Prius Healthcare USA Rhythm Multi User Manual

Alternating and low air loss pressure relief system

Advertisement

文件中心
正式發行
2020.10.20
文件中心
2022.09.30
正式發行
Rhythm Multi
Low Air Loss Pressure Relief System
User Manual
Prius Healthcare USA
160 Scarlet Blvd.
Oldsmar, FL 34677, USA
TEL: (813)854-5464
FAX: (813)854-5442
Alternating
and
2210

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Rhythm Multi and is the answer not in the manual?

Questions and answers

Summary of Contents for Prius Healthcare USA Rhythm Multi

  • Page 1 文件中心 2020.10.20 正式發行 Rhythm Multi Alternating Low Air Loss Pressure Relief System User Manual Prius Healthcare USA 160 Scarlet Blvd. Oldsmar, FL 34677, USA TEL: (813)854-5464 FAX: (813)854-5442 文件中心 2210 2022.09.30 正式發行...
  • Page 2: Table Of Contents

    3-1 Alternation Cycle Illustration Master Control Unit Features Mattress Features 3. Technical Data ................. 4 Master Control Unit Rhythm Multi Mattress Replacement Symbol Definition 4. Instruction for Proper Use ............5 5. Cleaning .................. 9 The Mattress The Master Control Unit...
  • Page 3 IMPORTANT SAFEGUARDS 文件中心 2020.10.20 正式發行 When using electrical products, especially when children are present, basic safety precautions should always be followed, including the following: CAUTION - READ ALL INSTRUCTIONS BEFORE USING THE APPLIANCE WARNING – To reduce the risk of electrocution: 1.
  • Page 4: The Purpose Of This Manual

    System. We recommend you keeping this manual handy to answer most of the question related to the system. 2. Product Description The Rhythm Multi Alternating and Low Air Loss Pressure Relief System is a unique and innovative specialized mattress replacement unit. The system utilizes low air loss technology with a high flow rate that provides pressure management for the treatment of pressure ulcers.
  • Page 5: Master Control Unit Features

    Master Control Unit Features 文件中心 2020.10.20 正式發行 • Alternating time can be adjusted in 3 min increments to 95 min. In addition, the caregiver can select the “Static Function to stop the alternation function and provide only low air loss therapy. •...
  • Page 6: Technical Data

    3. Technical Data 文件中心 2020.10.20 正式發行 Master Control Unit Model No. Rhythm Multi Control Unit Ref No. FC-PHR0025 Size (inch) 17.7” (L) x 6.8” (W) x 10.8” (H) Weight 12.8lbs (5.8 kg) Phase Time 3 ~ 95 minutes ≧35 mmHg...
  • Page 7: Symbol Definition

    Symbol Definition 文件中心 2020.10.20 正式發行 Type BF Protection Against Class II Equipment Electronic Shock Consult instructions for Waste Disposal Caution, Consult SGS product certification accompanying mark documents Date of Manufacture Batch code 4. Instructions for Proper Use 1. Remove the existing mattress from the bed frame. 2.
  • Page 8 9. Push the STANDBY/OPERATE button on the front panel. OPERATE LED 文件中心 2020.10.20 正式發行 will now be illuminated and the control unit will be in operation. 10.Push the AUTO FIRM button for fast inflation. Allow 4-7 min for full inflation. After the mattress is fully inflated, the caregiver can transfer the patient to the mattress.
  • Page 9 13.Pulsate Function: Push the Mode button to select pulsate mode and 文件中心 2020.10.20 正式發行 activate the pulsate function for additional pressure relief. Once the function is activated, the mattress automatically deflate for low air flow loss for about 1 minute, and after that, it will inflate to original press setting for one minute.
  • Page 10 16.The Master Control Unit is equipped with power failure alarm. This 文件中心 2020.10.20 正式發行 function enable the Control unit to generate a horning sound to notify the caregiver the main power failure. The alarm can be disabled by pushing the Alarm Reset Button on the front panel. Caution: Immediate response by the operator is required during power failure.
  • Page 11: Cleaning

    文件中心 2020.10.20 正式發行 Suggestions Ensure that there are no kinks in the hoses and the connectors are properly locked. The parts and accessories supplied are specifically designed for use with this control unit. Use other products in conjunction with the system is not recommended Caution Air outlet label is where Blower Exhausts, do not touch the blower exhaust...
  • Page 12: The Master Control Unit

    The Master Control Unit 文件中心 2020.10.20 正式發行 CAUTION SWITCH OFF THE ELECTRICAL SUPPLY TO THE PUMP AND DISCONNECT THE POWER CORD FROM THE MAIN SUPPLY BEFORE CLEANING AND INSPECTION The master control unit should be cleaned weekly by using a soft damp cloth and mild detergent.
  • Page 13: Storage And Care

    IP21 文件中心 2020.10.20 正式發行 The IP Code (or International Protection Rating, sometimes also interpreted as Ingress Protection Rating) consists of the letters IP followed by two digits and an optional letter. ⚫ First Digit: Solids The first digit indicates the level of protection that the enclosure provides against access to hazardous parts (e.g., electrical conductors, moving parts) and the ingress of solid foreign objects.
  • Page 14: Maintenance And Troubleshooting

    7. Maintenance & Troubleshooting 文件中心 2020.10.20 正式發行 Daily maintenance is not required. This equipment is intended to be serviced by qualified, authorized technical personnel. In case of minor troubleshooting issues, please refer to the following Troubleshooting table. Caregivers should check on the patient and the control unit setting once in every two hours.
  • Page 15: Emc Related Notification

    The Rhythm Multi Control Unit is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the Rhythm Multi Control Unit should assure that it is used in such an environment.
  • Page 16 The Rhythm Multi Control Unit is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the Rhythm Multi Control Unit should assure that it is used in such an environment.
  • Page 17 The Rhythm Multi Control Unit is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the Rhythm Multi Control Unit should assure that it is used in such and environment.
  • Page 18 The Rhythm Multi Control Unit is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the Rhythm Multi Control Unit should assure that it is used in such an environment.
  • Page 19: Expected Service Life

    EMC information provided. • Do NOT use unapproved accessories or attempt to modify, disassemble or otherwise misuse the Rhythm Multi Alternating and Low Air Loss Pressure Relief System or any of its components.
  • Page 20 2020.10.20 文件中心 正式發行 Prius Healthcare USA 160 Scarlet Blvd. Oldsmar, FL 34677, USA TEL: (813)854-5464 FAX: (813)854-5442 AL300580 V1.00 文件中心 2022.09.30 正式發行...