Prius Healthcare USA Salute RDX User Manual

3-1 alternating anti-decubitus system
Hide thumbs Also See for Salute RDX:

Advertisement

Quick Links

Salute RDX
3-1 Alternating
Anti-Decubitus System
User Manual
Prius Healthcare USA
160 Scarlet Blvd.
Oldsmar, FL 34677, USA
TEL: (813)854-5464
FAX: (813)854-5442
文件中心
2023.04.05
正式發行
2303

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Salute RDX and is the answer not in the manual?

Questions and answers

Summary of Contents for Prius Healthcare USA Salute RDX

  • Page 1 Salute RDX 3-1 Alternating Anti-Decubitus System User Manual Prius Healthcare USA 160 Scarlet Blvd. Oldsmar, FL 34677, USA TEL: (813)854-5464 FAX: (813)854-5442 文件中心 2023.04.05 正式發行 2303...
  • Page 2 文件中心 2023.04.05 正式發行...
  • Page 3: Table Of Contents

    Content 1. The Purpose of this Manual 2. Product Description Master Control Unit Features Mattress Features 3. Technical Data Master Control Unit Mattress Replacement Symbol Definition 4. Instruction for Proper Use 5. Cleaning The Mattress The Master Control Unit Replace Air Filter Waste Disposal IP21 6.
  • Page 4 14. Power cable & pump shall be placed at the foot-side of the patient to prevent any risk of strangulation due to cable. 15. Please ensure the Salute RDX Alternating Pressure System is used with stable power or in connection with UPS.
  • Page 5: The Purpose Of This Manual

    2. Product Description Salute RDX 3-1 Alternating Anti-Decubitus System is a pressure ulcer prevention mattress replacement system designed for use in the home, at a nursing facility or in a hospital environment. Use of this system on a standard bed frame is possible; however the use of this system on a bed frame designed for a healthcare environment is preferred.
  • Page 6: Mattress Features

    Alarm Mute allows you to mute the alarm while corrective action is being taken to determine the cause of alarm. Salute RDX Air Mattress Therapy System is recommended for use in the prevention and treatment of decubitus (pressure) ulcers stage I – III (medium risk).
  • Page 7: Technical Data

    文件中心 2023.04.05 3. Technical Data 正式發行 Master Control Unit Model Name Salute RDX Control Unit Model No. FC-PHR0032 Size (inch) LxWxH 13.5“ x 5.4“ x 8.3“ Weight(lbs) Cycle Time (min) 5, 10, 15, 20 min Min Operating Pressure 12 +/- 6mmHg...
  • Page 8: Symbol Definition

    Please make sure the designated Salute RDX control unit is connected with the designated Salute RDX mattress before use. Symbol Definition Type BF Protection Against Electronic Class II Equipment Shock Refer To Instruction Manual / Booklet Waste Disposal NOTE Background colour: blue...
  • Page 9 . The inflation time of the mattress can take 30 to 40 minutes. For quicker inflation a portable blower unit is available. Please contact Prius Healthcare USA for additional information, (813) 854-5464. 11. When the mattress is fully inflated , set the dial in accordance with the patient’s size and weight.
  • Page 10 Alarm Function The Salute RDX 3-1 Alternating Anti-Decubitus System is equipped with a visual and audible alarm in the event of low pressure. During the initial inflation period the system is in low pressure mode and the low pressure LED will illuminate.
  • Page 11: Cleaning

    5. Cleaning The Mattress The mattress should be cleaned weekly using a damp soft cloth and mild detergent. If the top cover or base cover becomes excessively soiled, put on clean gloves, plastic gown and eye protection and remove top and base covers from the air cells and tubing. Place soiled covers in an appropriate container in accordance with your facilities standard operating procedures for contaminated waste, replace mattress with clean covers.
  • Page 12: Replace Air Filter

    Replace Air Filter Remove air filter cover to access air filter. Pinch the center of air filter and pull outward, replace with a new filter. If air filter opening is soiled use a soft bristle brush to remove dust and dried-on soil.
  • Page 13: Storage And Care

    6. Storage and Care Master Control Unit: ● Check the power cord and plug for abrasions or excessive wear. ● Plug in the control unit and verify air flows from the units hose connection ports. ● Place in plastic bag for storage. Overlay Mattress: ●...
  • Page 14: Maintenance And Troubleshooting

    7. Maintenance & Troubleshooting Daily maintenance is not required. Caregivers should check the patient and control unit setting every two hours. This equipment should only be serviced by a qualified and authorized technician. For common troubleshooting tips please refer to the chart below. Symptom Inspection Procedures Possible Solution...
  • Page 15: Emc Related Notifications

    The Salute RDX is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the Salute RDX should assure that it is used in such an environment. Emission test Compliance...
  • Page 16 The Salute RDX is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the Salute RDX should assure that it is used in such an environment. Electromagnetic IEC 60601...
  • Page 17 The Salute RDX is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the Salute RDX should assure that it is used in such and environment. Immunity test IEC 60601 test level...
  • Page 18 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment The Salute RDX is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the Salute RDX should assure that it is used in such an environment. Compliance LEVEL...
  • Page 19: Expected Service Life

    9. Expected Service Life For maintain basic safety and essential performance in regards to EMC, the Salute RDX pump has an expected service life of five years. To maintain the condition of the pump, have the pump serviced regularly according to the schedule recommended by Prius Healthcare.
  • Page 20 Distributed by: Prius Healthcare USA 160 Scarlet Blvd. Oldsmar, FL 34677, USA TEL: (813)854-5464 FAX: (813)854-5442 AL300274 V1.05 文件中心 2023.04.05 正式發行...