Page 2
01458 Ottendorf-Okrilla Germany Legal notice This work is protected by copyright. Endress+Hauser SICK GmbH+Co KG retains the rights conferred by it. Reproduction of this work or parts thereof is only permitted within the limits of the statutory provisions of copyright law.
Page 3
3.11 Filter fan ......................36 3.12 Instrument air conditioning (option).............38 8030592/V2-0/2023-07| Endress+Hauser SICK O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 4
Spare parts list ..................49 Disposal..................... 51 O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 5
Other relevant publications of analyzer system components, see “Other relevant technical documentation/information”, page 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 6
● over with it in its entirety. O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 7
Container ventilation monitoring (l) = optional ● 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 8
Illustrations may differ from the actual design. ● O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 9
All information in this system description must be observed. 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 10
O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser...
Page 11
For more information: Circuit diagram ● 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 12
Refers to possible damage to property and materials. Note Important information and useful notes. O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 13
Corrosive substances warning Oxidizing materials warning Gas bottle warning 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 14
Put on head protection Put on breathing protection O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 15
Risk to health Hazardous to the aquatic environment 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 16
▸ Put on protective clothing if necessary. O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 17
Carefully open the probe cover and air out the housing with ambient air. 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 18
▸ Put on breathing protection if necessary. O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 19
8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 20
Fig. 1: Warning label on the analyzer cabinet door: DANGER - Hazardous Gases! O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser...
Page 21
● FIDOR. Operation of the FIDOR. ● 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 22
The two variants also differ due to the safety devices. O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 23
8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 24
25% LEL. – The ventilation fails or experiences a fault O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser...
Page 25
SICK operating instructions: GMS800 FIDOR hydrocarbon analyzer (FID) ● 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 26
Outside display Filter fan with louvered grills for ambient air intake O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 27
Option - air flow monitor Analyzer MCS200HW 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 28
Basic Control Unit (BCU) with display for FIDOR Analyzer FIDOR O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 29
Overview 4 - Inside right side view Legend 16 Test gas valves 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 30
SICK operating instructions - heated sample gas line type ELH ● Circuit diagram ● O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 31
Measured values are output in mg/Nm³ or percent by volume (% by vol). ● 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 32
SICK operating instructions - MCS200HW multi-component analyzer system ● Circuit diagram ● O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 33
SICK operating instructions: GMS800 FIDOR hydrocarbon analyzer (FID). ● 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 34
13. BImSchV and TA Luft ● 17. BImSchV. ● O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 35
SICK additional operating instructions - BCU control unit ● Circuit diagram ● 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 36
Press MEAS pushbutton for display of measured values Function buttons (BACK, ENTER, MENU, etc.) O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser...
Page 37
The following documents contain more information: Circuit diagram ● 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 38
Filter fan Air flow sensor (Stand Alone variant) O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 39
Mounting, installation and operating instructions of the filter fan ● Circuit diagram ● 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 40
Data sheet - VarioDry FRL SF 0010 - SF 0150 membrane dryer ● O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser...
Page 41
Stops H supply at H =25% LEL. 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 42
Data sheet - gas measurement and warning systems – GMC 8022. ● O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser...
Page 43
A test label on the detector specifies the date by which the device must be tested. 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 44
After a 25% LEL alarm, the alarm and the horns are switched off when an operator presses the reset pushbutton by hand. O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser...
Page 45
Operating instructions of the associated components ● 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 46
Operating instructions for the F2, F3, F4.2, F4.3 flow sensors ● O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 47
Operating instructions for the F2, F3, F4.2, F4.3 flow sensors ● 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 48
The following documents contain more information: Circuit diagram. ● O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 49
By carrying out a system test, the function test is not required, as this is included in the sys- tem test. 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 50
Periodic inspection by qualified persons 3 years O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 51
In case of doubt, contact SICK Service. 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...
Page 52
SPARE PARTS LIST O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR 8030592/V2-0/2023-07| Endress+Hauser Subject to change without notice...
Page 53
For detailed information to the disposal of analyzer modules, refer to the respec- tive Operating Instructions. 8030592/V2-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | MCS200HW & GMS800 FIDOR Subject to change without notice...