HP 3468A Operator's Manual page 74

Hide thumbs Also See for 3468A:
Table of Contents

Advertisement

REMOTE
Examples
REMOTE
9
REMOTE
901
Comments
As a general rule, the command
REMOTE
9 should be executed
at the beginning
of each
program.
This ensures
that the controller
has reassigned
addresses
for
each device. The REMOTE command is used to enable the 3468A to switch from
local front panel control to remote program control. The 3468A
must actually be
addressed
before it will go into its Remote state.
The Remote state for the 3468A means that the front panel keyboard (except the
LOCAL and SRQ keys) is disabled. The RMT annunciator in the display is turned
on, The 3468A remains in the same state after it receives the REMOTE command
that it was in before receiving it. To disable the LOCAL
and SRO
keys use the
LOCAL LOCKOUT command.