Page 1
TRACHEAL INTUBATION FIBERSCOPE OLYMPUS LF-TP OLYMPUS LF-DP OLYMPUS LF-GP Refer to the endoscope’s companion manual, the “OLYMPUS ENF/LF ENDOSCOPE REPROCESSING MANUAL” for reprocessing information. USA: CAUTION: Federal law restricts this device to sale by or on the order of a physician.
Preparation, Inspection and Connection of Ancillary Equipment ..Inspection of the Endoscopic System ..........Chapter 4 Operation ..............Insertion ..................Withdrawal of the Endoscope ............Transportation of the Endoscope............ Disposal ..................Chapter 5 Troubleshooting ............Troubleshooting Guide..............Returning the Endoscope for Repair..........LF-TP/DP/GP OPERATION MANUAL...
Refer to instructions. Caution Do not reuse. Lot number Type BF applied part Endoscope Lamp Caution, Hot Surface Position of Battery Power On Power Off Manufacturer Authorized representative in the European Community Serial number LF-TP/DP/GP OPERATION MANUAL...
Note that the complete instruction manual set for this endoscope consists of this manual and the “OLYMPUS ENF/LF ENDOSCOPE REPROCESSING MANUAL” which also accompanied the endoscope at shipment.
This instrument was not disinfected or sterilized before shipment. Before using this instrument for the first time, reprocess it according to the instructions given in the endoscope’s companion manual, the “OLYMPUS ENF/LF ENDOSCOPE REPROCESSING MANUAL”. After using this instrument, reprocess and store it according to the instructions given in the endoscope’s companion manual, the “OLYMPUS ENF/LF...
Otherwise, patient injury, bleeding and/or perforation can result. It may also become impossible to straighten the bending section during use and/or to withdraw this instrument from the patient. LF-TP/DP/GP OPERATION MANUAL...
Page 10
It may also cause patient and/or operator injury. • Be sure to prepare another endoscope to avoid that the examination be interrupted due to equipment failure or malfunction. LF-TP/DP/GP OPERATION MANUAL...
Bending Section. • The endoscope has not been designed for use in retroflexed observation. Performing retroflexed observation in a narrow lumen may make it impossible to straighten the bending section and/or withdraw the endoscope from the patient. LF-TP/DP/GP OPERATION MANUAL...
Dispose of used batteries as general waste according to local regulations. Body oils, dirt or dust on the battery terminals can cause poor contact. Wipe batteries carefully with a dry cloth before use. LF-TP/DP/GP OPERATION MANUAL...
Match all items in the package with the components shown below. Inspect each item for damage. If the instrument is damaged, a component is missing or you have any questions, do not use the instrument; immediately contact Olympus. This instrument was not disinfected or sterilized before shipment. Before using this instrument for the first time, reprocess it according to the instructions in the endoscope’s companion manual, the “OLYMPUS ENF/LF ENDOSCOPE...
Page 14
Chapter 1 Checking the Package Contents Endoscope Channel-opening Cleaning Brush (MH-507) Channel cleaning Brush (BW-15B for LF-TP) Suction Connector Cleaning Brush (BW-15SH for LF-DP/GP) Channel Cleaning Brush 3 (BW-7B for LF-DP/GP) Light guide adapter (MAJ-1413) 1 Single-use Biopsy Valve (MAJ-210, 20 pcs) Single Use Single-ended Cleaning 3...
Chapter 2 Instrument Nomenclature and Specifications Chapter 2 Instrument Nomenclature and Specifications Nomenclature Battery Holder Cap Attachment Ring CE marking Crossed-out wheeled bin’ symbol Socket 8. Miniature Light Source (MAJ-524, Optional) Lamp Holder LF-TP/DP/GP OPERATION MANUAL...
Page 17
Light Guide 5. Suction Channel 3. Suction Valve (MD-493) 2. UP/DOWN Angulation Suction Valve Housing Control Lever Suction Lever Serial Number Control Section ETO Cap (MB-156) Venting Connector Boot Insertion Tube/Working Length Distal End 1. Bending Section LF-TP/DP/GP OPERATION MANUAL...
Page 19
Distal End Outer Diameter ø 3.1 mm Insertion Tube Outer Diameter ø 3.1 mm Working Length 600 mm Suction Channel Inner Diameter ø 1.2 mm Channel Bending Angulation Range UP 120, DOWN 120 Section Total Length 855 mm LF-TP/DP/GP OPERATION MANUAL...
Page 20
93/42/EEC concerning medical devices. Classification: Class II a Year of manufacture 1912345 The last digit of the year of manufacture is the second digit of the serial number. Degree of protection against electric TYPE BF applied part shock LF-TP/DP/GP OPERATION MANUAL...
Page 21
Electrical and Electronic Equipment, this symbol indicates that the product must not be disposed of as unsorted municipal waste, but should be collected separately. Refer to your local Olympus distributor for return and/or collection systems available in your country. LF-TP/DP/GP OPERATION MANUAL...
