8 TECHNICAL INFORMATION
8.1 Respiro dimensions and weight
Height: approx. 81 mm.
Width: approx. 46 mm.
Depth: approx. 19 mm.
Weight: approx. 48 g.
8.2 Device IP classifications
Respiro: IP67
Charging dock: IP31
Pulse oximeter: IP33
8.3 Operating and storage conditions
Device
Storage temperature
range
Respiro
Pulse oximeter
Accessories
Pressure: 700 hPa-1060 hPa, operating.
Battery charging: Battery manufacturers restrict charging the battery above the defined battery tempera-
ture limit in order to avoid overheating the battery and to ensure safe user experience in all conditions. As
the device manufacturer, Bittium recommends to ensure that Respiro charging environment temperature
is max.+30°C to enable smooth and uninterrupted battery charging.
Always transport the equipment in the carry case. Protect the carry case from snow and rain. Remove the
batteries from the pulse oximeter when storing it.
D0000009279 Operating Instructions for healthcare professionals
Version 2.0
Table 8 Operating and storage conditions
Operating temperature
range
- 25... + 70 °C
+10...+ 30 °C
Humidity
+5... + 40 °C
Operating 15...90 % (non-
condensing)
Storage 10...90 % (non-
condensing)
© Bittium Biosignals Ltd 2023