Prius Healthcare USA Rhythm Turn User Manual

Prius Healthcare USA Rhythm Turn User Manual

Low air loss and dynamic turning mattress replacement system
Hide thumbs Also See for Rhythm Turn:

Advertisement

Quick Links

Rhythm Turn
Low Air Loss
and
Dynamic Turning
Mattress Replacement System
User Manual
Prius Healthcare USA
580 Corporate Center
4027 Tampa Road, #3200
Oldsmar, FL 34677, USA
TEL:(813)854-5464
FAX: (813)854-5442
1103

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Rhythm Turn and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Prius Healthcare USA Rhythm Turn

  • Page 1 Rhythm Turn Low Air Loss Dynamic Turning Mattress Replacement System User Manual Prius Healthcare USA 580 Corporate Center 4027 Tampa Road, #3200 Oldsmar, FL 34677, USA TEL:(813)854-5464 FAX: (813)854-5442 1103...
  • Page 2: Table Of Contents

    2. Product Description ....................1 Master Control Unit Features Mattress Features 3. Technical Data ......................2 Master Control Unit Rhythm Turn Mattress Replacement Symbol Definition 4. Instruction for Proper Use ..................3 5. Cleaning .......................... 6 The Mattress The Master Control Unit...
  • Page 3 Warning  Connect the Master Control unit to a proper power source.  Don’t use the system in the presence of any flammable gases. ( such as Anesthetic Agents)  Keep the blower and mattress away from source of liquid of open flame. ...
  • Page 5: The Purpose Of This Manual

    2. Product Description The Rhythm Turn mattress system is a specialized, unique and innovation mattress replacement system. The system is primarily designed for at risk patients or step-down intensive care units.
  • Page 6: Technical Data

    Operation Humidity 30% ~75% Mode of Operation Continuous IEC 60601-1, Standard CAN/CSA C22.2 No. 601.1, IEC 60601-1-2 Rhythm turn Mattress Replacement Model No FM-PHR0013 Size ( L x W x H ) 80” x36” x 10” Weight 37.5 lbs Cells Material...
  • Page 7: Instruction For Proper Use

    4. Instruction for Proper Use 1. Remove the existing mattress from the bed frame. 2. Replace the standard mattress with Low Air Loss Turning Mattress Replacement system and make sure the mattress is positioned so the air tube is at the foot of the bed.
  • Page 8 11. Turning time can be adjusted by the CYCLE button. The time can be adjusted from 3 minutes to 95 minutes. (When Static Function is selected, the timing window would not show any digits). 12. The Master Control Unit is equipped with power failure alarm. With this function enabled, the Control unit generates a horn sound to signal to caregiver that main power failure.
  • Page 9 15. 40 degrees turning function can be activated by pressing the “BT” button and follow the operation instruction on step 14. Mattress Turning Illustration...
  • Page 10: Cleaning

    5. Cleaning The Mattress The mattress should be cleaned weekly using a damp soft cloth and mild detergent. If top sheet (Top cover) or base (Bottom cover) becomes overly soiled, put on clean gloves, plastic gown and eye protection before removing top sheet or base and dispose of according to standard in function control procedures.
  • Page 11: Replace Air Filter

    Replace Air Filter 1. Remove air filter and replace with a new filter. 2. Use a soft bristle brush to remove dust and difficult dried-on soil. Filter Waste Disposal This Product has been supplied from an environmentally aware manufacturer that complies with the WEEE. This product may contain substances that can be harmful to the environment if disposed of in places that are not approved by your state, local or federal laws.
  • Page 12: Maintenance And Troubleshooting

    7. Maintenance & Troubleshooting No daily maintenance is required. It is intended this equipment should only be serviced by properly qualified, authorized technical personnel. In case of minor trouble please refer as following Troubleshooting. Symptom Inspection Procedure Possible Solution Air is blowing out from the control 1.
  • Page 13: Emc Related Notification

    8. EMC Related Notifications Guidance and manufacturer’s declaration – electromagnetic emissions The blower is intended for use in the electromagnetic environment specified below. The customer or the user of the blower should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment –...
  • Page 14 Guidance and manufacturer’s declaration – electromagnetic immunity The blower is intended for use in the electromagnetic environment specified below. The customer or the user of the blower should assure that it is used in such an environment. Immunity test IEC 60601 Compliance level Electromagnetic environment –...
  • Page 15 Guidance and manufacturer’s declaration – electromagnetic immunity The blower is intended for use in the electromagnetic environment specified below. The customer or the user of the unit should assure that it is used in such an environment. Immunity IEC 60601 test Compliance Electromagnetic environment –...
  • Page 16: Warranty

    9. Warranty  Prius Healthcare guarantees this equipment to be free from defects in material and workmanship for up to 12 months from the date of delivery.  All warranty work will be performed at the service address below, shipping charges prepaid.
  • Page 17 Prius Healthcare USA 580 Corporate Center 4027 Tampa Road, #3200 Oldsmar, FL 34677, USA TEL:(813)854-5464 FAX: (813)854-5442 AL300104 V1.00...

This manual is also suitable for:

Fc-phr0010

Table of Contents