Page 1
Item ref: 600.110UK MTS01 Smart Digital Multimeter User Manual...
Page 2
Please read this manual thoroughly and ensure all contents are fully understood before using the apparatus. Warning To avoid possible electric shock or personal injury, and to avoid possible damage to the tester or to the equipment under test, adhere to these following rules: •...
Page 3
• The internal circuit of the meter shall not be altered at will to avoid damage of the meter and any accident. • A soft cloth and mild detergent should be used to clean the surface of the tester on a regular basis. No abrasive and solvent should be used to prevent the surface of the tester from corrosion or damage.
Low Battery Indication: “ ” displayed Range select: auto or manual Operation Temperature: 0°C to 40°C, less than 80%RH Storage Temperature: -10°C to 50°C, less than 85%RH Battery Type: 1.5V x 2, AAA size Dimension (H×W×D): 153×73×35mm Weight: Approx 175g with battery Technical Specifications Accuracy is guaranteed for 1 year 23°C±5°C less than 80%RH DC Voltage (auto ranging)
Page 5
Response: True RMS Max. Input voltage: 600V AC rms DC Current RANGE RESOLUTION ACCURACY 600mA 0.1uA ±(1.2% of rdg + 8 digits) 10mA Overload Protection: F10A/600V fuse Max. Input Current: 10A Note: For measurements>400mA: duration<10seconds, interval >15 minutes AC Current RANGE RESOLUTION ACCURACY...
Page 6
Overload Protection: 250V DC/AC rms Diode and Continuity RANGE INTRODUCTION REMARK The approx. forward Open circuit voltage drop will be Voltage: displayed about 3V The built-in buzzer Open circuit will sound if the Voltage: resistance is less about 3V than about 50Ω Overload protect: 250V DC/AC rms Capacitance RANGE...
Connect the black test lead to the “COM” jack and the red to the “INPUT” jack. Connect the test leads across the source or load to be measured. The LCD display will show the voltage value and DC/AC symbol. Also the polarity of the red lead connection will be indicated when making a DC measurement.
Note: a. For resistance measurements >10MΩ, the meter may take a few seconds to stabilize the reading. This is normal for high-resistance measurement. b. Before measuring in-circuit resistance, be sure that the circuit under test has all the power removed and all capacitors are fully discharged. Continuity Test When the LCD displays “- - - -”, the meter can measure resistance.
Note: When the capacitance under measurement is above 100uF, the meter needs at least 10 seconds to display a stable reading. NCV (Non-Contact-Voltage) Test 1) Press NCV/ button, to select NCV function the LCD will display EF only. 2) "NCV" means "non-contact voltage detection" without the test leads. 3) Use the orange probe on top of the meter to test the target.
Page 10
damaged fuse with a new fuse of the specified ratings. Re-install the back cover. Accessories • Remote cable test unit • Set of test leads (red and black) • 2pcs AAA batteries • Instruction manual EN61010–1:2010 This product is classed as Electrical or Electronic equipment and should not be disposed with other household or commercial waste at the end of its useful life.