panel. Setting the switch on proportional position, for each external pulse corresponds one pump stroke. This pump can be driven by a CTFI or CWFI series water meters. This pump can also be driven by a digital signal coming from a “voltage free” contact. Driving signal is applied on the BNC plug on the left bottom of front cover. Dividing factor (N) value is obtained multiplying the value on the adjustment knob using the multiplying switch (x1, x10, x100). Capacity definition for “FPV” and “FAPV” pump Use the following formula to obtain minimum pump capacity: ppm x m ——————————————— = l/h 1000 x K l/h - minimum pump capacity required ppm - product amount to dose in p.p.m. (gr/m k - dosed product dilution factor (pure chemical k=1) - maximum capacity of system to be treated in m /h. Dividing factor (N) to set on adjustment knob: imp/l x cc ( ———————————— ) x 1000 = N ppm x K N - number of the external pulses are divided to be set on adjustment knob imp/l- pulse per liter given by the water meter cc - pump’s single stroke dosing quantity (in cc). Refer to following table...
Page 23
8. Models FPVM Proportional/constant pump driven by a water meter digital signal. Setting the switch on constant position, the pump has flow control (front panel knob may be set from 0% to 100% of pump nominal capacity). Flow control is electronically set and it operates on pump injections number. To avoid linearity problem do not set the knob of pump flow between 0% and 10%. It is furthermore possible to divide the maximum magnet strokes per minute by 1 (÷1), 10 (÷ 10) and 100 (÷100) using the selector on the front panel. The electronic capacity adjustment sets the injection per minute. Setting the switch on the proportional position and the selector on “multiplier” (X1), the pump gives a stroke each 10 external pulses sent. Setting the selector on “divider” (÷1 or ÷ 10), the pump gives at maximum a stroke each external pulse sent and at minimum a stroke each 100 pulses sensed (knob n=10, switch ÷ 10). This pump can be driven by a CTFI or CWFI series water meters. This pump can also be driven by a digital signal coming from a voltage free contact. Driving signal is applied on the BNC plug on the left of the bottom pump cover. “FPVM” and “FAPVM” capacities are defined by the same formulas used for “FPV” and “FAPV”. proportional constant stroke speed knob on/off button selector “multiplier/divider” (1, 10, 100)
Page 24
8. Models Timered/constant dosing pump. Setting the switch on the constant position, the pump has flow control (front panel knob may be set from 0% to 100% of pump nominal capacity). Flow control is electronically set and it operates on pump injections number. To avoid linearity problem do not set the knob of pump flow between 0% and 10%. Setting the switch on timered , the pump starts working when an external pulse is provided. Working active time for each external pulse is set on the adjustment knob in the range between 0 and 60 seconds (different time are available upon demand). The capacity can be also changed adjusting the stroke per minute, operating the % marked knob, at the same time. The double parameter adjustment (strokes and time) allows the use of this pump in high capacity systems, where the pulses are not enough to drive a PV type pump. TE pump is provided of a coaxial cable (RG58) with BNC plug to connect the command signal. Use following formula to evaluate the knob adjustment: 3600 —————————— = sec. imp/h imp/h - water meter output pulse per hour sec - time in seconds to be set on time knob Maximum tag capacity can be decreased 10 or 100 times in the “F” and “FA” pumps using the 1/10/100 switch (where available). In that case the % marked knob will work on the maximum set capacity. timered time regulation knob on/off button stroke speed knob...
Page 25
8. Models FPDR Intermittent constant metering pump with adjustable pause and work times. Metering is set using three knobs on the front panel: - (%) knob (red coloured range) sets speed stroke - (10÷180 seconds) Pause knob sets time between one metering and the next - (10÷180 seconds) Work knob sets metering time Once powered the pump, it always starts its working cycle metering for the set time. The pump can be provided with a N.O. (1A / 230Vac) service contact: this contact is active while metering. If the level input is active while the service contact is active, pump stops and service contact gets back to N.O. Service contact is available through a bipolar cable out of the pump. Once refilled the tank or recovered the level input, the pump starts to dose for the remaining time missing to complete the set work time once stopped.
8.1 Models FPDR PAUSE WORK Constant dosage pump with pause and operating phase. The dosage can be adjusted using the PAUSE knob (0 to 6 minutes) and the OPERATION knob (0 to 10 seconds). Press ON / OFF to switch the pump on or off.
10. Troubleshooting PROBLEM POSSIBLE CAUSE Pump isn’t powered. Connect it to main supply. Pump doesn’t turn Pump’s protection fuse is broken. Replace it. See page 28 for replacement procedure. Pump’s main board is broken. Replace it. See page 28 for replacement procedure. The foot filter is obstructed. Clean it. Suction hose is empty.