Should the slightest irregularity be suspected, do not use this instrument and see Chapter 5, “Troubleshooting”. If the irregularity is still suspected after consulting Chapter 5, contact Olympus. Damage or irregularity may compromise patient or user safety and may result in more-severe equipment damage.
Chapter 3 Preparation and Inspection Preparation and Inspection of the Endoscope Clean and disinfect or sterilize the endoscope as described in the “OLYMPUS ENF/LF ENDOSCOPE REPROCESSING MANUAL”. Inspection of the Endoscope Visually inspect the Control Section for excessive scratching. Visually inspect the Boot and the Insertion Tube near the Boot for bends, twists or other irregularities.
Page 25
Confirm that the Bending Section moves smoothly and correctly and that maximum angulation can be achieved. Move the UP/DOWN Angulation Control Lever slowly to its neutral position. Confirm that the Bending Section returns smoothly to an approximately straight condition. LF-TP/DP/GP OPERATION MANUAL...
Do not use an instrument after the expiration date displayed on the sterile package. Doing so may pose an infection control risk. • Do not attempt to sterilize the Biopsy Valve. This could pose an infection control risk or cause equipment damage. Slit Housing Figure 3.4 LF-TP/DP/GP OPERATION MANUAL...
Page 27
UK and all other countries where the Single-use Biopsy Valve (MAJ-210) is not included. • If using a valve which is provided non-sterile, sterilize it according to the Recommended ETO Gas Exposure Parameters given in the “OLYMPUS ENF/LF ENDOSCOPE REPROCESSING MANUAL”. LF-TP/DP/GP OPERATION MANUAL...
Page 28
It is recommended that you have a spare Miniature Light Source ready before use. If the Miniature Light Source is damaged or its battery runs out, you can continue the examination by replacing the Miniature Light Source with the spare. LF-TP/DP/GP OPERATION MANUAL...
Page 29
In case a pollutant enters the Miniature Light Source, remove the Battery Holder Cap and Lamp Holder according to Figure 3.15, and sterilize them together with the Main Body of the Light Source using ETO gas according to the “OLYMPUS ENF/LF ENDOSCOPE REPROCESSING MANUAL”. •...
(see Figure 3.9). Suction Valve Holder Figure 3.9 Turn the Suction Valve, with the Suction Valve Holder depressed, until it clicks into position (see Figure 3.10). Figure 3.10 Confirm that the Suction Lever can be operated smoothly. LF-TP/DP/GP OPERATION MANUAL...
Page 31
(see Figure 3.12). Mark#1 (red) Mark#2 (red) Attachment Ring Socket Light Guide Figure 3.12 While pushing the Miniature Light Source in the direction of the arrow shown in Figure 3.12, rotate the Attachment Ring clockwise until it stops. LF-TP/DP/GP OPERATION MANUAL...
Inspection of the Illumination To turn the lamp on, rotate the Miniature Light Source and align the “ON” mark with the red Mark#3 (see Figure 3.13). “ON” mark “OFF” mark Mark#3 (red) Not illuminated Illuminated Figure 3.13 LF-TP/DP/GP OPERATION MANUAL...
Page 33
Mark#3 to turn the switch off (see Figure 4.9). Replace the battery with a new one according to “Replacement of the Battery” on page 32. Inspect the image using the given above procedure. Turn the lamp off. LF-TP/DP/GP OPERATION MANUAL...
Page 34
If the Miniature Light Source is not attached to the endoscope, the lamp may be turned on when rotating the Lamp Holder to remove it. However, this is not a problem, the lamp will be turned off when the Lamp Holder is tightened. LF-TP/DP/GP OPERATION MANUAL...
Page 35
Electrical Contact Lamp Lamp Holder Figure 3.16 Prepare a new lamp. Put the lamp into the Lamp Holder and rotate it counterclockwise until it stops. Wipe the surface of the lamp tip using a clean, lint-free cloth. LF-TP/DP/GP OPERATION MANUAL...
Page 36
Remove the battery from the Miniature Light Source. Prepare a new battery. Insert it into the Battery Holder. Insert the (+) terminal first (see Figure 3.7). Attach the Battery Holder Cap and rotate it clockwise until it stops. LF-TP/DP/GP OPERATION MANUAL...
Page 37
Suction Lever is depressed during water feeding, water will be aspirated into the suction tube and not emitted from the endoscope’s Distal End. • If water is not emitted from the endoscope’s Distal End, flush air through the channel using the syringe. LF-TP/DP/GP OPERATION MANUAL...
If the endoscope cannot be withdrawn from the patient smoothly, do not attempt to forcibly withdraw it; leave it inside the patient and immediately contact Olympus. Forcibly withdrawing the endoscope may cause patient injury. •...
Lubricate the Insertion Tube with a medical-grade, water-soluble lubricant, taking care not to lubricate the Distal End. Pass the ETT over the Insertion Tube towards the Control Section (see Figure 4.2). Attach the ETT firmly to the Boot (tapered portion) of the Control Section. LF-TP/DP/GP OPERATION MANUAL...
