INSTRUCTIONS FOR USE
for the medical device Propeller
Instructions for use and care
Translation of Instructions for use
Item number: 23-11231-UK
Revision: 00
Valid from: 2023-05-12
Amended: -
UK

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Propeller and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Mercado Medic Propeller

  • Page 1 INSTRUCTIONS FOR USE for the medical device Propeller Instructions for use and care Translation of Instructions for use Item number: 23-11231-UK Revision: 00 Valid from: 2023-05-12 Amended: -...
  • Page 2 [Space for product label with serial number]...
  • Page 3: Table Of Contents

    9.1. Recycling ..... . . 26 3.1. Propeller ......9 3.2.
  • Page 4: Introduction

    We manufacture all our different medical chairs under our own management in Sweden. Mercado Medic is certified to ISO 13485, ISO 9001 and ISO 14001, and complies with applicable labour and environmental legislation.
  • Page 5: Contact Details

    Instruction manual for the use of medical work chairs Introduction 1.1. Contact details 1.2. Warranty Manufacturer All components of Mercado Medic's devices Street address: Mercado Medic AB, are covered by warranty, under the time Tryffelslingan 14, SE-181 57 Lidingö, Sweden periods and general terms and conditions...
  • Page 6: Getting Started

    Figure 2.1). The charging socket is located on the corresponding control button has the Propeller’s control box, which is mounted been released, the movement can be under the right or left armrest cushion. interrupted by pressing the button to stop movement in the opposite direction.
  • Page 7 Charging outlet Brake lever Figure 2.1. Activity chair equipped with the Propeller. 23-11231-UK-00...
  • Page 8: Charging

    2. First, insert the plug into the charging front wheels. outlet (see Figure 2.2). The charging outlet is located on the Propeller control Warning! The product must be box under the armrest cushion. handled with care to prevent hands 3.
  • Page 9: Caring For The Device

    PLEASE NOTE! Do not wash the device with water, other liquids or chemicals. 3.1. Propeller Do not clean the propeller actuator (see Figure 3.1). This is because there is a risk that the grease required for this mechanism to function as intended might be removed during cleaning.
  • Page 10: Intended Use

    Contraindications in Section 3. Caring for the device. There are no known contraindications for using REAL 9000 PLUS with Propeller. If the device is purchased without prescription from qualified healthcare professionals, the user should consult their doctor whether there are any contraindications.
  • Page 11: Configurations And Settings

    5.2. Basic model The activity chair with Propeller always comes with a seat, back support, armrest, base with four castors (Ø100 mm as standard) and manual brake with brake lever (350 mm as standard). Depending on the model, the device has either electric or manual height adjustment.
  • Page 12: Electrical Functions

    The Propeller is equipped with two electrical functions: an actuator and a drive motor. The actuator raises and lowers the drive motor between active and passive positions. The drive motor assists with the movement of the activity chair. The Propeller is driven forward by a control button.
  • Page 13: Important Safety Information

    The Propeller complies with protection rating with Regulation 1. The active properties of IP22. The first digit (2) means protection the Propeller are not deemed to affect the against electric shock for fingers, etc., by classification. preventing access to hazardous parts with fingers or similar.
  • Page 14: Warnings

    Medic may no longer have full carried out by personnel authorised product liability. Modifications may by Mercado Medic. If this is not affect the safety of the device and followed, the CE marking no longer lead to accidents.
  • Page 15: Expected Service Life

    Figure 6.1. Serial number, UDI-DI and date of manufacture on the label. Warning! The Propeller is not a PLEASE NOTE! If the serial number label brake, use the brake lever. and the date of manufacture are damaged...
  • Page 16: Reporting Of Adverse Events And Incidents

    Servicing and reconditioning, in addition to the overall assessment below. As a manufacturer of medical devices in the EU, Mercado Medic is obliged to have a After the expected service life, Mercado system to monitor how our devices work in Medic cannot guarantee the suitability and practical use.
  • Page 17: Technical Information

    Warning! Repairs and other technical measures may only be carried out by personnel authorised by Mercado Medic. If this is not followed, the CE marking no longer applies and Mercado Medic may no longer have full device liability.
  • Page 18: Symbols

    Instruction manual for the use of medical work chairs Technical information 7.1. Symbols The following symbols are used on the device controls, in the labelling or in this user manual. For an overview of the location of symbols, see Figures 7.1–7.2. No.
  • Page 19 Instruction manual for the use of medical work chairs Technical information 2-12 Figure 7.1. Overview of labelling symbols. 13-16 Figure 7.2. Overview of labelling symbols. 23-11231-UK-00...
  • Page 20: Servicing And Reconditioning

    Medic. If this is not followed, the device shows reduced or altered CE marking no longer applies and performance. A device showing Mercado Medic may no longer have reduced or altered performance full device liability. must immediately be removed from use to prevent incidents and accidents.
  • Page 21: Troubleshooting Guide

    8.2. Battery replacement Warning! Repairs and other technical measures may only be carried out by personnel authorised by Mercado Medic. If this is not followed, the CE marking no longer applies and Mercado Medic may no longer have full device liability.
  • Page 22: Reconditioning And Periodic Maintenance

    Long-term battery storage serial numbers are retained and expected The battery in the Propeller does not need to service life remains unaffected. The purpose be disconnected. In the event of prolonged of this reconditioning may be, for example, to...
  • Page 23 Check and verify the battery’s voltage level. If the voltage is low after charging, the battery needs to be replaced (see assembly instructions for the Propeller on www.mercado.se. Remember that the expected service life of the batteries is five (5) years if these are maintained according to this user manual.
  • Page 24 Listen for abnormal sounds in the drive motor. If abnormal sounds are heard, replace the drive motor. Bracket Check that the drive motor’s attachment to the Propeller is properly tightened. Replace any screws that have damaged sockets or threads. Cabling Inspect cables for any signs of wear or damage caused by pinching.
  • Page 25 Bracket Check that the Propeller’s attachment to the base is correctly assembled and tightened. Replace any screws that have damaged sockets or threads.
  • Page 26: Disposal Instructions

    Warning! Installation, connection or dismantling is not risk-free. If components are handled incorrectly, for example, crushing damage may occur. This type of work may therefore only be carried out by a Mercado Medic authorised technician. 9.1. Recycling Instructions for recycling the components used.
  • Page 27 Instruction manual for the use of medical work chairs Disposal instructions This page has been deliberately left blank 23-11231-UK-00...