11. Fuse and main board replacement Fuse or main board replacement is allowed to qualified personnel only. Before to operate disconnect the pump from main power and all hydraulic connections. For fuse replacement is necessary to use a 3x16 and 3x15 screwdriver and a new fuse (same model of old one). For main board replacement is necessary to use a 3x16 and 3x15 screwdriver and a new main board (same model of old one). Fuse replacement procedure: - Turn pump’s injection knob on 0%. - Remove 6 screws from pump’s back. - Pull pump’s back cover until it’s completed separated from pump’s front. Be careful of the knob’s spring. - Locate the blown fuse and replace it. - Reassemble the pump. Be careful to put back the knob’s spring. - Reinsert screws. Main board replacement procedure: - Turn pump’s injection knob on 0%. - Remove 6 screws from pump’s back. - Pull pump’s back cover until it’s completed separated from pump’s front. Be careful of the knob’s spring. - Remove board’s screws. - Completely disconnect wires from main board and replace it. Reinsert screws. - Reconnect wires to the main board (see enclosed picture). - Reassemble the pump. Be careful to put back the knob’s spring.
12. Main board Power supply Level Solenoid Solenoid Power supply Mod. FCO Mod. FCL Level Input Signal Water Power Power meter Level supply supply Solenoid Solenoid Level Input Signal Mod. FIC/FIS Mod. FPV/FPVM Level input Pulse input Solenoid Power Level signal Power Supply Supply Solenoid Service Signal N.O.* Mod. FPDR Mod. FTE * Service Signal N.O.: optional...
A Appendix. Maintenance. In order to ensure the requirements of potable drinking water treated and the Maintenance schedule maintenance of the improvements as declared by the manufacturer, this equipment must be checked at least once a month. OPERATOR PROTECTION Use safety equipment according to the company regulations. Use this safety equipment within the work area during installation, service and when handling chemicals: •...
Page 31
f the pump performance does not satisfy your process requirements, and the process requirements have not changed, then perform these steps: 1. Disassemble the pump. 2. Inspect it. 3. Replace worn parts. This procedure SHOULD BE CARRIED OUT BY AUTHORIZED AND QUALIFIED Shutdown procedure PERSONNEL...
C Appendix. Delivery Curves PCS pump head PCS pump head 01 09 05 07 l/h 07 l/h 09 ø ø Magnete Magnete bar 01 bar 05 PCS pump head C pump head 06 06 05 10 l/h 06 l/h 10 ø75 ø Magnete Magnete bar 06 bar 05 PCS pump head PCS pump head 05 05 07 03 l/h 05 l/h 03 ø ø Magnete Magnete bar 05 bar 07 PCS pump head PCS pump head 10 2...
C Appendix. Delivery Curves PCS pump head PCS pump head 07 05 01 07 l/h 05 l/h 07 ø70 ø60 Magnete Magnete bar 07 bar 01 PCS pump head PCS pump head 10 05 05 0,2 l/h 0,2 l/h 05 ø60 ø75 Magnete Magnete bar 05 bar 10 PCS pump head C pump head 03 11 03 6,5 l/h 11 l/h 6,5 ø60 ø75 Magnete Magnete bar 03 bar 03 PCS pump head 03 8,5 Flow rate indicated is for H...
E Appendix. Chemical Compatibility Table Solenoid driven metering pumps are widely used to dose chemical fluids and it is important that the most suitable material in contact with fluid is selected for each application. This compatibility table serves as a useful help in this respect. All the informations in this list are verified periodically and believed to be correct on the date of issuance.
G Appendix. “F” and “FA” series pump exploded view...
Page 42
MOD 7.5 B1 Q Ed. 1 - rev. 0 21/02/2012 PRODUCT SERVICE REPAIR FORM ENCLOSE THE PRESENT FORM TO THE DELIVERY NOTE DATE ..........SENDER Company name ............................Address ..............................Phone no..............................Contact person............................. PRODUCT TYPE (see product label) DEVICE CODE ..............................
SUMMARY GENERAL SAFETY GUIDELINES ................... 2 PURPOSE OF USE AND SAFETY ................... 3 ENVIRONMENTAL SAFETY ..................4 LABELS ........................4 SPARE PARTS ......................4 1. Introduction ......................5 2. Unpacking ......................6 3. General description ....................7 4. Before to Install warnings ..................8 5.
Page 44
Disposal of end-of-life equipment by users This symbol warns you not to dispose of the product with normal waste. Respect human health and the environment by giving the discarded equipment to a designated collection center for the recycling of electronic and electrical equipment. For more information visit the online site.
Need help?
Do you have a question about the F Series and is the answer not in the manual?
Questions and answers