Page 40
Bending Section is inside of the ETT. • Do not apply excessive force, when attaching the ETT to the Boot. Equipment damage may result. Some ETT models cannot be attached to the Boot. Boot Figure 4.2 LF-TP/DP/GP OPERATION MANUAL...
Page 41
Do not allow the Insertion Tube to bend in a radius of 10 cm or less. Insertion Tube damage can occur (see Figure 4.3). Must be 10 cm or more Figure 4.3 LF-TP/DP/GP OPERATION MANUAL...
Page 42
Olympus. Forcibly withdrawing the endoscope may cause patient injury. Operate the Angulation Control Lever as necessary to guide the Distal End for the insertion and observation.
Page 43
If the illumination unexpectedly turns dark or disappears, stop the examination immediately free the UP/DOWN Angulation Control Lever to the neutral position, and carefully withdraw the endoscope from the patient. Observe the endoscopic image at all times when operating the Angulation Control Lever. LF-TP/DP/GP OPERATION MANUAL...
Page 44
Chapter 4 Operation Observation With the Light Source • Only use the Olympus Light Guide Cable (WA03200A/WA03202A/A03200A/A3090/A3091/A3093/ A3290/A3291/A3293) with these instruments. If the another light guide cable is used, the Control Section of the endoscope may be hot and burns can be caused.
Page 45
Insert the Light Guide Connector into the output socket of the Light Source (see Figure 4.6). Turn the Light Source on, and adjust the brightness level according to the Light Source’s instruction manual. Light Source Light Guide Cable Figure 4.6 LF-TP/DP/GP OPERATION MANUAL...
Page 46
Before withdrawing the endoscope, set the Angulation Control Lever in the neutral position. Slowly withdraw the endoscope while observing the endoscopic image. Bronchoalveolar Lavage (LF-TP Only) When performing Bronchoalveolar Lavage, the BAL Kit or a syringe can be used. The syringe can be attached to the Single-use Biopsy Valve (MAJ-210) or Biopsy Valve (MD-495).
Page 47
Securely insert a syringe filled with lavage fluid (e.g. saline) into the Slit of the Biopsy Valve and press the plunger to feed lavage fluid (see Figure 4.8). With the syringe attached, slowly withdraw the plunger to aspirate lavage fluid (see Figure 4.8). Figure 4.8 LF-TP/DP/GP OPERATION MANUAL...
Slowly withdraw the endoscope while observing the endoscopic image. To turn the lamp off, rotate the Miniature Light Source and align the “OFF” mark with the red Mark#3 (see Figure 4.9). “ON” mark “OFF” Mark#3 mark (red) Illuminating Not illuminating Figure 4.9 LF-TP/DP/GP OPERATION MANUAL...
Attach the ETO Cap when transporting the endoscope, to avoid damage caused to the endoscope by changes in air pressure. Disposal When disposing of this product, or any of its components, follow all applicable national and local laws and guidelines. LF-TP/DP/GP OPERATION MANUAL...
Olympus for repair. Olympus does not repair accessory parts. If an accessory part becomes damaged, contact Olympus to purchase a replacement. Never use the endoscope on a patient if an abnormality is suspected.
Page 51
Install Biopsy or Syringe Valve installed. correctly. The Suction Lever is Valve is damaged. Replace it with a new Suction sticky. Valve. Suction Lever does not The aspiration pressure Lower the aspiration pressure. return to its original is excessive. position. LF-TP/DP/GP OPERATION MANUAL...
Olympus. Olympus is not liable for any injury or damage which occurs as a result of repairs attempted by non-Olympus personnel. When returning the endoscope for repair, contact Olympus. With the endoscope,...
Page 53
New products released after the introduction of this instrument may also be compatible for use in combination with this instrument. For further details, contact Olympus. If combinations of equipment other than those shown below are used, the full responsibility is assumed by the medical treatment facility.
Page 54
(MD-495) SCP-10 Adapter OES Instant Camera (AR-P10) (SCP-10) Light guide adapter Video Adapter (AR-TL08E/TF08E) Light guide cable Light guide cable OES Video System (WA03200A/WA03202A/A (A3090/A3091/A3093/ (OTV-S6/SC) 03200A) A3290/A3291/A3293) Video Adapter (AR-T10E/T12E) Video Adapter Video Coupler (AR-T10/T12/TZ2) (MH-999E) LF-TP/DP/GP OPERATION MANUAL...
Page 55
Appendix Cleaning and Disinfection Equipment Cleaning and Disinfection Equipment Single Use Single-ended Channel Cleaning Brush Channel Cleaning Brush 1 Cleaning Brush (BW-15B for LF-TP) (BW-7B for LF-DP/GP) 1 (BW-400B for LF-DP/GP) Suction Connector Channel-opening Cleaning Brush Cleaning Brush (BW-15SH for LF-DP/GP)